Entry  Thu Apr 13 16:31:55 2017, Johannes Grossmann, MaPSA, Teleskope Correlation, 9, 1000000, 0, No, Unknown 

Telescope 1237

Entry  Thu Apr 13 16:23:35 2017, Johannes Grossmann, MaPSA, Teleskope Correlation, 7, 1000000, 0, No, Unknown 

Telescope 1236

threshold 90 

 

Entry  Fri Apr 14 13:21:07 2017, Johannes Grossmann, MaPSA, Alignment Check, 66, 250000, 0, No, Unknown 

Telescope 1318

Threshold 100

Mapsa CLK Delay 0

Reference Plane Lost aber Anfang Gut

Entry  Fri Apr 14 12:58:08 2017, Johannes Grossmann, MaPSA, Alignment Check, 64, 250000, 0, No, Unknown 

Telescope 1316

Threshold 80

Mapsa CLK Delay 0

Reference Plane Lost aber Anfang Gut

Entry  Tue Apr 18 21:55:44 2017, Johannes Grossmann, MaPSA, Rotation, 635, 0, 0, No, Unknown 

HV 100 V Threshold 75 Winkel 15

16:14 17.4.2017

Telescope Runnumber 1887

Entry  Thu Apr 20 10:07:30 2017, Johannes Grossmann, MaPSA, Rotation, 635, 0, 0, No, Unknown 

HV 100 V Threshold 75 Winkel 25

16:14 17.4.2017

Telescope Runnumber 1887

Entry  Fri Apr 14 12:48:44 2017, Johannes Grossmann, MaPSA, Alignment Check, 61, 250000, 0, No, Unknown 

Telescope 1314

Threshold 80

Mapsa CLK Delay 0

Reference Plane Lost aber Anfang Gut

Entry  Tue Apr 18 16:16:59 2017, Johannes Grossmann, MaPSA, Std Settings, 604, 0, 0, No, Unknown 

Changed HV 100 V Threshold 75

16:14 17.4.2017

Telescope Runnumber 1855

Entry  Fri Apr 14 12:26:00 2017, Johannes Grossmann, MaPSA, Alignment Check, 60, 250000, 0, No, Unknown 

Telescope 1312

Threshold 80

Mapsa CLK Delay 0

Carbon fiber  broken at 2:20 AM. Alignement check.

Entry  Mon Apr 17 23:52:47 2017, Johannes Grossmann, MaPSA, HV_Scan, 523, 0, 0, No, Unknown 

Changed HV 80 V

23:50 17.4.2017

Entry  Mon Apr 17 23:50:59 2017, Johannes Grossmann, MaPSA, HV_Scan, 522, 0, 0, No, Unknown 

Changed HV from 90 V to 80 V

Kann ich wegwerfen!

23:49 17.4.2017

Entry  Fri Apr 14 01:33:35 2017, Johannes Grossmann, MaPSA, Teleskope Correlation, 52, 250000, 0, No, Unknown 

Telescope 1296

Threshold 90 

Mapsa CLK Delay 0 

Entry  Fri Apr 14 00:40:23 2017, Johannes Grossmann, MaPSA, Teleskope Correlation, 51, 50000, 0, No, Unknown 

Telescope 1294

Threshold 90 

Mapsa CLK Delay 24 ns

Entry  Fri Apr 14 00:22:30 2017, Johannes Grossmann, MaPSA, Teleskope Correlation, 50, 50000, 0, No, Unknown 

Telescope 1292

Threshold 90 

Mapsa CLK Delay 22 ns

Entry  Fri Apr 14 00:12:53 2017, Johannes Grossmann, MaPSA, Teleskope Correlation, 49, 50000, 0, No, Unknown 

Telescope 1292

Threshold 90 

Mapsa CLK Delay 20 ns

Entry  Mon Apr 17 17:55:55 2017, Johannes Grossmann, MaPSA, HV_Scan, 488, 0, 0, No, Unknown 

Changed HV from 120 V to 90 V

17:53 17.4.2017

Entry  Fri Apr 14 00:01:03 2017, Johannes Grossmann, MaPSA, Teleskope Correlation, 48, 50000, 0, No, Unknown 

Telescope 1288

Threshold 90 

Mapsa CLK Delay 18 ns

Entry  Fri Apr 14 00:12:05 2017, Johannes Grossmann, MaPSA, Teleskope Correlation, 48, 50000, 0, No, Unknown 

Telescope 1288

Threshold 90 

Mapsa CLK Delay 20 ns

Entry  Thu Apr 13 23:50:36 2017, Johannes Grossmann, MaPSA, Teleskope Correlation, 47, 50000, 0, No, Unknown 

Telescope 1286

Threshold 90 

Mapsa CLK Delay 16 ns

Entry  Thu Apr 13 23:40:26 2017, Johannes Grossmann, MaPSA, Teleskope Correlation, 46, 50000, 0, No, Unknown 

Telescope 1284

Threshold 90 

Mapsa CLK Delay 14 ns

Entry  Mon Apr 17 12:26:33 2017, Johannes Grossmann, MaPSA, HV_Scan, 456, 0, 0, No, Unknown 

Changed HV from 60 V to 120 V

12:24 17.4.2017 

Entry  Thu Apr 13 23:30:39 2017, Johannes Grossmann, MaPSA, Teleskope Correlation, 45, 50000, 0, No, Unknown 

Telescope 1282

Threshold 90 

Mapsa CLK Delay 12 ns

Entry  Thu Apr 13 23:27:01 2017, Johannes Grossmann, MaPSA, Teleskope Correlation, 44, 50000, 0, No, Unknown 

Telescope 1280

Threshold 90 

Mapsa CLK Delay 10 ns

Entry  Thu Apr 13 23:20:07 2017, Johannes Grossmann, MaPSA, Teleskope Correlation, 43, 50000, 0, No, Unknown 

Telescope 1278

Threshold 90 

Mapsa CLK Delay 10 ns

Reference out of sync

Entry  Thu Apr 13 23:09:37 2017, Johannes Grossmann, MaPSA, Teleskope Correlation, 42, 50000, 0, No, Unknown 

Telescope 1278

Threshold 90 

Mapsa CLK Delay 8 ns

Entry  Thu Apr 13 22:59:32 2017, Johannes Grossmann, MaPSA, Teleskope Correlation, 40, 50000, 0, No, Unknown 

Telescope 1276

Threshold 90 

Mapsa CLK Delay 8 ns

Reference out of sync

Entry  Thu Apr 13 14:54:15 2017, Johannes Grossmann, MaPSA, Teleskope Correlation, 4, 1000000, 0, No, Unknown 

Telescope 1209

threshold 90 

 

Entry  Thu Apr 13 22:52:18 2017, Johannes Grossmann, MaPSA, Teleskope Correlation, 39, 50000, 0, No, Unknown 

Telescope 1274

Threshold 90 

Mapsa CLK Delay 6 ns

Entry  Sun Apr 16 22:42:51 2017, Johannes Grossmann, MaPSA, HV_Scan, 387, 0, 0, No, Unknown 

Changed HV from 100V to 80 V

22:41 16.4.2017

Entry  Mon Apr 17 10:10:33 2017, Johannes Grossmann, MaPSA, HV_Scan, 387, 0, 0, No, Unknown 

Changed HV from 80 V to 60 V

Not in condition data applied in the settings

10:08 17.4.2017 

Entry  Thu Apr 13 22:41:52 2017, Johannes Grossmann, MaPSA, Teleskope Correlation, 38, 50000, 0, No, Unknown 

Telescope 1272

Threshold 90 

Mapsa CLK Delay 6 ns

Reference Plane out of sync

Entry  Thu Apr 13 22:31:22 2017, Johannes Grossmann, MaPSA, Teleskope Correlation, 37, 50000, 0, No, Unknown 

Telescope 1271

Threshold 90 

Mapsa CLK Delay 4 ns

Entry  Thu Apr 13 22:25:00 2017, Johannes Grossmann, MaPSA, Teleskope Correlation, 36, 50000, 0, No, Unknown 

Telescope 1270

keine Reference Plane ! 

