ID |
Date |
Author |
Project |
Subject |
Run Number |
Events |
StartTime |
EndTime |
147
|
Thu May 19 11:02:08 2011 |
Manfred Valentan | Micron Baby | DUT4, source run, irrad, postzupf | cal119 | 83800 | 19.05.2011 09:57:57 | 19.05.2011 10:06:06 |
DUT4: B2Micron_2826_19_baby4 (combined p-stop)
config file: D:/cern10/micronbaby/data_lab/DUT4/postzupf/micronbaby_single_irrad_postzupf.cfg
HV=+-50V
90Sr 1mCi ~centered
I_start=95µA
I_end=137µA
Every second bond zupfed... |
Attachment 1: cal119_goodstrips_apv4.png
|
|
Attachment 2: cal119_goodstrips_apv5.png
|
|
148
|
Thu May 19 11:04:04 2011 |
Manfred Valentan | Micron Baby | DUT4, source run, irrad, postzupf | run119 | 100k | 19.05.2011 10:08:14 | 19.05.2011 10:29:02 |
DUT4: B2Micron_2826_19_baby4 (combined p-stop)
config file: D:/cern10/micronbaby/data_lab/DUT4/postzupf/micronbaby_multi6_irrad_postzupf.cfg
HV=+-50V
90Sr 1mCi ~centered
I_start=141µA
I_end=167µA
Every second bond zupfed... |
Attachment 1: cal119_goodstrips_apv4.png
|
|
Attachment 2: cal119_goodstrips_apv5.png
|
|
Attachment 3: run119_all_hits_apv4.png
|
|
Attachment 4: run119_all_hits_apv5.png
|
|
Attachment 5: cal119_pednoise_apv4.png
|
|
Attachment 6: cal119_pednoise_apv5.png
|
|
149
|
Thu May 19 11:06:41 2011 |
Manfred Valentan | Micron Baby | DUT4, source run, irrad, prezupf | run114 | 100k | 18.05.2011 11:57:13 | 18.05.2011 12:13:12 |
DUT4: B2Micron_2826_19_baby4 (combined p-stop)
config file: D:/cern10/micronbaby/data_lab/DUT4/prezupf/micronbaby_multi6_irrad.cfg
HV=+-50V
90Sr 1mCi ~centered
I_start=118µA
I_end=168µA
keine hitmaps gespeichert... |
Attachment 1: cal114_pednoise_apv4.png
|
|
Attachment 2: cal114_pednoise_apv5.png
|
|
1
|
Wed Sep 15 10:25:13 2010 |
Marko Dragicevic | common | Equipment List | | | | |
Slow Control:
3x Source Measurement Unit Keithley K2410
-- Inv.Nr.: 540-702, 1000EUR
-- Inv.Nr.: 540-730, 1000EUR
-- Inv.Nr.: 540-709, 1000EUR
-- Koax-Kabel + Adapter auf Banane
Scanning Voltmeter Keithley K2700 , Inv.Nr.: 540-698, 1000EUR
-- Shunt-Widerstand-Platine
-- Flachbandkabel fuer Verbindung Shunt-Platine<->K2700
TRHX
-- 3 Sensoren
-- 2 lange Flachbandkabel, 1 serielles Kabel
LV-Power-Supplies, je 150EUR
-- GW Instek GPC-1850, SN.: 4080392
-- GW Instek GPC-1850D, Inv.Nr.: 545-059
-- GW Instek GPC-1850D, Inv.Nr.: 545-062
-- Systron Donner HR 20-10A, SN.: 781-092
-- Systron Donner HR 20-10A, SN.: 781-096
-- GW Instek GPS-3030D SN.: E871711
-- HAMEG MHP4030 SN.: 057910044 500EUR
Oszilloskop Tektronix TDS 3054, SN.: B011109, 2500EUR
-- 3 probes
Die Katze (Belle-HV-Supply)
-- 2x HV-Kabel 30m (rot/schwarz)
GBIP
USB Adapter NI GPIB-USB-HS, S/N 14693EB, 100EUR
-Kabel, 3x lang, 1x kurz
USB Temperaturlogger PicoTech TC-08, S/N.: VSR 08/519 100EUR
BELLE DAQ (CERN):
Chrisu-PC, Inv.Nr.: 401-925
-- Monitore
---- Belinea, Inv.Nr.: 401-696
---- Gericom, SN.: M1132GR5467345
-- Keyboard & Maus
Wiener 9U-VME-Crate, Inv.Nr.: 545-043, 2000EUR
-- Power Supply
-- VME-Controller
-- 4 VME-Module (1x Buffer, 1, Neco, 2x FADC)
DOCK-Box
-- Power Kabel
VME-Kabel (bockig) + Verlaengerung
30m-Kabel (8xCat7, 2xCat5, 2xgrau)
Lemo-Kabel
100pin-Flachkabel Neco-Buffer
Bananenkabel
Hybridkabel
Special TLU-Adapter (flipping pin order of RJ45)
Schrauben für Crate-Montage
Erdungskabel (liegen am CERN im Labor)
Magic Box
APVDAQ (Mol):
PC Inv.Nr: 401-943
-- Ilyama E2407, S/N 1105702000889
-- Keyboard & Maus
VME Crate Inv.Nr: 545-030
-- Create Controller Inv.Nr: 540-732
-- 3x APVDAQ
Repeater Board
-- 3x auf Platte montiert
Kabel
-- VME-Kabel (bockig)
-- 4x Netzwerk (gelb 1:1 verschalten)
-- 4x Netzwerk (grau 1:1 verschalten)
-- 1x Flachband
-- diverse Bananen
DUT:
Schraubenbox bau
-- diverse Schrauben
-- Gewindestangen
-- Fluegelmittern
-- 2x Sockel
4mm Steckschlauch fuer Trockenluft
Anschluss 1/8" auf 4mm Steckschlauch
Druckminderer
-- 6mm Snap-In
-- 6mm Schlauch
Chiller Huber ministat 230, Inv.Nr.: 482-021, 3000EUR
-- Manual
-- Netzteil
-- langes serielles Kabel
-- 5-6mm Schlauch
-- Adapter
-- 5 Liter dest. Wasser (falls noch Platz ist)
-- Frostschutz
-- LabView (falls ferngesteuert)
Box-Deckel
Baseplates (4?)
