ID |
Date |
Author |
Project |
Subject |
Run Number |
Events |
StartTime |
EndTime |
Data |
92
|
Wed Mar 7 15:44:25 2012 |
Christian Irmler | BELLE | Origami Module #4 | cal112 | 83800 | 07.03.2012 15:42:53 | | Unknown |
Sr90, small (new) scintillator, new trigger unit
Origami Mod. #4
HV=+-40V, Ibias=1,6µA
CO2 cooling on, Tin=-19,6°C, Tout=-21.3°C
I²C settings for cooled APV chips
e:\cvi-projects\apvdaq_fadc\origami6_V2_single_cooled.cfg
SMU: +HV=Inv.Nr. 545-77 -HV=Inv.Nr. 540-702
HVret configuration:
0 ... switch open / not connected
1 ... switch closed / connected
SMU |
Dock Box |
J5 |
J6 |
J7 |
0 |
0 (now isolated) |
1 |
1 |
0 |
J7... HV ret @ hybrid |
24
|
Tue Oct 4 00:11:31 2011 |
Manfred Valentan | Micron | Micron-Baby-Stack p-stop, zones w & h-w | run002 | 760000 | 04.10.2011 00:11:06 | 04.10.2011 09:01:06 | Good |
hit modules in bottom left corner -> Zones wide and half-wide for all modules
Position of HEPHY x-z-table:
x = 336
z = 89 |
68
|
Mon Oct 10 21:54:24 2011 |
Marko Dragicevic | Micron | Micron-Baby-Stack p-spray, run bottom right | run027 | 700k | 10.10.2011 22:40:31 | 11.10.2011 09:49:11 | Unknown |
hit modules in bottom left corner -> Zones narrow and half-narrow for COMB_IRRAD, Zones wide and half-wide for COMM_IRRAD
timing corrected w.r.t. run004
Position of HEPHY x-z-table:
x = 322
z = 70
Note: z-Position not identical with position of run004, x-z-table seemed to have moved vertically. This configuration now hits the same sensor position. |
61
|
Mon Oct 10 00:50:23 2011 |
Christian Irmler | HPK | baby_add | run023 | 680k | 10.10.2011 00:51:26 | 10.10.2011 09:15:35 | Good |
no cooling
beam on; xy table position x=330 y=74
HV=200V Istart_PY=1,9uA; Istart_N=0.27uA
std. I2C settings
same as run021, but taking more events |
60
|
Sun Oct 9 13:12:32 2011 |
Karl-Heinz Hoffmann | HPK | baby_add | run022 | 670k | 09.10.2011 13:09:32 | 10.10.2011 00:48:16 | Good |
no cooling
beam on; xy table position x=330 y=64
HV=200V Istart_PY=1,9uA; Istart_N=0.27uA
std. I2C settings |
72
|
Mon Feb 20 17:17:27 2012 |
Christian Irmler | BELLE | Origami Module #4 | run100 | 50k | 20.02.2012 17:07:21 | 20.02.2012 17:17:22 | Good |
Sr90, large scintillator, HVret @ Origami open
Origami Mod. #4
HV=+-40V, Ibias=2,03µA
CO2 cooling on, Tin=-16°C, Tout=-16,7°C
standard I²C settings for uncooled APV chips.
Lab measurement with new dark box but old, large scintillator.
20.02.2012 17:08:27 event=5771 ***WARNING*** Error detected in APV 6, subevent 0. Event discarded, reset sent. --> APV 6 recovered afterwards |
73
|
Mon Feb 20 17:20:20 2012 |
Christian Irmler | BELLE | Origami Module #4 | cal100 | 50k | 20.02.2012 17:19:03 | 20.02.2012 17:29:58 | Good |
Sr90, large scintillator, HVret @ Origami open
Origami Mod. #4
HV=+-40V, Ibias=2,03µA
CO2 cooling on, Tin=-16°C, Tout=-16,7°C
standard I²C settings for uncooled APV chips.
Lab measurement with new dark box but old, large scintillator.
