Wed Mar 7 16:03:30 2012, Christian Irmler, BELLE, Origami Module #4, run113, 100k, 07.03.2012 16:07:34, 07.03.2012 16:30:48 , Unknown
|
Sr90, small (new) scintillator, new trigger unit
Origami Mod. #4
HV=+-40V, Ibias=1,6µA
CO2 cooling on, Tin=-19,6°C, Tout=-21.3°C
I²C settings for cooled APV chips
e:\cvi-projects\apvdaq_fadc\origami6_V2_multi6_cooled.cfg
SMU: +HV=Inv.Nr. 545-77 -HV=Inv.Nr. 540-702
HVret configuration:
0 ... switch open / not connected
1 ... switch closed / connected
SMU |
Dock Box |
J5 |
J6 |
J7 |
0 |
0 (now isolated) |
1 |
1 |
0 |
J7... HV ret @ hybrid
same configuration as run112 but now the step motors of the xz table are powered. |
Wed Mar 7 15:12:42 2012, Christian Irmler, BELLE, Origami Module #4, run112, 100k, 07.03.2012 15:13:18, 07.03.2012 14:31:07, Unknown
|
Sr90, small (new) scintillator, new trigger unit
Origami Mod. #4
HV=+-40V, Ibias=1,6µA
CO2 cooling on, Tin=-19,6°C, Tout=-21.3°C
I²C settings for cooled APV chips
e:\cvi-projects\apvdaq_fadc\origami6_V2_multi6_cooled.cfg
SMU: +HV=Inv.Nr. 545-77 -HV=Inv.Nr. 540-702
HVret configuration:
0 ... switch open / not connected
1 ... switch closed / connected
SMU |
Dock Box |
J5 |
J6 |
J7 |
0 |
0 (now isolated) |
1 |
1 |
0 |
J7... HV ret @ hybrid |
Mon Mar 5 19:27:56 2012, Christian Irmler, BELLE, Origami Module #4, run111, 50k, 05.03.2012 19:26:28, , Unknown
|
Sr90, large scintillator,
Origami Mod. #4
HV=+-40V, Ibias=1,30µA
CO2 cooling on, Tin=-20.5°C, Tout=-21.5°C
I²C settings for cooled APV chips
e:\cvi-projects\apvdaq_fadc\origami6_V2_multi6_cooled.cfg
Lab measurement with new dark box but old, large scintillator
SMU: +HV=Inv.Nr. 545-77 -HV=Inv.Nr. 540-702
HVret configuration:
0 ... switch open / not connected
1 ... switch closed / connected
SMU |
Dock Box |
J5 |
J6 |
J7 |
0 |
0 (now isolated) |
1 |
0 |
1 |
J7... HV ret @ hybrid
same as run108
Noise of p-side APVs in run108 was very high --> repeat measurement! |
Mon Mar 5 18:41:23 2012, Christian Irmler, BELLE, Origami Module #4, run110, 100k, 05.03.2012 18:45:21, 05.03.2012 19:04:30, Unknown
|
Sr90, large scintillator,
Origami Mod. #4
HV=+-40V, Ibias=1,9µA
CO2 cooling on, Tin=-20.5°C, Tout=-21.5°C
I²C settings for cooled APV chips
e:\cvi-projects\apvdaq_fadc\origami6_V2_multi6_cooled.cfg
Lab measurement with new dark box but old, large scintillator
SMU: +HV=Inv.Nr. 545-77 -HV=Inv.Nr. 540-702
HVret configuration:
0 ... switch open / not connected
1 ... switch closed / connected
SMU |
Dock Box |
J5 |
J6 |
J7 |
0 |
0 (now isolated) |
1 |
1 |
0 |
J7... HV ret @ hybrid |
Mon Mar 5 17:36:46 2012, Christian Irmler, BELLE, Origami Module #4, run109, 100k, 05.03.2012 17:36:19 , 05.03.2012 17:55:24 , Unknown
|
Sr90, large scintillator,
Origami Mod. #4
HV=+-40V, Ibias=1,9µA
CO2 cooling on, Tin=-20.5°C, Tout=-21.5°C
I²C settings for cooled APV chips
e:\cvi-projects\apvdaq_fadc\origami6_V2_multi6_cooled.cfg
Lab measurement with new dark box but old, large scintillator
SMU: +HV=Inv.Nr. 545-77 -HV=Inv.Nr. 540-702
HVret configuration:
0 ... switch open / not connected
1 ... switch closed / connected
SMU |
Dock Box |
J5 |
J6 |
J7 |
0 |
0 (now isolated) |
1 |
1 |
1 |
J7... HV ret @ hybrid |
Mon Mar 5 17:34:43 2012, Christian Irmler, BELLE, Origami Module #4, run108, 100k, , 05.03.2012 16:24:12, Unknown
|
Sr90, large scintillator,
Origami Mod. #4
HV=+-40V, Ibias=1,9µA
CO2 cooling on, Tin=-20.5°C, Tout=-21.5°C
I²C settings for cooled APV chips
e:\cvi-projects\apvdaq_fadc\origami6_V2_multi6_cooled.cfg
Lab measurement with new dark box but old, large scintillator
SMU: +HV=Inv.Nr. 545-77 -HV=Inv.Nr. 540-702
HVret configuration:
