107
|
Mon Apr 17 17:02:34 2017 |
Axel Koenig | ALiBaVa (IFX) | run_003_HPK_CB_200V.dat | 003 | 250000 | 107, 2.6, 230 | No | Unknown |
Remarks: Telescope run 606. Angle scan.
Telescope chiller turned off (could be the root cause of runs crashing).
Parameter |
Value |
Alibava run |
run_003_HPK_CB_200V.dat |
Telescope run |
606 |
Triggers |
250000 |
Alibava trigger eff. |
~62% |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-200 |
Latency |
132 |
Table x |
107 mm |
Table z |
2.6 mm |
Tabe rot |
230 deg
|
|
109
|
Mon Apr 17 17:46:38 2017 |
Axel Koenig | ALiBaVa (IFX) | run_005_HPK_CB_50V.dat | 005 | 150000 | 107, 2.6, 240 | No | Unknown |
Remarks: Telescope run 609. Bias scan 50 V. Bias scan.
Telescope chiller turned off (could be the root cause of runs crashing).
Parameter |
Value |
Alibava run |
run_005_HPK_CB_50V.dat |
Telescope run |
609 |
Triggers |
150000 |
Alibava trigger eff. |
~62% |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-50 |
Latency |
132 |
Table x |
107 mm |
Table z |
2.6 mm |
Tabe rot |
240 deg
|
|
111
|
Mon Apr 17 19:19:14 2017 |
Axel Koenig | ALiBaVa (IFX) | run_006_HPK_CB_100V.dat | 006 | 150000 | 107, 2.6, 240 | No | Unknown |
Remarks: Telescope run 610. Bias scan 100 V. Bias scan.
Telescope chiller turned off (could be the root cause of runs crashing).
Parameter |
Value |
Alibava run |
run_006_HPK_CB_100V.dat |
Telescope run |
610 |
Triggers |
150000 |
Alibava trigger eff. |
~62% |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-100 |
Latency |
132 |
Table x |
107 mm |
Table z |
2.6 mm |
Tabe rot |
240 deg
|
|
113
|
Mon Apr 17 19:54:08 2017 |
Axel Koenig | ALiBaVa (IFX) | run_007_HPK_CB_150V.dat | 007 | 300000 | 107, 2.6, 240 | No | Unknown |
Remarks: Telescope run 611. Bias scan 150 V. Bias scan. Run crashed at 294000 triggers!
Telescope chiller turned off (could be the root cause of runs crashing).
Parameter |
Value |
Alibava run |
run_007_HPK_CB_150V.dat |
Telescope run |
611 |
Triggers |
300000 |
Alibava trigger eff. |
~62% |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-150 |
Latency |
132 |
Table x |
107 mm |
Table z |
2.6 mm |
Tabe rot |
240 deg
|
|
114
|
Mon Apr 17 21:52:45 2017 |
Axel Koenig | ALiBaVa (IFX) | run_008_HPK_CB_250V.dat | 008 | 150000 | 107, 2.6, 240 | No | Unknown |
Remarks: Telescope run 613. Bias scan 250 V.
Telescope chiller turned off (could be the root cause of runs crashing).
Parameter |
Value |
Alibava run |
run_008_HPK_CB_250V.dat |
Telescope run |
613 |
Triggers |
150000 |
Alibava trigger eff. |
~62% |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-250 |
Latency |
132 |
Table x |
107 mm |
Table z |
2.6 mm |
Tabe rot |
240 deg
|
|
115
|
Mon Apr 17 22:27:44 2017 |
Axel Koenig | ALiBaVa (IFX) | run_009_HPK_CB_200V.dat | 009 | 2400000 | 107, 2.6, 240 | No | Unknown |
Remarks: Telescope run 614. Long run over night. run crashed over night!
Telescope chiller turned off (could be the root cause of runs crashing).
Parameter |
Value |
Alibava run |
run_009_HPK_CB_200V.dat |
Telescope run |
614 |
Triggers |
2400000 |
Alibava trigger eff. |
~62% |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-200 |
Latency |
132 |
Table x |
107 mm |
Table z |
2.6 mm |
Tabe rot |
240 deg
|
|
119
|
Tue Apr 18 08:07:15 2017 |
Axel Koenig | ALiBaVa (IFX) | run_010_HPK_CB_200V.dat | 010 | 350000 | 107, 2.6, 240 | No | Unknown |
Remarks: Telescope run 616. Short run central position. Repeated due to crash over night.
