ID |
Date |
Author |
Project |
Subject |
Run Number |
Events |
StartTime |
EndTime |
Data |
48
|
Sat Oct 8 00:14:37 2011 |
Christian Irmler | BELLE | Origami stack | run015 | 180111 | 08.10.2011 00:06:41 | 08.10.2011 08:24:38 | Good | Origami6
HV=+-40V,
no cooling
beam on; xy table position x=306 y=74.5
Istart=57µA;
std. I2C settings
centered
currents:
8.10.2011 00:07:43 event=991 rate=0events/s temp/curr=33.0/-15.4 33.0/-17.3 33.0/-17.7 31.9/011.5 31.9/013.2 32.0/040.20
...
08.10.2011 06:32:34 event=179710 rate=0events/s temp/curr=33.1/-22.1 33.0/-25.6 33.1/-22.6 31.9/018.2 31.8/021.7 31.9/048.5
08.10.2011 06:33:28 event=180105 ***WARNING*** Error detected in APV 0, subevent 2. Event discarded, reset sent.
Permanent APV errors at several (all p-side, some n-side) APV chips after 180105 events, which could not be recovered by software resets. After restart of run, all APVs where alive again, hence hardware reset and I2C reconfiguration after 5 or 10 consecutive APV errors should be added to TUXDAQ software.
After APV errors occured, the cummulative bias current decrised from 78µA to 54µA. This effect can occures whenever the digital part of the APV is switched between idle and full operation. It seems that this is a thermal effect, caused by the slightly higher power consumption of the digital part of the chips.
|
50
|
Sat Oct 8 14:33:25 2011 |
Thomas Bergauer | HPK | baby_std | run016 | 250k | 08.10.2011 14:31:40 | | Crap | falsches config file (single statt multi6)
no cooling
beam on; xy table position x=322 y=80
HV=200V Istart_PY=7.6uA; Istart_N=0.42
std. I2C settings |
51
|
Sat Oct 8 16:35:04 2011 |
Thomas Bergauer | HPK | baby_std | run017 | | 08.10.2011 16:26:10 | | Crap | wrong timing
no cooling
beam on; xy table position x=322 y=80
HV=200V Istart_PY=7.6uA; Istart_N=0.42
std. I2C settings |
52
|
Sat Oct 8 16:45:14 2011 |
Thomas Bergauer | HPK | baby_std | run018 | 150000 | 08.10.2011 16:44:13 | 08.10.2011 19:33:33 | Good | no cooling
beam on; xy table position x=322 y=80
HV=200V Istart_PY=8.2uA; Istart_N=0.45uA
std. I2C settings, now with correct timing |
53
|
Sat Oct 8 20:33:58 2011 |
Karl-Heinz Hoffmann | HPK | baby_std | run019 | 150000 | 08.10.2011 20:22:45 | 08.10.2011 23:19:23 | Good | no cooling
beam on; xy table position x=333 y=80
HV=200V Istart_PY=8.2uA; Istart_N=0.45uA
std. I2C settings |
54
|
Sat Oct 8 23:27:44 2011 |
Karl-Heinz Hoffmann | HPK | baby_std | run020 | 382000 | 08.10.2011 23:24:16 | 09.10.2011 09:24:20 | Good | no cooling
beam on; xy table position x=327,5 y=65
HV=200V Istart_PY=8.2uA; Istart_N=0.45uA
std. I2C settings |
57
|
Sun Oct 9 11:00:33 2011 |
Karl-Heinz Hoffmann | HPK | baby_add | run021 | 150000 | 09.10.2011 10:56:47 | 09.10.2011 12:47:23 | Good | no cooling
beam on; xy table position x=330 y=74
HV=200V Istart_PY=1,9uA; Istart_N=0.27uA
std. I2C settings |
60
|
Sun Oct 9 13:12:32 2011 |
Karl-Heinz Hoffmann | HPK | baby_add | run022 | 670k | 09.10.2011 13:09:32 | 10.10.2011 00:48:16 | Good | no cooling
beam on; xy table position x=330 y=64
HV=200V Istart_PY=1,9uA; Istart_N=0.27uA
