Entry  Sat Mar 30 08:05:59 2024, Christian Irmler, RD50-MPW4, 999, , , , , , , , , , ,  

Tracker calibration run: mpw4-tj-monopix2_00-000999.xml
Generated global xml file for MPW4 shift: mpw4-tj-monopix2_TLU_00-000999.xml 

Previous local runs and xml files (mpw4-tj-monopix2_00-000212.xml) are unusable, because sensor L5.1.1.0 had a loose cotact and thus low gain.
Fixed by disconnecting and reconnecting.

Entry  Sat Jun 8 21:41:04 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs Calibrationcurve, 9, TOF-Cal GSI, 62.4 MeV, Low Flux High, Proton, Good, 20:55, 21:21, , , ,  

changed back to high due to beamsize vs LGAD size factor
650 khz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 230kHz on LGAD

TOF Calibration Curve

Entry  Sat Dec 14 12:56:38 2024, Felix Ulrich-Pur, Shift 23.11.24 low flux He test with RaMon, 9, RaMon, 177.8 MeV/u, Low Flux High, Helium, Good, 06:40, 06:41, 1, , ,  

Mean rate per spill~ 4.5e6 p/s (online estimation without proper analysis)

Entry  Sat Dec 14 15:50:19 2024, Felix Ulrich-Pur, LGAD+Diamond test with clinical carbon beam, 80, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 402MeV/u, , Carbon, Good, 05:15, 05:50, 35, , ,  

Carbon Deg10 (clinical rate)

TRB Setup: 260V bias and 250mA pre-amps, threshold 5000

Diamond: 900mA, threshold 1000 offset

Entry  Sat Jun 8 20:11:00 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs Calibrationcurve, 8, TOF-Cal GSI, 83 MeV, Low Flux Medium, Proton, Good, 19:42, 20:51, , , ,  

300 khz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 35kHz on LGAD

TOF Calibration Curve

Entry  Sat Dec 14 12:56:05 2024, Felix Ulrich-Pur, Shift 23.11.24 low flux He test with RaMon, 8, RaMon, 205.2 MeV/u, Low Flux High, Helium, Good, 06:39, 06:40, 1, , ,  

Mean rate per spill~ 5e6 p/s (online estimation without proper analysis)

Entry  Sat Dec 14 15:49:58 2024, Felix Ulrich-Pur, plastic mouse radiography, 79, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 04:51, 05:05, 14, , plastic mouse,  

pos x=250, posy=-175

TRB Setup: 260V bias and 250mA pre-amps, threshold 5000

Diamond: 900mA, threshold 1000 offset

Entry  Sat Dec 14 15:49:32 2024, Felix Ulrich-Pur, plastic mouse radiography, 78, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 04:36, 04:50, 14, , plastic mouse,  

pos x=175, posy=-175

TRB Setup: 260V bias and 250mA pre-amps, threshold 5000

Diamond: 900mA, threshold 1000 offset

Entry  Sat Dec 14 15:49:12 2024, Felix Ulrich-Pur, plastic mouse radiography, 77, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 04:22, 04:36, 14, , plastic mouse,  

pos x=100, posy=-175

TRB Setup: 260V bias and 250mA pre-amps, threshold 5000

Diamond: 900mA, threshold 1000 offset

Entry  Sat Dec 14 15:48:49 2024, Felix Ulrich-Pur, plastic mouse radiography, 76, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 04:06, 04:21, 15, , plastic mouse,  

pos x=25, posy=-175

TRB Setup: 260V bias and 250mA pre-amps, threshold 5000

Diamond: 900mA, threshold 1000 offset

Entry  Sat Dec 14 15:48:26 2024, Felix Ulrich-Pur, plastic mouse radiography, 75, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 03:52, 04:05, 13, , plastic mouse,  

pos x=-25, posy=-175

TRB Setup: 260V bias and 250mA pre-amps, threshold 5000

Diamond: 900mA, threshold 1000 offset

Entry  Sat Dec 14 15:47:59 2024, Felix Ulrich-Pur, plastic mouse radiography, 74, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 03:35, 03:51, 16, , plastic mouse,  

pos x=-100, posy=-175

TRB Setup: 260V bias and 250mA pre-amps, threshold 5000

Diamond: 900mA, threshold 1000 offset

Entry  Sat Dec 14 15:47:29 2024, Felix Ulrich-Pur, plastic mouse radiography, 73, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 03:20, 03:55, 25, , plastic mouse,  

pos x=-175, posy=-175

TRB Setup: 260V bias and 250mA pre-amps, threshold 5000

Diamond: 900mA, threshold 1000 offset

Entry  Sat Dec 14 15:46:54 2024, Felix Ulrich-Pur, plastic mouse radiography, 72, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 03:03, 03:19, 16, , plastic mouse,  

pos x=-175, posy=75

TRB Setup: 260V bias and 250mA pre-amps, threshold 5000

Diamond: 900mA, threshold 1000 offset

Entry  Sun Aug 25 12:51:10 2024, tb-crew, only base, 71, RD50-MPW4, 62.4 MeV, Low Flux Very Low, Proton, Unknown, 25.08.2024 12:40:41, 25.08.2024 12:51:37, 10, Normal, ,  peary_config.cfgmatrix_mpw4_base.txtmatrix_mpw4_piggy.txtmpw4.conf
automatic log for run 71
Comment:
E-scan
Entry  Sat Dec 14 15:46:16 2024, Felix Ulrich-Pur, plastic mouse radiography, 71, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 02:46, 03:03, 17, , plastic mouse,  

pos x=-100, posy=75

TRB Setup: 260V bias and 250mA pre-amps, threshold 5000

Diamond: 900mA, threshold 1000 offset

Entry  Sun Aug 25 12:39:45 2024, tb-crew, only base, 70, RD50-MPW4, 100.4 MeV, Low Flux Very Low, Proton, Unknown, 25.08.2024 12:24:36, 25.08.2024 12:40:12, 15, Normal, ,  peary_config.cfgmatrix_mpw4_base.txtmatrix_mpw4_piggy.txtmpw4.conf
automatic log for run 70
Comment:
E-scan
Entry  Sat Dec 14 15:45:36 2024, Felix Ulrich-Pur, plastic mouse radiography, 70, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 02:25, 02:45, 20, , plastic mouse,  

pos x=-25, posy=75

TRB Setup: 260V bias and 250mA pre-amps, threshold 5000

Diamond: 900mA, threshold 1000 offset

beam was shortly off at 02:30

Entry  Sat Jun 8 20:09:25 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs Calibrationcurve, 7, TOF-Cal GSI, 100.4 MeV, Low Flux Medium, Proton, Good, 19:00, 19:42, , , ,  

300 khz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 65kHz on LGAD

TOF Calibration Curve

Entry  Sat Dec 14 12:55:41 2024, Felix Ulrich-Pur, Shift 23.11.24 low flux He test with RaMon, 7, RaMon, 63.3 MeV/u, Low Flux Medium, Helium, Good, 06:35, 06:36, 1, , ,  

Mean rate per spill~ 2-5E5 p/s (online estimation without proper analysis)

Entry  Sun Aug 25 12:23:44 2024, tb-crew, only base, 69, RD50-MPW4, 175.3 MeV, Low Flux Very Low, Proton, Unknown, 25.08.2024 12:04:32, 25.08.2024 12:24:10, 19, Normal, ,  peary_config.cfgmatrix_mpw4_base.txtmatrix_mpw4_piggy.txtmpw4.conf
automatic log for run 69
Comment:
E-scan
Entry  Sat Dec 14 15:45:05 2024, Felix Ulrich-Pur, plastic mouse radiography, 69, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 02:07, 02:25, 18, , plastic mouse,  

pos x=25, posy=75

TRB Setup: 260V bias and 250mA pre-amps, threshold 5000

Diamond: 900mA, threshold 1000 offset

Entry  Sun Aug 25 12:00:00 2024, tb-crew, only base, 68, RD50-MPW4, 252.7 MeV, Low Flux Very Low, Proton, Unknown, 25.08.2024 11:37:41, 25.08.2024 12:00:27, 22, Normal, ,  peary_config.cfgmatrix_mpw4_base.txtmatrix_mpw4_piggy.txtmpw4.conf
automatic log for run 68
Comment:
E-scan
Entry  Sat Dec 14 15:43:45 2024, Felix Ulrich-Pur, plastic mouse radiography, 68, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 01:51, 02:07, 16, , plastic mouse,  

pos x=100, posy=75

TRB Setup: 260V bias and 250mA pre-amps, threshold 5000

Diamond: 900mA, threshold 1000 offset

Entry  Sun Aug 25 11:35:21 2024, tb-crew, only base, 67, RD50-MPW4, 252.7 MeV, Low Flux Very Low, Proton, Unknown, 25.08.2024 11:35:33, 25.08.2024 11:35:48, 0, Normal, ,  peary_config.cfgmatrix_mpw4_base.txtmatrix_mpw4_piggy.txtmpw4.conf
automatic log for run 67
Comment:
E-scan
Entry  Sat Dec 14 15:43:01 2024, Felix Ulrich-Pur, plastic mouse radiography, 67, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 01:35, 01:51, 16, , plastic mouse,  