No beam during run.Only ~10000 events.
Telescope chiller turned off (could be the root cause of runs crashing).
Parameter |
Value |
Alibava run |
run_010_HPK_CB_200V.dat |
Telescope run |
616 |
Triggers |
350000 |
Alibava trigger eff. |
~62% |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-200 |
Latency |
132 |
Table x |
107 mm |
Table z |
2.6 mm |
Tabe rot |
240 deg
|
|
118
|
Tue Apr 18 08:04:39 2017 |
Axel Koenig | ALiBaVa (IFX) | run_004_HPK_CB_200V.dat | 004 | 250000 | 107, 2.6, 250 | No | Unknown |
Remarks: Telescope run 607. Angle scan.
Telescope chiller turned off (could be the root cause of runs crashing).
Parameter |
Value |
Alibava run |
run_004_HPK_CB_200V.dat |
Telescope run |
607 |
Triggers |
250000 |
Alibava trigger eff. |
~62% |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-200 |
Latency |
132 |
Table x |
107 mm |
Table z |
2.6 mm |
Tabe rot |
250 deg
|
|
106
|
Mon Apr 17 16:36:56 2017 |
Axel Koenig | ALiBaVa (IFX) | run_002_HPK_CB_200V.dat | 002 | 450000 | 107, 5.8, 240 | No | Unknown |
Remarks: Telescope run 605. Bias scan 200 V. Only one chip.
Telescope chiller turned off (could be the root cause of runs crashing).
Parameter |
Value |
Alibava run |
run_002_HPK_CB_200V.dat |
Telescope run |
605 |
Triggers |
450000 |
Alibava trigger eff. |
~64% |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-200 |
Latency |
132 |
Table x |
107 mm |
Table z |
5.8 mm |
Tabe rot |
240 deg
|
|
68
|
Fri Apr 14 14:49:09 2017 |
Axel Koenig | ALiBaVa (IFX) | run_023_IFX_CB_100V.dat | 023 | 385000 | 109, 0, 240 | No | Crap |
Remarks: Telescope run 546. Bias voltage 100 V. Beam position on only one chip. Run destroyed. Intended trigger count 500k
Parameter |
Value |
Alibava run |
run_023_IFX_CB_100V.dat |
Telescope run |
546 |
Triggers |
385000 |
Alibava trigger eff |
~62 % |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-100 V |
Latency |
132 |
Table x |
109 mm |
table z |
0 mm |
table rot |
240 deg (parallel to tele. planes) |
|
4
|
Wed Apr 12 22:34:18 2017 |
Axel König | ALiBaVa (IFX) | run_002_IFX_CB_100V.dat | 002 | 100000 | 109, 0.5, 240 | No | Unknown |
Remarks: Beam position on only one chip.
Parameter |
Value |
Alibava run |
run_002_IFX_CB_100V.dat |
Telescope run |
498 |
triggers |
100000 |
Alibava trigger eff. |
? |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-100 V |
Latency |
132 |
Table x |
109 mm |
Table z |
0.5 mm |
Table rot |
240 deg (parallel to tele. planes) |
|
65
|
Fri Apr 14 13:06:30 2017 |
Axel Koenig | ALiBaVa (IFX) | run_022_IFX_CB_100V.dat | 022 | 500000 | 109, 10.5, 240 | No | Unknown |
Remarks: Telescope run 545. Bias voltage 100 V. Beam position on only one chip.
Parameter |
Value |
Alibava run |
run_022_IFX_CB_100V.dat |
Telescope run |
545 |
Triggers |
500000 |
Alibava trigger eff |
~62 % |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-100 V |
Latency |
132 |
Table x |
109 mm |
table z |
10.5 mm |
table rot |
240 deg (parallel to tele. planes) |
|
70
|
Fri Apr 14 18:22:58 2017 |
Axel Koenig | ALiBaVa (IFX) | run_026_IFX_CB_100V.dat | 026 | 350000 | 109, 4.5, 230 | No | Unknown |
Remarks: Telescope run 549. Angle shift 230 deg.
Parameter |
Value |
Alibava run |
run_026_IFX_CB_100V.dat |
Telescope run |
549 |
Triggers |
350000 |
Alibava trigger eff. |
~59.8% |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-100 |
Latency |
132 |
Table x |
109 mm |
Table z |
4.5 mm |
Tabe rot |
230 deg
|
|
1
|
Wed Apr 12 21:06:44 2017 |
Axel König | ALiBaVa (IFX) | run_001_IFX_CB_100V.dat | 001 | 200000 | 109, 4.5, 240 | No | Unknown |
First run for alibava modules with Sensor IFX_CB.