std. I2C settings |
61
|
Mon Oct 10 00:50:23 2011 |
Christian Irmler | HPK | baby_add | run023 | 680k | 10.10.2011 00:51:26 | 10.10.2011 09:15:35 | Good | no cooling
beam on; xy table position x=330 y=74
HV=200V Istart_PY=1,9uA; Istart_N=0.27uA
std. I2C settings
same as run021, but taking more events |
63
|
Mon Oct 10 12:15:52 2011 |
Manfred Valentan | CNM | CNM stack | run024 | 110000 | 10.10.2011 12:03:58 | 10.10.2011 14:05:56 | Good | no cooling
beam on; xy table position x=320.5 y=79.5
HV_CNM=20V, HV_BSTD1-N=200V
Istart_CNM= 28 / 1 uA; Istart_BSTD= 0.12 uA |
66
|
Mon Oct 10 14:11:53 2011 |
Christian Irmler | CNM | CNM stack | run025 | 110000 | 10.10.2011 14:09:00 | 10.10.2011 16:11:57 | Good | no cooling
beam on; xy table position x=335.5 y=79.5
HV_CNM=20V, HV_BSTD1-N=200V
Istart_CNM= 31 / 1.4 uA; Istart_BSTD= 0.2 uA |
67
|
Mon Oct 10 16:26:20 2011 |
Marko Dragicevic | CNM | CNM stack | run026 | 300000 | 10.10.2011 16:23:46 | 10.10.2011 20:30:00 | Good | no cooling
beam on; xy table position x=328 y=69.5
HV_CNM=20V, HV_BSTD1-N=200V
Istart_CNM= 27 / 1.5 uA; Istart_BSTD= 0.2 uA |
68
|
Mon Oct 10 21:54:24 2011 |
Marko Dragicevic | Micron | Micron-Baby-Stack p-spray, run bottom right | run027 | 700k | 10.10.2011 22:40:31 | 11.10.2011 09:49:11 | Unknown | hit modules in bottom left corner -> Zones narrow and half-narrow for COMB_IRRAD, Zones wide and half-wide for COMM_IRRAD
timing corrected w.r.t. run004
Position of HEPHY x-z-table:
x = 322
z = 70
Note: z-Position not identical with position of run004, x-z-table seemed to have moved vertically. This configuration now hits the same sensor position. |
72
|
Mon Feb 20 17:17:27 2012 |
Christian Irmler | BELLE | Origami Module #4 | run100 | 50k | 20.02.2012 17:07:21 | 20.02.2012 17:17:22 | Good | Sr90, large scintillator, HVret @ Origami open
Origami Mod. #4
HV=+-40V, Ibias=2,03µA
CO2 cooling on, Tin=-16°C, Tout=-16,7°C
standard I²C settings for uncooled APV chips.
Lab measurement with new dark box but old, large scintillator.
20.02.2012 17:08:27 event=5771 ***WARNING*** Error detected in APV 6, subevent 0. Event discarded, reset sent. --> APV 6 recovered afterwards |
74
|
Mon Feb 20 17:27:02 2012 |
Christian Irmler | BELLE | Origami Module #4 | run101 | 24497 | 20.02.2012 17:33:33 | 20.02.2012 17:38:21 | Good | Sr90, large scintillator, HVret @ Origami open
Origami Mod. #4
HV=+-40V, Ibias=2,03µA
CO2 cooling on, Tin=-16°C, Tout=-16,7°C
I²C settings for cooled APV chips --> same as used in run010.
Lab measurement with new dark box but old, large scintillator.
CO2 empty after 24497 events --> run stopped |
75
|
Wed Feb 22 12:15:11 2012 |
Christian Irmler | BELLE | Origami Module #4 | run102 | 170k | 22.02.2012 12:31:08 | 22.02.2012 13:03:27 | Good | same as run101
Sr90, large scintillator, HVret @ Origami open
Origami Mod. #4
HV=+-40V, Ibias=1,8µA
CO2 cooling on, Tin=-20.2°C, Tout=-21.3°C
I²C settings for cooled APV chips --> same as used in run010
e:\cvi-projects\apvdaq_fadc\origami6_V2_multi6_cooled.cfg
Lab measurement with new dark box but old, large scintillator.
