  HEPHY logbook for testbeam at SPS September/October 2010, Page 7 of 8  Not logged in ELOG logo
Entry  Wed Oct 6 16:33:55 2010, Markus Friedl, Micron Baby, run, run051, 90000, 06.10.2010 16:32:03, 06.10.2010 20:59:58 

8 x Micron Baby (3 irradiated), no cooling
HV=100V I=209uA (start of run, still rising); 203uA (end of run)
beam on, NIM trigger from TLU (no handshake)
telescope position 131 / 53.5 / 7.5
 

no telescope run - being tuned (some intervals during data taking are possible)

 

Entry  Wed Oct 6 23:48:49 2010, Markus Friedl, BELLE, run, run052, 165789, 06.10.2010 23:47:12, 07.10.2010 08:43:42 

Belle stack without Origami2
HV=80V I=217uA
telescope setting 103 / 30 / 277.5
telescope run 20184

temperature of chiller set to 15°C

beam stopped at 7.10. 08:00 sharp, no beam until 8.10 - 8:00 according to SPS Page 1. We keep DAQ and telescope running and cross our fingers for at least some spills of particles.

 

Irradiated micron module 2, n-side: excessive hit count on channels at the edges of the APV chips - no such thing on the p-side

Entry  Fri Oct 8 17:26:36 2010, Markus Friedl, BELLE, cal, cal015, 83800, 08.10.2010 17:22:44,  

Belle stack without Origami2
HV=11V I=32uA

beam on

2 suspicious APVs on n side of Micron due to burnt strip

trying to identify problematic Micron n side channels with low bias voltage (APV22 baseline saturates at higher bias)

Entry  Fri Oct 8 18:12:43 2010, Markus Friedl, BELLE, run, run054, 90000, 08.10.2010 18:08:48, 08.10.2010 21:49:36  

Belle stack w/o Origami2 after irradiation

HV=80V I=224uA (beginning of run)

2 suspicious APVs on Micron n side

xy-table position 7 / 30 / 277.5 (shoots at the smaller edge of the trapezoidal sensor)

 

telescope run 20195

Entry  Fri Oct 8 23:05:03 2010, Markus Friedl, BELLE, cal, cal016, 83800, 08.10.2010 23:03:16,  

Belle stack w/o Origami2 after irradiation
HV=80V / 258 µA
cooling: Origami only, chiller set to -25°C,
2 suspicious APVs on n side of Micron due to burnt strip
 

Bias connection of Origami repaired --> lower noise on p-side

Entry  Sat Oct 9 00:17:02 2010, Markus Friedl, BELLE, run, run056, 225k, 09.10.2010 00:15:53, 09.10.2010 09:54:28 

Belle stack w/o Origami2 after irradiation
HV=80V / I_start=255 µA, I_stop=243 µA  

cooling: Origami only, chiller set to -25°C,
2 suspicious APVs on n side of Micron (possibly due to burnt strip)

same position as in previous run

telescope run 20198 - now with correct telescope threshold settings

a lot of "Error detected in APV 21, subevent 2. Event discarded, reset sent." errors started at 09.10.2010 09:00:25 (after whole night without any error)

Entry  Sat Oct 9 15:29:09 2010, Markus Friedl, BELLE, run, run058, 30000, 09.10.2010 15:26:50, 09.10.2010 16:21:37 

Origami1 (n-side only) + Micron1 (non-irrad)
angle scan: 73 - now with multi6 (previous run was single)
HV=100V / 35uA (beginning of run) 34uA (end of run)
cooling: chiller set to 15C

telescope run 20201

Origami p-side does not receive triggers -> disabled

origami flex straight part needed to be folded due to severe rotation angle -> nanonics connector problem...

 

Entry  Sat Oct 9 19:44:36 2010, Markus Friedl, common, end of success, , , 09.10.2010 19:44:23, NaN 
 
Entry  Mon Sep 27 15:39:33 2010, Thomas Bergauer, common, PI-XY Table, , , ,  

M-521.dd: (links/rechts)

downstream rechts => pos. limit [0]

downstream links => neg. limit [207]

----------------------

M-511.dd: (oben/unten)

oben [102]

unten [0]

----------------------

M-060.dg: (rotation)

parallel beam [7,5]

 

-------------------------------------------

Remote Control:

lokal ssh tunnel starten:
ssh -g -C -L5959:zenpixelw:5900 eudet@eudetmac001.cern.ch

danach mit "Chicken of the VNC" (von lokalem Mac) VNC auf localhost (port 5959). 

