Home CMS Production Clean room MedAustron HEPHY testbeams old
testbeam-rd50-desy-Oct2024 testbeam-RD50-DESY-Apr2024 testbeam-RD50-DESY-Jul2023 testbeam-RD50-CERN-Oct2022 testbeam-DESY-Oct2017 testbeam-DESY-April2017 testbeam-SPS2015 testbeam-SPS2014 testbeam-DESY14 testbeam-SPS12 testbeam-SPS11 testbeam-SPS10 testbeam-SPS09 testbeam-SPS08 LP-TPC
  Alibava testbeam at DESY in Oct. 2017 (Dominic, Peter), Page 7 of 7  Not logged in ELOG logo
ID Date Author DUT Subject Run Numberdown Events XYPosition Cooling Data
  21   Fri Oct 20 14:11:03 2017 Peter PaulitschIFX_200_3e14run00012212211000075.93, 65.30, -72.55°YesGood
run settings
Parameter Value
Alibava filename  run000122.dat
Telescope filename  run000122.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current 39 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 110000
Telescope triggers 100000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency  210 Hz
Meas. time 15 min
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.7%

 

  20   Fri Oct 20 14:09:11 2017 Peter PaulitschIFX_200_3e14run00012012011000075.93, 64.80, -72.55°YesGood
run settings
Parameter Value
Alibava filename  run000120.dat
Telescope filename  run000120.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current 39 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 110000
Telescope triggers 100000
Data synced yes
Table position x/y 75.93, 64.80
Region of interest bias line
Trigger frequency  210 Hz
Meas. time 15 min
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.5%

 

  19   Fri Oct 20 14:05:39 2017 Peter PaulitschIFX_200_3e14run00011811811000075.93, 61.20, -72.55°YesGood
run settings
Parameter Value
Alibava filename  run000118.dat
Telescope filename  run000118.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current 39 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 110000
Telescope triggers 100000
Data synced yes
Table position x/y 75.93, 61.20
Region of interest bias line
Trigger frequency  210 Hz
Meas. time 15 min
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.2%

 

  18   Fri Oct 20 12:15:47 2017 Peter PaulitschIFX_200_3e14run00011611611000075.93, 63.00, -72.55°YesGood
run settings
Parameter Value
Alibava filename  run000116.dat
Telescope filename  run000116.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current 39 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 110000
Telescope triggers 100000
Data synced yes
Table position x/y 75.93, 63.00
Region of interest bias line
Trigger frequency  210 Hz
Meas. time 15 min
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.2%

 

  17   Wed Oct 18 13:38:47 2017 Peter PaulitschIFX_CenterBiasrun0001121123010000122.9, 24.25, -3.3°NoGood
run settings
Parameter Value
Alibava filename  run000112.dat
Telescope filename  run000112.raw
Pedestal filename pedestal_17102017
Bias voltage -100 V
Dark current -160 nA
Sensor temperature 24.6 °C
Latency 140 [a.u.]
Alibava triggers 3010000
Telescope triggers 3000000
Data synced yes
Table position x/y 122.9, 24.25
Region of interest bias line
Trigger frequency  157 Hz
Meas. time 320 min
Beam energy 5.6 GeV
Cooling no
Alibava efficiency 53.8%

 

  16   Wed Oct 18 13:36:50 2017 Peter PaulitschIFX_CenterBiasrun000109109110000122.9, 25.75, -3.3°NoGood
run settings
Parameter Value
Alibava filename  run000109.dat
Telescope filename  run000109.raw
Pedestal filename pedestal_17102017
Bias voltage -100 V
Dark current -160 nA
Sensor temperature 24.6 °C
Latency 140 [a.u.]
Alibava triggers 110000
Telescope triggers 100000
Data synced yes
Table position x/y 122.9, 25.75
Region of interest bias line
Trigger frequency  200 Hz
Meas. time 15 min
Beam energy 5.6 GeV
Cooling no
Alibava efficiency 53.4%

 

  15   Wed Oct 18 13:34:54 2017 Peter PaulitschIFX_CenterBiasrun000107107110000122.9, 23.5, -3.3°NoGood
run settings
Parameter Value
Alibava filename  run000107.dat
Telescope filename  run000107.raw
Pedestal filename pedestal_17102017
Bias voltage -100 V
Dark current -160 nA
Sensor temperature 24.6 °C
Latency 140 [a.u.]
Alibava triggers 110000
Telescope triggers 100000
Data synced yes
Table position x/y 122.9, 23.5
Region of interest bias line
Trigger frequency  200 Hz
Meas. time 15 min
Beam energy 5.6 GeV
Cooling no
Alibava efficiency 53.4%

 

