28
|
Tue Oct 4 15:16:06 2011 |
Manfred Valentan | Micron | Micron-Baby-Stack p-spray, run bottom left | run003 | 200k | 04.10.2011 11:29:19 | 04.10.2011 13:57:54 | Good | hit modules in bottom left corner -> Zones wide and half-wide for COMB_IRRAD, Zones narrow and half-narrow for BOMM_IRRAD
Position of HEPHY x-z-table:
x = 336
z = 89 |
29
|
Tue Oct 4 15:16:45 2011 |
Manfred Valentan | Micron | Micron-Baby-Stack p-spray, run bottom right NOT USABLE | run004 | 200k | 4.10.2011 15:28:48 | 04.10.2011 19:36:50 | Crap | HAS TO BE REPEATED!!! Before this run there were changes in the trigger line (vetoing the trigger when no spill), and thus the timing is incorrect.
hit modules in bottom right corner -> Zones narrow and half-narrow for COMB_IRRAD, Zones wide and half-wide for COMM_IRRAD
Position of HEPHY x-z-table:
x = 322
z = 89 |
44
|
Fri Oct 7 10:13:12 2011 |
Christian Irmler | BELLE | Origami stack cooled | run013 | 185k | 07.10.2011 10:02:20 | 07.10.2011 17:17:32 | Good | HV=+-40V,
CO2 cooling: P_TT_4: 18.2 bar, mf=1 g/s
beam on; xy table position x=321 y=99
I=18.6µA; old I2C settings; corrected trigger delay (htr); adjusted front-end voltage
The beam is directly focused onto APV chip #2 of n-side, so that it hits the area underneath chip and cooling pipe.
On p-side this location is attached to chip #4, which currently can not be read out, hence we only get data for the n-side.
currents:
07.10.2011 10:03:22 event=979 rate=0events/s temp/curr=33.2/-05.6 33.3/-06.0 33.2/-06.7 32.2/001.6 32.2/002.0 32.2/018.7
07.10.2011 17:17:18 event=184992 rate=0events/s temp/curr=33.8/-05.6 33.8/-06.0 33.8/-06.9 32.7/001.7 32.7/002.0 32.7/018.7
|
48
|
Sat Oct 8 00:14:37 2011 |
Christian Irmler | BELLE | Origami stack | run015 | 180111 | 08.10.2011 00:06:41 | 08.10.2011 08:24:38 | Good | Origami6
HV=+-40V,
no cooling
beam on; xy table position x=306 y=74.5
Istart=57µA;
std. I2C settings
centered
currents:
8.10.2011 00:07:43 event=991 rate=0events/s temp/curr=33.0/-15.4 33.0/-17.3 33.0/-17.7 31.9/011.5 31.9/013.2 32.0/040.20
...
08.10.2011 06:32:34 event=179710 rate=0events/s temp/curr=33.1/-22.1 33.0/-25.6 33.1/-22.6 31.9/018.2 31.8/021.7 31.9/048.5
08.10.2011 06:33:28 event=180105 ***WARNING*** Error detected in APV 0, subevent 2. Event discarded, reset sent.
Permanent APV errors at several (all p-side, some n-side) APV chips after 180105 events, which could not be recovered by software resets. After restart of run, all APVs where alive again, hence hardware reset and I2C reconfiguration after 5 or 10 consecutive APV errors should be added to TUXDAQ software.
After APV errors occured, the cummulative bias current decrised from 78µA to 54µA. This effect can occures whenever the digital part of the APV is switched between idle and full operation. It seems that this is a thermal effect, caused by the slightly higher power consumption of the digital part of the chips.
|
42
|
Thu Oct 6 17:36:47 2011 |
Christian Irmler | BELLE | Origami stack cooled | run011 | 180000 | 06.10.2011 17:06:38 | 07.10.2011 00:04:06 | Good | HV=+-40V,
CO2 cooling: P_TT_4: 18.2 bar, mf=1 g/s
beam on; xy table position x=260.8 y=60.25
I=19.4µA; old I2C settings; corrected trigger delay (htr); adjusted front-end voltages
beam focused on APV #0 of p- and n-side, respectively
currents:
06.10.2011 17:07:41 event=639 rate=3events/s temp/curr=35.2/-05.6 35.2/-06.0 35.2/-06.9 34.5/001.7 34.5/002.2 34.5/019.6 |
37
|
Wed Oct 5 19:57:15 2011 |
Christian Irmler | BELLE | Origami stack cooled | run007 | 17845 | 05.10.2011 18:27:51 | 05.10.2011 19:31:43 | Crap | Run stopped after 17845 events, because of bad trigger timing
Origami6
HV=+-40V
beam on
CO2 cooling:
preassure @ P_TT_4 = 18.77 bar
mass flow = 0.9 g/s
I=20µA
longterm files (same files used for cal005 and run006):
It-pos-0510111531.txt
It-neg-0510111531.txt
currents and temperatures:
05.10.2011 19:30:51 event=17469 rate=0events/s temp/curr=35.6/-05.6 35.6/-06.1 35.6/-07.1 35.1/001.9 35.1/002.4 35.1/020.9
|
75
|
Wed Feb 22 12:15:11 2012 |
Christian Irmler | BELLE | Origami Module #4 | run102 | 170k | 22.02.2012 12:31:08 | 22.02.2012 13:03:27 | Good | same as run101
Sr90, large scintillator, HVret @ Origami open
Origami Mod. #4
HV=+-40V, Ibias=1,8µA
CO2 cooling on, Tin=-20.2°C, Tout=-21.3°C
I²C settings for cooled APV chips --> same as used in run010
e:\cvi-projects\apvdaq_fadc\origami6_V2_multi6_cooled.cfg
Lab measurement with new dark box but old, large scintillator.