Threshold 90 

Delay 4 ns

Entry  Thu Apr 13 22:07:49 2017, Johannes Grossmann, MaPSA, Teleskope Correlation, 35, 50000, 0, No, Unknown 

Telescope 1269

Threshold 90 

Delay 2 ns

Entry  Thu Apr 13 22:04:40 2017, Johannes Grossmann, MaPSA, Teleskope Correlation, 34, 50000, 0, No, Unknown 

Telescope 1267

Threshold 90 

+2 ns 

Entry  Thu Apr 13 21:46:22 2017, Johannes Grossmann, MaPSA, Teleskope Correlation, 33, 1000000, 0, No, Unknown 

Telescope 1267

Threshold 90

4 mm +x 

Ort Gut aligned! 

Entry  Thu Apr 13 21:11:12 2017, Johannes Grossmann, MaPSA, Teleskope Correlation, 32, 1000000, 0, No, Unknown 

Telescope 1265

Threshold 90

Entry  Thu Apr 13 20:55:36 2017, Johannes Grossmann, MaPSA, Teleskope Correlation, 31, 100000, 0, No, Unknown 

Telescope 1263

Threshold 100

Sample 46_4

Entry  Thu Apr 13 18:14:00 2017, Johannes Grossmann, MaPSA, Teleskope Correlation, 18, 100000, 0, No, Unknown 

Telescope 1250

Threshold 130

Sample 46_4

Entry  Thu Apr 13 18:03:02 2017, Johannes Grossmann, MaPSA, Teleskope Correlation, 17, 100000, 0, No, Unknown 

Telescope 1241

Threshold 240

Entry  Thu Apr 13 17:51:44 2017, Johannes Grossmann, MaPSA, Teleskope Correlation, 16, 100000, 0, No, Unknown 

Telescope 1241

Threshold 200

Entry  Thu Apr 13 17:38:40 2017, Johannes Grossmann, MaPSA, Teleskope Correlation, 14, 100000, 0, No, Unknown 

Telescope 1241

Threshold 110

Entry  Thu Apr 13 17:29:09 2017, Johannes Grossmann, MaPSA, Teleskope Correlation, 13, 100000, 0, No, Unknown 

Telescope 1241

Threshold 110

Entry  Thu Apr 13 17:08:11 2017, Johannes Grossmann, MaPSA, Teleskope Correlation, 12, 100000, 0, No, Unknown 

Telescope 1241

Threshold 110

Entry  Thu Apr 13 17:04:52 2017, Johannes Grossmann, MaPSA, Teleskope Correlation, 11, 100000, 0, No, Unknown 

Telescope 1241

Threshold 90

Entry  Thu Apr 13 16:47:40 2017, Johannes Grossmann, MaPSA, Teleskope Correlation, 10, 1000000, 0, No, Unknown 

Telescope 1241

Threshold 90

Entry  Thu Apr 13 13:07:17 2017, Johannes Grossmann, MaPSA, Teleskope Correlation, 1, 1000000, 0, No, Unknown 

Telescope 1209

Entry  Sat Apr 15 12:12:00 2017, Axel Koenig, ALiBaVa (IFX), run_030_IFX_CB_100V.dat, 030, 1000000, 109, 4.5, 240, No, Unknown 

3Remarks: Telescope run 558. Bias scan 100 V. Telescope chiller turned off. Could be the root cause of crashing runs.

Parameter Value
Alibava run run_030_IFX_CB_100V.dat
Telescope run 558
Triggers 1000000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -100
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

Entry  Sat Apr 15 09:57:51 2017, Axel Koenig, ALiBaVa (IFX), run_029_IFX_CB_130V.dat, 029, 175000, 109, 4.5, 240, No, Unknown 

3Remarks: Telescope run 557. Bias scan 130 V.

Parameter Value
Alibava run run_029_IFX_CB_130V.dat
Telescope run 557
Triggers 175000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -130
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

Entry  Fri Apr 14 21:11:18 2017, Axel Koenig, ALiBaVa (IFX), run_028_IFX_CB_100V.dat, 028, 3000000, 109, 4.5, 240, No, Crap 

Remarks: Telescope run 551. Longrun over night. Run crashed. Events only up to 2288000 valid.

Parameter Value
Alibava run run_028_IFX_CB_100V.dat
Telescope run 551
Triggers 3000000
Alibava trigger eff. ~60 %
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -100
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg

Entry  Fri Apr 14 19:43:13 2017, Axel Koenig, ALiBaVa (IFX), run_027_IFX_CB_100V.dat, 027, 350000, 109, 4.5, 250, No, Unknown 

Remarks: Telescope run 550. Angle shift 250 deg.

Parameter Value
Alibava run run_027_IFX_CB_100V.dat
Telescope run 550
Triggers 350000
Alibava trigger eff. ~59.8%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -100
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

250 deg

Entry  Fri Apr 14 18:22:58 2017, Axel Koenig, ALiBaVa (IFX), run_026_IFX_CB_100V.dat, 026, 350000, 109, 4.5, 230, No, Unknown 

Remarks: Telescope run 549. Angle shift 230 deg.

Parameter Value
Alibava run run_026_IFX_CB_100V.dat
Telescope run 549
Triggers 350000
Alibava trigger eff. ~59.8%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -100
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

230 deg

Entry  Fri Apr 14 17:22:52 2017, Axel Koenig, ALiBaVa (IFX), run_025_IFX_CB_150V.dat, 025, 175000, 109, 4.5, 240, No, Unknown 

3Remarks: Telescope run 548. Bias scan 150 V.

Parameter Value
Alibava run run_025_IFX_CB_150V.dat
Telescope run 548
Triggers 175000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -150
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

Entry  Fri Apr 14 09:25:09 2017, Axel Koenig, ALiBaVa (IFX), run_024_IFX_CB_140V.dat, 024, 175000, 109, 4.5, 240, No, Unknown 

3Remarks: Telescope run 547. Bias scan 140 V.

Parameter Value
Alibava run run_024_IFX_CB_140V.dat
Telescope run 547
Triggers 175000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -140
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

Entry  Fri Apr 14 14:49:09 2017, Axel Koenig, ALiBaVa (IFX), run_023_IFX_CB_100V.dat, 023, 385000, 109, 0, 240, No, Crap 

Remarks: Telescope run 546. Bias voltage 100 V. Beam position on only one chip. Run destroyed. Intended trigger count 500k

Parameter Value
Alibava run run_023_IFX_CB_100V.dat
Telescope run 546
Triggers 385000
Alibava trigger eff ~62 %
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -100 V
Latency 132
Table x 109 mm
table z 0 mm
table rot 240 deg (parallel to tele. planes)
Entry  Mon Apr 17 12:06:11 2017, Axel Koenig, ALiBaVa (IFX), run_023_Baby_150V.dat, 023, 112000, 102.5, 22.5, 240, No, Unknown 

Remarks: Telescope run 600. Bias scan 150 V.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_023_Baby_150V.dat
Telescope run 600
Triggers 112000
Alibava trigger eff. ~90%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -150
Latency 132
Table x 102.5 mm
Table z 22.5 mm
Tabe rot

240 deg

Entry  Fri Apr 14 13:06:30 2017, Axel Koenig, ALiBaVa (IFX), run_022_IFX_CB_100V.dat, 022, 500000, 109, 10.5, 240, No, Unknown 

Remarks: Telescope run 545. Bias voltage 100 V. Beam position on only one chip.

Parameter Value
Alibava run run_022_IFX_CB_100V.dat
Telescope run 545
Triggers 500000
Alibava trigger eff ~62 %
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -100 V
Latency 132
Table x 109 mm
table z 10.5 mm
table rot 240 deg (parallel to tele. planes)
Entry  Mon Apr 17 11:34:55 2017, Axel Koenig, ALiBaVa (IFX), run_022_Baby_140V.dat, 022, 112000, 102.5, 22.5, 240, No, Unknown 

Remarks: Telescope run 599. Bias scan 140 V.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_022_Baby_140V.dat
Telescope run 599
Triggers 112000
Alibava trigger eff. ~90%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -140
Latency 132
Table x 102.5 mm
Table z 22.5 mm
Tabe rot

240 deg

Entry  Mon Apr 17 11:04:40 2017, Axel Koenig, ALiBaVa (IFX), run_021_Baby_130V.dat, 021, 112000, 102.5, 22.5, 240, No, Unknown 

Remarks: Telescope run 598. Bias scan 130 V.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_021_Baby_130V.dat
Telescope run 598
Triggers 112000
Alibava trigger eff. ~90%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -130
Latency 132
Table x 102.5 mm
Table z 22.5 mm
Tabe rot

240 deg

Entry  Fri Apr 21 09:14:45 2017, Axel Koenig, ALiBaVa (IFX), run_021_Irrad_500V.dat, 021, 200000, 96, 64, 360, No, Unknown 

Remarks: Telescope run 653. Bias voltage -500V, central position of sensor (DUT parallel to telescope planes at 360 deg due to different fixation compared to other DUTs). Telescope geometry identical to double irrad module.

