22
|
Mon Oct 3 21:01:00 2011 |
Manfred Valentan | Micron | Micron-Baby-Stack p-stop, zones n & h-n | run001 | 200k | 03.10.2011 20:54:22 | 03.10.2011 23:25:47 | Good |
hit modules in top right corner -> Zones narrow and half-narrow for all modules
Position of HEPHY x-z-table:
x = 324
z = 103 |
94
|
Wed Mar 7 16:58:36 2012 |
Christian Irmler | BELLE | Origami Module #4 | cal113 | 100k | 07.03.2012 16:40:04 | 07.03.2012 16:57:08 | Unknown |
Sr90, small (new) scintillator, new trigger unit
Origami Mod. #4
HV=+-40V, Ibias=1,6µA
CO2 cooling on, Tin=-19,6°C, Tout=-21.3°C
I²C settings for cooled APV chips
e:\cvi-projects\apvdaq_fadc\origami6_V2_single_cooled.cfg
SMU: +HV=Inv.Nr. 545-77 -HV=Inv.Nr. 540-702
HVret configuration:
0 ... switch open / not connected
1 ... switch closed / connected
SMU |
Dock Box |
J5 |
J6 |
J7 |
0 |
0 (now isolated) |
1 |
1 |
0 |
J7... HV ret @ hybrid
same configuration as run112 but now the step motors of the xz table are powered. |
92
|
Wed Mar 7 15:44:25 2012 |
Christian Irmler | BELLE | Origami Module #4 | cal112 | 83800 | 07.03.2012 15:42:53 | | Unknown |
Sr90, small (new) scintillator, new trigger unit
Origami Mod. #4
HV=+-40V, Ibias=1,6µA
CO2 cooling on, Tin=-19,6°C, Tout=-21.3°C
I²C settings for cooled APV chips
e:\cvi-projects\apvdaq_fadc\origami6_V2_single_cooled.cfg
SMU: +HV=Inv.Nr. 545-77 -HV=Inv.Nr. 540-702
HVret configuration:
0 ... switch open / not connected
1 ... switch closed / connected
SMU |
Dock Box |
J5 |
J6 |
J7 |
0 |
0 (now isolated) |
1 |
1 |
0 |
J7... HV ret @ hybrid |
89
|
Mon Mar 5 19:06:41 2012 |
Christian Irmler | BELLE | Origami Module #4 | cal110 | 83800 | 05.03.2012 19:05:52 | 05.03.2012 19:17:01 | Unknown |
Sr90, large scintillator,
Origami Mod. #4
HV=+-40V, Ibias=1,9µA
CO2 cooling on, Tin=-20.5°C, Tout=-21.5°C
I²C settings for cooled APV chips
e:\cvi-projects\apvdaq_fadc\origami6_V2_multi6_cooled.cfg
Lab measurement with new dark box but old, large scintillator
SMU: +HV=Inv.Nr. 545-77 -HV=Inv.Nr. 540-702
HVret configuration:
0 ... switch open / not connected
1 ... switch closed / connected
SMU |
Dock Box |
J5 |
J6 |
J7 |
0 |
0 (now isolated) |
1 |
1 |
0 |
J7... HV ret @ hybrid |
87
|
Mon Mar 5 18:08:02 2012 |
Christian Irmler | BELLE | Origami Module #4 | cal109 | 83800 | 05.03.2012 18:07:40 | 05.03.2012 18:18:32 | Unknown |
Sr90, large scintillator,
Origami Mod. #4
HV=+-40V, Ibias=1,9µA
CO2 cooling on, Tin=-20.5°C, Tout=-21.5°C
I²C settings for cooled APV chips
e:\cvi-projects\apvdaq_fadc\origami6_V2_multi6_cooled.cfg
Lab measurement with new dark box but old, large scintillator
SMU: +HV=Inv.Nr. 545-77 -HV=Inv.Nr. 540-702
HVret configuration:
0 ... switch open / not connected
1 ... switch closed / connected
SMU |
Dock Box |
J5 |
J6 |
J7 |
0 |
0 (now isolated) |
1 |
1 |
1 |
J7... HV ret @ hybrid |
85
|
Mon Mar 5 17:35:31 2012 |
Christian Irmler | BELLE | Origami Module #4 | cal108 | 83800 | 05.03.2012 16:47:51 | 05.03.2012 17:05:16 | Unknown |
Sr90, large scintillator,
Origami Mod. #4
HV=+-40V, Ibias=1,9µA
CO2 cooling on, Tin=-20.5°C, Tout=-21.5°C
I²C settings for cooled APV chips
e:\cvi-projects\apvdaq_fadc\origami6_V2_multi6_cooled.cfg
Lab measurement with new dark box but old, large scintillator
SMU: +HV=Inv.Nr. 545-77 -HV=Inv.Nr. 540-702
HVret configuration:
0 ... switch open / not connected
1 ... switch closed / connected
SMU |
Dock Box |
J5 |
J6 |
J7 |
0 |
0 (now isolated) |
1 |
0 |
1 |
J7... HV ret @ hybrid |
Attachment 1: run108-109.jpg
|
|
80
|
Wed Feb 22 17:25:34 2012 |
Christian Irmler | BELLE | Origami Module #4 | cal105 | 83800 | 22.02.2012 17:23:13 | 22.02.2012 17:34:27 | Good |
Sr90, large scintillator,
Origami Mod. #4
HV=+-40V, Ibias=1,6µA
CO2 cooling on, Tin=-20.5°C, Tout=-21.5°C
I²C settings for cooled APV chips --> same as used in run010
e:\cvi-projects\apvdaq_fadc\origami6_V2_single_cooled.cfg
Lab measurement with new dark box but old, large scintillator
SMU: +HV=Inv.Nr. 545-77 -HV=Inv.Nr. 540-702
HVret configuration:
