ID |
Date |
Author |
Project |
Subject |
Run Number |
Events |
StartTime |
EndTime |
Data |
42
|
Sat Oct 20 19:35:30 2012 |
Markus Friedl | MicronOrigami_gross | run | run020 | 300000 | 20.10.2012 19:34:19 | 21.10.2012 08:55:28 | Good | Belle Stack 4: 2 DSSD -z, Origami #3, #1, wedge new #2
HV = +-40V, Ibias=73A
Position: APVs #2, 7, 11, 15 | # 17/18, 21, 25, 29
= p: 2DSSD #2, Origami3 #3, Origami1 #3, wedgenew #3
= n: 2DSSD #1/2, Origami3 #1, Origami1 #1, wedgenew #1
|
70
|
Mon Oct 29 04:02:16 2012 |
Manfred Valentan | MicronBaby | run | run036 | 296294 | 29.10.2012 03:53:33 | 29.10.2012 11:05:35 | Unknown | Micron Baby stack 2 after irradiation (intermediate strips)
HV = -100 V
I = 36 uA
DESY table shifted by 2.5mm (beam down) after beam profile was changed (to get higher intensity)
beam hitting higher APVs
Stopped when long beam off period was announced |
51
|
Mon Oct 22 19:19:53 2012 |
Markus Friedl | MicronOrigami_gross | run | run024 | 260000 | 22.10.2012 19:18:16 | 23.10.2012 08:17:20 | Good | Belle Stack 4: 2 DSSD -z, Origami #3, #1, wedge new #2
HV = +-20V, Ibias=57uA
Position: APVs #2, 7, 11, 15 | # 17/18, 21, 25, 29
= p: 2DSSD #2, Origami3 #3, Origami1 #3, wedgenew #3
= n: 2DSSD #1/2, Origami3 #1, Origami1 #1, wedgenew #1
Beam shape on n side (horizontally) different than in previous runs. Number of triggers per spill is reduced to ~60k (was 200..250k before). Reason unknown. |
7
|
Sat Oct 13 06:55:47 2012 |
Manfred Valentan | MicronBaby | run | run002 | 234019 | 13.10.2012 06:50:00 | 13.10.2012 13:50:40 | Unknown | Micron baby stack
HV = -100V I=5.4uA
beam on
Final DESY position (center of vertical beam profile)
Module position: 0 / -13.0 -- hitting the lower half of the sensor (centered on even APVs)
Run interrupted due to RF transmitter fault. |
56
|
Tue Oct 23 23:12:03 2012 |
Thomas Bergauer | MicronOrigami_gross | run | run027 | 220847 | 23.10.2012 23:10:14 | 24.10.2012 08:19:51 | Unknown | Belle Stack 4: 2 DSSD -z, Origami #3, #1, wedge new #2
HV = +-5V, Ibias=23.5uA
Position: APVs #2, 7, 11, 15 | # 17/18, 21, 25, 29
= p: 2DSSD #2, Origami3 #3, Origami1 #3, wedgenew #3
= n: 2DSSD #1/2, Origami3 #1, Origami1 #1, wedgenew #1
SPS Status during whole run: Still problems with North Area steering. means that beam position was not stable. |
27
|
Tue Oct 16 11:23:43 2012 |
Christian Irmler | MicronOrigami_gross | run | run012 | 20k | 16.10.2012 11:16:01 | 16.10.2012 12:27:03 | Unknown | Origami Stack 2a: Origami #4, Wedge new1
reading out all 6 APV of p-side
HV = +-40V, Ibias=10.4µA
Origami 3 and Wedge old are still installed in the box, but not read out --> disconnected
beam:
p-side: centered between APV 4&5
n-side: centered between APV 2&3
--> upper left corner of Origami and wedge module --> we hit the APV chips and / or the PA0 of the Origami module
no cooling
Temperatur sensors:
- T11: Origami #4 in
- T12: Origami #4 out
- T13: Origami #3 in (wrong values!)
- T14: Origami #3 out
- T15: Wedge new 1 out
- T16: Wedge old out
- T17: Wedge new 1 cooling block
- T18: Wedge old cooling block
|
38
|
Fri Oct 19 00:56:16 2012 |
Markus Friedl | MicronOrigami_gross | run | run017 | 197899 | 19.10.2012 00:54:21 | 19.10.2012 14:10:02 | Good | Belle Stack 4: 2 DSSD -z, Origami #3, #1, wedge new #2
HV = +-50V, Ibias=76A
Position: APVs #2, 7, 11, 15 | # 17/18, 21, 25, 29
= p: 2DSSD #2, Origami3 #3, Origami1 #3, wedgenew #3
= n: 2DSSD #1/2, Origami3 #1, Origami1 #1, wedgenew #1
Same as run016, but timing shifted by 1 CLK to better center the peak of the shaped wafeforms for subsequent fitting during analysis.
