54
|
Sat Oct 8 23:27:44 2011 |
Karl-Heinz Hoffmann | HPK | baby_std | run020 | 382000 | 08.10.2011 23:24:16 | 09.10.2011 09:24:20 | Good |
no cooling
beam on; xy table position x=327,5 y=65
HV=200V Istart_PY=8.2uA; Istart_N=0.45uA
std. I2C settings |
53
|
Sat Oct 8 20:33:58 2011 |
Karl-Heinz Hoffmann | HPK | baby_std | run019 | 150000 | 08.10.2011 20:22:45 | 08.10.2011 23:19:23 | Good |
no cooling
beam on; xy table position x=333 y=80
HV=200V Istart_PY=8.2uA; Istart_N=0.45uA
std. I2C settings |
52
|
Sat Oct 8 16:45:14 2011 |
Thomas Bergauer | HPK | baby_std | run018 | 150000 | 08.10.2011 16:44:13 | 08.10.2011 19:33:33 | Good |
no cooling
beam on; xy table position x=322 y=80
HV=200V Istart_PY=8.2uA; Istart_N=0.45uA
std. I2C settings, now with correct timing |
55
|
Sun Oct 9 10:05:34 2011 |
Markus Friedl | HPK | baby_std | cal010 | 83800 | 08.10.2011 13:36:25 | 08.10.2011 13:54:07 | Good |
HV=+-40V,
no cooling
beam off; xy table position x= y=
HV= Istart=uA;
std. I2C settings
|
48
|
Sat Oct 8 00:14:37 2011 |
Christian Irmler | BELLE | Origami stack | run015 | 180111 | 08.10.2011 00:06:41 | 08.10.2011 08:24:38 | Good |
Origami6
HV=+-40V,
no cooling
beam on; xy table position x=306 y=74.5
Istart=57µA;
std. I2C settings
centered
currents:
8.10.2011 00:07:43 event=991 rate=0events/s temp/curr=33.0/-15.4 33.0/-17.3 33.0/-17.7 31.9/011.5 31.9/013.2 32.0/040.20
...
08.10.2011 06:32:34 event=179710 rate=0events/s temp/curr=33.1/-22.1 33.0/-25.6 33.1/-22.6 31.9/018.2 31.8/021.7 31.9/048.5
08.10.2011 06:33:28 event=180105 ***WARNING*** Error detected in APV 0, subevent 2. Event discarded, reset sent.
Permanent APV errors at several (all p-side, some n-side) APV chips after 180105 events, which could not be recovered by software resets. After restart of run, all APVs where alive again, hence hardware reset and I2C reconfiguration after 5 or 10 consecutive APV errors should be added to TUXDAQ software.
After APV errors occured, the cummulative bias current decrised from 78µA to 54µA. This effect can occures whenever the digital part of the APV is switched between idle and full operation. It seems that this is a thermal effect, caused by the slightly higher power consumption of the digital part of the chips.