pos x=175, posy=75

TRB Setup: 260V bias and 250mA pre-amps, threshold 5000

Diamond: 900mA, threshold 1000 offset

Entry  Sun Aug 25 11:31:49 2024, tb-crew, only base, 66, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, 25.08.2024 11:21:06, 25.08.2024 11:32:16, 11, Normal, ,  peary_config.cfgmatrix_mpw4_base.txtmatrix_mpw4_piggy.txtmpw4.conf
automatic log for run 66
Comment:
experimental DAC settings
Entry  Sat Dec 14 15:42:32 2024, Felix Ulrich-Pur, plastic mouse radiography, 66, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 01:21, 01:34, 13, , plastic mouse,  

pos x=250, posy=75

TRB Setup: 260V bias and 250mA pre-amps, threshold 5000

Diamond: 900mA, threshold 1000 offset

Entry  Sun Aug 25 11:19:53 2024, tb-crew, only base, 65, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, 25.08.2024 11:17:22, 25.08.2024 11:20:20, 2, Normal, ,  peary_config.cfgmatrix_mpw4_base.txtmatrix_mpw4_piggy.txtmpw4.conf
automatic log for run 65
Comment:
experimental DAC settings
Entry  Sat Dec 14 15:42:01 2024, Felix Ulrich-Pur, plastic mouse radiography, 65, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 01:08, 01:20, 12, , plastic mouse,  

pos x=250, posy=-75

TRB Setup: 260V bias and 250mA pre-amps, threshold 5000

Diamond: 900mA, threshold 1000 offset

Entry  Sun Aug 25 11:16:09 2024, tb-crew, only base, 64, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, 25.08.2024 11:14:19, 25.08.2024 11:16:36, 2, Normal, ,  peary_config.cfgmatrix_mpw4_base.txtmatrix_mpw4_piggy.txtmpw4.conf
automatic log for run 64
Comment:
experimental DAC settings
Entry  Sat Dec 14 15:41:43 2024, Felix Ulrich-Pur, plastic mouse radiography, 64, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 00:54, 01:08, 17, , plastic mouse,  

pos x=175, posy=-75

TRB Setup: 260V bias and 250mA pre-amps, threshold 5000

Diamond: 900mA, threshold 1000 offset

Entry  Sun Aug 25 11:09:44 2024, tb-crew, only base, 63, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, 25.08.2024 10:59:42, 25.08.2024 11:10:11, 10, Normal, ,  peary_config.cfgmatrix_mpw4_base.txtmatrix_mpw4_piggy.txtmpw4.conf
automatic log for run 63
Comment:
vnfb = 60
Entry  Sat Dec 14 15:41:19 2024, Felix Ulrich-Pur, plastic mouse radiography, 63, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 00:37, 00:54, 17, , plastic mouse,  

pos x=100, posy=-75

TRB Setup: 260V bias and 250mA pre-amps, threshold 5000

Diamond: 900mA, threshold 1000 offset

Entry  Sun Aug 25 10:58:26 2024, tb-crew, only base, 62, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, 25.08.2024 10:48:27, 25.08.2024 10:58:53, 10, Normal, ,  peary_config.cfgmatrix_mpw4_base.txtmatrix_mpw4_piggy.txtmpw4.conf
automatic log for run 62
Comment:
vnfb = 10
Entry  Sat Dec 14 15:40:45 2024, Felix Ulrich-Pur, plastic mouse radiography, 62, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 00:22, 00:37, 15, , plastic mouse,  

pos x=25, posy=-75

TRB Setup: 260V bias and 250mA pre-amps, threshold 5000

Diamond: 900mA, threshold 1000 offset

Entry  Sun Aug 25 10:44:10 2024, tb-crew, only base, 61, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, 25.08.2024 10:28:18, 25.08.2024 10:44:37, 16, Normal, ,  peary_config.cfgmatrix_mpw4_base.txtmatrix_mpw4_piggy.txtmpw4.conf
automatic log for run 61
Comment:
normal statistic run
Entry  Sat Dec 14 15:40:09 2024, Felix Ulrich-Pur, plastic mouse radiography, 61, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 00:09, 00:22, 13, , plastic mouse,  

pos x=-25, posy=-75

TRB Setup: 260V bias and 250mA pre-amps, threshold 5000

Diamond: 900mA, threshold 1000 offset

Entry  Sun Aug 25 10:26:57 2024, tb-crew, only base, 60, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, 25.08.2024 10:12:04, 25.08.2024 10:27:24, 15, Normal, ,  peary_config.cfgmatrix_mpw4_base.txtmatrix_mpw4_piggy.txtmpw4.conf
automatic log for run 60
Comment:
normal statistic run; !!! en_tb_mode was set to 0 -> possibly sync problem
Entry  Sat Dec 14 15:39:42 2024, Felix Ulrich-Pur, plastic mouse radiography, 60, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 23:50, 00:08, 18, , plastic mouse,  

pos x=-100, posy=-75

TRB Setup: 260V bias and 250mA pre-amps, threshold 5000

Diamond: 900mA, threshold 1000 offset

Entry  Sat Jun 8 20:08:55 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs Calibrationcurve, 6, TOF-Cal GSI, 115.7MeV, Low Flux Medium, Proton, Good, 18:24, 19:00, , , ,  

400 khz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 65kHz on LGAD

TOF Calibration Curve

Entry  Sat Dec 14 12:54:55 2024, Felix Ulrich-Pur, Shift 23.11.24 low flux He test with RaMon, 6, RaMon, 73.2 MeV/u, Low Flux Medium, Helium, Good, 06:34, 06:34, 1, , ,  

Mean rate per spill~ 4.5-8E5 p/s (online estimation without proper analysis)

Entry  Sun Aug 25 10:06:51 2024, tb-crew, only base, 59, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Crap, 25.08.2024 10:05:01, 25.08.2024 10:07:18, 2, Normal, ,  peary_config.cfgmatrix_mpw4_base.txtmatrix_mpw4_piggy.txtmpw4.conf
automatic log for run 59
Comment:
reducing base threshold
Entry  Sat Dec 14 15:39:11 2024, Felix Ulrich-Pur, plastic mouse radiography, 59, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 23:33, 23:46, 13, , plastic mouse,  

pos x=-175, posy=-75

TRB Setup: 260V bias and 250mA pre-amps, threshold 5000

Diamond: 900mA, threshold 1000 offset

Entry  Sun Aug 25 09:47:51 2024, tb-crew, VNFB scan, 58, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, 25.08.2024 09:37:13, 25.08.2024 09:48:18, 11, Normal, ,  peary_config.cfgmatrix_mpw4_base.txtmatrix_mpw4_piggy.txtmpw4.conf
automatic log for run 58
Comment:
vnfb = 60
Entry  Sat Dec 14 15:37:48 2024, Felix Ulrich-Pur, plastic mouse radiography, 58, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 23:17, 23:30, 13, , plastic mouse,  

pos x=-175, posy=0

TRB Setup: 260V bias and 250mA pre-amps, threshold 5000

Diamond: 900mA, threshold 1000 offset

Entry  Sun Aug 25 09:35:54 2024, tb-crew, VNFB scan, 57, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, 25.08.2024 09:25:43, 25.08.2024 09:36:22, 10, Normal, ,  peary_config.cfgmatrix_mpw4_base.txtmatrix_mpw4_piggy.txtmpw4.conf
automatic log for run 57
Comment:
vnfb = 50
Entry  Sat Dec 14 15:37:26 2024, Felix Ulrich-Pur, plastic mouse radiography, 57, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 23:02, 23:16, 14, , plastic mouse,  

pos x=-100, posy=0

TRB Setup: 260V bias and 250mA pre-amps, threshold 5000

Diamond: 900mA, threshold 1000 offset

Entry  Sun Aug 25 09:24:15 2024, tb-crew, VNFB scan, 56, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, 25.08.2024 09:13:35, 25.08.2024 09:24:42, 11, Normal, ,  peary_config.cfgmatrix_mpw4_base.txtmatrix_mpw4_piggy.txtmpw4.conf
automatic log for run 56
Comment:
vnfb = 40
Entry  Sat Dec 14 15:36:37 2024, Felix Ulrich-Pur, plastic mouse radiography, 56, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 22:46, 23:00, 14, , plastic mouse,  

pos x=-25, posy=0

TRB Setup: 260V bias and 250mA pre-amps, threshold 5000

Diamond: 900mA, threshold 1000 offset

Entry  Sun Aug 25 09:12:07 2024, tb-crew, VNFB scan, 55, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, 25.08.2024 09:01:26, 25.08.2024 09:12:34, 11, Normal, ,  peary_config.cfgmatrix_mpw4_base.txtmatrix_mpw4_piggy.txtmpw4.conf
automatic log for run 55
Comment:
vnfb = 30
Entry  Sat Dec 14 15:35:56 2024, Felix Ulrich-Pur, plastic mouse radiography, 55, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 22:12, 22:45, 33, , plastic mouse,  

