Home CMS Production Clean room MedAustron HEPHY testbeams old
testbeam-rd50-desy-Oct2024 testbeam-RD50-DESY-Apr2024 testbeam-RD50-DESY-Jul2023 testbeam-RD50-CERN-Oct2022 testbeam-DESY-Oct2017 testbeam-DESY-April2017 testbeam-SPS2015 testbeam-SPS2014 testbeam-DESY14 testbeam-SPS12 testbeam-SPS11 testbeam-SPS10 testbeam-SPS09 testbeam-SPS08 LP-TPC
  HEPHY logbook for testbeam at SPS September/October 2010, Page 2 of 8  Not logged in ELOG logo
New entries since:Thu Jan 1 01:00:00 1970
ID Date Authordown Project Subject Run Number Events StartTime EndTime
  18   Tue Sep 28 03:06:31 2010 Markus FriedlBELLErunrun00111951628.09.2010 03:06:0428.09.2010 12:18:33

first run with complete Belle stack: 2 x Origami, 2 x Micron

HV=100V  I=49uA

beam is on

chiller is set to 15°C (module temperature is ~17°C)

both Origamis and both Microns are serially cooled

 

position:centered in both axes  55 / 30 / 277.5

 

telescope: run20076 (in zero suppressed mode)

  19   Tue Sep 28 12:54:33 2010 Markus FriedlBELLErunrun002 28.09.2010 12:46:54 

same as run001 after shifting x by 48mm

103 / 30 / 277.5 ? unsure after crash of xy stage PC

  20   Tue Sep 28 13:04:02 2010 Markus FriedlBELLErunrun0032500028.09.2010 13:00:0528.09.2010 14:41:14

now for real (previous run was at wrong position)

all same as run001, except position now still centered on p-strips, but on high edge (last APVs) for n side

position: 103 / 30 / 277.5

corresponds to telescope run 20078

  22   Tue Sep 28 14:48:00 2010 Markus FriedlBELLErunrun004175528.09.2010 14:48:2828.09.2010 15:06:31

same as before, but now pos 1 is located on the other edge (-48mm from center)

telescope 7 / 30 / 277.5

telescope run 20079

  23   Tue Sep 28 15:09:16 2010 Markus FriedlBELLErunrun0054000028.09.2010 15:07:0828.09.2010 20:20:12

previous run stopped due to beam problem (shifted, broad, etc).

now recovered, restarting with same settings.

 

telescope run: 20080

  25   Tue Sep 28 22:07:16 2010 Markus FriedlBELLEcalcal002 DO NOT USE -> go to cal0038380028.09.2010 22:01:55 

Belle stack (Origami, Micron, Micron, Origami)

now cooling only for Origami, Micron are uncooled

chiller set point: -25°C, measured values on Origami: -10°C

HV=100V, Ibias=38µA, beam off

modified I2C settings for cold modules

temperature monitoring: Origami2 pipes in<->out swapped (to achieve shorter tubes)

 

PROBLEM: Origami n channels are swapped -> USE cal003 INSTEAD!!!

  26   Tue Sep 28 22:09:39 2010 Markus FriedlBELLETemperature sensors Belle stack    

PicoLogic TC-08 temperature sensor assignment Belle stack:

Sensor type: K

channel 1:  Origami_mod1_in
channel 2:  Origami_mod1_out
channel 3:  Origami_mod2_out    (swapped with 4)  - label not changed
channel 4:  Origami_mod2_in      (swapped with 3)  - label not changed
channel 5:  Micron_wedge_mod1_p
channel 6:  Micron_wedge_mod1_n
channel 7:  Micron_wedge_mod2_p
channel 8:  Micron_wedge_mod2_n
 

The cooling pipes of the two Origami modules are connected in series, respectively.

