Home CMS Production Clean room MedAustron HEPHY testbeams old
testbeam-RD50-DESY-Apr2024 testbeam-RD50-DESY-Jul2023 testbeam-RD50-CERN-Oct2022 testbeam-DESY-Oct2017 testbeam-DESY-April2017 testbeam-SPS2015 testbeam-SPS2014 testbeam-DESY14 testbeam-SPS12 testbeam-SPS11 testbeam-SPS10 testbeam-SPS09 testbeam-SPS08 LP-TPC
  HEPHY logbook for testbeam at SPS September/October 2010, Page 6 of 8  Not logged in ELOG logo
Entry  Sat Oct 9 00:17:02 2010, Markus Friedl, BELLE, run, run056, 225k, 09.10.2010 00:15:53, 09.10.2010 09:54:28 

Belle stack w/o Origami2 after irradiation
HV=80V / I_start=255 µA, I_stop=243 µA  

cooling: Origami only, chiller set to -25°C,
2 suspicious APVs on n side of Micron (possibly due to burnt strip)

same position as in previous run

telescope run 20198 - now with correct telescope threshold settings

a lot of "Error detected in APV 21, subevent 2. Event discarded, reset sent." errors started at 09.10.2010 09:00:25 (after whole night without any error)

Entry  Sat Oct 9 12:00:55 2010, Christian Irmler, BELLE, cal, cal017, 83800, 09.10.2010 11:59:37,  

Origami1 (n-side only) + Micron1 (non-irrad)
angle scan: 73°
HV=100V / 40 µA
cooling: chiller set to 15°C

Origami p-side does not receive triggers -> disabled

origami flex straight part needed to be folded due to severe rotation angle -> nanonics connector problem...

Entry  Sat Oct 9 12:25:36 2010, Thomas Bergauer, BELLE, run, run057, 276475, 09.10.2010 12:22:21, 09.10.2010 15:08:46 

 

Origami1 (n-side only) + Micron1 (non-irrad)

angle scan: 73 degree

HV=100V / 36 uA  (beginning), 35uA (end of run)

cooling: chiller set to 15C

 PROBLEM: wrong config file (single hit, not multi6)

telescope run 20200

Origami p-side does not receive triggers -> disabled

origami flex straight part needed to be folded due to severe rotation angle -> nanonics connector problem...

 

Entry  Sat Oct 9 15:29:09 2010, Markus Friedl, BELLE, run, run058, 30000, 09.10.2010 15:26:50, 09.10.2010 16:21:37 

Origami1 (n-side only) + Micron1 (non-irrad)
angle scan: 73 - now with multi6 (previous run was single)
HV=100V / 35uA (beginning of run) 34uA (end of run)
cooling: chiller set to 15C

telescope run 20201

Origami p-side does not receive triggers -> disabled

origami flex straight part needed to be folded due to severe rotation angle -> nanonics connector problem...

 

Entry  Sat Oct 9 18:56:35 2010, Thomas Bergauer, common, path to telescope data, , , ,  

eudetmac001.cern.ch

path is /Volumes/Data/eudet/Data/cern_2010_sept_silc_apvdaq (Vienna system)

and /Volumes/Data/eudet/Data/cern_2010_sept_silc_alibava (Santander system)

Entry  Sat Oct 9 19:44:36 2010, Markus Friedl, common, end of success, , , 09.10.2010 19:44:23, NaN 
 
    Reply  Sun Oct 10 15:45:04 2010, Thomas Bergauer, common, end of success, , , 09.10.2010 19:44:23, NaN 

 ;-) 

genau das gleiche wollte ich auch gerade schreiben!

    Reply  Mon Oct 11 11:18:17 2010, Christian Irmler, common, end of success, , , 09.10.2010 19:44:23, NaN IMG_9033.JPG

der Beiweis  :-)

Entry  Tue Oct 19 14:58:56 2010, Manfred Valentan, Micron Baby, geometry, , , ,  IMG_8888.JPGmicron_baby_stack_telescope_geometry.JPGmicron_baby_stack_telectope_geometry.dwg
 
Entry  Thu Mar 31 10:28:19 2011, Manfred Valentan, Micron Baby, DUT5, source run, post irrad, pre zupf, cal100, , 31.03.2011 10:24:59, 31.03.2011 10:33:15 

DUT5
cal-file: micronbaby_single_irrad_cal100.cfg


HV=+-50V
90Sr 1mCi ~centered
I_start=99muA
I_end=103muA

 

Entry  Thu Mar 31 11:05:39 2011, Manfred Valentan, Micron Baby, DUT5, source run, post irrad, pre zupf, run100, 100k, 31.03.2011 11:02:17, 31.03.2011 11:15:41 

DUT5
cal-file: micronbaby_multi6_irrad_run100.cfg


HV=+-50V
90Sr 1mCi ~centered
I_start=104µA
I_end=111µA

 

Entry  Thu Mar 31 12:11:04 2011, Manfred Valentan, Micron Baby, DUT4, source run, post irrad, pre zupf, cal101, , 31.03.2011 12:09:36, 31.03.2011 12:17:52 

