ID |
Date |
Author |
DUT |
Subject |
Run Number |
Events |
XYPosition |
Cooling |
Data |
1
|
Thu Oct 12 10:06:31 2017 |
Dominic Bloech | common | Packlist | | | | No | Unknown |
# Testbeam Packliste
### Alibava stuff
* Alibava Mainboard
* Alibava power supply
* Alibava Modules: IFX_CenterBias, HPK_CenterBias, IFX_baby
* Flatband cable 2x
* Alibava resetbox + Cable
* USB AtoB 2x
* BIAS Cable for BeetleChip
* 2410 SMU + Power + USBtoRS232
* Alibava mounting stuff
### Tools and Stuff
* Labeling machine
* Duct tape
* 2 pairs of wool gloves
* Seitenschneider klein
* USBtoRS232 2x
* Lan cable
* Schiebelehre
* Imbuschlüssel
* Screwdriver set
* Special screwdriver set
* Cutter 2x
* labeling cartridge
* Lemo 16 ns
* Lemo 8 ns 2x
* Lemo 4 ns
* Lemo 2 ns
### PC and stuff
* DAQ PC + Cables
* Monitor + Cables
* Mouse
* Keyboard
### Data Storage
* Toshiba 1TB USB3
* Maxtor 4TB USB3 |
Attachment 1: ALiBaVa-System_klein.jpg
|
|
Attachment 2: Stuff_klein.jpg
|
|
2
|
Mon Oct 16 15:01:21 2017 |
Dominic Bloech | common | Preparations | | | | No | Unknown |
### General
* Setup was set-up without problems. Pedestral and Calibration runs with ALIBAVA successfully.
** Warning: Trigger Out signal (busy signal) for the TLU is a TTL signal. Whereas the Trigger In signal can be a NIM signal. Either converting signal or adept in the TLU (we adepted the TLU matrix)
** TLU double triggered -> discriminated TLU signal beforehand
* Longterm BIAS Voltage for the IFX CenterBias to -100 V, slow control records IV-behavior in 10 sec steps.
* Security lectures taken
* 3 coincident measurements: 2 PM and the FEI4-Plane are connected to an "and" in the TLU
* Latency of Alibava without FEI4-Plane 133 [a.u] with FEI4 140 [a.u].
** Warning: This is due to the converter from the FEI4 which inverts the signal of the FEI4, this signal is then very noisy and would generate false positive triggers. -> Discriminate -> Long cables and latencies
### Other
* NI-Crate was broken (replaced by DESY) |
Attachment 1: IMG_20171016_093005514.jpg
|
|
Attachment 2: IMG_20171016_150215807.jpg
|
|
Attachment 3: IMG_20171016_150230304.jpg
|
|
Attachment 4: IMG_20171016_150238370.jpg
|
|
75
|
Tue Oct 24 18:02:47 2017 |
Peter Paulitsch | HPK_200 | run000264 | 264 | 110000 | 0° | No | Good |
Parameter |
Value |
Alibava filename |
run000264.dat |
Telescope filename |
run000264.raw |
Pedestal filename |
HEPHY_pedestal_HPK_CB_2.ped |
Bias voltage |
-200 V |
Dark current |
|
Sensor temperature |
-7 |
Latency |
140 [a.u.] |
Alibava triggers |
110000 |
Telescope triggers |
100000 |
Data synced |
yes |
Table position x/y |
|
Region of interest |
bias line |
Trigger frequency |
|
Meas. time |
|
Beam energy |
5.6 GeV |
Cooling |
no |
Alibava efficiency |
|
|
76
|
Tue Oct 24 18:04:09 2017 |
Peter Paulitsch | HPK_200 | run000266 | 266 | 110000 | 0° | No | Good |
Parameter |
Value |
Alibava filename |
run000266.dat |
Telescope filename |
run000266.raw |
Pedestal filename |
HEPHY_pedestal_HPK_CB_2.ped |
Bias voltage |
-200 V |
Dark current |
|
Sensor temperature |
|
Latency |
140 [a.u.] |
Alibava triggers |
110000 |
Telescope triggers |
100000 |
Data synced |
yes |
Table position x/y |
|
Region of interest |
bias line |
Trigger frequency |
|
