Thu Sep 23 15:54:20 2010, Manfred Valentan, Micron Baby, Modul B2Micron_2826_19_Baby3 -> DUT & TRACKING, , , , 6x
|
p-side: Hybrid 47
alles ok
n-side: Hybrid 38
APV5: Streifen 127 hoher Noise
ALS DUT VERWENDEN!
|
Thu Sep 23 16:00:31 2010, Manfred Valentan, Micron Baby, Modul B2Micron_2826_19_Baby2 -> TRACKING, , , , 6x
|
p-side: Hybrid 46
alles ok
n-side: Hybrid 37
APV4: 70-75 schlecht, nix zu sehen -> Sensorfehler
APV5: 115-128 schlecht, rebond fail?
FUER TRACKING GEEIGNET!
|
Thu Sep 23 16:01:10 2010, Manfred Valentan, Micron Baby, Modul B2Micron_2826_7_Baby3 -> DUT, , , , 6x
|
p-side: Hybrid 40
APV3: letzte 10 Streifennicht gebondet, ging sich beim Prototyp-Modul an der Stelle nicht aus
n-side: Hybrid 30
APV5: 57-64 schlecht, rebond fail?
ALS DUT GEEIGNET!
|
Thu Sep 23 16:12:54 2010, Manfred Valentan, Micron Baby, Modul B2Micron_2826_7_Baby2 -> DUT & TRACKING, , , , 6x
|
p-side: Hybrid 41
alles ok
n-side: Hybrid 31
APV4: Streifen 0, 83 schlecht
APV5: Streifen 64, 88, 126 schlecht
ALS DUT GEEIGNET!
FUER TRACKING GEEIGNET! |
Thu Sep 23 16:19:54 2010, Manfred Valentan, Micron Baby, Modul B2Micron_2826_1_Baby4, DUT EINGESCHRAENKT, TRACKING, , , , 6x
|
p-side: Hybrid 45
APV0: beide Randstreifen schlecht
APV1: Streifen 0 schlecht
APV3: beide Randstreifen etwas schlechter, nicht schlimm
n-side: Hybrid 36
APV4: 0, 9, 40, 79-85, 97, 102, 124 Sensorfehler?
APV5: 0-30 vereinzelt schlechte Streifen, 115-128 Bonds beruehren
ALS DUT AUGENZUDRUECKEN GEEIGNET!
FUER TRACKING GEEIGNET!
|
Thu Sep 23 16:22:33 2010, Manfred Valentan, Micron Baby, Modul B2Micron_2826_1_Baby3, TRACKING, , , , 6x
|
p-side: Hybrid 44
alles ok
n-side: Hybrid 34
APV4: zur Gaenze schlecht, APV-Fehler?
APV5: beide Randstreifen schlecht
FUER TRACKING GEEIGNET!
|
Thu Sep 23 16:23:07 2010, Manfred Valentan, Micron Baby, Modul B2Micron_2826_1_Baby2, DUT (eingeschraenkt) & TRACKING, , , , 6x
|
p-side: Hybrid 43
APV3: Streifen 9 schlecht
n-side: Hybrid 33
APV4: 0-10 andere Bond-Parameter?, 28&29 erhoehtes noise
APV5: 111-128: 4 schlechte Streifen + common mode correction?
ALS DUT GEEIGNET!
FUER TRACKING GEEIGNET!
|
Fri Sep 24 12:09:32 2010, Manfred Valentan, BELLE, gds-file of Micron wafer, , , ,
|
|
Mon Sep 27 15:39:33 2010, Thomas Bergauer, common, PI-XY Table, , , ,
|
M-521.dd: (links/rechts)
downstream rechts => pos. limit [0]
downstream links => neg. limit [207]
----------------------
M-511.dd: (oben/unten)
oben [102]
unten [0]
----------------------
M-060.dg: (rotation)
parallel beam [7,5]
-------------------------------------------
Remote Control:
lokal ssh tunnel starten:
ssh -g -C -L5959:zenpixelw:5900 eudet@eudetmac001.cern.ch
danach mit "Chicken of the VNC" (von lokalem Mac) VNC auf localhost (port 5959).