CO2 empty after 24497 events --> run stopped --> no corresponding calibration available, but for rought analysis cal100 can be used for this run, too |
52
|
Sat Oct 8 16:45:14 2011 |
Thomas Bergauer | HPK | baby_std | run018 | 150000 | 08.10.2011 16:44:13 | 08.10.2011 19:33:33 | Good | no cooling
beam on; xy table position x=322 y=80
HV=200V Istart_PY=8.2uA; Istart_N=0.45uA
std. I2C settings, now with correct timing |
53
|
Sat Oct 8 20:33:58 2011 |
Karl-Heinz Hoffmann | HPK | baby_std | run019 | 150000 | 08.10.2011 20:22:45 | 08.10.2011 23:19:23 | Good | no cooling
beam on; xy table position x=333 y=80
HV=200V Istart_PY=8.2uA; Istart_N=0.45uA
std. I2C settings |
57
|
Sun Oct 9 11:00:33 2011 |
Karl-Heinz Hoffmann | HPK | baby_add | run021 | 150000 | 09.10.2011 10:56:47 | 09.10.2011 12:47:23 | Good | no cooling
beam on; xy table position x=330 y=74
HV=200V Istart_PY=1,9uA; Istart_N=0.27uA
std. I2C settings |
63
|
Mon Oct 10 12:15:52 2011 |
Manfred Valentan | CNM | CNM stack | run024 | 110000 | 10.10.2011 12:03:58 | 10.10.2011 14:05:56 | Good | no cooling
beam on; xy table position x=320.5 y=79.5
HV_CNM=20V, HV_BSTD1-N=200V
Istart_CNM= 28 / 1 uA; Istart_BSTD= 0.12 uA |
66
|
Mon Oct 10 14:11:53 2011 |
Christian Irmler | CNM | CNM stack | run025 | 110000 | 10.10.2011 14:09:00 | 10.10.2011 16:11:57 | Good | no cooling
beam on; xy table position x=335.5 y=79.5
HV_CNM=20V, HV_BSTD1-N=200V
Istart_CNM= 31 / 1.4 uA; Istart_BSTD= 0.2 uA |
38
|
Wed Oct 5 20:28:00 2011 |
Markus Friedl | BELLE | Origami stack cooled | run008 | 10582 | 05.10.2011 20:25:10 | 05.10.2011 20:45:10 | Good | HV=+-40V,
CO2 cooling: P_TT_4: 18.2 bar, mf=0.9 g/s
beam on
I=20.5µA; new I2C settings; corrected trigger delay (htr)
New run with cold Origami and new I2C settings (shorter peaking time) and hopefully correct timing.
corresponding cal file: cal007 |
77
|
Wed Feb 22 13:37:11 2012 |
Christian Irmler | BELLE | Origami Module #4 | run103 | 100k | 22.02.2012 13:35:08 | 22.02.2012 13:54:08 | Crap | Sr90, large scintillator,
HVret @ Origami closed
HVret @ SMU connected
Origami Mod. #4
HV=+-40V, Ibias=1,6µA
CO2 cooling on, Tin=-20.5°C, Tout=-21.5°C
I²C settings for cooled APV chips --> same as used in run010
e:\cvi-projects\apvdaq_fadc\origami6_V2_multi6_cooled.cfg
Lab measurement with new dark box but old, large scintillator
Wires between HV connectors ground and HVret conntact at the mambo were open --> no HVret conntection between module and SMUs!
