| ID | Date  | Author | Project | Subject | Run Number | Events | XYPosition | Cooling | Data | 
| 85 | Sun Apr 16 18:25:59 2017 | Axel Koenig | ALiBaVa (IFX) | run_011_Baby_50V.dat | 011 | 135000 | 102.5, 22.5, 240 | No | Unknown | 
| Remarks: Telescope run 588. Bias scan 50 V. Telescope chiller turned off (could be the root cause of runs crashing). 
	
		
			| Parameter | Value |  
			| Alibava run | run_011_Baby_50V.dat |  
			| Telescope run | 588 |  
			| Triggers | 135000 |  
			| Alibava trigger eff. | ~60% |  
			| Beam energy | 5.6 GeV |  
			| Telescope threshold | 6 |  
			| Bias voltage | -50 |  
			| Latency | 132 |  
			| Table x | 102.5 mm |  
			| Table z | 22.5 mm |  
			| Tabe rot | 240 deg |  | 
| 84 | Sun Apr 16 17:42:06 2017 | Axel Koenig | ALiBaVa (IFX) | run_010_Baby_40V.dat | 010 | 135000 | 102.5, 22.5, 240 | No | Unknown | 
| Remarks: Telescope run 587. Bias scan 40 V. Telescope chiller turned off (could be the root cause of runs crashing). 
	
		
			| Parameter | Value |  
			| Alibava run | run_010_Baby_40V.dat |  
			| Telescope run | 587 |  
			| Triggers | 135000 |  
			| Alibava trigger eff. | ~60% |  
			| Beam energy | 5.6 GeV |  
			| Telescope threshold | 6 |  
			| Bias voltage | -40 |  
			| Latency | 132 |  
			| Table x | 102.5 mm |  
			| Table z | 22.5 mm |  
			| Tabe rot | 240 deg |  | 
| 83 | Sun Apr 16 15:44:35 2017 | Axel Koenig | ALiBaVa (IFX) | run_009_Baby_30V.dat | 009 | 135000 | 102.5, 22.5, 240 | No | Unknown | 
| Remarks: Telescope run 586. Bias scan 30 V. Telescope chiller turned off (could be the root cause of runs crashing). 
	
		
			| Parameter | Value |  
			| Alibava run | run_009_Baby_30V.dat |  
			| Telescope run | 586 |  
			| Triggers | 135000 |  
			| Alibava trigger eff. | ~60% |  
			| Beam energy | 5.6 GeV |  
			| Telescope threshold | 6 |  
			| Bias voltage | -30 |  
			| Latency | 132 |  
			| Table x | 102.5 mm |  
			| Table z | 22.5 mm |  
			| Tabe rot | 240 deg |  | 
| 82 | Sun Apr 16 13:15:55 2017 | Axel Koenig | ALiBaVa (IFX) | run_008_Baby_20V.dat | 008 | 175000 | 102.5, 22.5, 240 | No | Unknown | 
| Remarks: Telescope run 583. Bias scan 20 V. run crashed. Telescope chiller turned off (could be the root cause of runs crashing). 
	
		
			| Parameter | Value |  
			| Alibava run | run_008_Baby_20V.dat |  
			| Telescope run | 583 |  
			| Triggers | 175000 |  
			| Alibava trigger eff. | ~60% |  
			| Beam energy | 5.6 GeV |  
			| Telescope threshold | 6 |  
			| Bias voltage | -20 |  
			| Latency | 132 |  
			| Table x | 102.5 mm |  
			| Table z | 22.5 mm |  
			| Tabe rot | 240 deg |  | 
| 81 | Sat Apr 15 21:18:26 2017 | Axel Koenig | ALiBaVa (IFX) | run_007_Baby_10V.dat | 007 | 450000 | 102.5, 22.5, 240 | No | Crap | 
| Remarks: Telescope run 579. Bias scan 10 V. run crashed. Data vaild up to 274915. Telescope chiller turned off (could be the root cause of runs crashing). 
	
		
			| Parameter | Value |  
			| Alibava run | run_007_Baby_10V.dat |  
			| Telescope run | 579 |  
			| Triggers | 450000 |  
			| Alibava trigger eff. | ~24% |  
			| Beam energy | 5.6 GeV |  
			| Telescope threshold | 6 |  
			| Bias voltage | -10 |  
			| Latency | 132 |  
			| Table x | 102.5 mm |  
			| Table z | 22.5 mm |  
			| Tabe rot | 240 deg |  | 
| 80 | Sat Apr 15 20:12:41 2017 | Axel Koenig | ALiBaVa (IFX) | run_005_Baby_100V.dat | 005 | 250000 | 102.5, 22.5, 230 | No | Unknown | 
| Remarks: Telescope run 568. Telescope chiller turned off (could be the root cause of runs crashing). 
	