pos x=25, posy=0

TRB Setup: 260V bias and 250mA pre-amps, threshold 5000

Diamond: 900mA, threshold 1000 offset

Entry  Sun Aug 25 08:59:46 2024, tb-crew, VNFB scan, 54, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, 25.08.2024 08:53:16, 25.08.2024 09:00:13, 6, Normal, ,  peary_config.cfgmatrix_mpw4_base.txtmatrix_mpw4_piggy.txtmpw4.conf
automatic log for run 54
Comment:
vnfb = 20
Entry  Sat Dec 14 15:34:50 2024, Felix Ulrich-Pur, plastic mouse radiography, 54, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 21:58, 22:10, 22, , plastic mouse,  

pos x=100, posy=0

TRB Setup: 260V bias and 250mA pre-amps, threshold 5000

Diamond: 900mA, threshold 1000 offset

Entry  Sun Aug 25 08:51:02 2024, tb-crew, VNFB scan, 53, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, 25.08.2024 08:36:24, 25.08.2024 08:51:28, 15, Normal, ,  peary_config.cfgmatrix_mpw4_base.txtmatrix_mpw4_piggy.txtmpw4.conf
automatic log for run 53
Comment:
vnfb = 10
Entry  Sat Dec 14 15:33:35 2024, Felix Ulrich-Pur, plastic mouse radiography, 53, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 21:33, 21:57, 24, , plastic mouse,  

run 52 was crap

pos x=175, posy=0

TRB Setup: 260V bias and 250mA pre-amps, threshold 5000

Diamond: 900mA, threshold 1000 offset

Entry  Sun Aug 25 08:33:56 2024, tb-crew, Piggy test, 52, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, 25.08.2024 08:11:40, 25.08.2024 08:34:23, 22, Normal, ,  peary_config.cfgmatrix_mpw4_base.txtmatrix_mpw4_piggy.txtmpw4.conf
automatic log for run 52
Comment:
Entry  Sun Aug 25 08:10:54 2024, tb-crew, Piggy test, 51, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, 25.08.2024 07:45:15, 25.08.2024 08:11:20, 26, Normal, ,  peary_config.cfgmatrix_mpw4_base.txtmatrix_mpw4_piggy.txtmpw4.conf
automatic log for run 51
Comment:
stuck again?
Entry  Sat Dec 14 15:31:32 2024, Felix Ulrich-Pur, plastic mouse radiography, 51, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 21:14, 21:30, 16, , plastic mouse,  

before run, mouse was aligned with lasers and shift of beam w.r.t. sensor center was taken into account

pos x=250, posy=0

TRB Setup: 260V bias and 250mA pre-amps, threshold 5000

Diamond: 900mA, threshold 1000 offset

Entry  Sun Aug 25 07:43:36 2024, tb-crew, Piggy test, 50, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, 25.08.2024 07:19:46, 25.08.2024 07:44:03, 24, Normal, ,  peary_config.cfgmatrix_mpw4_base.txtmatrix_mpw4_piggy.txtmpw4.conf
automatic log for run 50
Comment:
Entry  Sat Dec 14 15:30:23 2024, Felix Ulrich-Pur, threshold and preamp tests, 50, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 20:19, 20:35, 16, , ,  

TRB Setup: 260V bias and 250mA pre-amps, threshold 5000 (looks good, clustersize max at ~3)

Diamond: 900mA, threshold 1000 offset

4MHz Ramon 370kHz TRB

Entry  Sat Jun 8 20:07:44 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs Calibrationcurve, 5, TOF-Cal GSI, 145.4 MeV, Low Flux Medium, Proton, Good, 17:55, 18:24, , , ,  

400 khz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 120kHz on LGAD

TOF Calibration Curve

Entry  Sat Dec 14 12:54:04 2024, Felix Ulrich-Pur, Shift 23.11.24 low flux He test with RaMon, 5, RaMon, 97.2MeV/u, Low Flux Medium, Helium, Good, 06:33, 06:33, 1, , ,  

Mean rate per spill~ 1-5E5 p/s (online estimation without proper analysis)

Entry  Sun Aug 25 07:03:33 2024, tb-crew, Piggy test, 49, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, 25.08.2024 07:02:24, 25.08.2024 07:04:00, 1, Normal, ,  peary_config.cfgmatrix_mpw4_base.txtmatrix_mpw4_piggy.txtmpw4.conf
automatic log for run 49
Comment:
data is correlated
Entry  Sat Dec 14 15:28:46 2024, Felix Ulrich-Pur, threshold and preamp tests, 49, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 20:04, 20:09, 5, , ,  

TRB Setup: 260V bias and 250mA pre-amps, threshold 10000 (was too high)

Diamond: 900mA, threshold 1000 offset

3.5MHz Ramon

Entry  Sun Aug 25 07:01:37 2024, tb-crew, Piggy test, 48, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, 25.08.2024 06:51:46, 25.08.2024 07:02:04, 10, Normal, ,  peary_config.cfgmatrix_mpw4_base.txtmatrix_mpw4_piggy.txtmpw4.conf
automatic log for run 48
Comment:
data is correlated
Entry  Sat Dec 14 15:27:08 2024, Felix Ulrich-Pur, WET measurement slabs, 48, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 147.2 MeV/u, Low Flux High, Helium, Good, 19:30, 19:50, 20, , Dense Bone Slabs, 5cm 

5cm dense bone

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

1.5MHz Ramon

Wave crashes if 1 TDC disappears in dog d, DABC is still running

Entry  Sun Aug 25 06:51:00 2024, tb-crew, Piggy test, 47, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, 25.08.2024 06:48:29, 25.08.2024 06:51:27, 2, Normal, ,  peary_config.cfgmatrix_mpw4_base.txtmatrix_mpw4_piggy.txtmpw4.conf
automatic log for run 47
Comment:
with tracker
Entry  Sat Dec 14 15:26:18 2024, Felix Ulrich-Pur, WET measurement slabs, 47, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 205.2 MeV/u, Low Flux High, Helium, Good, 19:16, 19:29, 13, , Dense Bone Slabs, 5cm 

5cm dense bone

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

3MHz Ramon 300kHz TRB

Entry  Sun Aug 25 06:45:12 2024, tb-crew, Piggy test, 46, RD50-MPW4, 800 MeV, Low Flux Low, Proton, Unknown, 25.08.2024 06:44:33, 25.08.2024 06:45:39, 1, Normal, ,  peary_config.cfgmatrix_mpw4_base.txtmatrix_mpw4_piggy.txtmpw4.conf
automatic log for run 46
Comment:
first tests, trying to get beam right beam
Entry  Sat Dec 14 15:25:12 2024, Felix Ulrich-Pur, WET measurement slabs, 46, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 147.2 MeV/u, Low Flux High, Helium, Good, 18:55, 19:11, 16, , Dense Bone Slabs, 4cm 

4cm dense bone

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

2MHz Ramon 120kHz TRB

Entry  Sun Aug 25 06:42:38 2024, tb-crew, Piggy test, 45, RD50-MPW4, 800 MeV, Low Flux Low, Proton, Unknown, 25.08.2024 06:40:52, 25.08.2024 06:43:04, 2, Normal, ,  peary_config.cfgmatrix_mpw4_base.txtmatrix_mpw4_piggy.txtmpw4.conf
automatic log for run 45
Comment:
Entry  Sat Dec 14 15:24:22 2024, Felix Ulrich-Pur, WET measurement slabs, 45, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 205.2 MeV/u, Low Flux High, Helium, Good, 18:37, 18:51, 14, , Dense Bone Slabs, 4cm 

4cm dense bone

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

3MHz Ramon 800kHz TRB

Entry  Sun Aug 25 06:34:56 2024, tb-crew, Piggy test, 44, RD50-MPW4, 800 MeV, Low Flux Low, Proton, Unknown, 25.08.2024 06:35:13, 25.08.2024 06:35:23, 0, Normal, ,  peary_config.cfgmatrix_mpw4_base.txtmatrix_mpw4_piggy.txtmpw4.conf
automatic log for run 44
Comment:
Entry  Sat Dec 14 15:23:21 2024, Felix Ulrich-Pur, WET measurement slabs, 44, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 147.2 MeV/u, Low Flux High, Helium, Good, 18:19, 18:33, 14, , Dense Bone Slabs, 2cm 

3cm dense bone

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

4MHz Ramon 200kHz TRB

Entry  Sat Dec 14 15:21:44 2024, Felix Ulrich-Pur, WET measurement slabs, 43, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 205.2 MeV/u, Low Flux High, Helium, Good, 18:04, 18:18, 14, , Dense Bone Slabs, 3cm 