Calibration run for run100. |
82
|
Wed Feb 22 18:10:29 2012 |
Christian Irmler | BELLE | Origami Module #4 | run106 | 50k | 22.02.2012 17:39:10 | 22.02.2012 18:23:47 | Good |
Sr90, large scintillator,
Origami Mod. #4
HV=+-40V, Ibias=1,6µA
CO2 cooling on, Tin=-20.5°C, Tout=-21.5°C
I²C settings for cooled APV chips --> same as used in run010
e:\cvi-projects\apvdaq_fadc\origami6_V2_multi6_cooled.cfg
Lab measurement with new dark box but old, large scintillator
SMU: +HV=Inv.Nr. 545-77 -HV=Inv.Nr. 540-702
HVret configuration:
0 ... switch open / not connected
1 ... switch closed / connected
SMU |
Dock Box |
J5 |
J6 |
J7 |
0 |
1 |
0 |
0 |
0 |
J7... HV ret @ hybrid
Same as run105, but now J5 and J7 open
|
Attachment 1: run106.jpg
|
|
83
|
Wed Feb 22 18:32:24 2012 |
Christian Irmler | BELLE | Origami Module #4 | run107 | 50k | 22.02.2012 18:30:33 | 22.02.2012 18:40:11 | Good |
Sr90, large scintillator,
Origami Mod. #4
HV=+-40V, Ibias=1,6µA
CO2 cooling on, Tin=-20.5°C, Tout=-21.5°C
I²C settings for cooled APV chips --> same as used in run010
e:\cvi-projects\apvdaq_fadc\origami6_V2_multi6_cooled.cfg
Lab measurement with new dark box but old, large scintillator
SMU: +HV=Inv.Nr. 545-77 -HV=Inv.Nr. 540-702
HVret configuration:
0 ... switch open / not connected
1 ... switch closed / connected
SMU |
Dock Box |
J5 |
J6 |
J7 |
0 |
1 |
0 |
0 |
1 |
J7... HV ret @ hybrid
Same as run106, but now J7 closed
APV pedestals slightly unstable. |
Attachment 1: run107_apv5_raw_data.jpg
|
|
Attachment 2: run107.jpg
|
|
90
|
Mon Mar 5 19:27:56 2012 |
Christian Irmler | BELLE | Origami Module #4 | run111 | 50k | 05.03.2012 19:26:28 | | Unknown |
Sr90, large scintillator,
Origami Mod. #4
HV=+-40V, Ibias=1,30µA
CO2 cooling on, Tin=-20.5°C, Tout=-21.5°C
I²C settings for cooled APV chips
e:\cvi-projects\apvdaq_fadc\origami6_V2_multi6_cooled.cfg
Lab measurement with new dark box but old, large scintillator
SMU: +HV=Inv.Nr. 545-77 -HV=Inv.Nr. 540-702
HVret configuration:
0 ... switch open / not connected
1 ... switch closed / connected
SMU |
Dock Box |
J5 |
J6 |
J7 |
0 |
0 (now isolated) |
1 |
0 |
1 |
J7... HV ret @ hybrid
same as run108
Noise of p-side APVs in run108 was very high --> repeat measurement! |
54
|
Sat Oct 8 23:27:44 2011 |
Karl-Heinz Hoffmann | HPK | baby_std | run020 | 382000 | 08.10.2011 23:24:16 | 09.10.2011 09:24:20 | Good |
no cooling
beam on; xy table position x=327,5 y=65
HV=200V Istart_PY=8.2uA; Istart_N=0.45uA
std. I2C settings |
39
|
Wed Oct 5 20:49:11 2011 |
Markus Friedl | BELLE | Origami stack cooled | run009 | 330k | 05.10.2011 20:46:30 | 06.10.2011 09:12:24 | Good |
HV=+-40V,
CO2 cooling: P_TT_4: 18.2 bar, mf=0.9 g/s
beam on; xy table position x=306 y=76.5
I=20.5µA; new I2C settings; corrected trigger delay (htr)
New run with cold Origami and new I2C settings (shorter peaking time) and hopefully correct timing - and centered position in both x and y.
corresponding cal file: cal007
currents:
06.10.2011 09:12:12 event=329714 rate=0events/s temp/curr=35.8/-05.5 36.0/-06.0 35.8/-06.9 33.6/001.7 33.5/002.2 33.6/019.3 |
40
|
Thu Oct 6 09:22:34 2011 |
Christian Irmler | BELLE | Origami stack cooled | run010 | 330k | 05.10.2011 20:46:30 | 06.10.2011 09:12:24 | Good |
HV=+-40V,
CO2 cooling: P_TT_4: 18.2 bar, mf=0.9 g/s
beam on; xy table position x=306 y=76.5
I=20.5µA; corrected trigger delay (htr)
New run with cold Origami and same I2C setting as used 2010, with longer shaping time for p-side
centered position in both x and y.
