Home CMS Production Clean room MedAustron HEPHY testbeams old
testbeam-rd50-desy-Oct2024 testbeam-RD50-DESY-Apr2024 testbeam-RD50-DESY-Jul2023 testbeam-RD50-CERN-Oct2022 testbeam-DESY-Oct2017 testbeam-DESY-April2017 testbeam-SPS2015 testbeam-SPS2014 testbeam-DESY14 testbeam-SPS12 testbeam-SPS11 testbeam-SPS10 testbeam-SPS09 testbeam-SPS08 LP-TPC
  Alibava testbeam at DESY in Oct. 2017 (Dominic, Peter), Page 7 of 7  Not logged in ELOG logo
Entry  Tue Oct 24 18:02:47 2017, Peter Paulitsch, HPK_200, run000264, 264, 110000, 0°, No, Good 



Parameter
Value




Alibava
filename
 run000264.dat


Telescope filename
 run000264.raw


Pedestal
filename
HEPHY_pedestal_HPK_CB_2.ped


Bias voltage
-200 V


Dark current
 


Sensor
Entry  Tue Oct 24 18:04:09 2017, Peter Paulitsch, HPK_200, run000266, 266, 110000, 0°, No, Good 



Parameter
Value




Alibava
filename
 run000266.dat


Telescope filename
 run000266.raw


Pedestal
filename
HEPHY_pedestal_HPK_CB_2.ped


Bias voltage
-200 V


Dark current
 


Sensor
Entry  Tue Oct 24 18:08:06 2017, Peter Paulitsch, HPK_200, run000271, 271, 110000, 0°, No, Good 



Parameter
Value




Alibava
filename
 run000271.dat


Telescope filename
 run000271.raw


Pedestal
filename
HEPHY_pedestal_HPK_CB_2.ped


Bias voltage
-200 V


Dark current
 


Sensor
Entry  Tue Oct 24 18:08:38 2017, Peter Paulitsch, HPK_200, run000273, 273, 110000, 0°, No, Good 



Parameter
Value




Alibava
filename
 run000273.dat


Telescope filename
 run000273.raw


Pedestal
filename
HEPHY_pedestal_HPK_CB_2.ped


Bias voltage
-200 V


Dark current
 


Sensor
Entry  Tue Oct 24 18:08:57 2017, Peter Paulitsch, HPK_200, run000275, 275, 1010000, 0°, No, Good 



Parameter
Value




Alibava
filename
 run000275.dat


Telescope filename
 run000275.raw


Pedestal
filename
HEPHY_pedestal_HPK_CB_2.ped


Bias voltage
-200 V


Dark current
 


Sensor
Entry  Tue Oct 24 18:09:24 2017, Peter Paulitsch, HPK_200, run000277, 277, 1010000, 0°, No, Good 



Parameter
Value




Alibava
filename
 run000277.dat


Telescope filename
 run000277.raw


Pedestal
filename
HEPHY_pedestal_HPK_CB_2.ped


Bias voltage
-200 V


Dark current
 


Sensor
Entry  Tue Oct 24 18:09:43 2017, Peter Paulitsch, HPK_200, run000279, 279, 110000, 0°, No, Good 



Parameter
Value




Alibava
filename
 run000279.dat


Telescope filename
 run000279.raw


Pedestal
filename
HEPHY_pedestal_HPK_CB_2.ped


Bias voltage
-200 V


Dark current
 


Sensor
Entry  Tue Oct 24 18:10:14 2017, Peter Paulitsch, HPK_200, run000281, 281, 110000, 0°, No, Good 



Parameter
Value




Alibava
filename
 run000281.dat


Telescope filename
 run000281.raw


Pedestal
filename
HEPHY_pedestal_HPK_CB_2.ped


Bias voltage
-200 V


Dark current
 


Sensor
Entry  Tue Oct 24 18:10:39 2017, Peter Paulitsch, HPK_200, run000283, 283, 1010000, 0°, No, Good 



Parameter
Value




Alibava
filename
 run000283.dat


Telescope filename
 run000283.raw


Pedestal
filename
HEPHY_pedestal_HPK_CB_2.ped


Bias voltage
-200 V


Dark current
 


Sensor
Entry  Tue Oct 24 18:11:13 2017, Peter Paulitsch, HPK_200, run000285, 285, 1010000, 0°, No, Good 



Parameter
Value




Alibava
filename
 run000285.dat


Telescope filename
 run000285.raw


Pedestal
filename
HEPHY_pedestal_HPK_CB_2.ped


Bias voltage
-200 V


Dark current
 


Sensor
Entry  Tue Oct 24 18:11:43 2017, Peter Paulitsch, HPK_200, run000287, 287, 1010000, 0°, No, Good 



Parameter
Value




Alibava
filename
 run000287.dat


Telescope filename
 run000287.raw


Pedestal
filename
HEPHY_pedestal_HPK_CB_2.ped


Bias voltage
-600 V


Dark current
 


Sensor
Entry  Thu Oct 12 10:06:31 2017, Dominic Bloech, common, Packlist, , , , No, Unknown ALiBaVa-System_klein.jpgStuff_klein.jpg
# Testbeam Packliste


### Alibava stuff
Entry  Mon Oct 16 15:01:21 2017, Dominic Bloech, common, Preparations, , , , No, Unknown IMG_20171016_093005514.jpgIMG_20171016_150215807.jpgIMG_20171016_150230304.jpgIMG_20171016_150238370.jpg
### General

* Setup was set-up without problems. Pedestral and Calibration runs with ALIBAVA successfully.

** Warning: Trigger Out signal (busy signal) for the TLU is a TTL signal. Whereas the Trigger In signal can be a NIM signal. Either converting
ELOG V3.1.5-fc6679b