3cm dense bone

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

4MHz Ramon 1MHz TRB

Entry  Sat Dec 14 15:20:52 2024, Felix Ulrich-Pur, WET measurement slabs, 42, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 147.2 MeV/u, Low Flux High, Helium, Good, 17:48, 18:00, 12, , Dense Bone Slabs, 2cm 

2cm dense bone

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

2.5MHz Ramon 250 kHz TRB

Entry  Sat Dec 14 15:19:57 2024, Felix Ulrich-Pur, WET measurement slabs, 41, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 205.2 MeV/u, Low Flux High, Helium, Good, 17:32, 17:44, 12, , Dense Bone Slabs, 2cm 

2cm dense bone

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

4MHz Ramon 600 kHz TRB

Entry  Sat Dec 14 15:18:48 2024, Felix Ulrich-Pur, WET measurement slabs, 40, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 147.2 MeV/u, Low Flux High, Helium, Good, 17:14, 17:28, 14, , Dense Bone Slabs, 1cm 

1cm dense bone

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

3.5MHz Ramon 400 kHz TRB

Entry  Sat Jun 8 20:05:27 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs Calibrationcurve, 4, TOF-Cal GSI, 175.3 MeV, Low Flux Medium, Proton, Good, 17:31, 17:54, , , ,  

500 khz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 170kHz on LGAD

TOF Calibration Curve

Entry  Sat Dec 14 12:53:28 2024, Felix Ulrich-Pur, Shift 23.11.24 low flux He test with RaMon, 4, RaMon, 124.4 MeV/u, Low Flux Medium, Helium, Good, 06:33, 06:33, 1, , ,  

Mean rate per spill~ 2-8E5 p/s (online estimation without proper analysis)

Entry  Sat Dec 14 15:17:57 2024, Felix Ulrich-Pur, WET measurement slabs, 39, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 205.2 MeV/u, Low Flux High, Helium, Good, 17:00, 17:14, 13, , Dense Bone Slabs, 1cm 

1cm dense bone

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

4MHz Ramon 700 kHz TRB

Entry  Sat Dec 14 15:17:10 2024, Felix Ulrich-Pur, WET measurement slabs, 38, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 147.2 MeV/u, Low Flux High, Helium, Good, 16:42, 16:55, 13, , Plastic Water Slabs, 5cm 

5cm plastic water

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

2.5MHz Ramon 200 kHz TRB

Entry  Sat Dec 14 15:15:46 2024, Felix Ulrich-Pur, WET measurement slabs, 37, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 205.2 MeV/u, Low Flux High, Helium, Good, 16:28, 16:42, 12, , Plastic Water Slabs, 5cm 

5cm plastic water

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

3-5MHz Ramon 500 kHz TRB

Entry  Sat Dec 14 15:15:04 2024, Felix Ulrich-Pur, WET measurement slabs, 36, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 147.2 MeV/u, Low Flux High, Helium, Good, 16:11, 16:23, 12, , Plastic Water Slabs, 4cm 

4cm plastic water

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

3MHz Ramon 300 kHz TRB

Entry  Sat Dec 14 15:12:46 2024, Felix Ulrich-Pur, WET measurement slabs, 35, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 205.2 MeV/u, Low Flux High, Helium, Good, 15:53, 16:10, 17, , Plastic Water Slabs, 4cm 

4cm plastic water

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

4MHz Ramon 600 kHz TRB

Entry  Sat Dec 14 15:12:01 2024, Felix Ulrich-Pur, WET measurement slabs, 34, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 147.2 MeV/u, Low Flux High, Helium, Good, 15:37, 15:48, 11, , Plastic Water Slabs, 3cm 

3cm plastic water

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

3MHz Ramon 300 kHz TRB

Entry  Sat Dec 14 15:10:35 2024, Felix Ulrich-Pur, WET measurement slabs, 33, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 205.2 MeV/u, Low Flux High, Helium, Good, 15:24, 15:36, 12, , Plastic Water Slabs, 3cm 

3cm plastic water

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

3-4MHz Ramon 600 kHz TRB

Entry  Sat Dec 14 15:09:34 2024, Felix Ulrich-Pur, WET measurement slabs, 32, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 147.2 MeV/u, Low Flux High, Helium, Good, 15:06, 15:19, 13, , Plastic Water Slabs, 2cm 

2cm plastic water

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

 

 

Entry  Sat Dec 14 13:56:16 2024, Felix Ulrich-Pur, WET measurement slabs, 31, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 205.2 MeV/u, Low Flux High, Helium, Good, 14:48, 15:06, 18, , Plastic Water Slabs, 2cm 

2cm plastic water

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

4MHz Ramon 1.4MHz TRB

Entry  Sat Dec 14 13:49:20 2024, Felix Ulrich-Pur, WET measurement slabs, 30, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 147.2 MeV/u, Low Flux High, Helium, Good, 14:02, 14:45, 43, , Plastic Water Slabs, 1cm 

1cm plastic water

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

4MHz Ramon 500 kHz TRB

Entry  Sat Jun 8 19:57:39 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs Calibrationcurve, 3, TOF-Cal GSI, 194.3 MeV, Low Flux Medium, Proton, Good, 16:51, 17:28, , , ,  

600 khz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 235kHz on LGAD

TOF Calibration Curve

Entry  Sat Dec 14 12:52:57 2024, Felix Ulrich-Pur, Shift 23.11.24 low flux He test with RaMon, 3, RaMon, 147.2 MeV/u, Low Flux Medium, Helium, Good, 06:32, 06:33, 1, , ,  

Mean rate per spill~ 2.5-4-5E5 p/s (online estimation without proper analysis)

Entry  Sun Jun 9 02:18:55 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Knochen, 29, TOF-Cal GSI, 194.3 MeV, Low Flux High, Proton, Good, 02:18, , , , ,  

5 cm dense bone (in isocentre)
1.8 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 350 kHz on LGAD

WET measurement

Entry  Sat Dec 14 13:46:11 2024, Felix Ulrich-Pur, WET measurement slabs, 29, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 205.2 MeV/u, Low Flux High, Helium, Good, 13:34, 14:03, 29, , Plastic Water Slabs, 1cm 

1cm Plastic Water

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000 (changed threshold)

Diamond: 900mA, threshold 1000 offset

DOGMA lost DiRICH after restarting Wave (happened several times)

Entry  Sun Jun 9 02:02:16 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Knochen, 28, TOF-Cal GSI, 145.4 MeV, Low Flux High, Proton, Good, 02:00, 02:14, , , ,  

4 cm dense bone (in isocentre)
1.4 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 175 kHz on LGAD

WET measurement

Entry  Sat Dec 14 13:45:34 2024, Felix Ulrich-Pur, threshold and preamp tests , 28, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 63.3 MeV/u, Low Flux High, Helium, Good, 13:20, 13:32, 2, , ,  

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000 (changed threshold)

Diamond: 900mA, threshold 1000 offset

Entry  Sun Jun 9 01:48:51 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Knochen, 27, TOF-Cal GSI, 194.3 MeV, Low Flux High, Proton, Good, 01:48, 01:59, , , ,  

4 cm dense bone (in isocentre)
2 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 470 kHz on LGAD

WET measurement

Entry  Sat Dec 14 13:42:52 2024, Felix Ulrich-Pur, threshold and preamp tests , 27, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 63.3 MeV/u, Low Flux High, Helium, Good, 12:52, 13:16, 24, , ,  

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000 (changed threshold)

Diamond: 900mA, threshold 50

Entry  Sun Jun 9 01:32:51 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Knochen, 26, TOF-Cal GSI, 145.4 MeV, Low Flux High, Proton, Good, 01:30, 01:45, , , ,  

3 cm dense bone (in isocentre)
1.7 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 230 kHz on LGAD

WET measurement

Entry  Sat Dec 14 13:40:47 2024, Felix Ulrich-Pur, threshold and preamp tests, 26, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 63.3 MeV/u, Low Flux High, Helium, Good, 12:35, 12:49, 14, , ,  

TRB Setup: 260V bias and 250mA pre-amps, threshold 10000 (changed threshold)

Diamond: 500mA, threshold 50

Entry  Sun Jun 9 01:27:38 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Knochen, 25, TOF-Cal GSI, 194.3 MeV, Low Flux High, Proton, Good, 01:16, 01:30, , , ,  

3 cm dense bone (in isocentre)
2.8 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 600 kHz on LGAD

WET measurement

Entry  Sat Dec 14 13:39:15 2024, Felix Ulrich-Pur, Energy vs TOF calibration, 25, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 63.3 MeV/u, Low Flux High, Helium, Good, 12:05, 12:23, 18, , ,  

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

380 kHz with Ramon and only 180kHz TRB, 42ps single channel res prae calib

Entry  Sun Jun 9 01:26:56 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Knochen, 24, TOF-Cal GSI, 145.4 MeV, Low Flux High, Proton, Good, 01:02, 01:14, , , ,  

2 cm dense bone (in isocentre)
2.1 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 315 kHz on LGAD