corresponding cal file: cal006
currents:
06.10.2011 09:20:29 event=988 rate=0events/s temp/curr=35.8/-05.5 36.0/-06.0 35.8/-06.7 33.5/001.7 33.6/002.0 33.6/019.3
|
31
|
Tue Oct 4 22:11:12 2011 |
Christian Irmler | BELLE | Origami stack | run005 | 330000 | 04.10.2011 22:09:37 | 05.10.2011 09:11:54 | Crap |
Origami6
HV=+-40V, no cooling
beam on
Istart=95,2µA
pos: x=306, y=87 (centered)
longterm files (same files used for cal004 and run005):
It-pos-0410112056.txt
It-neg-0410112056.txt
currents and temp:
05.10.2011 09:11:21 event=329649 rate=0events/s temp/curr=33.0/-23.9 33.0/-28.9 33.0/-28.9 32.0/020.1 32.0/025.0 32.0/060.0 |
67
|
Mon Oct 10 16:26:20 2011 |
Marko Dragicevic | CNM | CNM stack | run026 | 300000 | 10.10.2011 16:23:46 | 10.10.2011 20:30:00 | Good |
no cooling
beam on; xy table position x=328 y=69.5
HV_CNM=20V, HV_BSTD1-N=200V
Istart_CNM= 27 / 1.5 uA; Istart_BSTD= 0.2 uA |
43
|
Fri Oct 7 00:20:15 2011 |
Christian Irmler | BELLE | Origami stack cooled | run012 | 250k | 07.10.2011 00:15:54 | 07.10.2011 09:13:55 | Good |
HV=+-40V,
CO2 cooling: P_TT_4: 18.2 bar, mf=1 g/s
beam on; xy table position x=352 y=89
I=19.1µA; old I2C settings; corrected trigger delay (htr); adjusted front-end voltages
beam focused on APV #3 of p- and n-side, respectively
currents:
07.10.2011 00:16:57 event=987 rate=0events/s temp/curr=34.3/-05.6 34.3/-06.0 34.3/-06.9 33.3/001.7 33.5/002.2 33.3/019.2
07.10.2011 09:13:42 event=249992 rate=0events/s temp/curr=33.6/-05.5 33.5/-06.0 33.6/-06.7 32.5/001.7 32.5/002.0 32.5/018.8
|
50
|
Sat Oct 8 14:33:25 2011 |
Thomas Bergauer | HPK | baby_std | run016 | 250k | 08.10.2011 14:31:40 | | Crap |
falsches config file (single statt multi6)
no cooling
beam on; xy table position x=322 y=80
HV=200V Istart_PY=7.6uA; Istart_N=0.42
std. I2C settings |
33
|
Wed Oct 5 13:00:41 2011 |
Christian Irmler | BELLE | Origami stack cooled | run006 | 24692 | 05.10.2011 12:55:53 | 05.10.2011 14:05:49 | Crap |
Run stopped after 24692 events, because all n-side APV failed due to a shortcut of a capacitor (C223) on n-side rebo
Origami6
HV=+-40V
beam on
CO2 cooling:
preassure @ P_TT_4 = 18.77 bar
mass flow = 0.9 g/s
I=22.25µA
longterm files (same files used for cal005 and run006):
It-pos-0510111122.txt
It-neg-0510111122.txt
currents and temperatures:
05.10.2011 12:56:54 event=985 rate=0events/s temp/curr=35.0/-05.5 35.0/-06.1 34.9/-06.9 34.2/001.7 34.3/002.4 34.3/022.3
|
74
|
Mon Feb 20 17:27:02 2012 |
Christian Irmler | BELLE | Origami Module #4 | run101 | 24497 | 20.02.2012 17:33:33 | 20.02.2012 17:38:21 | Good |
Sr90, large scintillator, HVret @ Origami open
Origami Mod. #4
HV=+-40V, Ibias=2,03µA
CO2 cooling on, Tin=-16°C, Tout=-16,7°C
I²C settings for cooled APV chips --> same as used in run010.
Lab measurement with new dark box but old, large scintillator.
CO2 empty after 24497 events --> run stopped |
22
|
Mon Oct 3 21:01:00 2011 |
Manfred Valentan | Micron | Micron-Baby-Stack p-stop, zones n & h-n | run001 | 200k | 03.10.2011 20:54:22 | 03.10.2011 23:25:47 | Good |
hit modules in top right corner -> Zones narrow and half-narrow for all modules
Position of HEPHY x-z-table:
x = 324
z = 103 |