0 ... switch open / not connected
1 ... switch closed / connected
SMU |
Dock Box |
J5 |
J6 |
J7 |
0 |
0 (now isolated) |
1 |
0 |
1 |
J7... HV ret @ hybrid |
Wed Feb 22 18:32:24 2012, Christian Irmler, BELLE, Origami Module #4, run107, 50k, 22.02.2012 18:30:33, 22.02.2012 18:40:11, Good
|
Sr90, large scintillator,
Origami Mod. #4
HV=+-40V, Ibias=1,6µA
CO2 cooling on, Tin=-20.5°C, Tout=-21.5°C
I²C settings for cooled APV chips --> same as used in run010
e:\cvi-projects\apvdaq_fadc\origami6_V2_multi6_cooled.cfg
Lab measurement with new dark box but old, large scintillator
SMU: +HV=Inv.Nr. 545-77 -HV=Inv.Nr. 540-702
HVret configuration:
0 ... switch open / not connected
1 ... switch closed / connected
SMU |
Dock Box |
J5 |
J6 |
J7 |
0 |
1 |
0 |
0 |
1 |
J7... HV ret @ hybrid
Same as run106, but now J7 closed
APV pedestals slightly unstable. |
Wed Feb 22 18:10:29 2012, Christian Irmler, BELLE, Origami Module #4, run106, 50k, 22.02.2012 17:39:10 , 22.02.2012 18:23:47, Good
|
Sr90, large scintillator,
Origami Mod. #4
HV=+-40V, Ibias=1,6µA
CO2 cooling on, Tin=-20.5°C, Tout=-21.5°C
I²C settings for cooled APV chips --> same as used in run010
e:\cvi-projects\apvdaq_fadc\origami6_V2_multi6_cooled.cfg
Lab measurement with new dark box but old, large scintillator
SMU: +HV=Inv.Nr. 545-77 -HV=Inv.Nr. 540-702
HVret configuration:
0 ... switch open / not connected
1 ... switch closed / connected
SMU |
Dock Box |
J5 |
J6 |
J7 |
0 |
1 |
0 |
0 |
0 |
J7... HV ret @ hybrid
Same as run105, but now J5 and J7 open
|
Wed Feb 22 17:38:29 2012, Christian Irmler, BELLE, Origami Module #4, run105, 100k, 22.02.2012 17:39:10 , 22.02.2012 17:58:01, Good
|
Sr90, large scintillator,
Origami Mod. #4
HV=+-40V, Ibias=1,6µA
CO2 cooling on, Tin=-20.5°C, Tout=-21.5°C
I²C settings for cooled APV chips --> same as used in run010
e:\cvi-projects\apvdaq_fadc\origami6_V2_multi6_cooled.cfg
Lab measurement with new dark box but old, large scintillator
SMU: +HV=Inv.Nr. 545-77 -HV=Inv.Nr. 540-702
HVret configuration:
0 ... switch open / not connected
1 ... switch closed / connected
SMU |
Dock Box |
J5 |
J6 |
J7 |
0 |
1 |
1 |
0 |
0 |
J7... HV ret @ hybrid
Same as run104, but now J7 open
Again spikes in bias current!
|
Wed Feb 22 16:21:23 2012, Christian Irmler, BELLE, Origami Module #4, run104, 100k, 22.02.2012 16:22:34, 22.02.2012 16:41:26, Good
|
Sr90, large scintillator,
Origami Mod. #4
HV=+-40V, Ibias=1,6µA
CO2 cooling on, Tin=-20.5°C, Tout=-21.5°C
I²C settings for cooled APV chips --> same as used in run010
e:\cvi-projects\apvdaq_fadc\origami6_V2_multi6_cooled.cfg
Lab measurement with new dark box but old, large scintillator
SMU: +HV=Inv.Nr. 545-77 -HV=Inv.Nr. 540-702
HVret configuration:
0 ... switch open / not connected
1 ... switch closed / connected
SMU |
Dock Box |
J5 |
J6 |
J7 |
0 |
1 |
1 |
0 |
1 |
J7... HV ret @ hybrid
APV |
0 |
1 |
2 |
3 |
4 |
5 |
6 |
7 |
avg noise |
2.564 |
2.124 |
2.094 |
2.200 |
1.696 |
1.666 |
1.870 |
1.888 |
CM noise |
1.7 |
0.9 |
0.9 |
1.3 |
4.0 |
4.1 |
4.1 |
4.2 |
|
Wed Feb 22 13:37:11 2012, Christian Irmler, BELLE, Origami Module #4, run103, 100k, 22.02.2012 13:35:08, 22.02.2012 13:54:08 , Crap
|
Sr90, large scintillator,
HVret @ Origami closed
HVret @ SMU connected
Origami Mod. #4
HV=+-40V, Ibias=1,6µA
CO2 cooling on, Tin=-20.5°C, Tout=-21.5°C
I²C settings for cooled APV chips --> same as used in run010
e:\cvi-projects\apvdaq_fadc\origami6_V2_multi6_cooled.cfg
Lab measurement with new dark box but old, large scintillator
Wires between HV connectors ground and HVret conntact at the mambo were open --> no HVret conntection between module and SMUs!