WET measurement

Entry  Sat Dec 14 13:37:26 2024, Felix Ulrich-Pur, Energy vs TOF calibration, 24, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 73.2 MeV/u, Low Flux High, Helium, Good, 11:42, 12:00, 18, , ,  

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

2MHz with Ramon and only 200kHz TRB

Entry  Sun Jun 9 01:25:26 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Knochen, 23, TOF-Cal GSI, 194.3 MeV, Low Flux High, Proton, Good, 00:30, 01:02, , , ,  

2 cm dense bone (in isocentre)
- Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 1 MHz on LGAD

interlock at 00:31 till 00:50 --> 19min break

WET measurement

Entry  Sat Dec 14 13:35:59 2024, Felix Ulrich-Pur, Energy vs TOF calibration, 23, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 124.4 MeV/u, Low Flux High, Helium, Good, 11:24, 11:37, 13, , ,  

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

5MHz with Ramon and only 500kHz TRB

Entry  Sun Jun 9 01:22:48 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Knochen, 22, TOF-Cal GSI, 145.4 MeV, Low Flux High, Proton, Good, 00:14, 00:23, , , ,  

1 cm dense bone (in isocentre)
2.8 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 600 kHz on LGAD

WET measurement

Entry  Sun Jun 9 00:12:41 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Knochen, 21, TOF-Cal GSI, 194.3 MeV, Low Flux High, Proton, Good, 00:00, 00:14, , , ,  

1 cm dense bone (in isocentre)
4.5 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 1.2 MHz on LGAD

WET measurement

Entry  Sat Dec 14 13:31:49 2024, Felix Ulrich-Pur, Energy vs TOF calibration, 21, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 147.2 MeV/u, Low Flux High, Helium, Good, 10:41, 10:56, 15, , ,  

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

3.5MHz with Ramon and only 600kHz with TRB-52ps single channel res prae calib

Entry  Sat Dec 14 13:34:18 2024, Felix Ulrich-Pur, Energy vs TOF calibration, 21, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 10:58, 11:23, 25, , ,  

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

3MHz with Ramon and only 300kHz with TRB-44ps single channel res prae calib

Entry  Sun Jun 9 00:10:38 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser, 20, TOF-Cal GSI, 145.4 MeV, Low Flux High, Proton, Good, 23:31, 00:00, , , ,  

5 cm plastic water (in isocentre)
2 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 330 kHz on LGAD

14 min break with beam off (DAQ failed, had to power cycle TRB3s and all DIRICH boards-->out of synch DIRICH)

WET measurement

Entry  Sat Dec 14 13:28:57 2024, Felix Ulrich-Pur, Energy vs TOF calibration, 20, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 147.2 MeV/u, Low Flux Medium, Helium, Good, 10:33, 10:38, 5, , ,  

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

300kHz with Ramon and only 40kHz with TRB--> decision to go to low flux high was made

Entry  Sat Jun 8 19:56:37 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs Calibrationcurve, 2, TOF-Cal GSI, 252.7 MeV, Low Flux Medium, Proton, Good, 16:29, 16:47, , , ,  

1 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 350kHz on LGAD

TOF Calibration Curve

Entry  Sat Dec 14 12:52:02 2024, Felix Ulrich-Pur, Shift 23.11.24 low flux He test with RaMon, 2, RaMon, 177.8 MeV/u, Low Flux Medium, Helium, Good, 06:31, 06:32, 1, , ,  

Mean rate per spill~ 6-8E5 p/s (online estimation without proper analysis)

Entry  Sun Jun 9 00:08:07 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser, 19, TOF-Cal GSI, 194.3 MeV, Low Flux High, Proton, Good, 23:20, 23:30, , , ,  

5 cm plastic water (in isocentre)
3.7 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 795 kHz on LGAD

WET measurement

Entry  Sat Dec 14 13:26:29 2024, Felix Ulrich-Pur, Energy vs TOF calibration, 19, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 177.8 MeV/u, Low Flux Medium, Helium, Good, 10:08, 10:25, 17, , ,  

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

750kHz with Ramon and only 160kHz with TRB, 44ps single channel res preacalib

Entry  Sun Jun 9 00:07:17 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser, 18, TOF-Cal GSI, 145.4 MeV, Low Flux High, Proton, Good, 23:04, 23:20, , , ,  

4 cm plastic water (in isocentre)
2 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 500 kHz on LGAD

WET measurement

Entry  Sat Dec 14 13:16:33 2024, Felix Ulrich-Pur, Energy vs TOF calibration +Alignment of DOGMA setup, 18, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 205.2 MeV/u, Low Flux Medium, Helium, Good, 09:42, 09:48, 6, , ,  

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

DOGMA setup had to be shifted manually to align it with the beam

Entry  Sun Jun 9 00:06:01 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser, 17, TOF-Cal GSI, 194.3 MeV, Low Flux High, Proton, Good, 22:53, 23:04, , , ,  

4 cm plastic water (in isocentre)
3.8 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 900 kHz on LGAD

WET measurement

Entry  Sat Dec 14 13:07:15 2024, Felix Ulrich-Pur, Energy vs TOF calibration, 17, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 205.2 MeV/u, Low Flux Medium, Helium, Good, 08:46, 09:42, 56, , ,  

before run, setups had to be aligned (first slab in slabholder was put in isocentre. However, in this run, no slab was used). Room was closed at 08:10

Beam->Nozzle-> Slabholder->TOF-CAL (TRB)-> DOGMA Setup(Diamond+LGADs)-> RaMon

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

measured 400kHz with Ramon and only 100kHz with LGADs (third LGAD) and 60kHz accepted trrigger rate

beam was shifted vertically by x? cm? (need to lookup again). So TRB setup was shifted with motors such that the He beam was hitting the centre of the LGADs

Entry  Sun Jun 9 00:05:11 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser (wrong energy), 16, TOF-Cal GSI, 194.3 MeV, Low Flux High, Proton, Crap, 22:50, 22:52, , , ,  

wrong energy don't use

Entry  Sat Dec 14 13:04:57 2024, Felix Ulrich-Pur, Shift 23.11.24 low flux Carbon test with RaMon (did not work), 16, RaMon, , Low Flux High, Carbon, Crap, 07:02, 07:20, 18, , ,  

Low flux carbon did not work

5µs EFE, 10s Extraction, Deg100
we tried different energies (402.MeV/u)-> 238 MeV/u, but unfortunately we did not get any beam

Entry  Sun Jun 9 00:04:23 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser, 15, TOF-Cal GSI, 194.3 MeV, Low Flux High, Proton, Good, 22:34, 22:50, , , ,  

3 cm plastic water (in isocentre)
2.6 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 400 kHz on LGAD

WET measurement

Entry  Sat Dec 14 13:02:29 2024, Felix Ulrich-Pur, Shift 23.11.24 low flux Carbon test with RaMon (did not work), 15, RaMon, , Low Flux High, Carbon, Crap, 06:55, 07:01, 6, , ,  

Low flux carbon did not work

1µs EFE, 30s Extraction, Deg100
we tried different energies (402.MeV/u)-> 120 MeV/u, but unfortunately we did not get any beam

Entry  Sat Jun 8 22:38:41 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser, 14, TOF-Cal GSI, 194.3 MeV, Low Flux High, Proton, Good, 22:19, 22.34, , , ,  

3 cm plastic water (in isocentre)
4.3 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 1MHz on LGAD

WET measurement

Entry  Sat Dec 14 13:01:14 2024, Felix Ulrich-Pur, Shift 23.11.24 low flux He test with RaMon, 14, RaMon, 63.3 MeV/u, Low Flux High, Helium, Good, 06:46, 06:46, 1, , ,  

Mean rate per spill~ 5Ee6 p/s (online estimation without proper analysis)

Entry  Sat Jun 8 22:12:06 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser, 13, TOF-Cal GSI, 145.4 MeV, Low Flux High, Proton, Good, 22:05, 22:34, , , ,  

2cm plastic water (in isocentre)
3.2 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 600 kHz on LGAD

WET measurement

Entry  Sat Dec 14 13:00:26 2024, Felix Ulrich-Pur, Shift 23.11.24 low flux He test with RaMon, 13, RaMon, 73.2 MeV/u, Low Flux High, Helium, Good, 06:44, 06:45, 1, , ,  

Mean rate per spill~ 4Ee6 p/s (online estimation without proper analysis)

Entry  Sat Jun 8 21:54:59 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser, 12, TOF-Cal GSI, 194.3 MeV, Low Flux High, Proton, Good, 21:53, 22:05, , , ,  

2cm plastic water (in isocentre)
5 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 1.2 MHz on LGAD

WET measurement

Entry  Sat Dec 14 12:59:33 2024, Felix Ulrich-Pur, Shift 23.11.24 low flux He test with RaMon, 12, RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 06:43, 06:44, 1, , ,  

Mean rate per spill~ 6Ee6 p/s (online estimation without proper analysis)