Backside Biasing!

Beam energy scan.

Telescope chiller turned on (15 deg).

Parameter Value
Alibava run run_021_Irrad_500V.dat
Telescope run 653
Triggers 200000
Alibava trigger eff. ~70%
Beam energy 4.6 GeV
Telescope threshold 6
Bias voltage -500
Latency 132
Table x 96 mm
Table z 64 mm
Tabe rot

360 deg

Temperature at DUT -14.5 +- 1 deg. celsius
Current 39.2 uA
Entry  Fri Apr 14 08:59:26 2017, Axel Koenig, ALiBaVa (IFX), run_020_IFX_CB_100V.dat, 020, 190000, 109, 4.5, 240, No, Crap 

Remarks: Telescope run 535. Bias scan 100 V. Run crashed over night. Telescope sync only up to event 190000 (planned 2500000 events).

Parameter Value
Alibava run run_020_IFX_CB_100V.dat
Telescope run 535
Triggers 190000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -100
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

Entry  Mon Apr 17 10:36:22 2017, Axel Koenig, ALiBaVa (IFX), run_020_Baby_120V.dat, 020, 112000, 102.5, 22.5, 240, No, Unknown 

Remarks: Telescope run 597. Bias scan 120 V.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_020_Baby_120V.dat
Telescope run 597
Triggers 112000
Alibava trigger eff. ~90%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -120
Latency 132
Table x 102.5 mm
Table z 22.5 mm
Tabe rot

240 deg

Entry  Fri Apr 21 08:57:30 2017, Axel Koenig, ALiBaVa (IFX), run_020_Irrad_500V.dat, 020, 200000, 96, 64, 360, No, Unknown 

Remarks: Telescope run 652. Bias voltage -500V, central position of sensor (DUT parallel to telescope planes at 360 deg due to different fixation compared to other DUTs). Telescope geometry identical to double irrad module.

Backside Biasing!

Beam energy scan.

Telescope chiller turned on (15 deg).

Parameter Value
Alibava run run_020_Irrad_500V.dat
Telescope run 652
Triggers 200000
Alibava trigger eff. ~70%
Beam energy 4.8 GeV
Telescope threshold 6
Bias voltage -500
Latency 132
Table x 96 mm
Table z 64 mm
Tabe rot

360 deg

Temperature at DUT -14.5 +- 1 deg. celsius
Current 39.2 uA
Entry  Mon Apr 17 10:05:57 2017, Axel Koenig, ALiBaVa (IFX), run_019_Baby_110V.dat, 019, 112000, 102.5, 22.5, 240, No, Unknown 

Remarks: Telescope run 596. Bias scan 110 V.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_019_Baby_110V.dat
Telescope run 596
Triggers 112000
Alibava trigger eff. ~90%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -110
Latency 132
Table x 102.5 mm
Table z 22.5 mm
Tabe rot

240 deg

Entry  Fri Apr 21 08:38:17 2017, Axel Koenig, ALiBaVa (IFX), run_019_Irrad_500V.dat, 019, 200000, 96, 64, 360, No, Unknown 

Remarks: Telescope run 651. Bias voltage -500V, central position of sensor (DUT parallel to telescope planes at 360 deg due to different fixation compared to other DUTs). Telescope geometry identical to double irrad module.

Backside Biasing!

Beam energy scan.

Telescope chiller turned on (15 deg).

Parameter Value
Alibava run run_019_Irrad_500V.dat
Telescope run 651
Triggers 200000
Alibava trigger eff. ~70%
Beam energy 5 GeV
Telescope threshold 6
Bias voltage -500
Latency 132
Table x 96 mm
Table z 64 mm
Tabe rot

360 deg

Temperature at DUT -14.5 +- 1 deg. celsius
Current 39.2 uA
Entry  Thu Apr 13 21:23:11 2017, Axel Koenig, ALiBaVa (IFX), run_018_IFX_CB_120V.dat, 018, 175000, 109, 4.5, 240, No, Unknown 

Remarks: Telescope run 533. Bias scan 120 V.

Parameter Value
Alibava run run_018_IFX_CB_120V.dat
Telescope run 533
Triggers 175000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -120
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

Entry  Mon Apr 17 09:29:40 2017, Axel Koenig, ALiBaVa (IFX), run_018_Baby_90V.dat, 018, 112000, 102.5, 22.5, 240, No, Unknown 

Remarks: Telescope run 595. Bias scan 90 V.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_018_Baby_90V.dat
Telescope run 595
Triggers 112000
Alibava trigger eff. ~90%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -90
Latency 132
Table x 102.5 mm
Table z 22.5 mm
Tabe rot

240 deg

Entry  Fri Apr 21 08:15:03 2017, Axel Koenig, ALiBaVa (IFX), run_018_Irrad_500V.dat, 018, 200000, 96, 64, 360, No, Unknown 

Remarks: Telescope run 650. Bias voltage -500V, central position of sensor (DUT parallel to telescope planes at 360 deg due to different fixation compared to other DUTs). Telescope geometry identical to double irrad module.

Backside Biasing!

Beam energy scan.

Telescope chiller turned on (15 deg).

Parameter Value
Alibava run run_018_Irrad_500V.dat
Telescope run 650
Triggers 200000
Alibava trigger eff. ~70%
Beam energy 5.2 GeV
Telescope threshold 6
Bias voltage -500
Latency 132
Table x 96 mm
Table z 64 mm
Tabe rot

360 deg

Temperature at DUT -14.5 +- 1 deg. celsius
Current 39.2 uA
Entry  Thu Apr 13 20:49:30 2017, Axel Koenig, ALiBaVa (IFX), run_017_IFX_CB_110V.dat, 017, 175000, 109, 4.5, 240, No, Unknown 

Remarks: Telescope run 532. Bias scan 110 V.

Parameter Value
Alibava run run_017_IFX_CB_110V.dat
Telescope run 532
Triggers 175000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -110
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

Entry  Mon Apr 17 09:01:09 2017, Axel Koenig, ALiBaVa (IFX), run_017_Baby_80V.dat, 017, 112000, 102.5, 22.5, 240, No, Unknown 

Remarks: Telescope run 594. Bias scan 80 V.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_017_Baby_80V.dat
Telescope run 594
Triggers 112000
Alibava trigger eff. ~90%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -80
Latency 132
Table x 102.5 mm
Table z 22.5 mm
Tabe rot

240 deg

Entry  Fri Apr 21 07:46:57 2017, Axel Koenig, ALiBaVa (IFX), run_017_Irrad_500V.dat, 017, 200000, 96, 64, 360, No, Unknown 

Remarks: Telescope run 649. Bias voltage -500V, central position of sensor (DUT parallel to telescope planes at 360 deg due to different fixation compared to other DUTs). Telescope geometry identical to double irrad module.

Backside Biasing!

Beam energy scan.

Telescope chiller turned on (15 deg).

Parameter Value
Alibava run run_017_Irrad_500V.dat
Telescope run 649
Triggers 200000
Alibava trigger eff. ~70%
Beam energy 5.4 GeV
Telescope threshold 6
Bias voltage -500
Latency 132
Table x 96 mm
Table z 64 mm
Tabe rot

360 deg

Temperature at DUT -14.5 +- 1 deg. celsius
Current 39.2 uA
Entry  Thu Apr 13 18:26:57 2017, Axel Koenig, ALiBaVa (IFX), run_016_IFX_CB_100V.dat, 016, 750000, 109, 4.5, 240, No, Crap 

3Remarks: Telescope run 527-531. Bias scan 100 V. Telescope got out of sync. Match with telescope data only up to 175000 events.