Beam was turned off around here:
19.10.2012 08:56:12 event=193394
So better ignore the last ~5000 events for analysis.
|
49
|
Mon Oct 22 00:57:05 2012 |
Markus Friedl | MicronOrigami_gross | run | run023 | 176629 | 22.10.2012 00:55:59 | 22.10.2012 10:04:38 | Good | Belle Stack 4: 2 DSSD -z, Origami #3, #1, wedge new #2
HV = +-25V, Ibias=65uA
Position: APVs #2, 7, 11, 15 | # 17/18, 21, 25, 29
= p: 2DSSD #2, Origami3 #3, Origami1 #3, wedgenew #3
= n: 2DSSD #1/2, Origami3 #1, Origami1 #1, wedgenew #1
No more events since 22.10.2012 07:09:53 event=176629 - spill signals removed?
|
24
|
Mon Oct 15 23:26:52 2012 |
Marko Dragicevic | MicronOrigami_gross | run ATTENTION: wrong bias voltage | run010 | 158k | 15.10.2012 23:10:40 | 16.10.2012 05:05:35 | Crap | ATTENTION: Bias voltage was +-100 V instead of +- 50 V
Same as run009.
Origami Stack 2: Origami #4, #3, Wedge new1, old
HV = +-10V
Ibias=12.3µA
beam position:
vertical: between APV 1&2 (of 6)
horizontal: centered on APV#1 -> right side of sensor -> small side of wedge sensors
Position of xy table:
x = 23, y = -25
Temperatures:
Origami #4: In=-33 degC / Out=-34 degC
Origami #3: Out=-35 degC (temperatur sensor of inlet is not working)
Wedge new 1: Out = -17 degC, cooling block = -14 degC
Wedge old: Out = -16 degC, cooling block = -14 degC
15.10.2012 22:08:31 event=0 rate=0events/s temp/curr=31.3/089.2 31.3/-16.2 31.2/-17.9 31.3/-18.2 29.9/003.6 29.9/003.1 30.0/-01.1 29.9/-01.6
|
47
|
Sun Oct 21 16:04:06 2012 |
Markus Friedl | MicronOrigami_gross | run | run022 | 151282 | 21.10.2012 18:16:53 | 22.10.2012 00:04:42 | Good | Belle Stack 4: 2 DSSD -z, Origami #3, #1, wedge new #2
HV = +-30V, Ibias=70uA
Position: APVs #2, 7, 11, 15 | # 17/18, 21, 25, 29
= p: 2DSSD #2, Origami3 #3, Origami1 #3, wedgenew #3
= n: 2DSSD #1/2, Origami3 #1, Origami1 #1, wedgenew #1 |
25
|
Tue Oct 16 05:24:21 2012 |
Christian Irmler | MicronOrigami_gross | run ATTENTION: wrong bias voltage | run011 | 150k | 16.10.2012 05:19:28 | 16.10.2012 09:59:24 | Crap | ATTENTION: Bias voltage was +-100 V instead of +- 50 V
Origami Stack 2: Origami #4, #3, Wedge new1, old
HV = +-100V
Ibias=12.3µA
beam position:
vertical: between APV 1&2 (of 6)
horizontal: centered on APV#3 -> left side of sensor -> wide side of wedge sensors
Position of xy table:
x = 83, y = -25
Temperatures:
Origami #4: In=-33 degC / Out=-34 degC
Origami #3: Out=-35 degC (temperatur sensor of inlet is not working)
Wedge new 1: Out = -17 degC, cooling block = -16 degC
Wedge old: Out = -15 degC, cooling block = -15 degC
Stopped after 132658 events, because of empty gas bottle and thus insufficient cooling. |
28
|
Tue Oct 16 13:59:46 2012 |
Christian Irmler | MicronOrigami_gross | run | run013 | 150k | 16.10.2012 13:43:05 | 16.10.2012 20:21:04 | Unknown | Origami Stack 2: Origami #4, #3, Wedge new1, old
HV = +-50V
Ibias=5.2µA
beam position:
vertical: between APV 1&2 (of 6)
horizontal: centered on APV#3 -> left side of sensor -> wide side of wedge sensors
Position of xy table:
x = 83, y = -31
Temperatur sensors:
- T11: Origami #4 in
- T12: Origami #4 out
- T13: Origami #3 in (wrong values!)
- T14: Origami #3 out
- T15: Wedge new 1 out
- T16: Wedge old out
- T17: Wedge new 1 cooling block
- T18: Wedge old cooling block
|
31
|
Tue Oct 16 23:35:28 2012 |
Christian Irmler | MicronOrigami_gross | run | run014 | 150k | 16.10.2012 23:32:24 | 17.10.2012 04:59:02 | Unknown | Same as run011, but now with +- 50V bias voltage.
Origami Stack 2: Origami #4, #3, Wedge new1, old
HV = +-10V
Ibias=4.83µA
beam position:
vertical: between APV 1&2 (of 6)
horizontal: centered on APV#1 -> right side of sensor -> small side of wedge sensors
Position of xy table:
x = 23, y = -31
Temperatur sensors:
- T11: Origami #4 in
- T12: Origami #4 out
- T13: Origami #3 in (wrong values!)