Entry  Mon Sep 27 15:41:28 2010, Thomas Bergauer, common, Szintillatoren, , , ,  

1: breit  (1)

2: schmal (2)

3: breit  (4)

4: schmal  (8)

 

AND aus beiden breiten: 5

AND aus beiden schmalen: 10

 

Entry  Mon Sep 27 15:42:09 2010, Thomas Bergauer, common, TLU: wir stecken auf DUT 3, , , ,  

TLUControl.exe ist auf zenpixell3 unter /home/eudet/eudaq/bin

TLUControl.exe -d 14 -a 12
-dXX Mask for DUT:
1   abgeschnittenes Dummykabel
2   EUDRB hinten
4   EUDRB vorne
8   DUT3(wir)
-t100 internal trigger
-aXX Mask for Szintis:
1   4 mm (front)
2   4 mm (front)
4   9 mm (back)
8   9 mm (back)
Entry  Mon Sep 27 18:54:47 2010, Thomas Bergauer, BELLE, XY Position, , , ,  

center position: 55/40/277,5

Entry  Tue Sep 28 16:50:31 2010, Thomas Bergauer, common, serial connection chiller, , , ,  

 Der chiller braucht ein gekreuztes serielles Kabel!!

Entry  Wed Sep 29 16:28:17 2010, Thomas Bergauer, common, EUDET telescope pedestal run, 20089, 122850 (events built), 29.09.2010 16:27, 29.09.2010 17:33 

 EUDAQ config eudet2-test-6boards-in-2crates

rate: ~745.5 Hz

 

Entry  Wed Sep 29 17:38:15 2010, Thomas Bergauer, common, EUDET telescope pedestal run, 20097, 118125, 29.09.2010 17:37, 29.09.2010 18:09 

new pedestal run of EUDET telescope with new JTAG settings of plane 3 (sensor 11) [according to Ingrid]:


VRef2    98
VRef1A  87
VRef1B  143
VRef1C  108
VRef1D   153

 

Entry  Wed Sep 29 18:53:05 2010, Thomas Bergauer, Micron Baby, XY Positions, , , ,  

center position: 113/47/7.5

top left (direction jura): 121/53.5/7.5

bottom right (direction saleve): 105/40.5/7.5

Entry  Thu Sep 30 14:45:16 2010, Thomas Bergauer, Micron Baby, run, run015, 22.5k, 30.09.2010 13:18:11, 30.09.2010 14:38:01 

micron baby stack angle scan 15 grad mit 3 modulen

HV=100V, I=3.8uA

Beam on

 

telescope position: 129/76/187.5

telescope run 20109

Entry  Thu Sep 30 23:09:29 2010, Thomas Bergauer, Micron Baby, run, run019, 162000, 30.09.2010 23:08:03, 01.10.2010 9:33:38 run019_sensordisplay1.jpgEudet-X_Cluster_Corr-run020115.jpgEudet-Num_Clusters-run020115.jpg

micron baby: 4 DUTs
HV=100V I=12uA I_stop=15.4uA
Beam on
45° rotation

 

telescope run  20115 (after JTAG programming of telescope chips due to config errors)


Entry  Fri Oct 1 23:50:28 2010, Thomas Bergauer, CNM, run, run023, 2275, 01.10.2010 23:49:29, 01.10.2010 23:53:12 

CMN-SILC stack
HV_SILC (Katze)=75V, 0.25uA (6 SiLC)
HV_CNM (Keithley)=30V, CNM_1=84nA, CNM_2=4.3uA

too much optimized alignment (XY-table to 133/74/187.5)   ;-)

telescope run 20119

Entry  Fri Oct 1 23:55:12 2010, Thomas Bergauer, CNM, run, run024, 320k, 01.10.2010 23:53:12, 02.10.2010 09:47:37 cnm_2d_2.pngcnm_2d_1.png

CMN-SILC stack
HV_SILC (Katze)=75V, 0.25uA (6 SiLC)
HV_CNM (Keithley)=30V, CNM_1=84nA, CNM_2=4.3uA

optimized alignment (XY-table to 133/76/187.5)

telescope run 20120