Meas. time |
|
Beam energy |
5.6 GeV |
Cooling |
no |
Alibava efficiency |
|
|
77
|
Tue Oct 24 18:08:06 2017 |
Peter Paulitsch | HPK_200 | run000271 | 271 | 110000 | 0° | No | Good |
Parameter |
Value |
Alibava filename |
run000271.dat |
Telescope filename |
run000271.raw |
Pedestal filename |
HEPHY_pedestal_HPK_CB_2.ped |
Bias voltage |
-200 V |
Dark current |
|
Sensor temperature |
|
Latency |
140 [a.u.] |
Alibava triggers |
110000 |
Telescope triggers |
100000 |
Data synced |
yes |
Table position x/y |
|
Region of interest |
bias line |
Trigger frequency |
|
Meas. time |
|
Beam energy |
5.6 GeV |
Cooling |
no |
Alibava efficiency |
|
|
78
|
Tue Oct 24 18:08:38 2017 |
Peter Paulitsch | HPK_200 | run000273 | 273 | 110000 | 0° | No | Good |
Parameter |
Value |
Alibava filename |
run000273.dat |
Telescope filename |
run000273.raw |
Pedestal filename |
HEPHY_pedestal_HPK_CB_2.ped |
Bias voltage |
-200 V |
Dark current |
|
Sensor temperature |
|
Latency |
140 [a.u.] |
Alibava triggers |
110000 |
Telescope triggers |
100000 |
Data synced |
yes |
Table position x/y |
|
Region of interest |
bias line |
Trigger frequency |
|
Meas. time |
|
Beam energy |
5.6 GeV |
Cooling |
no |
Alibava efficiency |
|
|
79
|
Tue Oct 24 18:08:57 2017 |
Peter Paulitsch | HPK_200 | run000275 | 275 | 1010000 | 0° | No | Good |
Parameter |
Value |
Alibava filename |
run000275.dat |
Telescope filename |
run000275.raw |
Pedestal filename |
HEPHY_pedestal_HPK_CB_2.ped |
Bias voltage |
-200 V |
Dark current |
|
Sensor temperature |
|
Latency |
140 [a.u.] |
Alibava triggers |
1010000 |
Telescope triggers |
1000000 |
Data synced |
yes |
Table position x/y |
|
Region of interest |
bias line |
Trigger frequency |
|
Meas. time |
|
Beam energy |
5.6 GeV |
Cooling |
no |
Alibava efficiency |
|
|
80
|
Tue Oct 24 18:09:24 2017 |
Peter Paulitsch | HPK_200 | run000277 | 277 | 1010000 | 0° | No | Good |
Parameter |
Value |
Alibava filename |
run000277.dat |
Telescope filename |
run000277.raw |
Pedestal filename |
HEPHY_pedestal_HPK_CB_2.ped |
Bias voltage |
-200 V |
Dark current |
|
Sensor temperature |
|
Latency |
140 [a.u.] |
Alibava triggers |
1010000 |
Telescope triggers |
1000000 |
Data synced |
yes |
Table position x/y |
|
Region of interest |
bias line |
Trigger frequency |
|
Meas. time |
|
Beam energy |
5.6 GeV |
Cooling |
no |
Alibava efficiency |
|
|
81
|
Tue Oct 24 18:09:43 2017 |
Peter Paulitsch | HPK_200 | run000279 | 279 | 110000 | 0° | No | Good |
Parameter |
Value |
Alibava filename |
run000279.dat |
Telescope filename |
run000279.raw |
Pedestal filename |
HEPHY_pedestal_HPK_CB_2.ped |
Bias voltage |
-200 V |
Dark current |
|
Sensor temperature |
|
Latency |
140 [a.u.] |
Alibava triggers |
110000 |
Telescope triggers |
100000 |
Data synced |
yes |
Table position x/y |
|
Region of interest |
bias line |
Trigger frequency |
|
Meas. time |
|
Beam energy |
5.6 GeV |
Cooling |
no |
Alibava efficiency |
|
|
82
|
Tue Oct 24 18:10:14 2017 |
Peter Paulitsch | HPK_200 | run000281 | 281 | 110000 | 0° | No | Good |
Parameter |
Value |
Alibava filename |
run000281.dat |
Telescope filename |
run000281.raw |
Pedestal filename |
HEPHY_pedestal_HPK_CB_2.ped |
Bias voltage |
-200 V |
Dark current |
|
Sensor temperature |
|