  27   Tue Sep 28 23:23:47 2010 Markus FriedlBELLErunrun006200328.09.2010 23:13:2228.09.2010 23:22:26

channels swapped -> stopped

  28   Tue Sep 28 23:24:35 2010 Markus FriedlBELLErunrun007146028.09.2010 23:23:2928.09.2010 23:27:54

Belle stack: Origami-Micron-Micron-Origami
HV=100V, Ibias=38A, beam on
cooling: Origami only, chiller: -25C, Module: -10C
modified I2C settings for cooled modules
centered again

 

telescope run 20082

 

sensor config bug (only relevant for sensor display) found -> stopped

  29   Tue Sep 28 23:29:50 2010 Markus FriedlBELLErunrun00814000028.09.2010 23:29:0129.09.2010 09:36:58

Belle stack: Origami-Micron-Micron-Origami
HV=100V, Ibias=38A, beam on
cooling: Origami only, chiller: -25C, Module: -10C
modified I2C settings for cooled modules
centered again

 

telescope run 20083

Attachment 1: run008_sensordisplay.jpg
run008_sensordisplay.jpg
Attachment 2: run008_sensordisplay2.jpg
run008_sensordisplay2.jpg
  30   Wed Sep 29 09:39:33 2010 Markus FriedlBELLEcalcal0038380029.09.2010 09:38:00 

same as cal002, but now with corrected readout order

  31   Wed Sep 29 10:30:59 2010 Markus FriedlBELLErunrun0093000029.09.2010 10:27:2729.09.2010 11:06:35

zero suppressed (HIT4) @ FADC

Belle stack: Origami-Micron-Micron-Origami - HIT4
HV=100V, Ibias=28A, beam on
cooling: Origami only, chiller: -25C, Module: -10C
modified I2C settings for cooled modules
centered in both axes

telescope run 20084

  32   Wed Sep 29 11:09:10 2010 Markus FriedlBELLErunrun0103000029.09.2010 11:08:0329.09.2010 11:50:07

zero suppressed (HIT5) @ FADC

Belle stack: Origami-Micron-Micron-Origami - HIT5
HV=100V, Ibias=28A, beam on
cooling: Origami only, chiller: -25C, Module: -10C
modified I2C settings for cooled modules
centered in both axes

telescope run 20085

  34   Wed Sep 29 14:03:55 2010 Markus FriedlBELLEcalcal0048380029.09.2010 14:02:2529.09.2010 14:45:05

Belle stack: Origami-Micron-Micron-Origami
HV=100V, Ibias=100µA, beam off
no cooling

Temp. at modules: 
Origami 1: 50°C
Micron 1: 46°C
Micron 2: 48°C
Origami 2: 48°C

standard I2C settings
centered

  37   Wed Sep 29 17:38:19 2010 Markus FriedlMicron Babycalcal0058380029.09.2010 17:36:38 

micron baby stack
HV=100V I=6.5uA
Beam off

  39   Wed Sep 29 19:01:28 2010 Markus FriedlMicron Babyrunrun012224829.09.2010 18:59:0929.09.2010 19:07:02

micron baby stack
HV=100V I=11.0uA
Beam on
telescope position: 121/55/7.5

 

telescope run 20105

  40   Wed Sep 29 19:08:56 2010 Markus FriedlMicron Babyrunrun0138985329.09.2010 19:09:5730.09.2010 00:52:39

micron baby stack
HV=100V I=11.0uA
Beam on
telescope position: 121/53.5/7.5   (shifted in height by -1.5mm wrt run012 for better DUT centering)

 

telescope run 20107

Attachment 1: run013_hit.png
run013_hit.png
  41   Thu Sep 30 00:56:32 2010 Markus FriedlMicron Babyrunrun014100k30.09.2010 00:55:3030.09.2010 11:20:02

micron baby stack

 

HV=100V I=11.0uA

I_end = 16uA

 

Beam on

 

telescope position: 105/40.5/7.5 

telescope run 20108

  44   Thu Sep 30 15:06:28 2010 Markus FriedlMicron Babycalcal0068380030.09.2010 15:01:56 

micron baby: 4 DUTs - angle scan
HV=100V I=6.8uA
Beam off (turned on at ~8k events)
 

  45   Thu Sep 30 15:39:53 2010 Markus FriedlMicron Babyrunrun01631k30.09.2010 15:38:2630.09.2010 17:44:36

micron baby: 4 DUTs
HV=100V I=10.3uA (start of run);  I=16.2uA (end of run)
Beam on

 reference run @ 0°

telescope run 20110

Attachment 1: eventdisplay_schoen.png
eventdisplay_schoen.png
Attachment 2: event_display_nichsoschoen.png
event_display_nichsoschoen.png
ELOG V3.1.5-fc6679b