Beam energy = 5.6 GeV
Telescope threshold = 6
Telescope data file = 496
Bias voltage = -100V
Telescope synchornisation: yes
Position table (is center of sensor):
x=109mm
y=4.5mm
phi=240 deg
|
Attachment 1: 11.png
|
|
Attachment 2: 46.png
|
|
8
|
Thu Apr 13 09:23:20 2017 |
Axel König | ALiBaVa (IFX) | run_005_IFX_CB_10V.dat | 005 | 285000 | 109, 4.5, 240 | No | Unknown |
Remarks: Telescope run 508. Bias scan 10 V.
Parameter |
Value |
Alibava run |
run_005_IFX_CB_10V.dat |
Telescope run |
508 |
Triggers |
285000 |
Alibava trigger eff. |
~36.5% |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-10 |
Latency |
132 |
Table x |
109 mm |
Table z |
4.5 mm |
Tabe rot |
240 deg (parallel to tele. planes)
|
|
9
|
Thu Apr 13 10:03:29 2017 |
Axel König | ALiBaVa (IFX) | run_001_IFX_CB_100V.dat | 001 | 200000 | 109, 4.5, 240 | No | Unknown |
Remarks: Telescope run 496. Bias voltage 100 V. First run at DESY TB.
Parameter |
Value |
Alibava run |
run_001_IFX_CB_100V.dat |
Telescope run |
496 |
Triggers |
200000 |
Alibava trigger eff. |
? |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-100 |
Latency |
132 |
Table x |
109 mm |
Table z |
4.5 mm |
Tabe rot |
240 deg (parallel to tele. planes)
|
|
10
|
Thu Apr 13 10:58:57 2017 |
Axel Koenig | ALiBaVa (IFX) | run_006_IFX_CB_20V.dat | 006 | 185000 | 109, 4.5, 240 | No | Unknown |
Remarks: Telescope run 510. Bias scan 20 V.
Parameter |
Value |
Alibava run |
run_006_IFX_CB_10V.dat |
Telescope run |
510 |
Triggers |
185000 |
Alibava trigger eff. |
~55.6% |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-20 |
Latency |
132 |
Table x |
109 mm |
Table z |
4.5 mm |
Tabe rot |
240 deg (parallel to tele. planes)
|
|
11
|
Thu Apr 13 11:25:32 2017 |
Axel Koenig | ALiBaVa (IFX) | run_007_IFX_CB_30V.dat | 007 | 170000 | 109, 4.5, 240 | No | Unknown |
Remarks: Telescope run 513. Bias scan 30 V.
Parameter |
Value |
Alibava run |
run_007_IFX_CB_30V.dat |
Telescope run |
513 |
Triggers |
170000 |
Alibava trigger eff. |
~60% |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-30 |
Latency |
132 |
Table x |
109 mm |
Table z |
4.5 mm |
Tabe rot |
240 deg (parallel to tele. planes)
|
|
12
|
Thu Apr 13 12:00:40 2017 |
Axel Koenig | ALiBaVa (IFX) | run_008_IFX_CB_40V.dat | 008 | 175000 | 109, 4.5, 240 | No | Unknown |
Remarks: Telescope run 515. Bias scan 40 V.
Parameter |
Value |
Alibava run |
run_008_IFX_CB_40V.dat |
Telescope run |
515 |
Triggers |
175000 |
Alibava trigger eff. |
~58% |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-40 |
Latency |
132 |
Table x |
109 mm |
Table z |
4.5 mm |
Tabe rot |
240 deg (parallel to tele. planes)
|
|
13
|
Thu Apr 13 12:33:20 2017 |
Axel Koenig | ALiBaVa (IFX) | run_009_IFX_CB_50V.dat | 009 | 175000 | 109, 4.5, 240 | No | Unknown |
Remarks: Telescope run 517. Bias scan 50 V.
Parameter |
Value |
Alibava run |
run_009_IFX_CB_50V.dat |
Telescope run |
517 |
Triggers |
175000 |
Alibava trigger eff. |
~58% |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-50 |
Latency |
132 |
Table x |
109 mm |
Table z |
4.5 mm |
Tabe rot |
240 deg (parallel to tele. planes)
|
|