Moreover the SMU Inv. Nr. 545-078 has either an unstable output or problems with the current measurement. |
78
|
Wed Feb 22 16:21:23 2012 |
Christian Irmler | BELLE | Origami Module #4 | run104 | 100k | 22.02.2012 16:22:34 | 22.02.2012 16:41:26 | Good | Sr90, large scintillator,
Origami Mod. #4
HV=+-40V, Ibias=1,6µA
CO2 cooling on, Tin=-20.5°C, Tout=-21.5°C
I²C settings for cooled APV chips --> same as used in run010
e:\cvi-projects\apvdaq_fadc\origami6_V2_multi6_cooled.cfg
Lab measurement with new dark box but old, large scintillator
SMU: +HV=Inv.Nr. 545-77 -HV=Inv.Nr. 540-702
HVret configuration:
0 ... switch open / not connected
1 ... switch closed / connected
SMU |
Dock Box |
J5 |
J6 |
J7 |
0 |
1 |
1 |
0 |
1 |
J7... HV ret @ hybrid
APV |
0 |
1 |
2 |
3 |
4 |
5 |
6 |
7 |
avg noise |
2.564 |
2.124 |
2.094 |
2.200 |
1.696 |
1.666 |
1.870 |
1.888 |
CM noise |
1.7 |
0.9 |
0.9 |
1.3 |
4.0 |
4.1 |
4.1 |
4.2 |
|
81
|
Wed Feb 22 17:38:29 2012 |
Christian Irmler | BELLE | Origami Module #4 | run105 | 100k | 22.02.2012 17:39:10 | 22.02.2012 17:58:01 | Good | Sr90, large scintillator,
Origami Mod. #4
HV=+-40V, Ibias=1,6µA
CO2 cooling on, Tin=-20.5°C, Tout=-21.5°C
I²C settings for cooled APV chips --> same as used in run010
e:\cvi-projects\apvdaq_fadc\origami6_V2_multi6_cooled.cfg
Lab measurement with new dark box but old, large scintillator
SMU: +HV=Inv.Nr. 545-77 -HV=Inv.Nr. 540-702
HVret configuration:
0 ... switch open / not connected
1 ... switch closed / connected
SMU |
Dock Box |
J5 |
J6 |
J7 |
0 |
1 |
1 |
0 |
0 |
J7... HV ret @ hybrid
Same as run104, but now J7 open
Again spikes in bias current!
|
84
|
Mon Mar 5 17:34:43 2012 |
Christian Irmler | BELLE | Origami Module #4 | run108 | 100k | | 05.03.2012 16:24:12 | Unknown | Sr90, large scintillator,
Origami Mod. #4
HV=+-40V, Ibias=1,9µA
CO2 cooling on, Tin=-20.5°C, Tout=-21.5°C
I²C settings for cooled APV chips
e:\cvi-projects\apvdaq_fadc\origami6_V2_multi6_cooled.cfg
Lab measurement with new dark box but old, large scintillator
SMU: +HV=Inv.Nr. 545-77 -HV=Inv.Nr. 540-702
HVret configuration:
0 ... switch open / not connected
1 ... switch closed / connected
SMU |
Dock Box |
J5 |
J6 |
J7 |
0 |
0 (now isolated) |
1 |
0 |
1 |
J7... HV ret @ hybrid |
86
|
Mon Mar 5 17:36:46 2012 |
Christian Irmler | BELLE | Origami Module #4 | run109 | 100k | 05.03.2012 17:36:19 | 05.03.2012 17:55:24 | Unknown | Sr90, large scintillator,
Origami Mod. #4
HV=+-40V, Ibias=1,9µA
CO2 cooling on, Tin=-20.5°C, Tout=-21.5°C
I²C settings for cooled APV chips
e:\cvi-projects\apvdaq_fadc\origami6_V2_multi6_cooled.cfg
Lab measurement with new dark box but old, large scintillator
SMU: +HV=Inv.Nr. 545-77 -HV=Inv.Nr. 540-702
HVret configuration:
0 ... switch open / not connected
1 ... switch closed / connected
SMU |
Dock Box |
J5 |
J6 |
J7 |
0 |
0 (now isolated) |
1 |
1 |
1 |
J7... HV ret @ hybrid |
88
|
Mon Mar 5 18:41:23 2012 |
Christian Irmler | BELLE | Origami Module #4 | run110 | 100k | 05.03.2012 18:45:21 | 05.03.2012 19:04:30 | Unknown | Sr90, large scintillator,
Origami Mod. #4
HV=+-40V, Ibias=1,9µA
CO2 cooling on, Tin=-20.5°C, Tout=-21.5°C
I²C settings for cooled APV chips
e:\cvi-projects\apvdaq_fadc\origami6_V2_multi6_cooled.cfg
Lab measurement with new dark box but old, large scintillator
SMU: +HV=Inv.Nr. 545-77 -HV=Inv.Nr. 540-702
HVret configuration:
0 ... switch open / not connected
1 ... switch closed / connected
SMU |
Dock Box |
J5 |
J6 |
J7 |
0 |
0 (now isolated) |
1 |
1 |
0 |
J7... HV ret @ hybrid |
91
|
Wed Mar 7 15:12:42 2012 |
Christian Irmler | BELLE | Origami Module #4 | run112 | 100k | 07.03.2012 15:13:18 | 07.03.2012 14:31:07 | Unknown | Sr90, small (new) scintillator, new trigger unit
Origami Mod. #4
HV=+-40V, Ibias=1,6µA
CO2 cooling on, Tin=-19,6°C, Tout=-21.3°C
I²C settings for cooled APV chips
e:\cvi-projects\apvdaq_fadc\origami6_V2_multi6_cooled.cfg
SMU: +HV=Inv.Nr. 545-77 -HV=Inv.Nr. 540-702
HVret configuration:
0 ... switch open / not connected
1 ... switch closed / connected
SMU |
Dock Box |
J5 |
J6 |
J7 |
0 |
0 (now isolated) |
1 |
1 |
0 |
J7... HV ret @ hybrid |
|