Parameter Value
Alibava run run_016_IFX_CB_100V.dat
Telescope run 527-531
Triggers 750000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -100
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

Entry  Mon Apr 17 08:14:24 2017, Axel Koenig, ALiBaVa (IFX), run_016_Baby_75V.dat, 016, 112000, 102.5, 22.5, 240, No, Unknown 

Remarks: Telescope run 593. Bias scan 75 V.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_016_Baby_75V.dat
Telescope run 593
Triggers 112000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -75
Latency 132
Table x 102.5 mm
Table z 22.5 mm
Tabe rot

240 deg

Entry  Thu Apr 20 22:19:21 2017, Axel Koenig, ALiBaVa (IFX), run_016_Irrad_500V.dat, 016, 2500000, 96, 64, 370, No, Unknown 

Remarks: Telescope run 648. Bias voltage -500V, central position of sensor (DUT parallel to telescope planes at 360 deg due to different fixation compared to other DUTs). Telescope geometry identical to double irrad module.

Backside Biasing!

rotation scan.

Telescope chiller turned on (15 deg).

Run finished over night!

Parameter Value
Alibava run run_016_Irrad_500V.dat
Telescope run 648
Triggers 2500000
Alibava trigger eff. ~70%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -500
Latency 132
Table x 96 mm
Table z 64 mm
Tabe rot

370 deg

Temperature at DUT -14.5 +- 1 deg. celsius
Current 39.2 uA
Entry  Thu Apr 13 17:47:28 2017, Axel Koenig, ALiBaVa (IFX), run_015_IFX_CB_90V.dat, 015, 175000, 109, 4.5, 240, No, Unknown 

3Remarks: Telescope run 526. Bias scan 90 V.

Parameter Value
Alibava run run_015_IFX_CB_90V.dat
Telescope run 526
Triggers 175000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -90
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

Entry  Sun Apr 16 21:43:15 2017, Axel Koenig, ALiBaVa (IFX), run_015_Baby_100V.dat, 015, 2500000, 102.5, 22.5, 240, No, Unknown 

Remarks: Telescope run 592. Long run, bias voltage 100 V. Long run finished without any problems.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_015_Baby_100V.dat
Telescope run 592
Triggers 2500000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -100
Latency 132
Table x 102.5 mm
Table z 22.5 mm
Tabe rot

240 deg

Entry  Thu Apr 20 20:15:12 2017, Axel Koenig, ALiBaVa (IFX), run_015_Irrad_500V.dat, 015, 300000, 96, 64, 350, No, Unknown 

Remarks: Telescope run 647. Bias voltage -500V, central position of sensor (DUT parallel to telescope planes at 360 deg due to different fixation compared to other DUTs). Telescope geometry identical to double irrad module.

Backside Biasing!

rotation scan.

Telescope chiller turned on (15 deg).

Parameter Value
Alibava run run_015_Irrad_500V.dat
Telescope run 647
Triggers 300000
Alibava trigger eff. ~70%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -500
Latency 132
Table x 96 mm
Table z 64 mm
Tabe rot

350 deg

Temperature at DUT -14.5 +- 1 deg. celsius
Current 39.2 uA
Entry  Thu Apr 13 17:11:31 2017, Axel Koenig, ALiBaVa (IFX), run_014_IFX_CB_80V.dat, 014, 175000, 109, 4.5, 240, No, Unknown 

3Remarks: Telescope run 525. Bias scan 80 V.

Parameter Value
Alibava run run_014_IFX_CB_80V.dat
Telescope run 525
Triggers 175000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -80
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

Entry  Sun Apr 16 20:22:27 2017, Axel Koenig, ALiBaVa (IFX), run_014_Baby_70V.dat, 014, 250000, 102.5, 22.5, 240, No, Unknown 

Remarks: Telescope run 591. Bias scan 70 V.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_014_Baby_70V.dat
Telescope run 591
Triggers 250000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -70
Latency 132
Table x 102.5 mm
Table z 22.5 mm
Tabe rot

240 deg

Entry  Thu Apr 20 21:25:21 2017, Axel Koenig, ALiBaVa (IFX), run_014_Irrad_500V.dat, 014, 300000, 96, 69, 360, No, Unknown 

Remarks: Telescope run 646. Bias voltage -500V, central position of sensor (DUT parallel to telescope planes at 360 deg due to different fixation compared to other DUTs). Telescope geometry identical to double irrad module.

Backside Biasing!

Positoj scan vertical.

Telescope chiller turned on (15 deg).

Parameter Value
Alibava run run_014_Irrad_500V.dat
Telescope run 646
Triggers 300000
Alibava trigger eff. ~70%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -500
Latency 132
Table x 96 mm
Table z 69 mm
Tabe rot

360 deg

Temperature at DUT -14.5 +- 1 deg. celsius
Current 39.2 uA
Entry  Thu Apr 13 16:33:32 2017, Axel Koenig, ALiBaVa (IFX), run_013_IFX_CB_75V.dat, 013, 175000, 109, 4.5, 240, No, Unknown 

Remarks: Telescope run 524. Bias scan 75 V.

Parameter Value
Alibava run run_013_IFX_CB_75V.dat
Telescope run 524
Triggers 175000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -75
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

Entry  Sun Apr 16 19:46:01 2017, Axel Koenig, ALiBaVa (IFX), run_013_Baby_65V.dat, 013, 112000, 102.5, 22.5, 240, No, Unknown 

Remarks: Telescope run 590. Bias scan 65 V.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_013_Baby_65V.dat
Telescope run 590
Triggers 112000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -65
Latency 132
Table x 102.5 mm
Table z 22.5 mm
Tabe rot

240 deg

Entry  Thu Apr 20 16:05:14 2017, Axel Koenig, ALiBaVa (IFX), run_013_Irrad_500V.dat, 013, 750000, 96, 54, 360, No, Unknown 

Remarks: Telescope run 643. Bias voltage -500V, central position of sensor (DUT parallel to telescope planes at 360 deg due to different fixation compared to other DUTs). Telescope geometry identical to double irrad module.

Backside Biasing!

Positoj scan vertical.

Telescope chiller turned on (15 deg).

Parameter Value
Alibava run run_013_Irrad_500V.dat
Telescope run 643
Triggers 750000
Alibava trigger eff. ~70%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -500
Latency 132
Table x 96 mm
Table z 54 mm
Tabe rot

360 deg

Temperature at DUT -14.5 +- 1 deg. celsius
Current 39.2 uA
Entry  Thu Apr 13 16:00:25 2017, Axel Koenig, ALiBaVa (IFX), run_012_IFX_CB_70V.dat, 012, 175000, 109, 4.5, 240, No, Unknown 

Remarks: Telescope run 523. Bias scan 70 V.

Parameter Value
Alibava run run_012_IFX_CB_70V.dat
Telescope run 523
Triggers 175000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -70
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

Entry  Sun Apr 16 19:04:41 2017, Axel Koenig, ALiBaVa (IFX), run_012_Baby_60V.dat, 012, 135000, 102.5, 22.5, 240, No, Unknown 

Remarks: Telescope run 589. Bias scan 60 V.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_012_Baby_60V.dat
Telescope run 589
Triggers 135000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -60
Latency 132
Table x 102.5 mm
Table z 22.5 mm
Tabe rot

240 deg

Entry  Thu Apr 20 15:08:33 2017, Axel Koenig, ALiBaVa (IFX), run_012_Irrad_250V.dat, 012, 300000, 96, 64, 360, No, Unknown 

Remarks: Telescope run 642. Bias scan -250V, central position of sensor (DUT parallel to telescope planes at 360 deg due to different fixation compared to other DUTs). Telescope geometry identical to double irrad module.

Backside Biasing!

Telescope chiller turned on (15 deg).

Parameter Value
Alibava run run_012_Irrad_250V.dat
Telescope run 642
Triggers 300000
Alibava trigger eff. ~66%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -250
Latency 132
Table x 96 mm
Table z 64 mm
Tabe rot

360 deg

Temperature at DUT -14.5 +- 1 deg. celsius
Current 32.1 uA
Entry  Thu Apr 13 15:26:22 2017, Axel Koenig, ALiBaVa (IFX), run_011_IFX_CB_65V.dat, 011, 175000, 109, 4.5, 240, No, Unknown 

Remarks: Telescope run 522. Bias scan 65 V.