|
47
|
Fri Oct 7 19:35:14 2011 |
Christian Irmler | BELLE | Origami stack cooled | run014 | 90k | 07.10.2011 19:49:26 | 07.10.2011 23:26:52 | Good |
Origami6
HV=+-40V,
CO2 cooling: P_TT_4: 14.9 bar, mf=0.9 g/s
beam on; xy table position x=306 y=74.5
centered
I=17.8µA; old I2C settings; corrected trigger delay (htr); adjusted front-end voltages
CO2 pressure reduced to 14.9 bar --> temp is about 4K colder
Temp. at pipe inlet (K7) = -18°C
Temp. at pipe outlet (K8) = -21.8°C
currents:
07.10.2011 23:26:21 event=89786 rate=0events/s temp/curr=33.0/-05.3 33.0/-05.8 33.0/-06.6 31.9/001.4 32.0/001.9 31.9/017.7
|
46
|
Fri Oct 7 18:40:07 2011 |
Christian Irmler | BELLE | Origami stack cooled | cal009 | 83800 | 07.10.2011 18:35:04 | 07.10.2011 19:31:51 | Good |
Origami6
HV=+-40V,
CO2 cooling: P_TT_4: 14.9 bar, mf=0.9 g/s
beam on; xy table position x=306 y=76.5
I=18.6µA; old I2C settings; corrected trigger delay (htr); adjusted front-end voltages
CO2 pressure reduced to 14.9 bar --> temp is about 4K colder
Temp. at pipe inlet (K7) = -18°C
Temp. at pipe outlet (K8) = -21.8°C
used multi6 config instead of single hit config (cern11_origami_cooled2_multi6.cfg)
currents:
07.10.2011 18:36:07 event=1581 rate=31events/s temp/curr=33.9/-05.3 33.9/-05.6 33.9/-06.6 32.8/001.4 32.8/001.9 32.8/018.1 |
44
|
Fri Oct 7 10:13:12 2011 |
Christian Irmler | BELLE | Origami stack cooled | run013 | 185k | 07.10.2011 10:02:20 | 07.10.2011 17:17:32 | Good |
HV=+-40V,
CO2 cooling: P_TT_4: 18.2 bar, mf=1 g/s
beam on; xy table position x=321 y=99
I=18.6µA; old I2C settings; corrected trigger delay (htr); adjusted front-end voltage
The beam is directly focused onto APV chip #2 of n-side, so that it hits the area underneath chip and cooling pipe.
On p-side this location is attached to chip #4, which currently can not be read out, hence we only get data for the n-side.
currents:
07.10.2011 10:03:22 event=979 rate=0events/s temp/curr=33.2/-05.6 33.3/-06.0 33.2/-06.7 32.2/001.6 32.2/002.0 32.2/018.7
07.10.2011 17:17:18 event=184992 rate=0events/s temp/curr=33.8/-05.6 33.8/-06.0 33.8/-06.9 32.7/001.7 32.7/002.0 32.7/018.7
|
43
|
Fri Oct 7 00:20:15 2011 |
Christian Irmler | BELLE | Origami stack cooled | run012 | 250k | 07.10.2011 00:15:54 | 07.10.2011 09:13:55 | Good |
HV=+-40V,
CO2 cooling: P_TT_4: 18.2 bar, mf=1 g/s
beam on; xy table position x=352 y=89
I=19.1µA; old I2C settings; corrected trigger delay (htr); adjusted front-end voltages
beam focused on APV #3 of p- and n-side, respectively
currents:
07.10.2011 00:16:57 event=987 rate=0events/s temp/curr=34.3/-05.6 34.3/-06.0 34.3/-06.9 33.3/001.7 33.5/002.2 33.3/019.2
07.10.2011 09:13:42 event=249992 rate=0events/s temp/curr=33.6/-05.5 33.5/-06.0 33.6/-06.7 32.5/001.7 32.5/002.0 32.5/018.8
|
42
|
Thu Oct 6 17:36:47 2011 |
Christian Irmler | BELLE | Origami stack cooled | run011 | 180000 | 06.10.2011 17:06:38 | 07.10.2011 00:04:06 | Good |
HV=+-40V,
CO2 cooling: P_TT_4: 18.2 bar, mf=1 g/s
beam on; xy table position x=260.8 y=60.25
I=19.4µA; old I2C settings; corrected trigger delay (htr); adjusted front-end voltages
beam focused on APV #0 of p- and n-side, respectively
currents:
06.10.2011 17:07:41 event=639 rate=3events/s temp/curr=35.2/-05.6 35.2/-06.0 35.2/-06.9 34.5/001.7 34.5/002.2 34.5/019.6 |
