| ID | Date | Author | Project | Subject | Run Number | Events | StartTime | EndTime | Data  | Text |  | 
| 29 | Tue Oct  4 15:16:45 2011 | Manfred Valentan | Micron | Micron-Baby-Stack p-spray, run bottom right NOT USABLE | run004 | 200k | 4.10.2011 15:28:48 | 04.10.2011  19:36:50 | Crap | HAS TO BE REPEATED!!! Before this run there were changes in the trigger line (vetoing
 the trigger when no spill), and thus the
 |  | 
| 31 | Tue Oct  4 22:11:12 2011 | Christian Irmler | BELLE | Origami stack | run005 | 330000 | 04.10.2011  22:09:37 | 05.10.2011  09:11:54 | Crap | Origami6 HV=+-40V, no cooling
 beam on
 |  | 
| 32 | Wed Oct  5 12:09:00 2011 | Christian Irmler | BELLE | Origami stack cooled wrong I2C settings | cal005 | 83800 | 05.10.2011  12:04:50 | 05.10.2011  12:37:11 | Crap | This calibration file can not be used, due to wrong I2C settings!
 Origami6
 |  | 
| 33 | Wed Oct  5 13:00:41 2011 | Christian Irmler | BELLE | Origami stack cooled | run006 | 24692 | 05.10.2011  12:55:53 | 05.10.2011  14:05:49 | Crap | Run stopped after 24692 events, because all n-side APV failed due to a shortcut
 of a capacitor (C223) on n-side rebo
 |  | 
| 37 | Wed Oct  5 19:57:15 2011 | Christian Irmler | BELLE | Origami stack cooled | run007 | 17845 | 05.10.2011  18:27:51 | 05.10.2011  19:31:43 | Crap | Run stopped after 17845 events, because of bad trigger timing
 
 |  | 
| 50 | Sat Oct  8 14:33:25 2011 | Thomas Bergauer | HPK | baby_std | run016 | 250k | 08.10.2011  14:31:40 |  | Crap | falsches config file (single statt multi6)
 
 |  | 
| 51 | Sat Oct  8 16:35:04 2011 | Thomas Bergauer | HPK | baby_std | run017 |  | 08.10.2011  16:26:10 |  | Crap | wrong timing 
 no cooling
 |  | 
| 77 | Wed Feb 22 13:37:11 2012 | Christian Irmler | BELLE | Origami Module #4 | run103 | 100k | 22.02.2012  13:35:08 | 22.02.2012  13:54:08 | Crap | Sr90, large scintillator, HVret @ Origami closed
 HVret @ SMU connected
 |  | 
| 8 | Fri Sep 16 11:01:15 2011 | Manfred Valentan | Micron | Micron Baby Modules |  |  |  |  |  | Irradiated modules from last year: COMB-IRRAD: B2Micron_2826_19_Baby4:
 combined p-stop, intermediate strips (was
 |  | 
| 9 | Mon Sep 19 17:38:34 2011 | Manfred Valentan | Micron | Module ATOLL-2 |  |  |  |  |  | Sensor: B2Micron_2826_17_Baby3 p-stop: Atoll, with intermediate
 strips
 |  | 
| 10 | Mon Sep 19 17:39:09 2011 | Manfred Valentan | Micron | Module COMM-2 |  |  |  |  |  | Sensor: B2Micron_2826_17_Baby2 p-stop: Common, with intermediate
 strips
 |  | 
| 11 | Mon Sep 19 17:39:55 2011 | Manfred Valentan | Micron | Module COMB-2 |  |  |  |  |  | Sensor: B2Micron_2826_17_Baby4 p-stop: Combined with intermediate
 strips
 |  | 
| 12 | Mon Sep 19 17:40:47 2011 | Manfred Valentan | Micron | Module SPRAY-2 |  |  |  |  |  | Sensor: B2Micron_2825_05_Baby4 p-stop: p-spray WITHOUT intermediate
 strips
 |  | 
| 13 | Mon Sep 19 17:41:27 2011 | Manfred Valentan | Micron | Module SPRAY-I-2 |  |  |  |  |  | Sensor: B2Micron_2825_15_Baby3 p-stop: p-spray WITH intermediate
 strips
 |  | 
| 14 | Mon Sep 19 17:42:01 2011 | Manfred Valentan | Micron | Module SPRAY-1 |  |  |  |  |  | Sensor: B2Micron_2825_05_Baby3 p-stop: p-spray WITHOUT intermediate
 strips
 |  | 
| 15 | Mon Sep 19 17:42:39 2011 | Manfred Valentan | Micron | Module SPRAY-I-1 |  |  |  |  |  | Sensor: B2Micron_2825_15_Baby2 p-stop: p-spray WITH intermediate
 strips
 |  | 
| 16 | Mon Sep 19 17:44:00 2011 | Manfred Valentan | Micron | Module COMM-1 |  |  |  |  |  | Sensor: B2Micron_2826_19_Baby2 (reused from last year, was TRK7)
 p-stop: Common, with intermediate
 |  | 
| 17 | Mon Sep 19 17:44:39 2011 | Manfred Valentan | Micron | Module COMB-1 |  |  |  |  |  | Sensor: B2Micron_2826_01_Baby4 (reused from last year, was TRK6)
 p-stop: Combined, with intermediate
 |  | 
| 18 | Mon Sep 19 17:46:07 2011 | Manfred Valentan | Micron | Module ATOLL-1 |  |  |  |  |  | Sensor: B2Micron_2826_07_Baby3 (reused from last year, but was not used in beam
 test SPS2010)
 |  | 
| 19 | Mon Sep 19 17:47:11 2011 | Manfred Valentan | Micron | Module COMB-IRRAD |  |  |  |  |  | Sensor: B2Micron_2826_19_Baby4 (reused from last year, was DUT4, module is irradiated
 to 700 kGy)
 |  |