Latency |
140 [a.u.] |
Alibava triggers |
110000 |
Telescope triggers |
100000 |
Data synced |
yes |
Table position x/y |
|
Region of interest |
bias line |
Trigger frequency |
|
Meas. time |
|
Beam energy |
5.6 GeV |
Cooling |
no |
Alibava efficiency |
|
|
83
|
Tue Oct 24 18:10:39 2017 |
Peter Paulitsch | HPK_200 | run000283 | 283 | 1010000 | 0° | No | Good |
Parameter |
Value |
Alibava filename |
run000283.dat |
Telescope filename |
run000283.raw |
Pedestal filename |
HEPHY_pedestal_HPK_CB_2.ped |
Bias voltage |
-200 V |
Dark current |
|
Sensor temperature |
|
Latency |
140 [a.u.] |
Alibava triggers |
1010000 |
Telescope triggers |
1000000 |
Data synced |
yes |
Table position x/y |
|
Region of interest |
bias line |
Trigger frequency |
|
Meas. time |
|
Beam energy |
5.6 GeV |
Cooling |
no |
Alibava efficiency |
|
|
84
|
Tue Oct 24 18:11:13 2017 |
Peter Paulitsch | HPK_200 | run000285 | 285 | 1010000 | 0° | No | Good |
Parameter |
Value |
Alibava filename |
run000285.dat |
Telescope filename |
run000285.raw |
Pedestal filename |
HEPHY_pedestal_HPK_CB_2.ped |
Bias voltage |
-200 V |
Dark current |
|
Sensor temperature |
|
Latency |
140 [a.u.] |
Alibava triggers |
1010000 |
Telescope triggers |
1000000 |
Data synced |
yes |
Table position x/y |
|
Region of interest |
bias line |
Trigger frequency |
|
Meas. time |
|
Beam energy |
5.6 GeV |
Cooling |
no |
Alibava efficiency |
|
|
85
|
Tue Oct 24 18:11:43 2017 |
Peter Paulitsch | HPK_200 | run000287 | 287 | 1010000 | 0° | No | Good |
Parameter |
Value |
Alibava filename |
run000287.dat |
Telescope filename |
run000287.raw |
Pedestal filename |
HEPHY_pedestal_HPK_CB_2.ped |
Bias voltage |
-600 V |
Dark current |
|
Sensor temperature |
|
Latency |
140 [a.u.] |
Alibava triggers |
1010000 |
Telescope triggers |
1000000 |
Data synced |
yes |
Table position x/y |
|
Region of interest |
bias line |
Trigger frequency |
|
Meas. time |
|
Beam energy |
5.6 GeV |
Cooling |
no |
Alibava efficiency |
|
|
8
|
Wed Oct 18 12:12:55 2017 |
Peter Paulitsch | IFX_CenterBias | run000091 | 91 | 1010000 | 122.9, 21, -3.3° | No | Good |
run settings
Parameter |
Value |
Alibava filename |
run000091.dat |
Telescope filename |
run000091.raw |
Pedestal filename |
pedestal_17102017 |
Bias voltage |
-100 V |
Dark current |
-165 nA |
Sensor temperature |
24.6 °C |
Latency |
140 [a.u.] |
Alibava triggers |
1010000 |
Telescope triggers |
1000000 |
Data synced |
yes |
Table position x/y |
122.9, 21 |
Region of interest |
lower part |
Trigger frequency |
185 Hz |
Meas. time |
90 min |
Beam energy |
5.6 GeV |
Cooling |
no |
Alibava efficiency |
44.7% |
|
12
|
Wed Oct 18 13:29:39 2017 |
Peter Paulitsch | IFX_CenterBias | run000102 | 102 | 110000 | 122.9, 23.2, -3.3° | No | Good |
run settings
Parameter |
Value |
Alibava filename |
run000102.dat |
Telescope filename |
run000102.raw |
Pedestal filename |
pedestal_17102017 |
Bias voltage |
-100 V |
Dark current |
-160 nA |
Sensor temperature |
24.6 °C |
Latency |
140 [a.u.] |
Alibava triggers |
110000 |
Telescope triggers |
100000 |
Data synced |
yes |
Table position x/y |
122.9, 23.8 |
Region of interest |
bias line |
Trigger frequency |
185 Hz |
Meas. time |
17 min |
Beam energy |
5.6 GeV |
Cooling |
no |