|
Mon Sep 27 15:41:28 2010, Thomas Bergauer, common, Szintillatoren, , , ,
|
1: breit (1)
2: schmal (2)
3: breit (4)
4: schmal (8)
AND aus beiden breiten: 5
AND aus beiden schmalen: 10
|
Mon Sep 27 15:42:09 2010, Thomas Bergauer, common, TLU: wir stecken auf DUT 3, , , ,
|
TLUControl.exe ist auf zenpixell3 unter /home/eudet/eudaq/bin
TLUControl.exe -d 14 -a 12
-dXX Mask for DUT:
1 abgeschnittenes Dummykabel
2 EUDRB hinten
4 EUDRB vorne
8 DUT3(wir)
-t100 internal trigger
-aXX Mask for Szintis:
1 4 mm (front)
2 4 mm (front)
4 9 mm (back)
8 9 mm (back) |
Mon Sep 27 18:54:47 2010, Thomas Bergauer, BELLE, XY Position, , , ,
|
center position: 55/40/277,5 |
Tue Sep 28 01:13:57 2010, Markus Friedl, BELLE, cal, cal001, 83800, 28.09.2010 01:11:20,
|
first intcal with complete Belle stack: 2 x Origami, 2 x Micron
HV=100V I=49uA
beam is on
chiller is set to 15°C (module temperature is ~17°C)
both Origamis and both Microns are serially cooled |
Tue Sep 28 12:54:33 2010, Markus Friedl, BELLE, run, run002, , 28.09.2010 12:46:54,
|
same as run001 after shifting x by 48mm
103 / 30 / 277.5 ? unsure after crash of xy stage PC |
Tue Sep 28 14:38:24 2010, Christian Irmler, BELLE, Temperature sensors Belle stack, , , ,
|
PicoLogic TC-08 temperature sensor assignment Belle stack:
Sensor type: K
channel 1: Origami_mod1_in
channel 2: Origami_mod1_out
channel 3: Origami_mod2_in
channel 4: Origami_mod2_out
channel 5: Micron_wedge_mod1_p
channel 6: Micron_wedge_mod1_n
channel 7: Micron_wedge_mod2_p
channel 8: Micron_wedge_mod2_n
The cooling pipes of the two Origami modules and that of the Micron modules are connected in series, respectively. |
Tue Sep 28 16:50:31 2010, Thomas Bergauer, common, serial connection chiller, , , ,
|
Der chiller braucht ein gekreuztes serielles Kabel!! |
Tue Sep 28 22:07:16 2010, Markus Friedl, BELLE, cal, cal002 DO NOT USE -> go to cal003, 83800, 28.09.2010 22:01:55,
|
Belle stack (Origami, Micron, Micron, Origami)
now cooling only for Origami, Micron are uncooled
chiller set point: -25°C, measured values on Origami: -10°C
HV=100V, Ibias=38µA, beam off
modified I2C settings for cold modules
temperature monitoring: Origami2 pipes in<->out swapped (to achieve shorter tubes)
PROBLEM: Origami n channels are swapped -> USE cal003 INSTEAD!!! |
Tue Sep 28 22:09:39 2010, Markus Friedl, BELLE, Temperature sensors Belle stack, , , ,
|
PicoLogic TC-08 temperature sensor assignment Belle stack:
Sensor type: K
channel 1: Origami_mod1_in
channel 2: Origami_mod1_out
channel 3: Origami_mod2_out (swapped with 4) - label not changed
channel 4: Origami_mod2_in (swapped with 3) - label not changed
channel 5: Micron_wedge_mod1_p
channel 6: Micron_wedge_mod1_n
channel 7: Micron_wedge_mod2_p
channel 8: Micron_wedge_mod2_n
The cooling pipes of the two Origami modules are connected in series, respectively. |
Wed Sep 29 09:39:33 2010, Markus Friedl, BELLE, cal, cal003, 83800, 29.09.2010 09:38:00,
|
same as cal002, but now with corrected readout order |
Wed Sep 29 17:38:19 2010, Markus Friedl, Micron Baby, cal, cal005, 83800, 29.09.2010 17:36:38,
|
micron baby stack
HV=100V I=6.5uA
Beam off |
|