Entry  Tue Oct 24 18:02:47 2017, Peter Paulitsch, HPK_200, run000264, 264, 110000, 0°, No, Good 
Parameter Value
Alibava filename  run000264.dat
Telescope filename  run000264.raw
Pedestal filename HEPHY_pedestal_HPK_CB_2.ped
Bias voltage -200 V
Dark current  
Sensor temperature -7
Latency 140 [a.u.]
Alibava triggers 110000
Telescope triggers 100000
Data synced yes
Table position x/y  
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling no
Alibava efficiency  
Entry  Tue Oct 24 18:04:09 2017, Peter Paulitsch, HPK_200, run000266, 266, 110000, 0°, No, Good 
Parameter Value
Alibava filename  run000266.dat
Telescope filename  run000266.raw
Pedestal filename HEPHY_pedestal_HPK_CB_2.ped
Bias voltage -200 V
Dark current  
Sensor temperature  
Latency 140 [a.u.]
Alibava triggers 110000
Telescope triggers 100000
Data synced yes
Table position x/y  
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling no
Alibava efficiency  
Entry  Tue Oct 24 18:08:06 2017, Peter Paulitsch, HPK_200, run000271, 271, 110000, 0°, No, Good 
Parameter Value
Alibava filename  run000271.dat
Telescope filename  run000271.raw
Pedestal filename HEPHY_pedestal_HPK_CB_2.ped
Bias voltage -200 V
Dark current  
Sensor temperature  
Latency 140 [a.u.]
Alibava triggers 110000
Telescope triggers 100000
Data synced yes
Table position x/y  
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling no
Alibava efficiency  
Entry  Tue Oct 24 18:08:38 2017, Peter Paulitsch, HPK_200, run000273, 273, 110000, 0°, No, Good 
Parameter Value
Alibava filename  run000273.dat
Telescope filename  run000273.raw
Pedestal filename HEPHY_pedestal_HPK_CB_2.ped
Bias voltage -200 V
Dark current  
Sensor temperature  
Latency 140 [a.u.]
Alibava triggers 110000
Telescope triggers 100000
Data synced yes
Table position x/y  
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling no
Alibava efficiency  
Entry  Tue Oct 24 18:08:57 2017, Peter Paulitsch, HPK_200, run000275, 275, 1010000, 0°, No, Good 
Parameter Value
Alibava filename  run000275.dat
Telescope filename  run000275.raw
Pedestal filename HEPHY_pedestal_HPK_CB_2.ped
Bias voltage -200 V
Dark current  
Sensor temperature  
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y  
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling no
Alibava efficiency  
Entry  Tue Oct 24 18:09:24 2017, Peter Paulitsch, HPK_200, run000277, 277, 1010000, 0°, No, Good 
Parameter Value
Alibava filename  run000277.dat
Telescope filename  run000277.raw
Pedestal filename HEPHY_pedestal_HPK_CB_2.ped
Bias voltage -200 V
Dark current  
Sensor temperature  
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y  
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling no
Alibava efficiency  
Entry  Tue Oct 24 18:09:43 2017, Peter Paulitsch, HPK_200, run000279, 279, 110000, 0°, No, Good 
Parameter Value
Alibava filename  run000279.dat
Telescope filename  run000279.raw
Pedestal filename HEPHY_pedestal_HPK_CB_2.ped
Bias voltage -200 V
Dark current  
Sensor temperature  
Latency 140 [a.u.]
Alibava triggers 110000
Telescope triggers 100000
Data synced yes
Table position x/y  
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling no
Alibava efficiency  
Entry  Tue Oct 24 18:10:14 2017, Peter Paulitsch, HPK_200, run000281, 281, 110000, 0°, No, Good 
Parameter Value
Alibava filename  run000281.dat
Telescope filename  run000281.raw
Pedestal filename HEPHY_pedestal_HPK_CB_2.ped
Bias voltage -200 V
Dark current  
Sensor temperature  
Latency 140 [a.u.]
Alibava triggers 110000
Telescope triggers 100000
Data synced yes
Table position x/y  
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling no
Alibava efficiency  
Entry  Tue Oct 24 18:10:39 2017, Peter Paulitsch, HPK_200, run000283, 283, 1010000, 0°, No, Good 
Parameter Value
Alibava filename  run000283.dat
Telescope filename  run000283.raw
Pedestal filename HEPHY_pedestal_HPK_CB_2.ped
Bias voltage -200 V
Dark current  
Sensor temperature  
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y  
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling no
Alibava efficiency  
Entry  Tue Oct 24 18:11:13 2017, Peter Paulitsch, HPK_200, run000285, 285, 1010000, 0°, No, Good 
Parameter Value
Alibava filename  run000285.dat
Telescope filename  run000285.raw
Pedestal filename HEPHY_pedestal_HPK_CB_2.ped
Bias voltage -200 V
Dark current  
Sensor temperature  
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y  
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling no
Alibava efficiency  
Entry  Tue Oct 24 18:11:43 2017, Peter Paulitsch, HPK_200, run000287, 287, 1010000, 0°, No, Good 
Parameter Value
Alibava filename  run000287.dat
Telescope filename  run000287.raw
Pedestal filename HEPHY_pedestal_HPK_CB_2.ped
Bias voltage -600 V
Dark current  
Sensor temperature  
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y  
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling no
Alibava efficiency  
Entry  Thu Oct 12 10:06:31 2017, Dominic Bloech, common, Packlist, , , , No, Unknown ALiBaVa-System_klein.jpgStuff_klein.jpg