DUT4
config file: micronbaby_single_irrrad_cal101.cfg


HV=+-50V
90Sr 1mCi ~centered
I_start=128µA
I_end=140µA

 

Entry  Thu Mar 31 12:24:24 2011, Manfred Valentan, Micron Baby, DUT4, source run, post irrad, pre zupf, run101, 100k, 31.03.2011 12:23:00, 31.03.2011 12:36:48 

DUT4
config file: micronbaby_multi6_irrrad_run101.cfg


HV=+-50V
90Sr 1mCi ~centered
I_start=128µA
I_end=155µA

 

Entry  Fri Apr 8 13:25:58 2011, Manfred Valentan, Micron Baby, sensors and modules, , , ,  

TRK1: B2Micron_2826_1_baby2

TRK2: B2Micron_2826_1_baby3

DUT3: B2Micron_2826_7_baby2 (common p-stop)

DUT4: B2Micron_2826_19_baby4 (combined p-stop)

DUT5: B2Micron_2826_19_baby3 (atoll p-stop)

TRK6: B2Micron_2826_1_baby4

TRK7: B2Micron_2826_19_baby2

GEW: B2Micron_2826_7_baby4

not used: B2Micron_2826_7_baby3

Entry  Mon Apr 11 18:25:35 2011, Manfred Valentan, Micron Baby, DUT5, source run, post irrad, pre zupf, cal104, 83800, 11.04.2011 18:12:38, 11.04.2011 18:20:50 

DUT5 (atoll)
config file: D:/cern10/micronbaby/data_lab/DUT5/micronbaby_single_irrrad.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=26µA
I_end=48µA

 

Entry  Mon Apr 11 18:37:28 2011, Manfred Valentan, Micron Baby, DUT5, source run, post irrad, pre zupf, run104, 100k, 11.04.2011 18:25:28, 11.04.2011 18:32:11 

DUT5 (atoll)
config file: D:/cern10/micronbaby/data_lab/DUT5/micronbaby_multi6_irrrad.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=50µA
I_end=95µA

 

Entry  Wed Apr 13 12:13:09 2011, Manfred Valentan, Micron Baby, DUT5, source run, post irrad, post zupf ** DO NOT USE **, cal105, 83800, 13.04.2011 12:02:41, 13.04.2011 12:10:55  

DUT5 (atoll), each second bond removed to simulate intermediate strips
config file: D:/cern10/micronbaby/data_lab/DUT5/micronbaby_single_irrrad.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=18µA
I_end=21.5µA

data recorded with new switching power supplies and old intcal, which doesn't ignore bad strips!!!!!

Entry  Wed May 4 15:14:31 2011, Thomas Bergauer, Pt, info about module, , , ,  

pt-Modul hat jeweils zwei OB1 Sensoren von STM (mit 122um Pitch)

im Abstand von 2mm uebereinander

Entry  Wed May 11 16:28:07 2011, Manfred Valentan, Micron Baby, DUT5, source run, post irrad, post zupf, run106, 100k, 11.05.2011 16:00, 11.05.2011 16:25 

DUT5 (atoll)
config file: D:/cern10/micronbaby/data_lab/DUT5/micronbaby_multi6_irrrad.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=52µA
I_end=120µA

 DAQ modified so that only strips marked as good are considered when calculating the common mode correction.

Entry  Wed May 11 16:28:20 2011, Manfred Valentan, Micron Baby, DUT5, source run, post irrad, post zupf, cal106, 100k, 11.05.2011 16:00, 11.05.2011 16:25 

DUT5 (atoll)
config file: D:/cern10/micronbaby/data_lab/DUT5/micronbaby_multi6_irrrad.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=52µA
I_end=120µA

 DAQ modified so that only strips marked as good are considered when calculating the common mode correction.

Entry  Tue May 17 13:02:34 2011, Manfred Valentan, Micron Baby, TRK1, source run, unirrad, prezupf, cal107, 83800, 17.05.2011 12:20, 17.05.2011 12:35 

TRK1: B2Micron_2826_1_baby2 (common p-stop)
config file: D:/cern10/micronbaby/data_lab/TRK1/prezupf/micronbaby_single.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=1.1µA
I_end=2µA

Entry  Tue May 17 13:08:08 2011, Manfred Valentan, Micron Baby, TRK1, source run, unirrad, prezupf, run107, 100k, 17.05.2011 12:39:34, 17.05.2011 13:03:05 run107_all_hits_apv4.pngrun107_all_hits_apv5.pngcal107_pednoise_apv4.pngcal107_pednoise_apv5.png

DUT candidate common p-stop

TRK1: B2Micron_2826_1_baby2 (common p-stop)
config file: D:/cern10/micronbaby/data_lab/TRK1/prezupf/micronbaby_multi6.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=1.1µA
I_end=2µA

n-side: half of zones 1 and 4 obsolete due to noisy strips, rest is okay. DUT Candidate

ELOG V3.1.5-fc6679b