		
			| Parameter | Value |  
			| Alibava run | run_005_Baby_100V.dat |  
			| Telescope run | 568 |  
			| Triggers | 250000 |  
			| Alibava trigger eff. | ~93% |  
			| Beam energy | 5.6 GeV |  
			| Telescope threshold | 6 |  
			| Bias voltage | -100 |  
			| Latency | 132 |  
			| Table x | 102.5 mm |  
			| Table z | 22.5 mm |  
			| Tabe rot | 230 deg |  | 
| 79 | Sat Apr 15 19:20:34 2017 | Axel Koenig | ALiBaVa (IFX) | run_004_Baby_100V.dat | 004 | 250000 | 102.5, 22.5, 250 | No | Unknown | 
| Remarks: Telescope run 565. Telescope chiller turned off (could be the root cause of runs crashing). 
	
		
			| Parameter | Value |  
			| Alibava run | run_004_Baby_100V.dat |  
			| Telescope run | 565 |  
			| Triggers | 250000 |  
			| Alibava trigger eff. | ~93% |  
			| Beam energy | 5.6 GeV |  
			| Telescope threshold | 6 |  
			| Bias voltage | -100 |  
			| Latency | 132 |  
			| Table x | 102.5 mm |  
			| Table z | 22.5 mm |  
			| Tabe rot | 250 deg |  | 
| 78 | Sat Apr 15 18:31:14 2017 | Axel Koenig | ALiBaVa (IFX) | run_003_Baby_100V.dat | 003 | 250000 | 102.5, 32.5, 240 | No | Unknown | 
| Remarks: Telescope run 564. Telescope chiller turned off (could be the root cause of runs crashing). 
	
		
			| Parameter | Value |  
			| Alibava run | run_003_Baby_100V.dat |  
			| Telescope run | 564 |  
			| Triggers | 250000 |  
			| Alibava trigger eff. | ~93% |  
			| Beam energy | 5.6 GeV |  
			| Telescope threshold | 6 |  
			| Bias voltage | -100 |  
			| Latency | 132 |  
			| Table x | 102.5 mm |  
			| Table z | 32.5 mm |  
			| Tabe rot | 240 deg (parallel to tele. planes) |  | 
| 77 | Sat Apr 15 17:38:58 2017 | Axel Koenig | ALiBaVa (IFX) | run_002_Baby_100V.dat | 002 | 250000 | 102.5, 12.5, 240 | No | Unknown | 
| Remarks: Telescope run 563. Telescope chiller turned off (could be the root cause of runs crashing). 
	
		
			| Parameter | Value |  
			| Alibava run | run_002_Baby_100V.dat |  
			| Telescope run | 563 |  
			| Triggers | 250000 |  
			| Alibava trigger eff. | ~93% |  
			| Beam energy | 5.6 GeV |  
			| Telescope threshold | 6 |  
			| Bias voltage | -100 |  
			| Latency | 132 |  
			| Table x | 102.5 mm |  
			| Table z | 12.5 mm |  
			| Tabe rot | 240 deg (parallel to tele. planes) |  | 
| 76 | Sat Apr 15 16:55:19 2017 | Axel Koenig | common | Chiller turned off! |  |  |  | No | Unknown | 
| The chiller has been turned off beginning of run_30_IFX_CB_100V.dat. Set temperature was 15 C. Now it should settle at 20 +- 1 C (room temperature). | 
| 75 | Sat Apr 15 16:52:06 2017 | Axel Koenig | ALiBaVa (IFX) | run_001_Baby_100V.dat | 001 | 250000 | 102.5, 22.5, 240 | No | Unknown | 
| Remarks: Telescope run 562. Telescope chiller turned off (could be the root cause of runs crashing). Position: 102.5, 22.5, 240 (should be the center of the sensor). 
	
		
			| Parameter | Value |  
			| Alibava run | run_001_Baby_100V.dat |  
			| Telescope run | 562 |  
			| Triggers | 250000 |  
			| Alibava trigger eff. | ~93% |  
			| Beam energy | 5.6 GeV |  
			| Telescope threshold | 6 |  
			| Bias voltage | -100 |  
			| Latency | 132 |  
			| Table x | 102.5 mm |  
			| Table z | 22.5 mm |  
			| Tabe rot | 240 deg (parallel to tele. planes) |  | 
| 74 | Sat Apr 15 12:12:00 2017 | Axel Koenig | ALiBaVa (IFX) | run_030_IFX_CB_100V.dat | 030 | 1000000 | 109, 4.5, 240 | No | Unknown | 
| 3Remarks: Telescope run 558. Bias scan 100 V. Telescope chiller turned off. Could be the root cause of crashing runs. 
	