Entry  Sat Jun 8 21:43:26 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser, 11, TOF-Cal GSI, 145.4 MeV, Low Flux High, Proton, Good, 21:39, 21:48, , , ,  

1cm plastic water (in isocentre)
4 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 700 kHz on LGAD

WET measurement

Entry  Sat Dec 14 12:58:35 2024, Felix Ulrich-Pur, Shift 23.11.24 low flux He test with RaMon, 11, RaMon, 124.4 MeV/u, Low Flux High, Helium, Good, 06:43, 06:43, 1, , ,  

Mean rate per spill~ 4Ee6 p/s (online estimation without proper analysis)

Entry  Sat Mar 30 20:51:31 2024, hh, mb, ps, High statistics, 1054, TJ-Monopix2, 800 MeV, Low Flux High, Proton, Unknown, 30.03.2024 20:50:53, 30.03.2024 20:53:58, 3, HV, ,  tjmonopix2.conf
automatic log for run 1054
Comment:
Litle high settings run
Entry  Sat Mar 30 20:46:41 2024, hh, mb, ps, High statistics, 1053, TJ-Monopix2, 800 MeV, Low Flux High, Proton, Unknown, 30.03.2024 20:31:14, 30.03.2024 20:49:08, 17, HV, ,  tjmonopix2.conf
automatic log for run 1053
Comment:
Back to stock veto period of 7us (100)
Lets try with high
Entry  Sat Mar 30 20:27:35 2024, hh, mb, ps, High statistics, 1052, TJ-Monopix2, 800 MeV, Low Flux Medium, Proton, Unknown, 30.03.2024 20:17:11, 30.03.2024 20:30:02, 12, HV, ,  tjmonopix2.conf
automatic log for run 1052
Comment:
Back to stock veto period of 7us (100)
Entry  Sat Mar 30 20:13:29 2024, hh, mb, ps, High statistics, 1051, TJ-Monopix2, 800 MeV, Low Flux Medium, Proton, Unknown, 30.03.2024 19:48:36, 30.03.2024 20:15:55, 27, HV, ,  tjmonopix2.conf
automatic log for run 1051
Comment:
Try with increased veto period of 20us (800)
Entry  Sat Mar 30 19:44:43 2024, hh, mb, ps, High statistics, 1050, TJ-Monopix2, 800 MeV, Low Flux Medium, Proton, Unknown, 30.03.2024 19:40:21, 30.03.2024 19:47:10, 6, HV, ,  tjmonopix2.conf
automatic log for run 1050
Comment:
Try with increased veto period of 20us (800)
Entry  Sat Mar 30 19:36:57 2024, hh, mb, ps, High statistics, 1049, TJ-Monopix2, 800 MeV, Low Flux Medium, Proton, Unknown, 30.03.2024 19:28:44, 30.03.2024 19:39:24, 10, HV, ,  tjmonopix2.conf
automatic log for run 1049
Comment:
Try with increased veto period of 20us (800)
Entry  Sat Mar 30 19:24:57 2024, hh, mb, ps, High statistics, 1048, TJ-Monopix2, 800 MeV, Low Flux Medium, Proton, Unknown, 30.03.2024 19:02:52, 30.03.2024 19:27:24, 24, HV, ,  tjmonopix2.conf
automatic log for run 1048
Comment:
Run with medium settings
Entry  Sat Mar 30 18:59:16 2024, hh, mb, ps, 1st tests, 1047, TJ-Monopix2, 800 MeV, Low Flux Low, Proton, Crap, 30.03.2024 19:00:11, 30.03.2024 19:01:43, 1, HV, ,  tjmonopix2.conf
automatic log for run 1047
Comment:
Test if we get beam, MAUS has problems
Entry  Sat Mar 30 17:31:59 2024, hh, mb, ps, High statistics, 1046, TJ-Monopix2, 800 MeV, Low Flux Low, Proton, Unknown, 30.03.2024 17:13:09, 30.03.2024 17:34:25, 21, HV, ,  tjmonopix2.conftestbench_AB.yamltestbench_CD.yaml
automatic log for run 1046
Comment:
High statistics run, from now we fix the geometry

Veto: 7us (100, so 7 us is handshake time)
Entry  Sat Mar 30 17:07:45 2024, hh, mb, ps, 1st tests, 1045, TJ-Monopix2, 800 MeV, Low Flux Low, Proton, Unknown, 30.03.2024 17:07:01, 30.03.2024 17:10:12, 3, HV, ,  tjmonopix2.conf
automatic log for run 1045
Comment:
Entry  Sat Mar 30 17:00:53 2024, hh, mb, ps, 1st tests, 1044, TJ-Monopix2, 800 MeV, Low Flux Low, Proton, Unknown, 30.03.2024 16:57:08, 30.03.2024 17:03:20, 6, HV, ,  tjmonopix2.conf
automatic log for run 1044
Comment:
Entry  Sat Mar 30 16:50:35 2024, hh, mb, ps, 1st tests, 1043, TJ-Monopix2, 800 MeV, Low Flux Low, Proton, Unknown, 30.03.2024 16:48:39, 30.03.2024 16:53:01, 4, HV, ,  tjmonopix2.conf
automatic log for run 1043
Comment:
Entry  Sat Mar 30 16:38:41 2024, hh, mb, ps, 1st tests, 1042, TJ-Monopix2, 800 MeV, Low Flux Low, Proton, Unknown, 30.03.2024 16:32:12, 30.03.2024 16:41:08, 8, HV, ,  tjmonopix2.conf
automatic log for run 1042
Comment: First good run
Entry  Sat Mar 30 16:24:39 2024, hh, mb, ps, 1st tests, 1041, TJ-Monopix2, 800 MeV, Low Flux Low, Proton, Unknown, 30.03.2024 16:20:34, 30.03.2024 16:27:05, 6, HV, ,  tjmonopix2.conf
automatic log for run 1041
Comment:
Entry  Sat Mar 30 16:15:20 2024, hh, mb, ps, 1st tests, 1040, TJ-Monopix2, 800 MeV, Low Flux Low, Proton, Unknown, 30.03.2024 16:15:22, 30.03.2024 16:17:46, 2, HV, ,  tjmonopix2.conf
automatic log for run 1040
Comment:
Entry  Sat Mar 30 16:09:54 2024, hh, mb, ps, 1st tests, 1039, TJ-Monopix2, 800 MeV, Low Flux Low, Proton, Unknown, 30.03.2024 16:11:32, 30.03.2024 16:12:21, 0, HV, ,  tjmonopix2.conf
automatic log for run 1039
Comment:
Entry  Sat Mar 30 16:03:35 2024, hh, mb, ps, 1st tests, 1038, TJ-Monopix2, 800 MeV, Low Flux Low, Proton, Unknown, 30.03.2024 16:03:18, 30.03.2024 16:06:01, 2, HV, ,  tjmonopix2.conf
automatic log for run 1038
Comment:
Entry  Sat Mar 30 15:55:32 2024, hh, mb, ps, 1st tests, 1037, TJ-Monopix2, 800 MeV, Low Flux Low, Proton, Unknown, 30.03.2024 15:54:44, 30.03.2024 15:57:59, 3, HV, ,  tjmonopix2.conf
automatic log for run 1037
Comment:
Entry  Sat Mar 30 15:51:21 2024, hh, mb, ps, 1st tests, 1036, TJ-Monopix2, 800 MeV, Low Flux Low, Proton, Unknown, 30.03.2024 15:43:53, 30.03.2024 15:53:47, 9, HV, ,  tjmonopix2.conf
automatic log for run 1036
Comment:
Entry  Sat Mar 30 15:39:06 2024, hh, mb, ps, 1st tests, 1035, TJ-Monopix2, 800 MeV, Low Flux Low, Proton, Unknown, 30.03.2024 15:35:06, 30.03.2024 15:41:33, 6, HV, ,  tjmonopix2.conf
automatic log for run 1035

