| ID | 
Date | 
Author | 
Project | 
Subject | 
Run Number | 
Events | 
StartTime  | 
EndTime | 
Data | 
 
| 
  38  
 | 
Fri Oct 19 00:56:16 2012 | 
Markus Friedl | MicronOrigami_gross | run | run017 | 197899 | 19.10.2012  00:54:21 | 19.10.2012  14:10:02  | Good |  Belle Stack 4: 2 DSSD -z, Origami #3, #1, wedge new #2 
HV = +-50V, Ibias=76A 
Position: APVs #2, 7, 11, 15 | # 17/18, 21, 25, 29 
= p: 2DSSD #2, Origami3 #3, Origami1 #3, wedgenew #3 
= n: 2DSSD #1/2, Origami3 #1, Origami1 #1, wedgenew #1 
Same as run016, but timing shifted by 1 CLK to better center the peak of the shaped wafeforms for subsequent fitting during analysis. 
  
Beam was turned off around here:  
19.10.2012  08:56:12  event=193394 
So better ignore the last ~5000 events for analysis. 
   |  
| 
  37  
 | 
Fri Oct 19 00:45:52 2012 | 
Markus Friedl | MicronOrigami_gross | run | run016 | 11855 | 19.10.2012  00:24:05 | 19.10.2012  00:48:24 | Crap |    
Belle Stack 4: 2 DSSD -z, Origami #3, #1, wedge new #2 
HV = +-50V, Ibias=77A 
Position: APVs #2, 7, 11, 15 | # 17/18, 21, 25, 29 
= p: 2DSSD #2, Origami3 #3, Origami1 #3, wedgenew #3 
= n: 2DSSD #1/2, Origami3 #1, Origami1 #1, wedgenew #1 
  
Stack is now positioned on globally best performing position which should also enable reasonable tracking. 
Trigger timing not perfect, needs adjustment (-1 CLK) -> new run  |  
| 
  35  
 | 
Wed Oct 17 21:30:39 2012 | 
Markus Friedl | MicronOrigami_gross | cal | cal008 | 83800 | 17.10.2012  17:15:13 | 17.10.2012  18:08:11 | Good |  Internal calibration for "stealth parasitic" mode Belle stack; taken without beam 
  
Belle Stack 4: 2 DSSD -z, Origami #3, #1, wedge new #2 
HV = +-50V, Ibias=80A 
Position: approx. centered 
no beam 
  
  
Total bias current history: 
  
At the beginning, the system was already running for a few hours and the APV25 chips were initialized before, so thermal equilibrium was reached. 
The actual calibration run happened during the first ~250 measurement points (~1 hour). It seems like the APVs draw a bit more current during internal calibration. The overall decay is supposedly due to a temperature decrease towards night. 
  
Now tried the same thing with a short software run - same effect: 
  
Need to investigate supply currents while data taking - they should actually go up a bit - and their dependency on the trigger rate. 
   |  
| 
  33  
 | 
Wed Oct 17 05:16:22 2012 | 
Christian Irmler | MicronOrigami_gross | run | run015 | 54005 | 17.10.2012  05:10:06 | 17.10.2012  07:27:20 | Unknown |  Origami Stack 2: Origami #4, #3, Wedge new1, old 
HV = +-50V 
Ibias=4.43µA 
  
beam position: 
vertical: between APV 1&2 (of 6) 
horizontal: centered on APV#2 -> center of measureable area 
Position of xy table: 
x = 53, y = -31 
Temperatur sensors: 
    - T11: Origami #4 in
 
    - T12: Origami #4 out
 
    - T13: Origami #3 in (wrong values!)
 
    - T14: Origami #3 out
 
    - T15: Wedge new 1 out
 
    - T16: Wedge old out
 
    - T17: Wedge new 1 cooling block
 
    - T18: Wedge old cooling block
 
 
CO2 was empty @ 6 am.  |  
| Attachment 1: run015_temperatures.JPG
 |    |  
| 
  31  
 | 
Tue Oct 16 23:35:28 2012 | 
Christian Irmler | MicronOrigami_gross | run | run014 | 150k | 16.10.2012  23:32:24 | 17.10.2012  04:59:02  | Unknown |  Same as run011, but now with +- 50V bias voltage. 
Origami Stack 2: Origami #4, #3, Wedge new1, old 
HV = +-10V 
Ibias=4.83µA 
  
beam position: 
vertical: between APV 1&2 (of 6) 
horizontal: centered on APV#1 ->  right side of sensor -> small side of wedge sensors 
Position of xy table: 
x = 23, y = -31 
Temperatur sensors: 
    - T11: Origami #4 in
 
    - T12: Origami #4 out
 
    - T13: Origami #3 in (wrong values!)
 