		
			| Parameter | Value |  
			| Alibava run | run_030_IFX_CB_100V.dat |  
			| Telescope run | 558 |  
			| Triggers | 1000000 |  
			| Alibava trigger eff. | ~60% |  
			| Beam energy | 5.6 GeV |  
			| Telescope threshold | 6 |  
			| Bias voltage | -100 |  
			| Latency | 132 |  
			| Table x | 109 mm |  
			| Table z | 4.5 mm |  
			| Tabe rot | 240 deg (parallel to tele. planes) |  | 
| 73 | Sat Apr 15 09:57:51 2017 | Axel Koenig | ALiBaVa (IFX) | run_029_IFX_CB_130V.dat | 029 | 175000 | 109, 4.5, 240 | No | Unknown | 
| 3Remarks: Telescope run 557. Bias scan 130 V. 
	
		
			| Parameter | Value |  
			| Alibava run | run_029_IFX_CB_130V.dat |  
			| Telescope run | 557 |  
			| Triggers | 175000 |  
			| Alibava trigger eff. | ~60% |  
			| Beam energy | 5.6 GeV |  
			| Telescope threshold | 6 |  
			| Bias voltage | -130 |  
			| Latency | 132 |  
			| Table x | 109 mm |  
			| Table z | 4.5 mm |  
			| Tabe rot | 240 deg (parallel to tele. planes) |  | 
| 72 | Fri Apr 14 21:11:18 2017 | Axel Koenig | ALiBaVa (IFX) | run_028_IFX_CB_100V.dat | 028 | 3000000 | 109, 4.5, 240 | No | Crap | 
| Remarks: Telescope run 551. Longrun over night. Run crashed. Events only up to 2288000 valid. 
	
		
			| Parameter | Value |  
			| Alibava run | run_028_IFX_CB_100V.dat |  
			| Telescope run | 551 |  
			| Triggers | 3000000 |  
			| Alibava trigger eff. | ~60 % |  
			| Beam energy | 5.6 GeV |  
			| Telescope threshold | 6 |  
			| Bias voltage | -100 |  
			| Latency | 132 |  
			| Table x | 109 mm |  
			| Table z | 4.5 mm |  
			| Tabe rot | 240 deg |  | 
| 71 | Fri Apr 14 19:43:13 2017 | Axel Koenig | ALiBaVa (IFX) | run_027_IFX_CB_100V.dat | 027 | 350000 | 109, 4.5, 250 | No | Unknown | 
| Remarks: Telescope run 550. Angle shift 250 deg. 
	
		
			| Parameter | Value |  
			| Alibava run | run_027_IFX_CB_100V.dat |  
			| Telescope run | 550 |  
			| Triggers | 350000 |  
			| Alibava trigger eff. | ~59.8% |  
			| Beam energy | 5.6 GeV |  
			| Telescope threshold | 6 |  
			| Bias voltage | -100 |  
			| Latency | 132 |  
			| Table x | 109 mm |  
			| Table z | 4.5 mm |  
			| Tabe rot | 250 deg |  | 
| 70 | Fri Apr 14 18:22:58 2017 | Axel Koenig | ALiBaVa (IFX) | run_026_IFX_CB_100V.dat | 026 | 350000 | 109, 4.5, 230 | No | Unknown | 
| Remarks: Telescope run 549. Angle shift 230 deg. 
	
		
			| Parameter | Value |  
			| Alibava run | run_026_IFX_CB_100V.dat |  
			| Telescope run | 549 |  
			| Triggers | 350000 |  
			| Alibava trigger eff. | ~59.8% |  
			| Beam energy | 5.6 GeV |  
			| Telescope threshold | 6 |  
			| Bias voltage | -100 |  
			| Latency | 132 |  
			| Table x | 109 mm |  
			| Table z | 4.5 mm |  
			| Tabe rot | 230 deg |  | 
| 69 | Fri Apr 14 17:22:52 2017 | Axel Koenig | ALiBaVa (IFX) | run_025_IFX_CB_150V.dat | 025 | 175000 | 109, 4.5, 240 | No | Unknown | 
| 3Remarks: Telescope run 548. Bias scan 150 V. 
	
