Entry  Tue Sep 28 14:48:00 2010, Markus Friedl, BELLE, run, run004, 1755, 28.09.2010 14:48:28, 28.09.2010 15:06:31 

same as before, but now pos 1 is located on the other edge (-48mm from center)

telescope 7 / 30 / 277.5

telescope run 20079

Entry  Tue Sep 28 15:09:16 2010, Markus Friedl, BELLE, run, run005, 40000, 28.09.2010 15:07:08, 28.09.2010 20:20:12 

previous run stopped due to beam problem (shifted, broad, etc).

now recovered, restarting with same settings.

 

telescope run: 20080

Entry  Tue Sep 28 16:50:31 2010, Thomas Bergauer, common, serial connection chiller, , , ,  

 Der chiller braucht ein gekreuztes serielles Kabel!!

Entry  Tue Sep 28 22:07:16 2010, Markus Friedl, BELLE, cal, cal002 DO NOT USE -> go to cal003, 83800, 28.09.2010 22:01:55,  

Belle stack (Origami, Micron, Micron, Origami)

now cooling only for Origami, Micron are uncooled

chiller set point: -25°C, measured values on Origami: -10°C

HV=100V, Ibias=38µA, beam off

modified I2C settings for cold modules

temperature monitoring: Origami2 pipes in<->out swapped (to achieve shorter tubes)

 

PROBLEM: Origami n channels are swapped -> USE cal003 INSTEAD!!!

Entry  Tue Sep 28 22:09:39 2010, Markus Friedl, BELLE, Temperature sensors Belle stack, , , ,  

PicoLogic TC-08 temperature sensor assignment Belle stack:

Sensor type: K

channel 1:  Origami_mod1_in
channel 2:  Origami_mod1_out
channel 3:  Origami_mod2_out    (swapped with 4)  - label not changed
channel 4:  Origami_mod2_in      (swapped with 3)  - label not changed
channel 5:  Micron_wedge_mod1_p
channel 6:  Micron_wedge_mod1_n
channel 7:  Micron_wedge_mod2_p
channel 8:  Micron_wedge_mod2_n
 

The cooling pipes of the two Origami modules are connected in series, respectively.

Entry  Tue Sep 28 23:23:47 2010, Markus Friedl, BELLE, run, run006, 2003, 28.09.2010 23:13:22, 28.09.2010 23:22:26 

channels swapped -> stopped

Entry  Tue Sep 28 23:24:35 2010, Markus Friedl, BELLE, run, run007, 1460, 28.09.2010 23:23:29, 28.09.2010 23:27:54 

Belle stack: Origami-Micron-Micron-Origami
HV=100V, Ibias=38A, beam on
cooling: Origami only, chiller: -25C, Module: -10C
modified I2C settings for cooled modules
centered again

 

telescope run 20082

 

sensor config bug (only relevant for sensor display) found -> stopped

Entry  Tue Sep 28 23:29:50 2010, Markus Friedl, BELLE, run, run008, 140000, 28.09.2010 23:29:01, 29.09.2010 09:36:58 run008_sensordisplay.jpgrun008_sensordisplay2.jpg

Belle stack: Origami-Micron-Micron-Origami
HV=100V, Ibias=38A, beam on
cooling: Origami only, chiller: -25C, Module: -10C
modified I2C settings for cooled modules
centered again