All four chips enabled now


Comment:
Entry  Sat Mar 30 15:31:41 2024, hh, mb, ps, 1st tests, 1034, TJ-Monopix2, 800 MeV, Low Flux Low, Proton, Unknown, 30.03.2024 15:29:49, 30.03.2024 15:34:08, 4, HV, ,  tjmonopix2.conf
automatic log for run 1034
Comment:
Entry  Sat Mar 30 15:24:44 2024, hh, mb, ps, 1st tests, 1033, TJ-Monopix2, 800 MeV, Low Flux Low, Proton, Unknown, 30.03.2024 15:25:06, 30.03.2024 15:27:10, 2, HV, ,  tjmonopix2.conf
automatic log for run 1033
Comment:
Entry  Sat Mar 30 15:22:15 2024, hh, mb, ps, 1st tests, 1032, TJ-Monopix2, 800 MeV, Low Flux Low, Proton, Unknown, 30.03.2024 15:24:02, 30.03.2024 15:24:41, 0, HV, ,  tjmonopix2.conf
automatic log for run 1032
Comment:
Entry  Sat Mar 30 15:20:37 2024, hh, mb, ps, 1st tests, 1031, TJ-Monopix2, 800 MeV, Low Flux Low, Proton, Unknown, 30.03.2024 15:21:51, 30.03.2024 15:23:03, 1, HV, ,  tjmonopix2.conf
automatic log for run 1031
Comment:
Entry  Sat Mar 30 15:18:29 2024, hh, mb, ps, 1st tests, 1030, TJ-Monopix2, 800 MeV, Low Flux Low, Proton, Unknown, 30.03.2024 15:18:21, 30.03.2024 15:20:55, 2, HV, ,  tjmonopix2.conf
automatic log for run 1030
Comment:
Entry  Sat Mar 30 15:08:33 2024, hh, mb, ps, 1st tests, 1029, TJ-Monopix2, 800 MeV, Low Flux Low, Proton, Unknown, 30.03.2024 15:08:26, 30.03.2024 15:11:00, 2, HV, ,  tjmonopix2.conf
automatic log for run 1029
Comment:
Entry  Sat Mar 30 15:05:28 2024, hh, mb, ps, 1st tests, 1028, TJ-Monopix2, 800 MeV, Low Flux Low, Proton, Unknown, 30.03.2024 15:04:47, 30.03.2024 15:07:55, 3, HV, ,  tjmonopix2.conf
automatic log for run 1028
Comment:
Entry  Sat Mar 30 15:02:07 2024, hh, mb, ps, 1st tests, 1027, TJ-Monopix2, 800 MeV, Low Flux Low, Proton, Unknown, 30.03.2024 14:52:54, 30.03.2024 15:04:32, 11, HV, ,  tjmonopix2.conf
automatic log for run 1027
Comment:
Entry  Sat Mar 30 13:28:15 2024, bp, ci, hh, useful datataking, 1026, RD50-MPW4, 800 MeV, Low Flux Low, Proton, Good, 30.03.2024 12:58:02, 30.03.2024 13:30:42, 32, Normal, ,  peary_config.cfgmatrix_mpw4.txtmpw4.conf
automatic log for run 1026
Comment:
changed thr to 0.98V
Entry  Sat Mar 30 12:54:24 2024, bp, ci, hh, useful datataking, 1025, RD50-MPW4, 800 MeV, Low Flux Low, Proton, Good, 30.03.2024 12:32:27, 30.03.2024 12:56:51, 24, Normal, ,  peary_config.cfgmatrix_mpw4.txtmpw4.conf
automatic log for run 1025
Comment:
changed thr to 0.95V
Entry  Sat Mar 30 12:28:29 2024, bp, ci, hh, useful datataking, 1024, RD50-MPW4, 800 MeV, Low Flux Low, Proton, Good, 30.03.2024 11:58:45, 30.03.2024 12:30:56, 32, Normal, ,  peary_config.cfgmatrix_mpw4.txtmpw4.conf
automatic log for run 1024
Comment:
changed thr to 1V (not 100% sure, configure was not executed)
Entry  Sat Mar 30 11:55:55 2024, bp, ci, hh, useful datataking, 1023, RD50-MPW4, 800 MeV, Low Flux Low, Proton, Good, 30.03.2024 11:06:49, 30.03.2024 11:58:22, 51, Normal, ,  peary_config.cfgmatrix_mpw4.txtmpw4.conf
automatic log for run 1023
Comment:
Entry  Sat Mar 30 11:04:10 2024, bp, ci, hh, RD50-MPW4, 1022, RD50-MPW4, 800 MeV, Low Flux Low, Proton, Good, 30.03.2024 11:03:05, 30.03.2024 11:06:37, 3, Normal, ,  peary_config.cfgmatrix_mpw4.txtmpw4.conf
automatic log for run 1022
Comment:
changed alignment again
Entry  Sat Mar 30 10:39:45 2024, bp, ci, hh, RD50-MPW4, 1021, RD50-MPW4, 800 MeV, Low Flux Low, Proton, Good, 30.03.2024 10:37:01, 30.03.2024 10:42:12, 5, Normal, ,  peary_config.cfgmatrix_mpw4.txtmpw4.conf
automatic log for run 1021
Comment:
no tracker at start of run
Entry  Sat Mar 30 10:34:27 2024, bp, ci, hh, RD50-MPW4, 1020, RD50-MPW4, 800 MeV, Low Flux Low, Proton, Good, 30.03.2024 10:33:01, 30.03.2024 10:36:53, 3, Normal, ,  peary_config.cfgmatrix_mpw4.txtmpw4.conf
automatic log for run 1020
Comment:
Entry  Sat Mar 30 10:27:56 2024, bp, ci, hh, RD50-MPW4, 1019, RD50-MPW4, 800 MeV, Low Flux Low, Proton, Crap, 30.03.2024 10:28:28, 30.03.2024 10:30:22, 1, Normal, ,  peary_config.cfgmatrix_mpw4.txtmpw4.conf
automatic log for run 1019
Comment:
Entry  Sat Mar 30 10:25:55 2024, bp, ci, hh, RD50-MPW4, 1018, RD50-MPW4, 800 MeV, Low Flux Low, Proton, Crap, 30.03.2024 10:26:17, 30.03.2024 10:28:21, 2, Normal, ,  peary_config.cfgmatrix_mpw4.txtmpw4.conf
automatic log for run 1018
Comment:
Entry  Sat Mar 30 10:23:06 2024, bp, ci, hh, RD50-MPW4, 1017, RD50-MPW4, 800 MeV, Low Flux Low, Proton, Crap, 30.03.2024 10:25:09, 30.03.2024 10:25:33, 0, Normal, ,  peary_config.cfgmatrix_mpw4.txtmpw4.conf
automatic log for run 1017
Comment:
changed rate to low again, too much "noise"
also moved table down and a bit to the left

no HV, is crap
Entry  Sat Mar 30 10:10:57 2024, bp, ci, hh, RD50-MPW4, 1016, RD50-MPW4, 800 MeV, Low Flux Medium, Proton, Good, 30.03.2024 10:07:51, 30.03.2024 10:13:23, 5, Normal, ,  peary_config.cfgmatrix_mpw4.txtmpw4.conf
automatic log for run 1016
Comment:
changed rate
Entry  Sat Mar 30 10:05:09 2024, bp, ci, hh, RD50-MPW4, 1015, RD50-MPW4, 800 MeV, Low Flux Medium, Proton, Good, 30.03.2024 10:02:17, 30.03.2024 10:07:36, 5, Normal, ,  peary_config.cfgmatrix_mpw4.txtmpw4.conf
automatic log for run 1015
Comment:
changed rate
Entry  Sat Mar 30 09:59:25 2024, bp, ci, hh, RD50-MPW4, 1014, RD50-MPW4, 800 MeV, Low Flux Low, Proton, Good, 30.03.2024 09:56:17, 30.03.2024 10:01:51, 5, Normal, ,  peary_config.cfgmatrix_mpw4.txtmpw4.conf
automatic log for run 1014
Comment:
Entry  Sat Mar 30 09:53:41 2024, bp, ci, hh, RD50-MPW4, 1013, RD50-MPW4, 800 MeV, Low Flux Low, Proton, Good, 30.03.2024 09:46:27, 30.03.2024 09:56:08, 9, Normal, ,  peary_config.cfgmatrix_mpw4.txtmpw4.conf
automatic log for run 1013
Comment:
Entry  Sat Mar 30 09:42:55 2024, bp, ci, hh, RD50-MPW4, 1012, RD50-MPW4, 800 MeV, Low Flux Low, Proton, Good, 30.03.2024 09:39:08, 30.03.2024 09:45:21, 6, Normal, ,  peary_config.cfgmatrix_mpw4.txtmpw4.conf
automatic log for run 1012
Comment:
changed rate
Entry  Sat Mar 30 09:36:22 2024, bp, ci, hh, RD50-MPW4, 1011, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Good, 30.03.2024 09:30:18, 30.03.2024 09:38:48, 8, Normal, ,  peary_config.cfgmatrix_mpw4.txtmpw4.conf
automatic log for run 1011
Comment:
testing new tablealignment
Entry  Sat Mar 30 09:27:33 2024, bp, ci, hh, RD50-MPW4, 1010, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Crap, 30.03.2024 09:23:42, 30.03.2024 09:29:59, 6, Normal, ,  peary_config.cfgmatrix_mpw4.txtmpw4.conf
automatic log for run 1010
Comment:
Entry  Sat Mar 30 09:03:54 2024, bp, ci, hh, RD50-MPW4, 1009, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Crap, 30.03.2024 08:43:22, 30.03.2024 09:06:20, 22, Normal, ,  peary_config.cfgmatrix_mpw4.txtmpw4.conf
automatic log for run 1009
Comment:
waiting for beam
Performing MTS restart
Entry  Sat Mar 30 08:40:10 2024, bp, ci, hh, RD50-MPW4, 1008, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Crap, 30.03.2024 08:37:03, 30.03.2024 08:42:37, 5, Normal, ,  peary_config.cfgmatrix_mpw4.txtmpw4.conf
automatic log for run 1008
Comment:
waiting for beam
Entry  Sat Mar 30 08:21:04 2024, bp, ci, hh, RD50-MPW4, 1007, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Good, 30.03.2024 08:02:46, 30.03.2024 08:23:30, 20, Normal, ,  peary_config.cfgmatrix_mpw4.txtmpw4.conf
automatic log for run 1007
Comment:
waiting for eam
Entry  Sat Mar 30 07:59:41 2024, bp, ci, hh, RD50-MPW4, 1006, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Good, 30.03.2024 07:46:50, 30.03.2024 08:02:08, 15, Normal, ,  peary_config.cfgmatrix_mpw4.txtmpw4.conf
automatic log for run 1006
Comment:
beautifully correlated Big grin
Entry  Sat Mar 30 07:44:15 2024, bp, ci, hh, RD50-MPW4, 1005, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, 30.03.2024 07:25:51, 30.03.2024 07:46:41, 20, Normal, ,  peary_config.cfgmatrix_mpw4.txtmpw4.conf
automatic log for run 1005
Comment:
beautifully correlated Big grin
Entry  Sat Mar 30 07:22:49 2024, bp, ci, hh, RD50-MPW4, 1004, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, 30.03.2024 07:14:51, 30.03.2024 07:25:16, 10, Normal, ,  peary_config.cfgmatrix_mpw4.txtmpw4.conf
automatic log for run 1004
Comment:
1st test runs
Entry  Sat Mar 30 07:11:30 2024, bp, ci, hh, RD50-MPW4, 1003, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, 30.03.2024 07:12:19, 30.03.2024 07:13:57, 1, Normal, ,  peary_config.cfgmatrix_mpw4.txtmpw4.conf
automatic log for run 1003
Comment:
1st test runs
Entry  Sat Mar 30 07:09:36 2024, bp, ci, hh, RD50-MPW4, 1002, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, 30.03.2024 07:11:15, 30.03.2024 07:12:03, 0, Normal, ,  peary_config.cfgmatrix_mpw4.txtmpw4.conf
automatic log for run 1002
Comment:
1st test runs
Entry  Sat Mar 30 07:00:10 2024, bp, ci, hh, RD50-MPW4, 1001, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, 30.03.2024 06:58:03, 30.03.2024 07:02:36, 4, Normal, ,  peary_config.cfgmatrix_mpw4.txtmpw4.conf
automatic log for run 1001
Comment:
1st test runs
Entry  Sat Mar 30 06:54:51 2024, bp, ci, hh, RD50-MPW4, 1000, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, 30.03.2024 06:56:23, 30.03.2024 06:57:18, 0, Normal, ,  peary_config.cfgmatrix_mpw4.txtmpw4.conf
automatic log for run 1000
Comment:
1st test runs
Entry  Sat Jun 8 21:42:13 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser, 10, TOF-Cal GSI, 194.3 MeV, Low Flux High, Proton, Good, 21:27, 21:39, , , ,  