  14   Wed Oct 18 13:33:05 2017 Peter PaulitschIFX_CenterBiasrun00010610660000122.9, 23.5, -3.3°NoGood
run settings
Parameter Value
Alibava filename  run000106.dat
Telescope filename  run000106.raw
Pedestal filename pedestal_17102017
Bias voltage -100 V
Dark current -160 nA
Sensor temperature 24.6 °C
Latency 140 [a.u.]
Alibava triggers 60000
Telescope triggers 50000
Data synced yes
Table position x/y 122.9, 23.5
Region of interest bias line
Trigger frequency  200 Hz
Meas. time 8 min
Beam energy 5.6 GeV
Cooling no
Alibava efficiency 53.4%

 

  13   Wed Oct 18 13:30:58 2017 Peter PaulitschIFX_CenterBiasrun000104104110000122.9, 23.5, -3.3°NoUnknown
run settings
Parameter Value
Alibava filename  run000104.dat
Telescope filename  run000104.raw
Pedestal filename pedestal_17102017
Bias voltage -100 V
Dark current -160 nA
Sensor temperature 24.6 °C
Latency 140 [a.u.]
Alibava triggers 110000
Telescope triggers 100000
Data synced yes
Table position x/y 122.9, 23.5
Region of interest bias line
Trigger frequency  200 Hz
Meas. time 15 min
Beam energy 5.6 GeV
Cooling no
Alibava efficiency 52.3%

 

  12   Wed Oct 18 13:29:39 2017 Peter PaulitschIFX_CenterBiasrun000102102110000122.9, 23.2, -3.3°NoGood
run settings
Parameter Value
Alibava filename  run000102.dat
Telescope filename  run000102.raw
Pedestal filename pedestal_17102017
Bias voltage -100 V
Dark current -160 nA
Sensor temperature 24.6 °C
Latency 140 [a.u.]
Alibava triggers 110000
Telescope triggers 100000
Data synced yes
Table position x/y 122.9, 23.8
Region of interest bias line
Trigger frequency  185 Hz
Meas. time 17 min
Beam energy 5.6 GeV
Cooling no
Alibava efficiency 52.3%

 

  11   Wed Oct 18 12:28:08 2017 Peter PaulitschIFX_CenterBiasrun00010110110000122.9, 23.8, -3.3°NoGood
run settings
Parameter Value
Alibava filename  run000101.dat
Telescope filename  run000101.raw
Pedestal filename pedestal_17102017
Bias voltage -100 V
Dark current -160 nA
Sensor temperature 24.6 °C
Latency 140 [a.u.]
Alibava triggers 10000
Telescope triggers 14500
Data synced yes
Table position x/y 122.9, 23.8
Region of interest bias line
Trigger frequency  185 Hz
Meas. time 2 min
Beam energy 5.6 GeV
Cooling no
Alibava efficiency 52.3%

 

  1   Thu Oct 12 10:06:31 2017 Dominic BloechcommonPacklist   NoUnknown

# Testbeam Packliste


### Alibava stuff

* Alibava Mainboard
* Alibava power supply 
* Alibava Modules: IFX_CenterBias, HPK_CenterBias, IFX_baby
* Flatband cable 2x
* Alibava resetbox + Cable
* USB AtoB 2x
* BIAS Cable for BeetleChip
* 2410 SMU + Power + USBtoRS232
* Alibava mounting stuff


### Tools and Stuff

* Labeling machine
* Duct tape
* 2 pairs of wool gloves
* Seitenschneider klein
* USBtoRS232 2x
* Lan cable
* Schiebelehre
* Imbuschlüssel
* Screwdriver set
* Special screwdriver set
* Cutter 2x
* labeling cartridge
* Lemo 16 ns
* Lemo 8 ns 2x
* Lemo 4 ns
* Lemo 2 ns

 

 

### PC and stuff

* DAQ PC + Cables

* Monitor + Cables
* Mouse
* Keyboard

 

### Data Storage
* Toshiba 1TB USB3
* Maxtor 4TB USB3

  2   Mon Oct 16 15:01:21 2017 Dominic BloechcommonPreparations   NoUnknown

### General

* Setup was set-up without problems. Pedestral and Calibration runs with ALIBAVA successfully.

** Warning: Trigger Out signal (busy signal) for the TLU is a TTL signal. Whereas the Trigger In signal can be a NIM signal. Either converting signal or adept in the TLU (we adepted the TLU matrix)

** TLU double triggered -> discriminated TLU signal beforehand 

*  Longterm BIAS Voltage for the IFX CenterBias to -100 V, slow control records IV-behavior in 10 sec steps.

* Security lectures taken

* 3 coincident measurements: 2 PM and the FEI4-Plane are connected to an "and" in the TLU

* Latency of Alibava without FEI4-Plane 133 [a.u] with FEI4 140 [a.u].

** Warning: This is due to the converter from the FEI4 which inverts the signal of the FEI4, this signal is then very noisy and would generate false positive triggers. -> Discriminate -> Long cables and latencies

### Other

* NI-Crate was broken (replaced by DESY)

ELOG V3.1.5-fc6679b