    - T14: Origami #3 out
 
    - T15: Wedge new 1 out
 
    - T16: Wedge old out
 
    - T17: Wedge new 1 cooling block
 
    - T18: Wedge old cooling block
 
 
   |  
| Attachment 1: run014_temperatures.jpg
 |    |  
| 
  29  
 | 
Tue Oct 16 22:19:21 2012 | 
Christian Irmler | MicronOrigami_gross | cal | cal007 | 83800 | 16.10.2012  20:25:17  | 16.10.2012  21:12:01  | Unknown |  Origami Stack 2: Origami #4, #3, Wedge new1, old 
HV = +-50V, Ibias=12.3µA 
Position: verticallbtween APV 1&2 of 6,  left side of sensor,  
wide side of wedge sensors 
calibration for run013 
Temperatures @ 21:00: 
T11: -32 degC 
T12: -33 degC 
T13: -16 degC (obviously wrong value) 
T14: -35 degC 
T15: -18 degC 
T16: -17 degC 
T17: -15 degC 
T18: -15 degC  |  
| 
  28  
 | 
Tue Oct 16 13:59:46 2012 | 
Christian Irmler | MicronOrigami_gross | run | run013 | 150k | 16.10.2012  13:43:05 | 16.10.2012  20:21:04 | Unknown |  Origami Stack 2: Origami #4, #3, Wedge new1, old 
HV = +-50V 
Ibias=5.2µA 
  
beam position: 
vertical: between APV 1&2 (of 6) 
horizontal: centered on APV#3 ->  left side of sensor -> wide side of wedge sensors 
Position of xy table: 
x = 83, y = -31 
  
Temperatur sensors: 
    - T11: Origami #4 in
 
    - T12: Origami #4 out
 
    - T13: Origami #3 in (wrong values!)
 
    - T14: Origami #3 out
 
    - T15: Wedge new 1 out
 
    - T16: Wedge old out
 
    - T17: Wedge new 1 cooling block
 
    - T18: Wedge old cooling block
 
 
  
   |  
| Attachment 1: belle_2_temperatures.jpg
 |    |  
| 
  27  
 | 
Tue Oct 16 11:23:43 2012 | 
Christian Irmler | MicronOrigami_gross | run | run012 | 20k | 16.10.2012  11:16:01 | 16.10.2012  12:27:03  | Unknown |  Origami Stack 2a: Origami #4, Wedge new1 
reading out all 6 APV of p-side 
HV = +-40V, Ibias=10.4µA 
  
Origami 3 and Wedge old are still installed in the box, but not read out --> disconnected 
beam: 
p-side: centered between APV 4&5 
n-side: centered between APV 2&3 
--> upper left corner of Origami and wedge module --> we hit the APV chips and / or the PA0 of the Origami module 
no cooling 
Temperatur sensors: 
    - T11: Origami #4 in
 
    - T12: Origami #4 out
 
    - T13: Origami #3 in (wrong values!)
 
    - T14: Origami #3 out
 
    - T15: Wedge new 1 out
 
    - T16: Wedge old out
 
    - T17: Wedge new 1 cooling block
 
    - T18: Wedge old cooling block
 
  |  
| Attachment 1: belle_2a_temperatures.jpg
 |    |  
| 
  26  
 | 
Tue Oct 16 11:07:17 2012 | 
Christian Irmler | MicronOrigami_gross | cal | cal006 | 83800 | 16.10.2012  10:30:34 | 16.10.2012  11:00:51 | Unknown |  Origami Stack 2a: Origami #4, Wedge new1 
reading out all 6 APV of p-side 
HV = +-40V, Ibias=5.12µA 
  
no beam  
no cooling  |  
| 
  25  
 | 
Tue Oct 16 05:24:21 2012 | 
Christian Irmler | MicronOrigami_gross | run ATTENTION: wrong bias voltage | run011 | 150k | 16.10.2012  05:19:28  | 16.10.2012  09:59:24 | Crap |  ATTENTION: Bias voltage was +-100 V instead of +- 50 V 
Origami Stack 2: Origami #4, #3, Wedge new1, old 
HV = +-100V 
Ibias=12.3µA 
  
beam position: 
vertical: between APV 1&2 (of 6) 
horizontal: centered on APV#3 ->  left side of sensor -> wide side of wedge sensors 
Position of xy table: 
x = 83, y = -25 
 
Temperatures: 
Origami #4: In=-33 degC / Out=-34 degC 
Origami #3: Out=-35 degC (temperatur sensor of inlet is not working) 
Wedge new 1: Out =  -17 degC, cooling block = -16 degC 
Wedge old: Out = -15 degC, cooling block = -15 degC 
  
Stopped after 132658 events, because of empty gas bottle and thus insufficient cooling.  |  
| 
  32  
 | 
Wed Oct 17 01:21:50 2012 | 
Marko Dragicevic | Infineon | run | run009 | 150k | 15.10.2012 08:20:15 | 15.10.2012 11:01:11 | Unknown |   This run was not recorded in original sequence in the ELOG!!! 
  