# Testbeam Packliste


### Alibava stuff

* Alibava Mainboard
* Alibava power supply 
* Alibava Modules: IFX_CenterBias, HPK_CenterBias, IFX_baby
* Flatband cable 2x
* Alibava resetbox + Cable
* USB AtoB 2x
* BIAS Cable for BeetleChip
* 2410 SMU + Power + USBtoRS232
* Alibava mounting stuff


### Tools and Stuff

* Labeling machine
* Duct tape
* 2 pairs of wool gloves
* Seitenschneider klein
* USBtoRS232 2x
* Lan cable
* Schiebelehre
* Imbuschlüssel
* Screwdriver set
* Special screwdriver set
* Cutter 2x
* labeling cartridge
* Lemo 16 ns
* Lemo 8 ns 2x
* Lemo 4 ns
* Lemo 2 ns

 

 

### PC and stuff

* DAQ PC + Cables

* Monitor + Cables
* Mouse
* Keyboard

 

### Data Storage
* Toshiba 1TB USB3
* Maxtor 4TB USB3

Entry  Mon Oct 16 15:01:21 2017, Dominic Bloech, common, Preparations, , , , No, Unknown IMG_20171016_093005514.jpgIMG_20171016_150215807.jpgIMG_20171016_150230304.jpgIMG_20171016_150238370.jpg

### General

* Setup was set-up without problems. Pedestral and Calibration runs with ALIBAVA successfully.

** Warning: Trigger Out signal (busy signal) for the TLU is a TTL signal. Whereas the Trigger In signal can be a NIM signal. Either converting signal or adept in the TLU (we adepted the TLU matrix)

** TLU double triggered -> discriminated TLU signal beforehand 

*  Longterm BIAS Voltage for the IFX CenterBias to -100 V, slow control records IV-behavior in 10 sec steps.

* Security lectures taken

* 3 coincident measurements: 2 PM and the FEI4-Plane are connected to an "and" in the TLU

* Latency of Alibava without FEI4-Plane 133 [a.u] with FEI4 140 [a.u].

** Warning: This is due to the converter from the FEI4 which inverts the signal of the FEI4, this signal is then very noisy and would generate false positive triggers. -> Discriminate -> Long cables and latencies

### Other

* NI-Crate was broken (replaced by DESY)

ELOG V3.1.5-fc6679b