94
|
Wed Mar 7 16:58:36 2012 |
Christian Irmler | BELLE | Origami Module #4 | cal113 | 100k | 07.03.2012 16:40:04 | 07.03.2012 16:57:08 | Unknown |
Sr90, small (new) scintillator, new trigger unit
Origami Mod. #4
HV=+-40V, Ibias=1,6µA
CO2 cooling on, Tin=-19,6°C, Tout=-21.3°C
I²C settings for cooled APV chips
e:\cvi-projects\apvdaq_fadc\origami6_V2_single_cooled.cfg
SMU: +HV=Inv.Nr. 545-77 -HV=Inv.Nr. 540-702
HVret configuration:
0 ... switch open / not connected
1 ... switch closed / connected
SMU |
Dock Box |
J5 |
J6 |
J7 |
0 |
0 (now isolated) |
1 |
1 |
0 |
J7... HV ret @ hybrid
same configuration as run112 but now the step motors of the xz table are powered. |
93
|
Wed Mar 7 16:03:30 2012 |
Christian Irmler | BELLE | Origami Module #4 | run113 | 100k | 07.03.2012 16:07:34 | 07.03.2012 16:30:48 | Unknown |
Sr90, small (new) scintillator, new trigger unit
Origami Mod. #4
HV=+-40V, Ibias=1,6µA
CO2 cooling on, Tin=-19,6°C, Tout=-21.3°C
I²C settings for cooled APV chips
e:\cvi-projects\apvdaq_fadc\origami6_V2_multi6_cooled.cfg
SMU: +HV=Inv.Nr. 545-77 -HV=Inv.Nr. 540-702
HVret configuration:
0 ... switch open / not connected
1 ... switch closed / connected
SMU |
Dock Box |
J5 |
J6 |
J7 |
0 |
0 (now isolated) |
1 |
1 |
0 |
J7... HV ret @ hybrid
same configuration as run112 but now the step motors of the xz table are powered. |
41
|
Thu Oct 6 15:57:51 2011 |
Christian Irmler | BELLE | Origami stack cooled | cal008 | 83800 | 06.10.2011 15:53:20 | 06.10.2011 16:25:29 | Good |
HV=+-40V,
CO2 cooling: P_TT_4: 18.2 bar, mf=1 g/s
beam on; xy table position x=306 y=76.5
I=19.5µA; old I2C settings; corrected trigger delay (htr); adjusted front-end voltages
Temp. at pipe inlet (K7) = -14°C
Temp. at pipe outlet (K8) = -17°C
Before this run, the front-end voltages (+2.5 and +1.25) were readjusted so that the nominal values appear on MAMBO. (Previously, both were about 50mV too low.) Not sure if this makes a big difference, let's see.
currents:
06.10.2011 15:54:24 event=2901 rate=36events/s temp/curr=35.5/-05.6 35.5/-06.0 35.4/-06.9 34.9/001.9 35.0/002.2 35.0/019.6
06.10.2011 16:24:25 event=81350 rate=50events/s temp/curr=34.9/-05.6 34.9/-06.1 34.9/-06.9 34.3/001.7 34.3/002.2 34.3/019.6
|
39
|
Wed Oct 5 20:49:11 2011 |
Markus Friedl | BELLE | Origami stack cooled | run009 | 330k | 05.10.2011 20:46:30 | 06.10.2011 09:12:24 | Good |
HV=+-40V,
CO2 cooling: P_TT_4: 18.2 bar, mf=0.9 g/s
beam on; xy table position x=306 y=76.5
I=20.5µA; new I2C settings; corrected trigger delay (htr)
New run with cold Origami and new I2C settings (shorter peaking time) and hopefully correct timing - and centered position in both x and y.
corresponding cal file: cal007
currents:
06.10.2011 09:12:12 event=329714 rate=0events/s temp/curr=35.8/-05.5 36.0/-06.0 35.8/-06.9 33.6/001.7 33.5/002.2 33.6/019.3 |
40
|
Thu Oct 6 09:22:34 2011 |
Christian Irmler | BELLE | Origami stack cooled | run010 | 330k | 05.10.2011 20:46:30 | 06.10.2011 09:12:24 | Good |
HV=+-40V,
CO2 cooling: P_TT_4: 18.2 bar, mf=0.9 g/s
beam on; xy table position x=306 y=76.5
I=20.5µA; corrected trigger delay (htr)
New run with cold Origami and same I2C setting as used 2010, with longer shaping time for p-side
centered position in both x and y.