 

telescope run 20083

Entry  Wed Sep 29 09:39:33 2010, Markus Friedl, BELLE, cal, cal003, 83800, 29.09.2010 09:38:00,  

same as cal002, but now with corrected readout order

Entry  Wed Sep 29 10:30:59 2010, Markus Friedl, BELLE, run, run009, 30000, 29.09.2010 10:27:27, 29.09.2010 11:06:35 

zero suppressed (HIT4) @ FADC

Belle stack: Origami-Micron-Micron-Origami - HIT4
HV=100V, Ibias=28A, beam on
cooling: Origami only, chiller: -25C, Module: -10C
modified I2C settings for cooled modules
centered in both axes

telescope run 20084

Entry  Wed Sep 29 11:09:10 2010, Markus Friedl, BELLE, run, run010, 30000, 29.09.2010 11:08:03, 29.09.2010 11:50:07 

zero suppressed (HIT5) @ FADC

Belle stack: Origami-Micron-Micron-Origami - HIT5
HV=100V, Ibias=28A, beam on
cooling: Origami only, chiller: -25C, Module: -10C
modified I2C settings for cooled modules
centered in both axes

telescope run 20085

Entry  Wed Sep 29 11:53:16 2010, Erik Huemer, BELLE, run, run011, 19003, 29.09.2010 12:19:57, 29.09.2010 14:00:57 

Belle stack: Origami-Micron-Micron-Origami
HV=100V, Ibias=83.22µA, beam on
no cooling

Temp. at modules: 
Origami 1: 50°C
Micron 1: 46°C
Micron 2: 48°C
Origami 2: 48°C

standard I2C settings
centered

telescope run 20085 (continued)

Entry  Wed Sep 29 14:03:55 2010, Markus Friedl, BELLE, cal, cal004, 83800, 29.09.2010 14:02:25, 29.09.2010 14:45:05 

Belle stack: Origami-Micron-Micron-Origami
HV=100V, Ibias=100µA, beam off
no cooling

Temp. at modules: 
Origami 1: 50°C
Micron 1: 46°C
Micron 2: 48°C
Origami 2: 48°C

standard I2C settings
centered

Entry  Wed Sep 29 16:28:17 2010, Thomas Bergauer, common, EUDET telescope pedestal run, 20089, 122850 (events built), 29.09.2010 16:27, 29.09.2010 17:33 

 EUDAQ config eudet2-test-6boards-in-2crates

rate: ~745.5 Hz

 

Entry  Wed Sep 29 17:38:15 2010, Thomas Bergauer, common, EUDET telescope pedestal run, 20097, 118125, 29.09.2010 17:37, 29.09.2010 18:09 

new pedestal run of EUDET telescope with new JTAG settings of plane 3 (sensor 11) [according to Ingrid]:


VRef2    98
VRef1A  87
VRef1B  143
VRef1C  108
VRef1D   153

 

Entry  Wed Sep 29 17:38:19 2010, Markus Friedl, Micron Baby, cal, cal005, 83800, 29.09.2010 17:36:38,  

micron baby stack
HV=100V I=6.5uA
Beam off

Entry  Wed Sep 29 18:53:05 2010, Thomas Bergauer, Micron Baby, XY Positions, , , ,  

center position: 113/47/7.5

top left (direction jura): 121/53.5/7.5

bottom right (direction saleve): 105/40.5/7.5

Entry  Wed Sep 29 19:01:28 2010, Markus Friedl, Micron Baby, run, run012, 2248, 29.09.2010 18:59:09, 29.09.2010 19:07:02 

micron baby stack
HV=100V I=11.0uA
Beam on
telescope position: 121/55/7.5

 

telescope run 20105

Entry  Wed Sep 29 19:08:56 2010, Markus Friedl, Micron Baby, run, run013, 89853, 29.09.2010 19:09:57, 30.09.2010 00:52:39 run013_hit.png

micron baby stack
HV=100V I=11.0uA
Beam on
telescope position: 121/53.5/7.5   (shifted in height by -1.5mm wrt run012 for better DUT centering)

 

telescope run 20107

Entry  Thu Sep 30 00:56:32 2010, Markus Friedl, Micron Baby, run, run014, 100k, 30.09.2010 00:55:30, 30.09.2010 11:20:02 

micron baby stack

 

HV=100V I=11.0uA

I_end = 16uA

 

Beam on

 

telescope position: 105/40.5/7.5 

telescope run 20108