Moreover the SMU Inv. Nr. 545-078 has either an unstable output or problems with the current measurement. |
Wed Feb 22 12:15:11 2012, Christian Irmler, BELLE, Origami Module #4, run102, 170k, 22.02.2012 12:31:08, 22.02.2012 13:03:27 , Good
|
same as run101
Sr90, large scintillator, HVret @ Origami open
Origami Mod. #4
HV=+-40V, Ibias=1,8µA
CO2 cooling on, Tin=-20.2°C, Tout=-21.3°C
I²C settings for cooled APV chips --> same as used in run010
e:\cvi-projects\apvdaq_fadc\origami6_V2_multi6_cooled.cfg
Lab measurement with new dark box but old, large scintillator.
CO2 empty after 24497 events --> run stopped --> no corresponding calibration available, but for rought analysis cal100 can be used for this run, too |
Mon Feb 20 17:27:02 2012, Christian Irmler, BELLE, Origami Module #4, run101, 24497, 20.02.2012 17:33:33, 20.02.2012 17:38:21, Good
|
Sr90, large scintillator, HVret @ Origami open
Origami Mod. #4
HV=+-40V, Ibias=2,03µA
CO2 cooling on, Tin=-16°C, Tout=-16,7°C
I²C settings for cooled APV chips --> same as used in run010.
Lab measurement with new dark box but old, large scintillator.
CO2 empty after 24497 events --> run stopped |
Mon Feb 20 17:17:27 2012, Christian Irmler, BELLE, Origami Module #4, run100, 50k, 20.02.2012 17:07:21, 20.02.2012 17:17:22, Good
|
Sr90, large scintillator, HVret @ Origami open
Origami Mod. #4
HV=+-40V, Ibias=2,03µA
CO2 cooling on, Tin=-16°C, Tout=-16,7°C
standard I²C settings for uncooled APV chips.
Lab measurement with new dark box but old, large scintillator.
20.02.2012 17:08:27 event=5771 ***WARNING*** Error detected in APV 6, subevent 0. Event discarded, reset sent. --> APV 6 recovered afterwards |
Mon Oct 10 21:54:24 2011, Marko Dragicevic, Micron, Micron-Baby-Stack p-spray, run bottom right, run027, 700k, 10.10.2011 22:40:31, 11.10.2011 09:49:11, Unknown
|
hit modules in bottom left corner -> Zones narrow and half-narrow for COMB_IRRAD, Zones wide and half-wide for COMM_IRRAD
timing corrected w.r.t. run004
Position of HEPHY x-z-table:
x = 322
z = 70
Note: z-Position not identical with position of run004, x-z-table seemed to have moved vertically. This configuration now hits the same sensor position. |
Mon Oct 10 16:26:20 2011, Marko Dragicevic, CNM, CNM stack, run026, 300000, 10.10.2011 16:23:46, 10.10.2011 20:30:00, Good
|
no cooling
beam on; xy table position x=328 y=69.5
HV_CNM=20V, HV_BSTD1-N=200V
Istart_CNM= 27 / 1.5 uA; Istart_BSTD= 0.2 uA |
Mon Oct 10 14:11:53 2011, Christian Irmler, CNM, CNM stack, run025, 110000, 10.10.2011 14:09:00, 10.10.2011 16:11:57, Good
|
no cooling
beam on; xy table position x=335.5 y=79.5
HV_CNM=20V, HV_BSTD1-N=200V
Istart_CNM= 31 / 1.4 uA; Istart_BSTD= 0.2 uA |
Mon Oct 10 12:15:52 2011, Manfred Valentan, CNM, CNM stack, run024, 110000, 10.10.2011 12:03:58, 10.10.2011 14:05:56, Good
|
no cooling
beam on; xy table position x=320.5 y=79.5
HV_CNM=20V, HV_BSTD1-N=200V
Istart_CNM= 28 / 1 uA; Istart_BSTD= 0.12 uA |
Mon Oct 10 00:50:23 2011, Christian Irmler, HPK, baby_add, run023, 680k, 10.10.2011 00:51:26, 10.10.2011 09:15:35, Good
|
no cooling
beam on; xy table position x=330 y=74
HV=200V Istart_PY=1,9uA; Istart_N=0.27uA
std. I2C settings
same as run021, but taking more events |
Sun Oct 9 13:12:32 2011, Karl-Heinz Hoffmann, HPK, baby_add, run022, 670k, 09.10.2011 13:09:32, 10.10.2011 00:48:16, Good
|
no cooling
beam on; xy table position x=330 y=64
HV=200V Istart_PY=1,9uA; Istart_N=0.27uA
std. I2C settings |
|