0 ... switch open / not connected
1 ... switch closed / connected
SMU |
Dock Box |
J5 |
J6 |
J7 |
0 |
1 |
1 |
0 |
0 |
J7... HV ret @ hybrid
Same as cal104, but now J7 open
|
79
|
Wed Feb 22 16:49:29 2012 |
Christian Irmler | BELLE | Origami Module #4 | cal104 | 83800 | 22.02.2012 16:51:13 | 22.02.2012 17:02:11 | Good |
Sr90, large scintillator,
Origami Mod. #4
HV=+-40V, Ibias=1,6µA
CO2 cooling on, Tin=-20.5°C, Tout=-21.5°C
I²C settings for cooled APV chips --> same as used in run010
e:\cvi-projects\apvdaq_fadc\origami6_V2_multi6_cooled.cfg
Lab measurement with new dark box but old, large scintillator
SMU: +HV=Inv.Nr. 545-77 -HV=Inv.Nr. 540-702
HVret configuration:
0 ... switch open / not connected
1 ... switch closed / connected
SMU |
Dock Box |
J5 |
J6 |
J7 |
0 |
1 |
1 |
0 |
1 |
J7... HV ret @ hybrid
|
76
|
Wed Feb 22 13:17:34 2012 |
Christian Irmler | BELLE | Origami Module #4 | cal102 | 83800 | 22.02.2012 13:11:30 | 22.02.2012 13:25:18 | Good |
Sr90, large scintillator, HVret @ Origami open
Origami Mod. #4
HV=+-40V, Ibias=1,8µA
CO2 cooling on, Tin=-20.2°C, Tout=-21.3°C
I²C settings for cooled APV chips --> same as used in run010
e:\cvi-projects\apvdaq_fadc\origami6_V2_single_cooled.cfg
Lab measurement with new dark box but old, large scintillator. |
73
|
Mon Feb 20 17:20:20 2012 |
Christian Irmler | BELLE | Origami Module #4 | cal100 | 50k | 20.02.2012 17:19:03 | 20.02.2012 17:29:58 | Good |
Sr90, large scintillator, HVret @ Origami open
Origami Mod. #4
HV=+-40V, Ibias=2,03µA
CO2 cooling on, Tin=-16°C, Tout=-16,7°C
standard I²C settings for uncooled APV chips.
Lab measurement with new dark box but old, large scintillator.
Calibration run for run100. |
69
|
Mon Oct 10 21:55:30 2011 |
Marko Dragicevic | Micron | Internal Calibration | cal013 | 83800 | 10.10.2011 21:52:32 | | Unknown |
micronbaby p-spray/irrad
HV = -100V
beam on, baseline of irrad APV reduced
APV #2 (I²C addr. 40) deactivated, latency error after initialization, reset line is open |
62
|
Mon Oct 10 10:37:52 2011 |
Christian Irmler | CNM | CNM stack | cal012 | 83800 | 10.10.2011 11:29:55 | 10.10.2011 11:40:55 | Unknown |
no cooling
beam on; xy table position x=330 y=64
HV_CNM=20V, HV_BSTD1=200V
Istart_CNM= 28 / 1 uA; Istart_BSTD= 0.12 uA |
56
|
Sun Oct 9 10:13:11 2011 |
Karl-Heinz Hoffmann | HPK | baby_add | cal011 | 83800 | 09.10.2011 10:05:01 | | Good |
no cooling
beam on; xy table position x= y=
HV=200V Istart_P=1.5uA; Istart_N=0.22uA
std. I2C settings |
55
|
Sun Oct 9 10:05:34 2011 |
Markus Friedl | HPK | baby_std | cal010 | 83800 | 08.10.2011 13:36:25 | 08.10.2011 13:54:07 | Good |
HV=+-40V,
no cooling
beam off; xy table position x= y=
HV= Istart=uA;
std. I2C settings
|
46
|
Fri Oct 7 18:40:07 2011 |
Christian Irmler | BELLE | Origami stack cooled | cal009 | 83800 | 07.10.2011 18:35:04 | 07.10.2011 19:31:51 | Good |
Origami6
HV=+-40V,
CO2 cooling: P_TT_4: 14.9 bar, mf=0.9 g/s
beam on; xy table position x=306 y=76.5
I=18.6µA; old I2C settings; corrected trigger delay (htr); adjusted front-end voltages
CO2 pressure reduced to 14.9 bar --> temp is about 4K colder
Temp. at pipe inlet (K7) = -18°C
Temp. at pipe outlet (K8) = -21.8°C
used multi6 config instead of single hit config (cern11_origami_cooled2_multi6.cfg)
currents:
07.10.2011 18:36:07 event=1581 rate=31events/s temp/curr=33.9/-05.3 33.9/-05.6 33.9/-06.6 32.8/001.4 32.8/001.9 32.8/018.1 |
41
|
Thu Oct 6 15:57:51 2011 |
Christian Irmler | BELLE | Origami stack cooled | cal008 | 83800 | 06.10.2011 15:53:20 | 06.10.2011 16:25:29 | Good |
HV=+-40V,
CO2 cooling: P_TT_4: 18.2 bar, mf=1 g/s
beam on; xy table position x=306 y=76.5
I=19.5µA; old I2C settings; corrected trigger delay (htr); adjusted front-end voltages
Temp. at pipe inlet (K7) = -14°C
Temp. at pipe outlet (K8) = -17°C
Before this run, the front-end voltages (+2.5 and +1.25) were readjusted so that the nominal values appear on MAMBO. (Previously, both were about 50mV too low.) Not sure if this makes a big difference, let's see.