- T14: Origami #3 out
- T15: Wedge new 1 out
- T16: Wedge old out
- T17: Wedge new 1 cooling block
- T18: Wedge old cooling block
|
32
|
Wed Oct 17 01:21:50 2012 |
Marko Dragicevic | Infineon | run | run009 | 150k | 15.10.2012 08:20:15 | 15.10.2012 11:01:11 | Unknown | This run was not recorded in original sequence in the ELOG!!!
Infineon Stack Version 2 (no height information, BabyStrips vertical Strixel vertical)
HV = 300V I=4uA
beam on
Final DESY position (center of vertical beam profile)
Module position: 15 / 3 -- hitting lower part of baby and bottom strixels
07_baby |
0,0 |
APV1 lower, APV2 upper |
Strixel 2 |
20,8 |
Hybrid1/APV1 top-left, Hybrid1/APV2 top right, Hybrid2/APV1 bottom left, Hybrid2/APV2 bottom right |
Strixel 1 |
39,4 |
Hybrid1/APV1 top-left, Hybrid1/APV2 top right, Hybrid2/APV1 bottom left, Hybrid2/APV2 bottom right |
12_baby |
60,6 |
APV1 lower, APV2 upper |
|
|
64
|
Sat Oct 27 21:49:08 2012 |
Markus Friedl | MicronOrigami_gross | run | run031 | 150k | 27.10.2012 21:42:46 | 28.10.2012 02:35:55 (+1h due to winter time) | Good | Belle stack 2 after gamma irrdiation
HV = +50V -50 V
I = 102 uA
CO2 cooling
|
65
|
Sun Oct 28 02:51:49 2012 |
Markus Friedl | MicronOrigami_gross | run | run032 | 150k | 28.10.2012 02:47:24 | 28.10.2012 08:31:19 | Good | Belle stack 2 after gamma irradiation
HV = +50V -50V
I = 101 uA
Bias currents absolutely stable during the whole run.
CO2 cooling
hitting APV3 (n side) = wide side of wedge sensor |
76
|
Tue Oct 30 11:05:32 2012 |
Markus Friedl | HPK_baby | run | run039 | 150k | 30.10.2012 11:03:19 | 30.10.2012 14:57:57 | Unknown | HPK baby stack after irradiation - no cooling
I = 420A and growing (warming up)
positioned centered on p and n
|
37
|
Fri Oct 19 00:45:52 2012 |
Markus Friedl | MicronOrigami_gross | run | run016 | 11855 | 19.10.2012 00:24:05 | 19.10.2012 00:48:24 | Crap |
Belle Stack 4: 2 DSSD -z, Origami #3, #1, wedge new #2
HV = +-50V, Ibias=77A
Position: APVs #2, 7, 11, 15 | # 17/18, 21, 25, 29
= p: 2DSSD #2, Origami3 #3, Origami1 #3, wedgenew #3
= n: 2DSSD #1/2, Origami3 #1, Origami1 #1, wedgenew #1
Stack is now positioned on globally best performing position which should also enable reasonable tracking.
Trigger timing not perfect, needs adjustment (-1 CLK) -> new run |
17
|
Sun Oct 14 20:12:14 2012 |
Marko Dragicevic | Infineon | run | run007 | 110k | 14.10.2012 20:07:12 | 14.10.2012 21:04:12 | Unknown | Infineon Stack Version 1 (with height information, BabyStrips horizontal Strixel vertical)
HV = 300V I=3uA
beam on
Final DESY position (center of vertical beam profile)
Module position: 0 / 0 -- hitting the center of babies, between strixels
07_baby |
0,0 |
APV1 lower, APV2 upper |
Strixel 2 |
20,8 |
Hybrid1/APV1 top-left, Hybrid1/APV2 top right, Hybrid2/APV1 bottom left, Hybrid2/APV2 bottom right |
Strixel 1 |
39,4 |
Hybrid1/APV1 top-left, Hybrid1/APV2 top right, Hybrid2/APV1 bottom left, Hybrid2/APV2 bottom right |
12_baby |
60,6 |
APV1 lower, APV2 upper |
|
55
|
Tue Oct 23 16:06:09 2012 |
Markus Friedl | MicronOrigami_gross | run | run026 | 102124 | 23.10.2012 16:04:47 | 23.10.2012 23:04:56 | Unknown | Belle Stack 4: 2 DSSD -z, Origami #3, #1, wedge new #2
HV = +-10V, Ibias=36uA
Position: APVs #2, 7, 11, 15 | # 17/18, 21, 25, 29
= p: 2DSSD #2, Origami3 #3, Origami1 #3, wedgenew #3
= n: 2DSSD #1/2, Origami3 #1, Origami1 #1, wedgenew #1
SPS Status during whole run: Still problems with North Area steering. means that beam position was not stable. |
|
|