  Belle II and AIDA/Infineon Testbeam at DESY in Jan2014, all entries  Not logged in ELOG logo
New entries since:Thu Jan 1 01:00:00 1970
ID Date Authordown Project Subject Run Number Events StartTime EndTime Data Text Attachments
  71   Fri Jan 24 11:03:14 2014 Wolfgang TreberspurgAIDAAIDAHW_AIDA002024.01.2014 10:02 CrapTuxDaq crashed   
  73   Fri Jan 24 18:02:46 2014 Wolfgang TreberspurgInfineonNew Infineon Stack    GoodExchanged AIDA stack with modified Infineon
Stack 
Configuration: 
  
  74   Sat Jan 25 18:24:03 2014 Wolfgang TreberspurgInfineonSingle mode Run Stack 2HW_Run_Infineon_Stack2_0011.05 M25.01.2014 10:3025.01.2014 15:40UnknownHV at 300V; I at app. 11µA 4GeV,
Single mode
Modules centered to beam but no hits
  
  75   Sat Jan 25 18:28:38 2014 Wolfgang TreberspurgInfineonCalibration run Infineon Stack 2Calibration_Run_Infineon_Stack2_0018290025.01.2014 10:1525.01.2014 10:25UnknownHV 300V, I ca. 11µA single mode   
  76   Sat Jan 25 18:46:32 2014 Wolfgang TreberspurgInfineonMulit 6 mode Run Stack 2HW_Run_Infineon_Stack2_Multi_001340k25.01.2014 16:3025.01.2014 18:40UnknownHV 300V, I 10µA
Beam Energy changed during the run
between 2GeV - 5GeV - 4GeV
  
  77   Sat Jan 25 21:14:46 2014 Wolfgang TreberspurgInfineonMulit 6 mode Run Stack 2HW_Run_Infineon_Stack2_Multi_002230k25.01.2014 19:0025.01.2014 21:15GoodHV 300V, I 10µA
Beam Energy  4GeV
Latency changed to 93
  
  78   Sun Jan 26 19:12:42 2014 Wolfgang TreberspurgInfineonSingle mode Run Stack 2HW_Run_Infineon_Stack2_0024 M25.01.2014 22:3026.01.2014 15:15GoodHV at 300V; I at app. 11µA 4GeV,
Single mode
 
  
  79   Sun Jan 26 19:14:07 2014 Wolfgang TreberspurgInfineonSingle mode Run Stack 2HW_Run_Infineon_Stack2_003140 k26.01.2014 18:1026.01.2014 19:15GoodHV at 300V; I at app. 10µA 5GeV,
Single mode
 
  
  80   Mon Jan 27 12:20:53 2014 Wolfgang TreberspurgInfineonSingle mode Run Stack 2HW_Run_Infineon_Stack2_004800k27.01.2014 09:1027.01.2014 12:15GoodHV at 300V; I at app. 10µA 3GeV,
Single mode
 
  
  81   Wed Jan 29 12:02:52 2014 Wolfgang TreberspurgInfineonSingle mode Run Stack 2 Bias ScanHW_Run_Infineon_Stack2_BiasScan_001100k28.01.2014 16:3028.01.2014 16:50GoodHV at 0V; I at app. 5nA 3GeV, Single mode
 
  
  82   Wed Jan 29 12:04:53 2014 Wolfgang TreberspurgInfineonSingle mode Run Stack 2 Bias ScanHW_Run_Infineon_Stack2_BiasScan_002140k28.01.2014 16:5028.01.2014 17:20GoodHV at 100V; I at app. 8.5µA 3GeV,
Single mode
 
  
  83   Wed Jan 29 12:07:48 2014 Wolfgang TreberspurgInfineonSingle mode Run Stack 2 Bias ScanHW_Run_Infineon_Stack2_BiasScan_003200k28.01.2014 17:4028.01.2014 18:20GoodHV at 150V; I at app. 9µA 3GeV, Single
mode
 
  
  84   Wed Jan 29 12:08:52 2014 Wolfgang TreberspurgInfineonSingle mode Run Stack 2 Bias ScanHW_Run_Infineon_Stack2_BiasScan_004200k28.01.2014 19:3028.01.2014 20:00GoodHV at 200V; I at app. 9.5µA 3GeV,
Single mode
 
  
  85   Wed Jan 29 12:10:06 2014 Wolfgang TreberspurgInfineonSingle mode Run Stack 2 Bias ScanHW_Run_Infineon_Stack2_BiasScan_005200k28.01.2014 20:0028.01.2014 23:00GoodHV at 210V; I at app. 8.5µA 3GeV,
Single mode
 
  
  86   Wed Jan 29 12:11:00 2014 Wolfgang TreberspurgInfineonSingle mode Run Stack 2 Bias ScanHW_Run_Infineon_Stack2_BiasScan_006200k28.01.2014 20:3028.01.2014 21:00GoodHV at 220V; I at app. 8.5µA 3GeV,
Single mode
 
  
  87   Wed Jan 29 21:55:09 2014 Wolfgang TreberspurgInfineonSingle mode Run Stack 2 Bias ScanHW_Run_Infineon_Stack2_BiasScan_007200k29.01.2014 11:5029.01.2014 12:15GoodHV at 230V; I at app. 8µA 3GeV, Single
mode
 
  
  88   Wed Jan 29 21:56:15 2014 Wolfgang TreberspurgInfineonSingle mode Run Stack 2 Bias ScanHW_Run_Infineon_Stack2_BiasScan_008200k29.01.2014 12:1529.01.2014 12:45GoodHV at 240V; I at app. 7.5µA 3GeV,
Single mode
 