currents:
06.10.2011 15:54:24 event=2901 rate=36events/s temp/curr=35.5/-05.6 35.5/-06.0 35.4/-06.9 34.9/001.9 35.0/002.2 35.0/019.6
06.10.2011 16:24:25 event=81350 rate=50events/s temp/curr=34.9/-05.6 34.9/-06.1 34.9/-06.9 34.3/001.7 34.3/002.2 34.3/019.6
|
35
|
Wed Oct 5 17:54:09 2011 |
Christian Irmler | BELLE | Origami stack cooled | cal007 | 83800 | 05.10.2011 17:51:57 | | Good |
I2C setting modified to improve the shape of the p-side signal --> shaping time reduced, pulse height still low
Origami6
HV=+-40V
beam on
CO2 cooling:
preassure @ P_TT_4 = 18.2 bar
mass flow = 0.9 g/s
I=18.0µA
longterm files
It-pos-0510111531.txt
It-neg-0510111531.txt
currents and temperatures:
05.10.2011 17:53:00 event=2900 rate=47events/s temp/curr=35.3/-05.6 35.3/-06.0 35.3/-06.9 34.6/001.9 34.8/002.2 34.8/020.7
|
34
|
Wed Oct 5 16:43:00 2011 |
Christian Irmler | BELLE | Origami stack cooled | cal006 | 83800 | 05.10.2011 16:40:18 | 05.10.2011 17:12:27 | Good |
Origami6
HV=+-40V
beam on
CO2 cooling:
preassure @ P_TT_4 = 18.2 bar
mass flow = 0.9 g/s
I=21.8µA
longterm files (same files used for cal005 and run006):
It-pos-0510111531.txt
It-neg-0510111531.txt
currents and temperatures:
05.10.2011 16:41:21 event=2850 rate=50events/s temp/curr=35.4/-05.8 35.3/-06.3 35.4/-07.2 34.9/002.0 34.9/002.5 34.9/021.7
05.10.2011 17:11:20 event=81237 rate=37events/s temp/curr=35.1/-05.8 35.1/-06.1 35.0/-07.1 34.5/002.0 34.5/002.5 34.5/021.4
|
32
|
Wed Oct 5 12:09:00 2011 |
Christian Irmler | BELLE | Origami stack cooled wrong I2C settings | cal005 | 83800 | 05.10.2011 12:04:50 | 05.10.2011 12:37:11 | Crap |
This calibration file can not be used, due to wrong I2C settings!
Origami6
HV=+-40V
beam on
CO2 cooling:
preassure @ P_TT_4 = 18.77 bar
mass flow = 0.9 g/s
I=22.45µA
longterm files (same files used for cal005 and run006):
It-pos-0510111122.txt
It-neg-0510111122.txt
currents and temperatures:
05.10.2011 12:05:53 event=2883 rate=37events/s temp/curr=35.0/-05.6 35.0/-06.3 35.0/-06.9 34.2/001.9 34.1/002.5 34.1/022.6
05.10.2011 12:36:53 event=83350 rate=48events/s temp/curr=34.4/-05.5 34.5/-06.1 34.5/-06.9 33.9/001.7 33.9/002.4 33.9/022.5
|
30
|
Tue Oct 4 21:40:52 2011 |
Christian Irmler | BELLE | Origami stack | cal004 | 83800 | 04.10.2011 21:18:09 | 04.10.2011 21:50:23 | Good |
Origami6
HV=+-40V, no cooling
beam on
I=91µA
longterm files (same files used for cal004 and run005):
It-pos-0410112056.txt
It-neg-0410112056.txt
currents and temperatures:
04.10.2011 21:50:13 event=83750 rate=50events/s temp/curr=35.2/-26.1 35.2/-28.7 35.2/-31.4 34.6/022.6 34.6/025.0 34.6/060.4
|