    
        
            | 
              Infineon Stack Version 2 (no height information, BabyStrips vertical Strixel vertical) 
            HV = 300V   I=4uA 
            beam on 
              
            Final DESY position (center of vertical beam profile) 
            Module position: 15 / 3 -- hitting lower part of baby and bottom strixels 
            
                
                    
                        | 07_baby | 
                        0,0 | 
                        APV1 lower, APV2 upper | 
                     
                    
                        | Strixel 2 | 
                        20,8 | 
                        Hybrid1/APV1 top-left, Hybrid1/APV2 top right, Hybrid2/APV1 bottom left, Hybrid2/APV2 bottom right  | 
                     
                    
                        | Strixel 1 | 
                        39,4 | 
                        Hybrid1/APV1 top-left, Hybrid1/APV2 top right, Hybrid2/APV1 bottom left, Hybrid2/APV2 bottom right  | 
                     
                    
                        | 12_baby | 
                        60,6 | 
                        APV1 lower, APV2 upper | 
                     
                
             
              
              
             | 
         
    
  |  
| 
  19  
 | 
Mon Oct 15 00:08:26 2012 | 
Marko Dragicevic | Infineon | run | run008 | 339075 | 15.10.2012 01:49:33 | 15.10.2012 08:04:24 | Unknown |   Infineon Stack Version 2 (no height information, BabyStrips vertical Strixel vertical) 
HV = 300V   I=4uA 
beam on 
  
Final DESY position (center of vertical beam profile) 
Module position: 0 / 3 -- hitting upper part of baby and top strixels 
    
        
            | 07_baby | 
            0,0 | 
            APV1 lower, APV2 upper | 
         
        
            | Strixel 2 | 
            20,8 | 
            Hybrid1/APV1 top-left, Hybrid1/APV2 top right, Hybrid2/APV1 bottom left, Hybrid2/APV2 bottom right  | 
         
        
            | Strixel 1 | 
            39,4 | 
            Hybrid1/APV1 top-left, Hybrid1/APV2 top right, Hybrid2/APV1 bottom left, Hybrid2/APV2 bottom right  | 
         
        
            | 12_baby | 
            60,6 | 
            APV1 lower, APV2 upper | 
         
    
 
   |  
| 
  24  
 | 
Mon Oct 15 23:26:52 2012 | 
Marko Dragicevic | MicronOrigami_gross | run ATTENTION: wrong bias voltage | run010 | 158k | 15.10.2012  23:10:40  | 16.10.2012  05:05:35 | Crap |  ATTENTION: Bias voltage was +-100 V instead of +- 50 V 
Same as run009. 
Origami Stack 2: Origami #4, #3, Wedge new1, old 
HV = +-10V 
Ibias=12.3µA 
  
beam position: 
vertical: between APV 1&2 (of 6) 
horizontal: centered on APV#1 ->  right side of sensor -> small side of wedge sensors 
Position of xy table: 
x = 23, y = -25 
 
Temperatures: 
Origami #4: In=-33 degC / Out=-34 degC 
Origami #3: Out=-35 degC (temperatur sensor of inlet is not working) 
Wedge new 1: Out =  -17 degC, cooling block = -14 degC 
Wedge old: Out = -16 degC, cooling block = -14 degC 
  
15.10.2012  22:08:31  event=0  rate=0events/s temp/curr=31.3/089.2 31.3/-16.2 31.2/-17.9 31.3/-18.2 29.9/003.6 29.9/003.1 30.0/-01.1 29.9/-01.6  
  |  
| 
  23  
 | 
Mon Oct 15 22:13:28 2012 | 
Marko Dragicevic | MicronOrigami_gross | cal ATTENTION: wrong bias voltage  | cal005 | 83800 | 15.10.2012  22:16:26 | 15.10.2012  23:07:37  | Unknown |  ATTENTION: Bias voltage was +-100 V instead of +- 50 V 
Origami Stack 2: Origami #4, #3, Wedge new1, old 
HV = +-100V 
Ibias=12.3µA 
  
beam position: 
vertical: between APV 1&2 (of 6) 
horizontal: centered on APV#1 ->  right side of sensor 
Position of xy table: 
x = 23, y = -25 
 