1cm plastic water (in isocentre)
6 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 1.5 MHz on LGAD

WET measurement

Entry  Sat Dec 14 12:57:13 2024, Felix Ulrich-Pur, Shift 23.11.24 low flux He test with RaMon, 10, RaMon, 147.2 MeV/u, Low Flux High, Helium, Good, 06:42, 06:42, 1, , ,  

Mean rate per spill~ 3..5e6 p/s (online estimation without proper analysis)

Entry  Sat Jun 8 19:53:47 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs Test High Rate, 1, TOF-Cal GSI, 252.7 MeV, Low Flux High, Proton, Good, 16:00, 16:18, , , ,  

7.5 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 2.5 MHz on LGAD

Entry  Fri Dec 13 11:48:21 2024, Felix Ulrich-Pur, Shift 23.11.24 low flux He test with RaMon, 1, RaMon, 205.2 MeV/u, Low Flux Medium, Helium, Good, 06:28, 06:30, 2, , ,  

Mean rate per spill~ 4-6E5 p/s (online estimation without proper analysis)

Entry  Sat Mar 30 06:34:39 2024, Bernhard Pilsl, And a new testbeam starts :D, , , , , , , , , , , ,  
 
Entry  Sat Mar 30 06:53:31 2024, bp, ci, hh, RD50-MPW4, , RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, , , , Normal, ,  mpw4.inidetectors.geo
about to start

geo and eudaq ini attached
Entry  Sat Mar 30 07:59:55 2024, bp, ci, hh, RD50-MPW4, , RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Crap, , , , Normal, ,  
currently no beam
Entry  Sat Mar 30 08:34:12 2024, bp, ci, hh, RD50-MPW4, , RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Good, , , , Normal, ,  
beam should be back, there was something in the nozzle too
Entry  Sat Mar 30 08:38:21 2024, bp, ci, hh, RD50-MPW4, , RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Good, , , , Normal, ,  
beam not back, fake news
Entry  Sat Mar 30 08:56:25 2024, bp, ci, hh, RD50-MPW4, , RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Crap, , , , Normal, ,  
waiting for beam
Performing MTS restart
Entry  Sat Mar 30 09:15:13 2024, bp, ci, hh, RD50-MPW4, , RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Crap, , , , Normal, ,  
moved entire setup a bit in order too better focus beam on MPW4 (hopefully)
Entry  Sat Mar 30 09:25:05 2024, bp, ci, hh, RD50-MPW4, , RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Crap, , , , Normal, ,  
bea is back Big grin
Entry  Sat Mar 30 09:43:29 2024, bp, ci, hh, RD50-MPW4, , RD50-MPW4, 800 MeV, Low Flux Low, Proton, Good, , , , Normal, ,  
max triggerof tracker set to 100
Entry  Sat Mar 30 09:54:37 2024, bp, ci, hh, RD50-MPW4, , RD50-MPW4, 800 MeV, Low Flux Low, Proton, Good, , , , Normal, ,  
correlations still there after rate change
Entry  Sat Mar 30 09:59:40 2024, bp, ci, hh, RD50-MPW4, , RD50-MPW4, 800 MeV, Low Flux Medium, Proton, Good, , , , Normal, ,  
changed rate
Entry  Sat Mar 30 10:23:04 2024, bp, ci, hh, RD50-MPW4, , RD50-MPW4, 800 MeV, Low Flux Low, Proton, Good, , , , Normal, ,  
changed rate to low again, too much "noise"
also moved table down and a bit to the left
Entry  Sat Mar 30 10:28:42 2024, bp, ci, hh, RD50-MPW4, , RD50-MPW4, 800 MeV, Low Flux Low, Proton, Good, , , , Normal, ,  
tracker FADC down, HV not on earlier, restart happening
Entry  Sat Mar 30 11:20:13 2024, bp, ci, hh, RD50-MPW4, , RD50-MPW4, 800 MeV, Low Flux Low, Proton, Good, , , , Normal, ,  
alignment good now
Entry  Sat Mar 30 13:35:09 2024, bp, ci, hh, useful datataking, , RD50-MPW4, 800 MeV, Low Flux Low, Proton, Good, , , , Normal, ,  
new users arrived
Entry  Sun Aug 25 06:43:53 2024, tb-crew, Piggy test, , RD50-MPW4, 800 MeV, Low Flux Low, Proton, Unknown, , , , Normal, ,  
first tests, trying to get beam right beam
Entry  Sun Aug 25 06:47:14 2024, tb-crew, Piggy test, , RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, , , , Normal, ,  
rate should be fine now
Entry  Sun Aug 25 06:57:58 2024, tb-crew, Piggy test, , RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, , , , Normal, ,  
data is correlated
Entry  Sun Aug 25 07:04:31 2024, tb-crew, Piggy test, , RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, , , , Normal, ,  
adjusting geometry
was a lil schief
Entry  Sun Aug 25 07:43:57 2024, tb-crew, Piggy test, , RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, , , , Normal, ,  
DUT might have been stuck at end of last run
Entry  Sun Aug 25 09:58:51 2024, tb-crew, only base, , RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, , , , Normal, ,  
removed Piggy connections
Entry is currently edited by RD50 Generic User on 193.170.243.131  Entry  Sun Aug 25 10:02:35 2024, RD50 Generic User, init_geo_file, , , , , , , , , , , ,  detectors.geo

position of duts needs to be switched

Entry  Sun Aug 25 11:33:21 2024, tb-crew, only base, , RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, , , , Normal, ,  
back to nominal DACs and thr
Entry  Sun Aug 25 13:04:26 2024, tb-crew, only base, , RD50-MPW4, 62.4 MeV, Low Flux Very Low, Proton, Unknown, , , , Normal, ,  
End of (promising) shift
Entry  Mon Aug 26 09:18:25 2024, RD50 Generic User, geo, , , , , , , , , , , ,  detectors_new.geo

new geo file with adjusted dut position