  93   Wed Mar 7 16:03:30 2012 Christian IrmlerBELLEOrigami Module #4run113100k07.03.2012 16:07:3407.03.2012 16:30:48 Unknown

Sr90, small (new) scintillator, new trigger unit
Origami Mod. #4
HV=+-40V, Ibias=1,6µA
CO2 cooling on, Tin=-19,6°C, Tout=-21.3°C

I²C settings for cooled APV chips

e:\cvi-projects\apvdaq_fadc\origami6_V2_multi6_cooled.cfg

SMU:  +HV=Inv.Nr. 545-77    -HV=Inv.Nr. 540-702

HVret configuration:

0 ... switch open / not connected
1 ... switch closed / connected
 

SMU Dock Box J5 J6 J7
0 0 (now isolated) 1 1 0

J7... HV ret @ hybrid

same configuration as run112 but now the step motors of the xz table are powered.

  94   Wed Mar 7 16:58:36 2012 Christian IrmlerBELLEOrigami Module #4cal113100k07.03.2012 16:40:04 07.03.2012 16:57:08Unknown

Sr90, small (new) scintillator, new trigger unit
Origami Mod. #4
HV=+-40V, Ibias=1,6µA
CO2 cooling on, Tin=-19,6°C, Tout=-21.3°C

I²C settings for cooled APV chips

e:\cvi-projects\apvdaq_fadc\origami6_V2_single_cooled.cfg

SMU:  +HV=Inv.Nr. 545-77    -HV=Inv.Nr. 540-702

HVret configuration:

0 ... switch open / not connected
1 ... switch closed / connected
 

SMU Dock Box J5 J6 J7
0 0 (now isolated) 1 1 0

J7... HV ret @ hybrid

same configuration as run112 but now the step motors of the xz table are powered.

  8   Fri Sep 16 11:01:15 2011 Manfred ValentanMicronMicron Baby Modules     

Irradiated modules from last year:

COMB-IRRAD: B2Micron_2826_19_Baby4: combined p-stop, intermediate strips (was DUT4)
COMM-IRRAD: B2Micron_2826_07_Baby2: common p-stop, intermediate strips (was DUT3)
(the irradiated atoll p-stop sensor was destroyed for electron microscopy and SRP)

Unirradiated modules from last year:

COMB-1: B2Micron_2826_01_Baby4: combined p-stop, intermediate strips (was TRK6)
COMM-1: B2Micron_2826_19_Baby2: common p-stop, intermediate strips (was TRK7)
ATOLL-1: B2Micron_2826_07_Baby3: atoll p-stop, intermediate strips (was prepared, but not used in the beam test)

New modules:

SPRAY-I-1: B2Micron_2825_15_Baby2: p-spray, intermediate strips
SPRAY-I-2: B2Micron_2825_15_Baby3: p-spray, intermediate strips
SPRAY-1: B2Micron_2825_05_Baby3: p-spray, no intermediate strips
SPRAY-2: B2Micron_2825_05_Baby4: p-spray, no intermediate strips
COMB-2: B2Micron_2826_17_Baby4: combined p-stop, intermediate strips
COMM-2: B2Micron_2826_17_Baby2: common p-stop, intermediate strips
ATOLL-2: B2Micron_2826_17_Baby3: atoll p-stop, intermediate strips

 