CO2 empty after 24497 events --> run stopped --> no corresponding calibration available, but for rought analysis cal100 can be used for this run, too |
77
|
Wed Feb 22 13:37:11 2012 |
Christian Irmler | BELLE | Origami Module #4 | run103 | 100k | 22.02.2012 13:35:08 | 22.02.2012 13:54:08 | Crap | Sr90, large scintillator,
HVret @ Origami closed
HVret @ SMU connected
Origami Mod. #4
HV=+-40V, Ibias=1,6µA
CO2 cooling on, Tin=-20.5°C, Tout=-21.5°C
I²C settings for cooled APV chips --> same as used in run010
e:\cvi-projects\apvdaq_fadc\origami6_V2_multi6_cooled.cfg
Lab measurement with new dark box but old, large scintillator
Wires between HV connectors ground and HVret conntact at the mambo were open --> no HVret conntection between module and SMUs!
Moreover the SMU Inv. Nr. 545-078 has either an unstable output or problems with the current measurement. |
78
|
Wed Feb 22 16:21:23 2012 |
Christian Irmler | BELLE | Origami Module #4 | run104 | 100k | 22.02.2012 16:22:34 | 22.02.2012 16:41:26 | Good | Sr90, large scintillator,
Origami Mod. #4
HV=+-40V, Ibias=1,6µA
CO2 cooling on, Tin=-20.5°C, Tout=-21.5°C
I²C settings for cooled APV chips --> same as used in run010
e:\cvi-projects\apvdaq_fadc\origami6_V2_multi6_cooled.cfg
Lab measurement with new dark box but old, large scintillator
SMU: +HV=Inv.Nr. 545-77 -HV=Inv.Nr. 540-702
HVret configuration:
0 ... switch open / not connected
1 ... switch closed / connected
SMU |
Dock Box |
J5 |
J6 |
J7 |
0 |
1 |
1 |
0 |
1 |
J7... HV ret @ hybrid
APV |
0 |
1 |
2 |
3 |
4 |
5 |
6 |
7 |
avg noise |
2.564 |
2.124 |
2.094 |
2.200 |
1.696 |
1.666 |
1.870 |
1.888 |
CM noise |
1.7 |
0.9 |
0.9 |
1.3 |
4.0 |
4.1 |
4.1 |
4.2 |
|
Attachment 1: run104.jpg
| |
Attachment 2: run104_current.xls
|
81
|
Wed Feb 22 17:38:29 2012 |
Christian Irmler | BELLE | Origami Module #4 | run105 | 100k | 22.02.2012 17:39:10 | 22.02.2012 17:58:01 | Good | Sr90, large scintillator,
Origami Mod. #4
HV=+-40V, Ibias=1,6µA
CO2 cooling on, Tin=-20.5°C, Tout=-21.5°C
I²C settings for cooled APV chips --> same as used in run010
e:\cvi-projects\apvdaq_fadc\origami6_V2_multi6_cooled.cfg
Lab measurement with new dark box but old, large scintillator
SMU: +HV=Inv.Nr. 545-77 -HV=Inv.Nr. 540-702
HVret configuration:
0 ... switch open / not connected
1 ... switch closed / connected
SMU |
Dock Box |
J5 |
J6 |
J7 |
0 |
1 |
1 |
0 |
0 |
J7... HV ret @ hybrid
Same as run104, but now J7 open
Again spikes in bias current!
|
Attachment 1: run105.jpg
| |
82
|
Wed Feb 22 18:10:29 2012 |
Christian Irmler | BELLE | Origami Module #4 | run106 | 50k | 22.02.2012 17:39:10 | 22.02.2012 18:23:47 | Good | Sr90, large scintillator,
Origami Mod. #4
HV=+-40V, Ibias=1,6µA
CO2 cooling on, Tin=-20.5°C, Tout=-21.5°C
I²C settings for cooled APV chips --> same as used in run010
e:\cvi-projects\apvdaq_fadc\origami6_V2_multi6_cooled.cfg
Lab measurement with new dark box but old, large scintillator
SMU: +HV=Inv.Nr. 545-77 -HV=Inv.Nr. 540-702
HVret configuration:
0 ... switch open / not connected
1 ... switch closed / connected
SMU |
Dock Box |
J5 |
J6 |
J7 |
0 |
1 |
0 |
0 |
0 |
J7... HV ret @ hybrid
Same as run105, but now J5 and J7 open
|
Attachment 1: run106.jpg
| |
|