Schwarze Selbstklebefolie
Alufolie
GROSSES SCHWARZES LICHTDICHTES TUCH !!!!
Module
-- 2 x Micron Trapezoid
-- 2 x 6" Origami
-- 9 x Micron Baby
-- 2 x CNM 2D Sensoren
-- 8x SiLC (No. 0-7)
-- 2x Pt
EDV:
Netzwerk-Switch:
-- D-Link 8-Port GBit-Switch DGS1008D S/N F37v89B010104, 40 EUR
WLAN-Router: Linksys WRT54GL-DE, SN.: CL7B1H502248, 60 EUR
1TB externe Festplatte: Western Digital MyBook, SN.: WCAJ1469094, 40 EUR
-- Netzgeraet
-- USB-Kabel
-- Firewire 400/800 Kabel
Webcam
USB-Stick
Etikettiergeraet: Brother P-touch 1000
Laptops
-- Apple MacBook (MV) Inv.Nr.: 401-810
(-- Apple MacBook Pro (MD) Inv Nr.: 401-831)
2x USB-Seriell Adapter
Netzwerkkabel
1 Trommel Ethernet-Kabel (100m, vom Walzel)
einige RJ-45 Stecker
2.5" externe HD
CD/DVD Rohlinge
Tools:
Diverse Adaptoren Lemo-BNC, Lemo-Lemo
Taschenlampe + 2x Hirnbirn
Massband
Multimeter
Werkzeug (Schraubenzieher & Co)
Pinzetten
Laserpointer
Schwarze Fetzen
Gaffa
Doppelscotch
Kabelbinder
Druckluft zur Reinigung
Adapter f. Steckdosen
Verteiler-Leisten
Kabeltrommel
Schaltbare Steckerleisten
Tixo klein & gross
Loetstation Advanced JBC, SN.: 76841165, inkl. Zubehoer
Ethernet-Crimp-Zange
Eisensagl
Schublehre
Wasserwaage
Misc:
Fondueset
Spiegelreflex Digitalkamera
-- Stativ
-- Reflexionszelt
Scheinwerfer
-- 3 x 500 W
-- 2 x Stativ
Akkus+Ladegerät
Dickes Logbook
Kugelschreiber/Bleistifte
Pflaster
Carnet ATA |
64
|
Sun Oct 3 14:35:12 2010 |
Marko Dragicevic | CNM | cal | cal009 | 83800 | 03.10.2010 13:50:54 | |
CMN-SILC stack
HV_SILC (Katze)=75V, 0.25uA (6 SiLC)
HV_CNM (Keithley)=30V, CNM_1=84nA, CNM_2=4.3uA
DUT outside of beam
calibration started after 1 hour warmup |
Attachment 1: cal009_cal_apv6.png
|
|
Attachment 2: cal009_cal_apv7.png
|
|
Attachment 3: slowcontrol_CNM_cal009.gif
|
|
73
|
Mon Oct 4 10:43:45 2010 |
Marko Dragicevic | Pt | run | run036 | 25600 | 04.10.2010 10:33:05 | 04.10.2010 11:35:51 |
Pt modules, angle scan
h01-Pt and h02-Pt
currents: h01=24.9A h02=12.5A
HV=200V beam on
angle: 39 deg
telescope run 20142 |
89
|
Wed Oct 6 10:15:26 2010 |
Marko Dragicevic | Micron Baby | XY Positions After Irradiation | | | | |
center position: 119/50/(7.5)
top left (direction jura, bottom right on sensor): 127/56.5/7.5
bottom right (direction saleve, top left on sensor): 111/43.5/7.5 |
90
|
Wed Oct 6 11:03:59 2010 |
Marko Dragicevic | Micron Baby | cal | cal013 | 83800 | 06.10.2010 11:03:12 | |
8 x Micron Baby (3 irradiated), no cooling
HV=100V I=115uA (start of run)
beam off
# VADJ/VPSP increased to 48 (was 30) for irradiated DUTs to restore baseline position (higher after irradiation)
ADC delay timing did not change after irradiation (verified)
|
94
|
Wed Oct 6 23:02:09 2010 |
Marko Dragicevic | common | Beam Configuration | | | | |
Beam Settings and Profile on 5.10.2010 - 17:00h
Saved as: H6B.801 |
Attachment 1: BeamProfile.gif
|
|
Attachment 2: BeamMagnets.gif
|
|
Attachment 3: BeamColli.gif
|
|
95
|
Wed Oct 6 23:03:15 2010 |
Marko Dragicevic | BELLE | cal | cal014 | 83800 | 06.10.2010 23:01:56 | 06.10.2010 23:34:07 |
Belle stack without Origami2
HV=80V I=227uA
beam on |
17
|
Tue Sep 28 01:13:57 2010 |
Markus Friedl | BELLE | cal | cal001 | 83800 | 28.09.2010 01:11:20 | |