corresponding cal file: cal006
currents:
06.10.2011 09:20:29 event=988 rate=0events/s temp/curr=35.8/-05.5 36.0/-06.0 35.8/-06.7 33.5/001.7 33.6/002.0 33.6/019.3
|
38
|
Wed Oct 5 20:28:00 2011 |
Markus Friedl | BELLE | Origami stack cooled | run008 | 10582 | 05.10.2011 20:25:10 | 05.10.2011 20:45:10 | Good |
HV=+-40V,
CO2 cooling: P_TT_4: 18.2 bar, mf=0.9 g/s
beam on
I=20.5µA; new I2C settings; corrected trigger delay (htr)
New run with cold Origami and new I2C settings (shorter peaking time) and hopefully correct timing.
corresponding cal file: cal007 |
37
|
Wed Oct 5 19:57:15 2011 |
Christian Irmler | BELLE | Origami stack cooled | run007 | 17845 | 05.10.2011 18:27:51 | 05.10.2011 19:31:43 | Crap |
Run stopped after 17845 events, because of bad trigger timing
Origami6
HV=+-40V
beam on
CO2 cooling:
preassure @ P_TT_4 = 18.77 bar
mass flow = 0.9 g/s
I=20µA
longterm files (same files used for cal005 and run006):
It-pos-0510111531.txt
It-neg-0510111531.txt
currents and temperatures:
05.10.2011 19:30:51 event=17469 rate=0events/s temp/curr=35.6/-05.6 35.6/-06.1 35.6/-07.1 35.1/001.9 35.1/002.4 35.1/020.9
|
34
|
Wed Oct 5 16:43:00 2011 |
Christian Irmler | BELLE | Origami stack cooled | cal006 | 83800 | 05.10.2011 16:40:18 | 05.10.2011 17:12:27 | Good |
Origami6
HV=+-40V
beam on
CO2 cooling:
preassure @ P_TT_4 = 18.2 bar
mass flow = 0.9 g/s
I=21.8µA
longterm files (same files used for cal005 and run006):
It-pos-0510111531.txt
It-neg-0510111531.txt
currents and temperatures:
05.10.2011 16:41:21 event=2850 rate=50events/s temp/curr=35.4/-05.8 35.3/-06.3 35.4/-07.2 34.9/002.0 34.9/002.5 34.9/021.7
05.10.2011 17:11:20 event=81237 rate=37events/s temp/curr=35.1/-05.8 35.1/-06.1 35.0/-07.1 34.5/002.0 34.5/002.5 34.5/021.4
|
33
|
Wed Oct 5 13:00:41 2011 |
Christian Irmler | BELLE | Origami stack cooled | run006 | 24692 | 05.10.2011 12:55:53 | 05.10.2011 14:05:49 | Crap |
Run stopped after 24692 events, because all n-side APV failed due to a shortcut of a capacitor (C223) on n-side rebo
Origami6
HV=+-40V
beam on
CO2 cooling:
preassure @ P_TT_4 = 18.77 bar
mass flow = 0.9 g/s
I=22.25µA
longterm files (same files used for cal005 and run006):
It-pos-0510111122.txt
It-neg-0510111122.txt
currents and temperatures:
05.10.2011 12:56:54 event=985 rate=0events/s temp/curr=35.0/-05.5 35.0/-06.1 34.9/-06.9 34.2/001.7 34.3/002.4 34.3/022.3
|
32
|
Wed Oct 5 12:09:00 2011 |
Christian Irmler | BELLE | Origami stack cooled wrong I2C settings | cal005 | 83800 | 05.10.2011 12:04:50 | 05.10.2011 12:37:11 | Crap |
This calibration file can not be used, due to wrong I2C settings!
Origami6
HV=+-40V
beam on
CO2 cooling:
preassure @ P_TT_4 = 18.77 bar
mass flow = 0.9 g/s
I=22.45µA
longterm files (same files used for cal005 and run006):
It-pos-0510111122.txt
It-neg-0510111122.txt
currents and temperatures:
05.10.2011 12:05:53 event=2883 rate=37events/s temp/curr=35.0/-05.6 35.0/-06.3 35.0/-06.9 34.2/001.9 34.1/002.5 34.1/022.6
05.10.2011 12:36:53 event=83350 rate=48events/s temp/curr=34.4/-05.5 34.5/-06.1 34.5/-06.9 33.9/001.7 33.9/002.4 33.9/022.5
|