Parameter Value
Alibava run run_011_IFX_CB_65V.dat
Telescope run 522
Triggers 175000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -65
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

Entry  Sun Apr 16 18:25:59 2017, Axel Koenig, ALiBaVa (IFX), run_011_Baby_50V.dat, 011, 135000, 102.5, 22.5, 240, No, Unknown 

Remarks: Telescope run 588. Bias scan 50 V.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_011_Baby_50V.dat
Telescope run 588
Triggers 135000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -50
Latency 132
Table x 102.5 mm
Table z 22.5 mm
Tabe rot

240 deg

Entry  Thu Apr 20 14:09:44 2017, Axel Koenig, ALiBaVa (IFX), run_011_Irrad_350V.dat, 011, 300000, 96, 64, 360, No, Unknown 

Remarks: Telescope run 641. Bias scan -350V, central position of sensor (DUT parallel to telescope planes at 360 deg due to different fixation compared to other DUTs). Telescope geometry identical to double irrad module.

Backside Biasing!

Telescope chiller turned on (15 deg).

Parameter Value
Alibava run run_011_Irrad_350V.dat
Telescope run 641
Triggers 300000
Alibava trigger eff. ~70%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -350
Latency 132
Table x 96 mm
Table z 64 mm
Tabe rot

360 deg

Temperature at DUT -14.5 +- 1 deg. celsius
Current 37.9 uA
Entry  Thu Apr 13 13:07:39 2017, Axel Koenig, ALiBaVa (IFX), run_010_IFX_CB_60V.dat, 010, 175000, 109, 4.5, 240, No, Unknown 

Remarks: Telescope run 521. Bias scan 60 V.

Parameter Value
Alibava run run_010_IFX_CB_60V.dat
Telescope run 521
Triggers 175000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -60
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

Entry  Sun Apr 16 17:42:06 2017, Axel Koenig, ALiBaVa (IFX), run_010_Baby_40V.dat, 010, 135000, 102.5, 22.5, 240, No, Unknown 

Remarks: Telescope run 587. Bias scan 40 V.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_010_Baby_40V.dat
Telescope run 587
Triggers 135000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -40
Latency 132
Table x 102.5 mm
Table z 22.5 mm
Tabe rot

240 deg

Entry  Tue Apr 18 08:07:15 2017, Axel Koenig, ALiBaVa (IFX), run_010_HPK_CB_200V.dat, 010, 350000, 107, 2.6, 240, No, Unknown 

Remarks: Telescope run 616. Short run central position. Repeated due to crash over night.

No beam during run.Only ~10000 events.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_010_HPK_CB_200V.dat
Telescope run 616
Triggers 350000
Alibava trigger eff. ~62%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -200
Latency 132
Table x 107 mm
Table z 2.6 mm
Tabe rot

240 deg

Entry  Thu Apr 20 12:55:02 2017, Axel Koenig, ALiBaVa (IFX), run_010_Irrad_650V.dat, 010, 300000, 96, 64, 360, No, Unknown 

Remarks: Telescope run 640. Bias scan -650V, central position of sensor (DUT parallel to telescope planes at 360 deg due to different fixation compared to other DUTs). Telescope geometry identical to double irrad module.

Backside Biasing!

Telescope chiller turned on (15 deg).

Parameter Value
Alibava run run_010_Irrad_650V.dat
Telescope run 640
Triggers 300000
Alibava trigger eff. ~70%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -650
Latency 132
Table x 96 mm
Table z 64 mm
Tabe rot

360 deg

Temperature at DUT -14.5 +- 1 deg. celsius
Current 45.4 uA
Entry  Thu Apr 13 12:33:20 2017, Axel Koenig, ALiBaVa (IFX), run_009_IFX_CB_50V.dat, 009, 175000, 109, 4.5, 240, No, Unknown 

Remarks: Telescope run 517. Bias scan 50 V.

Parameter Value
Alibava run run_009_IFX_CB_50V.dat
Telescope run 517
Triggers 175000
Alibava trigger eff. ~58%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -50
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

Entry  Sun Apr 16 15:44:35 2017, Axel Koenig, ALiBaVa (IFX), run_009_Baby_30V.dat, 009, 135000, 102.5, 22.5, 240, No, Unknown 

Remarks: Telescope run 586. Bias scan 30 V.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_009_Baby_30V.dat
Telescope run 586
Triggers 135000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -30
Latency 132
Table x 102.5 mm
Table z 22.5 mm
Tabe rot

240 deg

Entry  Mon Apr 17 22:27:44 2017, Axel Koenig, ALiBaVa (IFX), run_009_HPK_CB_200V.dat, 009, 2400000, 107, 2.6, 240, No, Unknown 

Remarks: Telescope run 614. Long run over night. run crashed over night!

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_009_HPK_CB_200V.dat
Telescope run 614
Triggers 2400000
Alibava trigger eff. ~62%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -200
Latency 132
Table x 107 mm
Table z 2.6 mm
Tabe rot

240 deg

Entry  Thu Apr 20 10:29:10 2017, Axel Koenig, ALiBaVa (IFX), run_009_Irrad_600V.dat, 009, 600000, 96, 64, 360, No, Unknown 

Remarks: Telescope run 639. Bias scan -600V, central position of sensor (DUT parallel to telescope planes at 360 deg due to different fixation compared to other DUTs). Telescope geometry identical to double irrad module.

Backside Biasing!

Telescope chiller turned on (15 deg).

Parameter Value
Alibava run run_009_Irrad_600V.dat
Telescope run 639
Triggers 600000
Alibava trigger eff. ~70%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -600
Latency 132
Table x 96 mm
Table z 64 mm
Tabe rot

360 deg

Temperature at DUT -14.5 +- 1 deg. celsius
Current 43.4 uA
Entry  Thu Apr 13 12:00:40 2017, Axel Koenig, ALiBaVa (IFX), run_008_IFX_CB_40V.dat, 008, 175000, 109, 4.5, 240, No, Unknown 

Remarks: Telescope run 515. Bias scan 40 V.

Parameter Value
Alibava run run_008_IFX_CB_40V.dat
Telescope run 515
Triggers 175000
Alibava trigger eff. ~58%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -40
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

Entry  Sun Apr 16 13:15:55 2017, Axel Koenig, ALiBaVa (IFX), run_008_Baby_20V.dat, 008, 175000, 102.5, 22.5, 240, No, Unknown 

Remarks: Telescope run 583. Bias scan 20 V. run crashed.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_008_Baby_20V.dat
Telescope run 583
Triggers 175000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -20
Latency 132
Table x 102.5 mm
Table z 22.5 mm
Tabe rot

240 deg

Entry  Mon Apr 17 21:52:45 2017, Axel Koenig, ALiBaVa (IFX), run_008_HPK_CB_250V.dat, 008, 150000, 107, 2.6, 240, No, Unknown 

Remarks: Telescope run 613. Bias scan 250 V.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_008_HPK_CB_250V.dat
Telescope run 613
Triggers 150000
Alibava trigger eff. ~62%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -250
Latency 132
Table x 107 mm
Table z 2.6 mm
Tabe rot

240 deg

Entry  Thu Apr 20 09:31:58 2017, Axel Koenig, ALiBaVa (IFX), run_008_Irrad_550V.dat, 008, 300000, 96, 64, 360, No, Unknown 

Remarks: Telescope run 638. Bias scan -550V, central position of sensor (DUT parallel to telescope planes at 360 deg due to different fixation compared to other DUTs). Telescope geometry identical to double irrad module.

Backside Biasing!

Telescope chiller turned on (15 deg).

Parameter Value
Alibava run run_008_Irrad_550V.dat
Telescope run 638
Triggers 300000
Alibava trigger eff. ~70%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -550
Latency 132
Table x 96 mm
Table z 64 mm
Tabe rot

360 deg

Temperature at DUT -14.5 +- 1 deg. celsius
Current 41.8 uA
Entry  Thu Apr 13 11:25:32 2017, Axel Koenig, ALiBaVa (IFX), run_007_IFX_CB_30V.dat, 007, 170000, 109, 4.5, 240, No, Unknown 

Remarks: Telescope run 513. Bias scan 30 V.