		
			| Parameter | Value |  
			| Alibava run | run_025_IFX_CB_150V.dat |  
			| Telescope run | 548 |  
			| Triggers | 175000 |  
			| Alibava trigger eff. | ~60% |  
			| Beam energy | 5.6 GeV |  
			| Telescope threshold | 6 |  
			| Bias voltage | -150 |  
			| Latency | 132 |  
			| Table x | 109 mm |  
			| Table z | 4.5 mm |  
			| Tabe rot | 240 deg (parallel to tele. planes) |  | 
| 68 | Fri Apr 14 14:49:09 2017 | Axel Koenig | ALiBaVa (IFX) | run_023_IFX_CB_100V.dat | 023 | 385000 | 109, 0, 240 | No | Crap | 
| Remarks: Telescope run 546. Bias voltage 100 V. Beam position on only one chip. Run destroyed. Intended trigger count 500k 
	
		
			| Parameter | Value |  
			| Alibava run | run_023_IFX_CB_100V.dat |  
			| Telescope run | 546 |  
			| Triggers | 385000 |  
			| Alibava trigger eff | ~62 % |  
			| Beam energy | 5.6 GeV |  
			| Telescope threshold | 6 |  
			| Bias voltage | -100 V |  
			| Latency | 132 |  
			| Table x | 109 mm |  
			| table z | 0 mm |  
			| table rot | 240 deg (parallel to tele. planes) |  | 
| 67 | Fri Apr 14 14:21:35 2017 | Axel Koenig | common | Run summary IFX_CB |  |  |  | No | Unknown | 
| 
	
		
			| Alibava run | Telescope run | File existence check | Run type | Triggers | Remarks |  
			| 001 | 496 | yes | position check | 200000 | none |  
			| 002 | 498 | yes | beam focused on one strip segment only | 100000 | none |  
			| 003 | 500 | yes | beam focused on one strip segment only | 100000 | none |  
			| 005 | 508 | yes | Bias scan. Voltage = 10 V | 285000 | none |  
			| 006 | 510 | yes | Bias scan. Voltage = 20 V | 185000 | none |  
			| 007 | 513 | yes | Bias scan. Voltage = 30 V | 170000 | none |  
			| 008 | 515 | yes | Bias scan. Voltage = 40 V | 175000 | none |  
			| 009 | 517 | yes | Bias scan. Voltage = 50 V | 175000 | none |  
			| 010 | 521 | yes | Bias scan. Voltage = 60 V | 175000 | none |  
			| 011 | 522 | yes | Bias scan. Voltage = 65 V | 175000 | none |  
			| 012 | 523 | yes | Bias scan. Voltage = 70 V | 175000 | none |  
			| 013 | 524 | yes | Bias scan. Voltage = 75 V | 175000 | none |  
			| 014 | 525 | yes | Bias scan. Voltage = 80 V | 175000 | none |  
			| 015 | 526 | yes | Bias scan. Voltage = 90 V | 175000 | none |  
			| 016 | 527-531 | yes | Bias scan. Voltage = 100 V | 750000 | Telescope files 527 to 531, run crashed! |  
			| 017 | 532 | yes | Bias scan. Voltage = 110 V | 175000 | none |  
			| 018 | 533 | yes | Bias scan. Voltage = 120 V | 175000 | none |  
			| 020 | 535 | yes | overnight run at 100 V | 190000 | run crashed data may be corrupted |  
			| 021 | 540 | yes | CRAP! | ? | CRAP! |  
			| 022 | 545 | yes | beam focused on one strip segment only | 500000 | none |  
			| 023 | 546 | yes | beam focused on one strip segment only | 385000 | run crashed data may be corrupted |  
			| 024 | 547 | yes | Bias scan. Voltage = 140 V | 175000 | none |  
			| 025 | 548 | yes | Bias scan. Voltage = 150 V | 175000 | none |  
			| 026 | 549 | yes | Rotation -10 deg | 350000 | none |  
			| 027 | 550 | yes | Rotation +10 deg | 350000 | none |  
			| 028 | 551 | yes | overnight run at 100 V | 3000000 | run crashed data may be corrupted |  
			| 029 | 557 | yes | Bias scan. Voltage = 130 V | 175000 | none |  
			| 030 | 558 | yes | long run | 1000000 | telescope chiller turned off (keept at this state unless otherwisde stated |  Central beam position: x = 109 mm, y = 4.5 mm, rot = 240 deg Current was stable for all runs i.e. 140 +- 10 nA at 100 V Trigger efficiency ~ 60% for all runs (except bias scan). Temperature in area 20+-1 deg. | 
| 66 | Fri Apr 14 13:21:07 2017 | Johannes Grossmann | MaPSA | Alignment Check | 66 | 250000 | 0 | No | Unknown | 
| Telescope 1318 Threshold 100 Mapsa CLK Delay 0 Reference Plane Lost aber Anfang Gut |