Alibava efficiency |
52.3% |
|
10
|
Wed Oct 18 12:22:03 2017 |
Peter Paulitsch | IFX_CenterBias | run000099 | 99 | 110000 | 122.9, 23.5, -3.3° | No | Good |
run settings
Parameter |
Value |
Alibava filename |
run000099.dat |
Telescope filename |
run000099.raw |
Pedestal filename |
pedestal_17102017 |
Bias voltage |
-100 V |
Dark current |
-160 nA |
Sensor temperature |
24.6 °C |
Latency |
140 [a.u.] |
Alibava triggers |
110000 |
Telescope triggers |
100000 |
Data synced |
yes |
Table position x/y |
122.9, 23.5 |
Region of interest |
bias line |
Trigger frequency |
185 Hz |
Meas. time |
17 min |
Beam energy |
5.6 GeV |
Cooling |
no |
Alibava efficiency |
52.3% |
|
13
|
Wed Oct 18 13:30:58 2017 |
Peter Paulitsch | IFX_CenterBias | run000104 | 104 | 110000 | 122.9, 23.5, -3.3° | No | Unknown |
run settings
Parameter |
Value |
Alibava filename |
run000104.dat |
Telescope filename |
run000104.raw |
Pedestal filename |
pedestal_17102017 |
Bias voltage |
-100 V |
Dark current |
-160 nA |
Sensor temperature |
24.6 °C |
Latency |
140 [a.u.] |
Alibava triggers |
110000 |
Telescope triggers |
100000 |
Data synced |
yes |
Table position x/y |
122.9, 23.5 |
Region of interest |
bias line |
Trigger frequency |
200 Hz |
Meas. time |
15 min |
Beam energy |
5.6 GeV |
Cooling |
no |
Alibava efficiency |
52.3% |
|
14
|
Wed Oct 18 13:33:05 2017 |
Peter Paulitsch | IFX_CenterBias | run000106 | 106 | 60000 | 122.9, 23.5, -3.3° | No | Good |
run settings
Parameter |
Value |
Alibava filename |
run000106.dat |
Telescope filename |
run000106.raw |
Pedestal filename |
pedestal_17102017 |
Bias voltage |
-100 V |
Dark current |
-160 nA |
Sensor temperature |
24.6 °C |
Latency |
140 [a.u.] |
Alibava triggers |
60000 |
Telescope triggers |
50000 |
Data synced |
yes |
Table position x/y |
122.9, 23.5 |
Region of interest |
bias line |
Trigger frequency |
200 Hz |
Meas. time |
8 min |
Beam energy |
5.6 GeV |
Cooling |
no |
Alibava efficiency |
53.4% |
|
15
|
Wed Oct 18 13:34:54 2017 |
Peter Paulitsch | IFX_CenterBias | run000107 | 107 | 110000 | 122.9, 23.5, -3.3° | No | Good |
run settings
Parameter |
Value |
Alibava filename |
run000107.dat |
Telescope filename |
run000107.raw |
Pedestal filename |
pedestal_17102017 |
Bias voltage |
-100 V |
Dark current |
-160 nA |
Sensor temperature |
24.6 °C |
Latency |
140 [a.u.] |
Alibava triggers |
110000 |
Telescope triggers |
100000 |
Data synced |
yes |
Table position x/y |
122.9, 23.5 |
Region of interest |
bias line |
Trigger frequency |
200 Hz |
Meas. time |
15 min |
Beam energy |
5.6 GeV |
Cooling |
no |
Alibava efficiency |
53.4% |
|
11
|
Wed Oct 18 12:28:08 2017 |
Peter Paulitsch | IFX_CenterBias | run000101 | 101 | 10000 | 122.9, 23.8, -3.3° | No | Good |
run settings
Parameter |
Value |
Alibava filename |
run000101.dat |
Telescope filename |
run000101.raw |
Pedestal filename |
pedestal_17102017 |
Bias voltage |
-100 V |
Dark current |
-160 nA |
Sensor temperature |
24.6 °C |
Latency |
140 [a.u.] |
Alibava triggers |
10000 |
Telescope triggers |
14500 |
Data synced |
yes |
Table position x/y |
122.9, 23.8 |
Region of interest |
bias line |
Trigger frequency |
185 Hz |
Meas. time |
2 min |
Beam energy |
5.6 GeV |
Cooling |
no |
Alibava efficiency |
52.3% |
|