Parameter Value
Alibava run run_007_IFX_CB_30V.dat
Telescope run 513
Triggers 170000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -30
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

Entry  Sat Apr 15 21:18:26 2017, Axel Koenig, ALiBaVa (IFX), run_007_Baby_10V.dat, 007, 450000, 102.5, 22.5, 240, No, Crap 

Remarks: Telescope run 579. Bias scan 10 V. run crashed. Data vaild up to 274915.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_007_Baby_10V.dat
Telescope run 579
Triggers 450000
Alibava trigger eff. ~24%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -10
Latency 132
Table x 102.5 mm
Table z 22.5 mm
Tabe rot

240 deg

Entry  Mon Apr 17 19:54:08 2017, Axel Koenig, ALiBaVa (IFX), run_007_HPK_CB_150V.dat, 007, 300000, 107, 2.6, 240, No, Unknown 

Remarks: Telescope run 611. Bias scan 150 V. Bias scan. Run crashed at 294000 triggers!

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_007_HPK_CB_150V.dat
Telescope run 611
Triggers 300000
Alibava trigger eff. ~62%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -150
Latency 132
Table x 107 mm
Table z 2.6 mm
Tabe rot

240 deg

Entry  Thu Apr 20 08:42:29 2017, Axel Koenig, ALiBaVa (IFX), run_007_Irrad_500V.dat, 007, 250000, 96, 64, 360, No, Unknown 

Remarks: Telescope run 637. Bias scan -500V, central position of sensor (DUT parallel to telescope planes at 360 deg due to different fixation compared to other DUTs). Telescope geometry identical to double irrad module.

Backside Biasing!

Telescope chiller turned on (15 deg).

Parameter Value
Alibava run run_007_Irrad_500V.dat
Telescope run 637
Triggers 250000
Alibava trigger eff. ~70%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -500
Latency 132
Table x 96 mm
Table z 64 mm
Tabe rot

360 deg

Temperature at DUT -14.5 +- 1 deg. celsius
Current 41.5 uA
Entry  Thu Apr 13 10:58:57 2017, Axel Koenig, ALiBaVa (IFX), run_006_IFX_CB_20V.dat, 006, 185000, 109, 4.5, 240, No, Unknown 

Remarks: Telescope run 510. Bias scan 20 V.

Parameter Value
Alibava run run_006_IFX_CB_10V.dat
Telescope run 510
Triggers 185000
Alibava trigger eff. ~55.6%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -20
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

Entry  Mon Apr 17 09:04:02 2017, Axel Koenig, ALiBaVa (IFX), run_006_Baby_100V.dat, 006, k.a., 102.5, 22.5, 230, No, Unknown 

Remarks: Telescope run 570. Telescope chiller turned off (could be the root cause of runs crashing).

Forgot ro record data in elog.

Parameter Value
Alibava run run_006_Baby_100V.dat
Telescope run 570
Triggers k.a.
Alibava trigger eff. ~93%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -100
Latency 132
Table x 102.5 mm
Table z 22.5 mm
Tabe rot

230 deg

Entry  Mon Apr 17 19:19:14 2017, Axel Koenig, ALiBaVa (IFX), run_006_HPK_CB_100V.dat, 006, 150000, 107, 2.6, 240, No, Unknown 

Remarks: Telescope run 610. Bias scan 100 V. Bias scan.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_006_HPK_CB_100V.dat
Telescope run 610
Triggers 150000
Alibava trigger eff. ~62%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -100
Latency 132
Table x 107 mm
Table z 2.6 mm
Tabe rot

240 deg

Entry  Thu Apr 20 07:55:02 2017, Axel Koenig, ALiBaVa (IFX), run_006_Irrad_450V.dat, 006, 250000, 96, 64, 360, No, Unknown 

Remarks: Telescope run 636. Bias scan -450V, central position of sensor (DUT parallel to telescope planes at 360 deg due to different fixation compared to other DUTs). Telescope geometry identical to double irrad module.

Backside Biasing!

Telescope chiller turned on (15 deg).

Parameter Value
Alibava run run_006_Irrad_450V.dat
Telescope run 636
Triggers 250000
Alibava trigger eff. ~70%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -450
Latency 132
Table x 96 mm
Table z 64 mm
Tabe rot

360 deg

Temperature at DUT -14.5 +- 1 deg. celsius
Current 38.8 uA
Entry  Thu Apr 13 09:23:20 2017, Axel König, ALiBaVa (IFX), run_005_IFX_CB_10V.dat, 005, 285000, 109, 4.5, 240, No, Unknown 

Remarks: Telescope run 508. Bias scan 10 V.

Parameter Value
Alibava run run_005_IFX_CB_10V.dat
Telescope run 508
Triggers 285000
Alibava trigger eff. ~36.5%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -10
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

Entry  Sat Apr 15 20:12:41 2017, Axel Koenig, ALiBaVa (IFX), run_005_Baby_100V.dat, 005, 250000, 102.5, 22.5, 230, No, Unknown 

Remarks: Telescope run 568. Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_005_Baby_100V.dat
Telescope run 568
Triggers 250000
Alibava trigger eff. ~93%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -100
Latency 132
Table x 102.5 mm
Table z 22.5 mm
Tabe rot

230 deg

Entry  Mon Apr 17 17:46:38 2017, Axel Koenig, ALiBaVa (IFX), run_005_HPK_CB_50V.dat, 005, 150000, 107, 2.6, 240, No, Unknown 

Remarks: Telescope run 609. Bias scan 50 V. Bias scan.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_005_HPK_CB_50V.dat
Telescope run 609
Triggers 150000
Alibava trigger eff. ~62%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -50
Latency 132
Table x 107 mm
Table z 2.6 mm
Tabe rot

240 deg

Entry  Wed Apr 19 22:56:07 2017, Axel Koenig, ALiBaVa (IFX), run_005_Irrad_450V.dat, 005, 2800000, 96, 64, 360, No, Unknown 

Remarks: Telescope run 635. Bias scan -450V, central position of sensor (DUT parallel to telescope planes at 360 deg due to different fixation compared to other DUTs). Telescope geometry identical to double irrad module.

Backside Biasing! Run crashed over night. Triggers valid up to 2032867.

Telescope chiller turned on (15 deg).

Parameter Value
Alibava run run_005_Irrad_450V.dat
Telescope run 635
Triggers 2800000
Alibava trigger eff. ~70%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -450
Latency 132
Table x 96 mm
Table z 64 mm
Tabe rot

360 deg

Temperature at DUT -14.5 +- 1 deg. celsius
Current 38.8 uA
Entry  Sat Apr 15 19:20:34 2017, Axel Koenig, ALiBaVa (IFX), run_004_Baby_100V.dat, 004, 250000, 102.5, 22.5, 250, No, Unknown 

Remarks: Telescope run 565. Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_004_Baby_100V.dat
Telescope run 565
Triggers 250000
Alibava trigger eff. ~93%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -100
Latency 132
Table x 102.5 mm
Table z 22.5 mm
Tabe rot

250 deg

Entry  Tue Apr 18 08:04:39 2017, Axel Koenig, ALiBaVa (IFX), run_004_HPK_CB_200V.dat, 004, 250000, 107, 2.6, 250, No, Unknown 

Remarks: Telescope run 607. Angle scan.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_004_HPK_CB_200V.dat
Telescope run 607
Triggers 250000
Alibava trigger eff. ~62%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -200
Latency 132
Table x 107 mm
Table z 2.6 mm
Tabe rot

250 deg

Entry  Wed Apr 19 22:01:57 2017, Axel Koenig, ALiBaVa (IFX), run_004_Irrad_400V.dat, 004, 250000, 96, 64, 360, No, Unknown 

Remarks: Telescope run 633. Bias scan -400V, central position of sensor (DUT parallel to telescope planes at 360 deg due to different fixation compared to other DUTs). Telescope geometry identical to double irrad module.

Backside Biasing!

Telescope chiller turned on (15 deg).

Parameter Value
Alibava run run_004_Irrad_400V.dat
Telescope run 633
Triggers 250000
Alibava trigger eff. ~70%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -400
Latency 132
Table x 96 mm
Table z 64 mm
Tabe rot

360 deg

Temperature at DUT -14.5 +- 1 deg. celsius
Current 37.5 uA
Entry  Wed Apr 12 22:41:08 2017, Axel König, ALiBaVa (IFX), run_003_IFX_CB_100V.dat, 003, 100000, 109, 8.5, 240, No, Unknown 

Remarks: Beam position on only one chip.