Temperatures: 
Origami #4: In=-33 degC / Out=-34 degC 
Origami #3: Out=-35 degC (temperatur sensor of inlet is not working) 
Wedge new 1: Out =  -17 degC, cooling block = -14 degC 
Wedge old: Out = -16 degC, cooling block = -14 degC 
   |  
| 
  22  
 | 
Mon Oct 15 19:40:03 2012 | 
Marko Dragicevic | MicronOrigami_gross | run ATTENTION: wrong bias voltage | run009 | 31650 | 15.10.2012  19:28:30  | 15.10.2012  22:08:31  | Crap |  ATTENTION! Bias voltage was +-100 V instead of +-50 V  
Origami Stack 2: Origami #4, #3, Wedge new1, old 
HV = +-100V 
Ibias=12.3µA 
  
beam position: 
vertical: between APV 1&2 (of 6) 
horizontal: centered on APV#1 ->  right side of sensor 
Position of xy table: 
x = 23, y = -25 
 
Temperatures: 
Origami #4: In=-33 degC / Out=-34 degC 
Origami #3: Out=-35 degC (temperatur sensor of inlet is not working) 
Wedge new 1: Out =  -17 degC, cooling block = -14 degC 
Wedge old: Out = -16 degC, cooling block = -14 degC 
  
15.10.2012  22:08:31  event=0  rate=0events/s temp/curr=31.3/089.2 31.3/-16.2 31.2/-17.9 31.3/-18.2 29.9/003.6 29.9/003.1 30.0/-01.1 29.9/-01.6  
  
Stopped after 31650 events, because of PS problems and no beam.  |  
| 
  18  
 | 
Sun Oct 14 23:01:11 2012 | 
Marko Dragicevic | Infineon | cal | cal004 | 83800 | 14.10.2012 21:44:00 | 14.10.2012 21:03:00 | Unknown |  Infineon Stack Version 1 (with height information, BabyStrips horizontal Strixel vertical) 
HV = 300V   I=3.1 uA  |  
| 
  17  
 | 
Sun Oct 14 20:12:14 2012 | 
Marko Dragicevic | Infineon | run | run007 | 110k | 14.10.2012 20:07:12 | 14.10.2012 21:04:12 | Unknown |  Infineon Stack Version 1 (with height information, BabyStrips horizontal Strixel vertical) 
HV = 300V   I=3uA 
beam on 
  
Final DESY position (center of vertical beam profile) 
Module position: 0 / 0 -- hitting the center of babies, between strixels 
    
        
            | 07_baby | 
            0,0 | 
            APV1 lower, APV2 upper | 
         
        
            | Strixel 2 | 
            20,8 | 
            Hybrid1/APV1 top-left, Hybrid1/APV2 top right, Hybrid2/APV1 bottom left, Hybrid2/APV2 bottom right  | 
         
        
            | Strixel 1 | 
            39,4 | 
            Hybrid1/APV1 top-left, Hybrid1/APV2 top right, Hybrid2/APV1 bottom left, Hybrid2/APV2 bottom right  | 
         
        
            | 12_baby | 
            60,6 | 
            APV1 lower, APV2 upper | 
         
    
 
  
   |  
| 
  16  
 | 
Sun Oct 14 09:08:42 2012 | 
Manfred Valentan | MicronBaby | run | run006 | 300k | 14.10.2012  09:01:15 |   | Unknown |  Micron baby intermediate stack 
HV = -100V   I=6.9uA 
beam on 
  
Final DESY position (center of vertical beam profile) 
Module position: 0 / -31.0 -- hitting the lower half of the sensor (centered on even APVs)  |  
| 
  15  
 | 
Sun Oct 14 08:25:10 2012 | 
Manfred Valentan | MicronBaby | cal | cal003 | 83800 | 14.10.2012  08:21:46 | 14.10.2012  08:51:58 | Unknown |  Micron baby intermediate stack 
HV = -100 V 
I = 6.9 uA  |  
| 
  14  
 | 
Sun Oct 14 02:02:39 2012 | 
Manfred Valentan | MicronBaby | run | run005 | 300k | 14.10.2012  02:00:08 | 14.10.2012  08:17:02 | Unknown |  Micron baby intermediate stack 
HV = -100V   I=6.3uA 
beam on 
  
Final DESY position (center of vertical beam profile) 
Module position: 0 / -12.0 -- hitting the upper half of the sensor (centered on odd APVs)  |  
 
 |