Sun Jun 7 23:04:06 2015, Hao Yin, Belle II FADC (H6A), Hardware, 013, 50000, , , Good
|
cooled (see attachment)
FIRRun002, PedestalRun006
target: CE, pos 3 |
Sun Jun 7 23:09:29 2015, Hao Yin, Belle II FADC (H6A), Hardware, 014, 224k, , , Good
|
cooled (see attachment) in - 20,5 out -7,8
FIRRun002, PedestalRun006
target: fw, pos 4 |
Mon Jun 8 11:33:24 2015, Thomas Bergauer, common, beam profile, , , , , Unknown
|
120GeV hadrons, beam config H6A.BELLE.003 |
Mon Jun 8 15:03:53 2015, Benedikt Würkner, SVD3 (H6B), hardware run with TLU and FEI4 WORKING, 59, 19000, , , Unknown
|
Testrun with the FEI4 working for the first time.
Lessons learned about the configuration:
EnableDUTVeto needs to be set for the FEI4 to allow it to work |
Mon Jun 8 16:53:13 2015, Benedikt Würkner, SVD3 (H6B), EUTelescope Fitter not working investigation, , , , , Unknown
|
Apparently every problem can be attributed to a misalignment problem. The telescope has troubles with heavily misaligned DUTs. Was resolved by creating
different gear-files for each run with pretty precise coordinates of the DUT position calculated for the telescope coordinate system.
Additionally one has to watch out for misaligned planes of the telescope itself. It was required to allow for more than 2mm offset for the telescope |
Mon Jun 8 18:08:49 2015, Thomas Bergauer, SVD3 (H6B), Telescope online monitor, , , , , Unknown
|
a working version of the EUDET online monitor (which works also with the FE-I4) is at the following location
~/BELLETEST/eudaq-neu/build/monitors/onlinemon$ ./OnlineMon.exe -sc 0 -f /home/telescope/eudaq/data/run000060.raw |
Tue Jun 9 09:32:41 2015, Hao Yin, Belle II FADC (H6A), Pedestal, 007, 10000, , , Good
|
Room temp, fir002 |
Tue Jun 9 09:33:33 2015, Hao Yin, Belle II FADC (H6A), Calibration, 003, , , , Good
|
room temp
FIRRun002,
pedestalrun006; |
Tue Jun 9 10:01:03 2015, Lukas Bulla, SVD3 (H6B), TLU setup info, , , , , Unknown
|
Channels:
Upstream vertical -> channel 0; (bitmask 1)
Upstream horizontal -> channel 1; (bitmask 2)
Downstream
vertical -> channel 2; (bitmask 4)
Downstream horizontal ->channel 3; (bitmask 8)
|
Tue Jun 9 20:30:37 2015, Hao Yin, Belle II FADC (H6A), FIR, 004, , , , Good
|
FIR filer and ADC delay with all sensor and telescope.
Telescope alignment |
Tue Jun 9 20:32:46 2015, Hao Yin, Belle II FADC (H6A), Pedestal, 008, 10000, , No, Good
|
Room temp, fir004 |
Tue Jun 9 20:36:39 2015, Hao Yin, Belle II FADC (H6A), Hardware, 020, 62k, , No, Good
|
room temp
FIRRun004, PedestalRun008
telescope alignment |
Wed Jun 10 10:40:24 2015, Hao Yin, Belle II FADC (H6A), Sixlet Calibration, 004, , , No, Good
|
room temp
pedestalrun 008
FIRRun004 |
Wed Jun 10 11:19:33 2015, Hao Yin, Belle II FADC (H6A), Sixlet Calibration, 005, , , No, Good
|
room temp
pedestalrun 008
FIRRun004 |
Wed Jun 10 12:03:53 2015, Hao Yin, Belle II FADC (H6A), Sixlet Calibration, 006, , , No, Good
|
room temp
pedestalrun 008
FIRRun004 |
Wed Jun 10 12:04:27 2015, Hao Yin, Belle II FADC (H6A), Sixlet Calibration, 007, , , No, Good
|
room temp
pedestalrun 008
FIRRun004 |
Wed Jun 10 13:12:33 2015, Hao Yin, Belle II FADC (H6A), Sixlet Calibration, 010, , , Yes, Good
|
cooled
pedestalrun 008
FIRRun004 |
Wed Jun 10 18:05:03 2015, Benedikt Würkner, SVD3 (H6B), FEI4 configuration, , , , No, Unknown
|
There are two planes, one fixed upstream and one fixed downstream.
CIS3WII_6 upstream and CIS2WI_IBL2 downstream.
They are configured using the pc in H6B which can be accessed via microsoft remote desktop from the network 192.168.5.x using the ip 192.168.5.6. |
Wed Jun 10 18:52:34 2015, Hao Yin, Belle II FADC (H6A), Pedestal, 009, 10000, , Yes, Good
|
temp: in -22 out -18,
FIRRun004, |
Thu Jun 11 11:41:50 2015, Hao Yin, Belle II FADC (H6A), Sixlet Calibration, 011, , , Yes, Good
|
cooled
pedestalrun 008
FIRRun004 |
|