Parameter Value
Alibava run run_003_IFX_CB_100V.dat
Telescope run 500
Triggers 100000
Alibava trigger eff ?
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -100 V
Latency 132
Table x 109 mm
table z 8.5 mm
table rot 240 deg (parallel to tele. planes)
Entry  Sat Apr 15 18:31:14 2017, Axel Koenig, ALiBaVa (IFX), run_003_Baby_100V.dat, 003, 250000, 102.5, 32.5, 240, No, Unknown 

Remarks: Telescope run 564. Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_003_Baby_100V.dat
Telescope run 564
Triggers 250000
Alibava trigger eff. ~93%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -100
Latency 132
Table x 102.5 mm
Table z 32.5 mm
Tabe rot

240 deg (parallel to tele. planes)

Entry  Mon Apr 17 17:02:34 2017, Axel Koenig, ALiBaVa (IFX), run_003_HPK_CB_200V.dat, 003, 250000, 107, 2.6, 230, No, Unknown 

Remarks: Telescope run 606. Angle scan.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_003_HPK_CB_200V.dat
Telescope run 606
Triggers 250000
Alibava trigger eff. ~62%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -200
Latency 132
Table x 107 mm
Table z 2.6 mm
Tabe rot

230 deg

Entry  Tue Apr 18 21:03:42 2017, Axel Koenig, ALiBaVa (IFX), run_003_double_irrad_100V.dat, 003, 3500000, 103, 1, 240, No, Unknown 

Remarks: Telescope run 625. Both chips. Central position. Run crashed over night.

Collected triggers 2710991.

Telescope chiller turned on!

Parameter Value
Alibava run run_003_double_irrad_100V.dat
Telescope run 625
Triggers 3500000
Alibava trigger eff. ~70%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -100
Latency 132
Table x 103 mm
Table z 1 mm
Tabe rot

240 deg

Entry  Wed Apr 19 20:55:36 2017, Axel Koenig, ALiBaVa (IFX), run_003_Irrad_300V.dat, 003, 300000, 96, 64, 360, No, Unknown 

Remarks: Telescope run 632. Bias scan -300V, central position of sensor (DUT parallel to telescope planes at 360 deg due to different fixation compared to other DUTs). Telescope geometry identical to double irrad module.

Backside Biasing!

Telescope chiller turned on (15 deg).

Parameter Value
Alibava run run_003_Irrad_300V.dat
Telescope run 632
Triggers 300000
Alibava trigger eff. ~70%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -300
Latency 132
Table x 96 mm
Table z 64 mm
Tabe rot

360 deg

Temperature at DUT -14.5 +- 1 deg. celsius
Current 35.1 uA
Entry  Wed Apr 12 22:34:18 2017, Axel König, ALiBaVa (IFX), run_002_IFX_CB_100V.dat, 002, 100000, 109, 0.5, 240, No, Unknown 

Remarks: Beam position on only one chip.

Parameter Value
Alibava run run_002_IFX_CB_100V.dat
Telescope run 498
triggers 100000
Alibava trigger eff. ?
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -100 V
Latency 132
Table x 109 mm
Table z 0.5 mm
Table rot 240 deg (parallel to tele. planes)
Entry  Sat Apr 15 17:38:58 2017, Axel Koenig, ALiBaVa (IFX), run_002_Baby_100V.dat, 002, 250000, 102.5, 12.5, 240, No, Unknown 

Remarks: Telescope run 563. Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_002_Baby_100V.dat
Telescope run 563
Triggers 250000
Alibava trigger eff. ~93%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -100
Latency 132
Table x 102.5 mm
Table z 12.5 mm
Tabe rot

240 deg (parallel to tele. planes)

Entry  Mon Apr 17 16:36:56 2017, Axel Koenig, ALiBaVa (IFX), run_002_HPK_CB_200V.dat, 002, 450000, 107, 5.8, 240, No, Unknown 

Remarks: Telescope run 605. Bias scan 200 V. Only one chip.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_002_HPK_CB_200V.dat
Telescope run 605
Triggers 450000
Alibava trigger eff. ~64%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -200
Latency 132
Table x 107 mm
Table z 5.8 mm
Tabe rot

240 deg

Entry  Tue Apr 18 18:48:27 2017, Axel Koenig, ALiBaVa (IFX), run_002_double_irrad_100V.dat, 002, 700000, 94, 1, 240, No, Unknown 

Remarks: Telescope run 623. 1nd chip only. Central position.

Telescope chiller turned on!

Parameter Value
Alibava run run_002_double_irrad_100V.dat
Telescope run 623
Triggers 700000
Alibava trigger eff. ~62%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -100
Latency 132
Table x 94 mm
Table z 1 mm
Tabe rot

240 deg

Entry  Wed Apr 19 19:27:38 2017, Axel Koenig, ALiBaVa (IFX), run_002_Irrad_200V.dat, 002, 400000, 96, 64, 360, No, Unknown 

Remarks: Telescope run 630. Bias scan -200V, central position of sensor (DUT parallel to telescope planes at 360 deg due to different fixation compared to other DUTs). Telescope geometry identical to double irrad module.

Backside Biasing!

Telescope chiller turned on (15 deg).

Parameter Value
Alibava run run_002_Irrad_200V.dat
Telescope run 630
Triggers 400000
Alibava trigger eff. ~65%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -200
Latency 132
Table x 96 mm
Table z 64 mm
Tabe rot

360 deg

Temperature at DUT -14.5 +- 1 deg. celsius
Current 29.7 uA
Entry  Wed Apr 12 21:06:44 2017, Axel König, ALiBaVa (IFX), run_001_IFX_CB_100V.dat, 001, 200000, 109, 4.5, 240, No, Unknown 11.png46.png

First run for alibava modules with Sensor IFX_CB.

Beam energy = 5.6 GeV

Telescope threshold = 6

Telescope data file = 496

Bias voltage = -100V

Telescope synchornisation: yes

Position table (is center of sensor):

x=109mm

y=4.5mm

phi=240 deg

 

Entry  Thu Apr 13 10:03:29 2017, Axel König, ALiBaVa (IFX), run_001_IFX_CB_100V.dat, 001, 200000, 109, 4.5, 240, No, Unknown 

Remarks: Telescope run 496. Bias voltage 100 V. First run at DESY TB.

Parameter Value
Alibava run run_001_IFX_CB_100V.dat
Telescope run 496
Triggers 200000
Alibava trigger eff. ?
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -100
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

Entry  Sat Apr 15 16:52:06 2017, Axel Koenig, ALiBaVa (IFX), run_001_Baby_100V.dat, 001, 250000, 102.5, 22.5, 240, No, Unknown 

Remarks: Telescope run 562. Telescope chiller turned off (could be the root cause of runs crashing).

Position: 102.5, 22.5, 240 (should be the center of the sensor).

Parameter Value
Alibava run run_001_Baby_100V.dat
Telescope run 562
Triggers 250000
Alibava trigger eff. ~93%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -100
Latency 132
Table x 102.5 mm
Table z 22.5 mm
Tabe rot

240 deg (parallel to tele. planes)

Entry  Mon Apr 17 13:30:01 2017, Axel Koenig, ALiBaVa (IFX), run_001_HPK_CB_200V.dat, 001, 450000, 107, 0, 240, No, Unknown 

Remarks: Telescope run 604. Bias scan 200 V. Only one chip.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_001_HPK_CB_200V.dat
Telescope run 604
Triggers 450000
Alibava trigger eff. ~64%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -200
Latency 132
Table x 107 mm
Table z 0 mm
Tabe rot

240 deg

Entry  Tue Apr 18 18:19:33 2017, Axel Koenig, ALiBaVa (IFX), run_001_double_irrad_100V.dat, 001, 350000, 111, 1, 240, No, Unknown 

Remarks: Telescope run 620. 2nd chip only. Central position.

Telescope chiller turned on!

Parameter Value
Alibava run run_001_double_irrad_100V.dat
Telescope run 620
Triggers 350000
Alibava trigger eff. ~53%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -100
Latency 132
Table x 111 mm
Table z 1 mm
Tabe rot

240 deg

Entry  Wed Apr 19 17:31:43 2017, Axel Koenig, ALiBaVa (IFX), run_001_Irrad_100V.dat, 001, 400000, 96, 64, 360, No, Unknown 

Remarks: Telescope run 628. Bias scan -100V, central position of sensor (DUT parallel to telescope planes at 360 deg due to different fixation compared to other DUTs). Telescope geometry identical to double irrad module.

Backside Biasing!

Telescope chiller turned on (15 deg).