  9   Mon Sep 19 17:38:34 2011 Manfred ValentanMicronModule ATOLL-2     

Sensor: B2Micron_2826_17_Baby3

p-stop: Atoll, with intermediate strips

Hybrid p-side: H35 (reused from last year)
Hybrid n-side: MB3

  10   Mon Sep 19 17:39:09 2011 Manfred ValentanMicronModule COMM-2     

Sensor: B2Micron_2826_17_Baby2

p-stop: Common, with intermediate strips

Hybrid p-side: H34 (reused from last year)
Hybrid n-side: MB4

  11   Mon Sep 19 17:39:55 2011 Manfred ValentanMicronModule COMB-2     

Sensor: B2Micron_2826_17_Baby4

p-stop: Combined with intermediate strips

Hybrid p-side: H42 (reused from last year)
Hybrid n-side: MB3

  12   Mon Sep 19 17:40:47 2011 Manfred ValentanMicronModule SPRAY-2     

Sensor: B2Micron_2825_05_Baby4

p-stop: p-spray WITHOUT intermediate strips

Hybrid p-side: H44 (reused from last year)
Hybrid n-side: MB6

  13   Mon Sep 19 17:41:27 2011 Manfred ValentanMicronModule SPRAY-I-2     

Sensor: B2Micron_2825_15_Baby3

p-stop: p-spray WITH intermediate strips

Hybrid p-side: H43 (reused from last year)
Hybrid n-side: MB7

  14   Mon Sep 19 17:42:01 2011 Manfred ValentanMicronModule SPRAY-1     

Sensor: B2Micron_2825_05_Baby3

p-stop: p-spray WITHOUT intermediate strips

Hybrid p-side: H32 (reused from last year)
Hybrid n-side: MB8

  15   Mon Sep 19 17:42:39 2011 Manfred ValentanMicronModule SPRAY-I-1     

Sensor: B2Micron_2825_15_Baby2

p-stop: p-spray WITH intermediate strips

Hybrid p-side: MB1
Hybrid n-side: MB9

  16   Mon Sep 19 17:44:00 2011 Manfred ValentanMicronModule COMM-1     

Sensor: B2Micron_2826_19_Baby2 (reused from last year, was TRK7)

p-stop: Common, with intermediate strips

Hybrid p-side: H46
Hybrid n-side: M37

  17   Mon Sep 19 17:44:39 2011 Manfred ValentanMicronModule COMB-1     

Sensor: B2Micron_2826_01_Baby4 (reused from last year, was TRK6)

p-stop: Combined, with intermediate strips

Hybrid p-side: H45
Hybrid n-side: M36

  18   Mon Sep 19 17:46:07 2011 Manfred ValentanMicronModule ATOLL-1     

Sensor: B2Micron_2826_07_Baby3 (reused from last year, but was not used in beam test SPS2010)

p-stop: Atoll, with intermediate strips

Hybrid p-side: H40
Hybrid n-side: M30

  19   Mon Sep 19 17:47:11 2011 Manfred ValentanMicronModule COMB-IRRAD     

Sensor: B2Micron_2826_19_Baby4 (reused from last year, was DUT4, module is irradiated to 700 kGy)

p-stop: Combined, with intermediate strips, irradiated

Hybrid p-side: H48
Hybrid n-side: M35

  20   Mon Sep 19 17:47:50 2011 Manfred ValentanMicronModule COMM-IRRAD     

Sensor: B2Micron_2826_07_Baby2 (reused from last year, was DUT3, module is irradiated to 700 kGy)

p-stop: Common, with intermediate strips, irradiated

Hybrid p-side: H41
Hybrid n-side: M31

  21   Mon Oct 3 20:30:14 2011 Thomas BergauercommonXY Tisch     

Grosser (DESY) Tisch:  x = -342.0    y = 343.0  (Einheiten unbekannt, möglicherweise mm)

Raten-Scan siehe gescannter Zettel im Attachment.

 

erster Wert: X, also horizontal (positive Zahlen: Saleve side = Restaurant Nr. 3)

zweiter Wert: Z, also Höhe (positive Zahlen nach unten)

Micron DUT: X:324, Z:103 [Einheiten in mm]

 

configuration: IP-Address 192.168.0.13; 2 Achsen

  23   Mon Oct 3 21:08:27 2011 Manfred ValentanMicronMicron-Baby-Stack p-stop, configuration     

Photo of hand-drawn stack configuration, config files

  36   Wed Oct 5 19:33:40 2011 Markus FriedlcommonTrigger condition  4.10.2011 15:00  

Changed trigger such that off-spill scintillator triggers are vetoed.

Inserted Flip-Flop which is set by begin of spill (WWE) and reset by end of spill signal (EE). Out-bar is put in coincidence with the scintillator trigger.

This also changes the timing, which was unfortunately not detected until 30 hours later. Thus, runs 4-7 have bad (wrong) timing -> UNUSABLE

  45   Fri Oct 7 18:28:53 2011 Christian Irmlercommonrun schedule    Unknown
  49   Sat Oct 8 09:30:44 2011 Christian IrmlerBELLECooling plant tests  08.10.2011 09:30 Unknown

Software run with 6 samples to ensure that APV chips dissipate full power.

DAQ data not saved, bias current monitored

new longterm files:
It-neg-0810110926.txt
It-pos-0810110926.txt