first intcal with complete Belle stack: 2 x Origami, 2 x Micron
HV=100V I=49uA
beam is on
chiller is set to 15°C (module temperature is ~17°C)
both Origamis and both Microns are serially cooled |
18
|
Tue Sep 28 03:06:31 2010 |
Markus Friedl | BELLE | run | run001 | 119516 | 28.09.2010 03:06:04 | 28.09.2010 12:18:33 |
first run with complete Belle stack: 2 x Origami, 2 x Micron
HV=100V I=49uA
beam is on
chiller is set to 15°C (module temperature is ~17°C)
both Origamis and both Microns are serially cooled
position:centered in both axes 55 / 30 / 277.5
telescope: run20076 (in zero suppressed mode) |
19
|
Tue Sep 28 12:54:33 2010 |
Markus Friedl | BELLE | run | run002 | | 28.09.2010 12:46:54 | |
same as run001 after shifting x by 48mm
103 / 30 / 277.5 ? unsure after crash of xy stage PC |
20
|
Tue Sep 28 13:04:02 2010 |
Markus Friedl | BELLE | run | run003 | 25000 | 28.09.2010 13:00:05 | 28.09.2010 14:41:14 |
now for real (previous run was at wrong position)
all same as run001, except position now still centered on p-strips, but on high edge (last APVs) for n side
position: 103 / 30 / 277.5
corresponds to telescope run 20078 |
22
|
Tue Sep 28 14:48:00 2010 |
Markus Friedl | BELLE | run | run004 | 1755 | 28.09.2010 14:48:28 | 28.09.2010 15:06:31 |
same as before, but now pos 1 is located on the other edge (-48mm from center)
telescope 7 / 30 / 277.5
telescope run 20079 |
23
|
Tue Sep 28 15:09:16 2010 |
Markus Friedl | BELLE | run | run005 | 40000 | 28.09.2010 15:07:08 | 28.09.2010 20:20:12 |
previous run stopped due to beam problem (shifted, broad, etc).
now recovered, restarting with same settings.
telescope run: 20080 |
25
|
Tue Sep 28 22:07:16 2010 |
Markus Friedl | BELLE | cal | cal002 DO NOT USE -> go to cal003 | 83800 | 28.09.2010 22:01:55 | |
Belle stack (Origami, Micron, Micron, Origami)
now cooling only for Origami, Micron are uncooled
chiller set point: -25°C, measured values on Origami: -10°C
HV=100V, Ibias=38µA, beam off
modified I2C settings for cold modules
temperature monitoring: Origami2 pipes in<->out swapped (to achieve shorter tubes)
PROBLEM: Origami n channels are swapped -> USE cal003 INSTEAD!!! |
26
|
Tue Sep 28 22:09:39 2010 |
Markus Friedl | BELLE | Temperature sensors Belle stack | | | | |
PicoLogic TC-08 temperature sensor assignment Belle stack:
Sensor type: K
channel 1: Origami_mod1_in
channel 2: Origami_mod1_out
channel 3: Origami_mod2_out (swapped with 4) - label not changed
channel 4: Origami_mod2_in (swapped with 3) - label not changed
channel 5: Micron_wedge_mod1_p
channel 6: Micron_wedge_mod1_n
channel 7: Micron_wedge_mod2_p
channel 8: Micron_wedge_mod2_n
The cooling pipes of the two Origami modules are connected in series, respectively. |
27
|
Tue Sep 28 23:23:47 2010 |
Markus Friedl | BELLE | run | run006 | 2003 | 28.09.2010 23:13:22 | 28.09.2010 23:22:26 |
channels swapped -> stopped |
28
|
Tue Sep 28 23:24:35 2010 |
Markus Friedl | BELLE | run | run007 | 1460 | 28.09.2010 23:23:29 | 28.09.2010 23:27:54 |
Belle stack: Origami-Micron-Micron-Origami
HV=100V, Ibias=38A, beam on
cooling: Origami only, chiller: -25C, Module: -10C
modified I2C settings for cooled modules
centered again
telescope run 20082