Parameter Value
Alibava run run_001_Irrad_100V.dat
Telescope run 628
Triggers 400000
Alibava trigger eff. ~49%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -100
Latency 132
Table x 96 mm
Table z 64 mm
Tabe rot

360 deg

Temperature at DUT -14.5 +- 1 deg. celsius
Current 23.5 uA
Entry  Wed Apr 12 21:55:13 2017, Axel König, common, pulse shape , , , , No, Unknown 21.png

pulse shape looks good for latency 132. This value is now taken by default for all runs unless otherwise stated.

Entry  Wed Apr 12 21:56:48 2017, Axel König, common, Kazu beetle settings, , , , No, Unknown 

Beetle setting for a better defined pulse shape. These parameters were found by kazu, a beetle developer at lhcb.

CompCtrl=9
Ibuf=10
Icurrbuf=119
Ipipe=13
Ipre=75
Isf=29
Isha=10
Itp=0
Ivoltbuf=29
Latency=133
ROCtrl=26
RclkDiv=0
Vd=140
Vdcl=107
Vfp=31
Vfs=102

Entry  Fri Apr 14 14:21:35 2017, Axel Koenig, common, Run summary IFX_CB , , , , No, Unknown 
Alibava run Telescope run File existence check Run type Triggers Remarks
001 496 yes position check 200000 none
002 498 yes beam focused on one strip segment only 100000 none
003 500 yes beam focused on one strip segment only 100000 none
005 508 yes Bias scan. Voltage = 10 V 285000 none
006 510 yes Bias scan. Voltage = 20 V 185000 none
007 513 yes Bias scan. Voltage = 30 V 170000 none
008 515 yes Bias scan. Voltage = 40 V 175000 none
009 517 yes Bias scan. Voltage = 50 V 175000 none
010 521 yes Bias scan. Voltage = 60 V 175000 none
011 522 yes Bias scan. Voltage = 65 V 175000 none
012 523 yes Bias scan. Voltage = 70 V 175000 none
013 524 yes Bias scan. Voltage = 75 V 175000 none
014 525 yes Bias scan. Voltage = 80 V 175000 none
015 526 yes Bias scan. Voltage = 90 V 175000 none
016 527-531 yes Bias scan. Voltage = 100 V 750000 Telescope files 527 to 531, run crashed!
017 532 yes Bias scan. Voltage = 110 V 175000 none
018 533 yes Bias scan. Voltage = 120 V 175000 none
020 535 yes overnight run at 100 V 190000 run crashed data may be corrupted
021 540 yes CRAP! ? CRAP!
022 545 yes beam focused on one strip segment only 500000 none
023 546 yes beam focused on one strip segment only 385000 run crashed data may be corrupted
024 547 yes Bias scan. Voltage = 140 V 175000 none
025 548 yes Bias scan. Voltage = 150 V 175000 none
026 549 yes Rotation -10 deg 350000 none
027 550 yes Rotation +10 deg 350000 none
028 551 yes overnight run at 100 V 3000000 run crashed data may be corrupted
029 557 yes Bias scan. Voltage = 130 V 175000 none
030 558 yes long run 1000000 telescope chiller turned off (keept at this state unless otherwisde stated

Central beam position: x = 109 mm, y = 4.5 mm, rot = 240 deg

Current was stable for all runs i.e. 140 +- 10 nA at 100 V

Trigger efficiency ~ 60% for all runs (except bias scan).

Temperature in area 20+-1 deg.

Entry  Sat Apr 15 16:55:19 2017, Axel Koenig, common, Chiller turned off!, , , , No, Unknown 

The chiller has been turned off beginning of run_30_IFX_CB_100V.dat.

Set temperature was 15 C. Now it should settle at 20 +- 1 C (room temperature).

   
Entry  Mon Apr 17 10:14:15 2017, Axel Koenig, common, Run summary Baby, , , , No, Unknown 
Alibava run Telescope run File existence check Run type Triggers Remarks
001 562 yes first run 250000 none
002 563 yes central position shifted by -1 cm vertically 250000 none
003 564 yes central position shifted by + 1cm 250000 none
004 565 yes central position angle +10 deg 250000 none
005 568 yes central position angle -10 deg 250000 none
006 570 yes unknown (maybe long run over night that crashed) ? data may be corrupted
007 579 yes bias scan -10V 450000 run crashed data only valid up to 274915
008 583 yes bias scan -20V 175000 none
009 586 yes bias scan -30V 135000 none
010 587 yes bias scan -40V 135000 none
011 588 yes bias scan -50V 135000 none
012 589 yes bias scan -60V 135000 none
013 590 yes bias scan -65V 112000 none
014 591 yes bias scan -70V 112000 none
015 592 yes long run over night 2500000 none
016 593 yes bias scan -75V 112000 none
017 594 yes bias scan -80V 112000 none
018 595 yes bias scan -90V 112000 none
019 596 yes bias scan -110V 112000 none
020 597 yes bias scan -120V 112000 none
021 598 yes bias scan -130V 112000 none
022 599 yes bias scan -140V 112000 none
023 600 yes bias scan -150V 112000 none

Central beam position: x = 102 mm, y = 22.5 mm, rot = 240 deg

Current was stable for all runs i.e. 500 +- 10 nA at 100 V

Trigger efficiency ~ 90% for all runs (except bias scan).

Temperature in area 20+-1 deg.

Telescope chiller off for all runs.

Entry  Mon Apr 17 17:25:24 2017, Axel Koenig, common, Telescope geometry ALiBaVa, , , , No, Unknown IMG_1522.JPGbeam.pdfbeam1.pdfbeam2.pdf

The plane distances are measured at the center of the planes.

The distance of the DUT to the neighbouring planes is measured from the position of the sensor to the center of the planes.

   
Entry  Tue Apr 18 17:55:19 2017, Axel Koenig, common, Run summary HPK_CB, , , , No, Unknown 
Alibava run Telescope run File existence check Run type Triggers Remarks
001 562 yes first run 250000 none
002 563 yes central position shifted by -1 cm vertically 250000 none
003 564 yes central position shifted by + 1cm 250000 none
004 565 yes central position angle +10 deg 250000 none
005 568 yes central position angle -10 deg 250000 none
006 570 yes unknown (maybe long run over night that crashed) ? data may be corrupted
007 579 yes bias scan -10V 450000 run crashed data only valid up to 274915
008 583 yes bias scan -20V 175000 none
009 586 yes bias scan -30V 135000 none
010 587 yes bias scan -40V 135000 none
011 588 yes bias scan -50V 135000 none
012 589 yes bias scan -60V 135000 none
013 590 yes bias scan -65V 112000 none
014 591 yes bias scan -70V 112000 none
015 592 yes long run over night 2500000 none
016 593 yes bias scan -75V 112000 none
017 594 yes bias scan -80V 112000 none
018 595 yes bias scan -90V 112000 none
019 596 yes bias scan -110V 112000 none
020 597 yes bias scan -120V 112000 none
021 598 yes bias scan -130V 112000 none
022 599 yes bias scan -140V 112000 none
023 600 yes bias scan -150V 112000 none

Central beam position: x = 102 mm, y = 22.5 mm, rot = 240 deg

Current was stable of all runs i.e. 500 +- 10 nA at 100 V

Trigger efficiency ~ 90% for all runs (except bias scan).

Temperature in area 20+-1 deg.

Telescope chiller off for all runs.

Entry  Tue Apr 18 17:56:37 2017, Axel Koenig, common, Run summary HPK_CB, , , , No, Unknown 
Alibava run Telescope run File existence check Run type Triggers Remarks
001 604 yes first run, only one chip 450000 none
002 605 yes only one chip 450000 none
003 606 yes angle scan -10 deg 250000 none
004 607 yes angle scan +10 deg 250000 none
005 609 yes Bias scan -50V 150000 none
006 610 yes Bias scan -100V 150000 none
007 611 yes Bias scan -150V 300000 none
008 613 yes Bias scan -250V 150000 none
009 614 yes Long run over night 2400000 run crashed over night
010 616 yes repeated long run ~10000 run crashed

Central beam position: x = 107 mm, y = 2.6 mm, rot = 240 deg

Current was stable for all runs i.e. 80 +- 5 nA at 200 V

Trigger efficiency ~ 65% for all runs (except bias scan).

Temperature in area 20+-1 deg.

Telescope chiller off for all runs.