  
  89   Wed Jan 29 21:56:58 2014 Wolfgang TreberspurgInfineonSingle mode Run Stack 2 Bias ScanHW_Run_Infineon_Stack2_BiasScan_009110k29.01.2014 12:5029.01.2014 13:15GoodHV at 250V; I at app. 7.5µA 3GeV,
Single mode
 
  
  90   Wed Jan 29 21:57:55 2014 Wolfgang TreberspurgInfineonSingle mode Run Stack 2 Bias ScanHW_Run_Infineon_Stack2_BiasScan_010100k29.01.2014 13:2029.01.2014 13:35GoodHV at 260V; I at app. 7.6µA 3GeV,
Single mode
 
  
  91   Wed Jan 29 21:59:51 2014 Wolfgang TreberspurgInfineonSingle mode Run Stack 2 Bias ScanHW_Run_Infineon_Stack2_BiasScan_011200k29.01.2014 13:4529.01.2014 14:20GoodHV at 270V; I at app. 7.6µA 3GeV,
Single mode
 
  
  92   Wed Jan 29 22:01:08 2014 Wolfgang TreberspurgInfineonSingle mode Run Stack 2 Bias ScanHW_Run_Infineon_Stack2_BiasScan_012200k29.01.2014 14:2529.01.2014 15:26GoodHV at 280V; I at app. 8µA 3GeV, Single
mode
 
  
  93   Wed Jan 29 22:02:04 2014 Wolfgang TreberspurgInfineonSingle mode Run Stack 2 Bias ScanHW_Run_Infineon_Stack2_BiasScan_013200k29.01.2014 15:2629.01.2014 17:00GoodHV at 290V; I at app. 8µA 3GeV, Single
mode
 
  
  94   Wed Jan 29 22:03:15 2014 Wolfgang TreberspurgInfineonSingle mode Run Stack 2 Bias ScanHW_Run_Infineon_Stack2_BiasScan_014200k29.01.2014 18:1529.01.2014 18:45GoodHV at 260V; I at app. 7.5µA 3GeV,
Single mode
 
  
  95   Wed Jan 29 22:06:12 2014 Wolfgang TreberspurgInfineonSingle mode Run Stack 2 Bias ScanHW_Run_Infineon_Stack2_BiasScan_015200k29.01.2014 19:1529.01.2014 19:45GoodHV at 300V; I at app. 7.5µA 3GeV,
Single mode
 
  
  96   Wed Jan 29 22:08:48 2014 Wolfgang TreberspurgInfineonSingle mode Run Stack 2 Bias Scan  29.01.2014 22:0029.01.2014 22:00GoodAll Infineon Stack 2 runs have been taken
at table position: 1.089030, -12.800610
 
  
  97   Wed Jan 29 22:36:35 2014 Wolfgang TreberspurgInfineonSingle mode Run Stack 2 Bias ScanHW_Run_Infineon_Stack2_pos1_001200k29.01.2014 22:1529.01.2014 22:35GoodHV at 300V; I at app. 7.5µA 3GeV,
Single mode
Position changed to -3.910970 -17.800610
  
  98   Wed Jan 29 22:40:09 2014 Wolfgang TreberspurgInfineonSingle mode Run Stack 2 Bias ScanHW_Run_Infineon_Stack2_pos2_002200k29.01.2014 22:4029.01.2014 23:15GoodHV at 300V; I at app. 7.5µA 3GeV,
Single mode
Position changed to --8.91097 -22.80061
  
  7   Mon Jan 6 19:10:36 2014 Thomas BergauercommonBeam Momentum 3GeV/crun001 to run012   UnknownWe have set the beam momentum to 3 GeV   IMG_1517.jpgIMG_1516.jpg 
  17   Tue Jan 7 02:17:14 2014 Thomas Bergauer beam momentum 6GeV/crun013   Unknown    
  19   Tue Jan 7 08:36:29 2014 Thomas BergauerInfineonrunrun01420k07.01.2014 08:34:0607.01.2014 09:05:47GoodHV=290V  I=9.55uA
position
centered horizontally (+5mm)
DESY
table z=-346 x=-255 
  
  20   Tue Jan 7 09:12:24 2014 Thomas BergauerBelle SVDrunrun01520k07.01.2014 09:10:5207.01.2014 09:15:39Good HV=290V  I=9.55uA
position centered horizontally
(+5mm)
DESY table z=-346 x=-255
 
energy
  
  21   Tue Jan 7 09:22:09 2014 Thomas Bergauer runrun01650k07.01.2014 09:19:1007.01.2014 09:30:03GoodHV=290V  I=9.00uA
position
move back to original position (0mm/0mm)
(not hitting the right APVs anymore)
DESY
  
  22   Tue Jan 7 09:33:16 2014 Thomas Bergauer runrun017100k07.01.2014 09:31:4507.01.2014 09:52:58GoodHV=300V  I=9.08uA
position
adapted to center beam better (+10mm horizonal)
DESY
table z=-346 x=-255 
  
  23   Tue Jan 7 09:33:51 2014 Thomas Bergauercommonbeam momentum back to 3Gev/crun014 and later ones   Unknown    
  28   Tue Jan 7 14:51:45 2014 Thomas BergauerBelle SVDTLU control    Unknown1. ssh bastion
2. ssh belle-control
3. ssh 
  
  41   Wed Jan 8 12:40:30 2014 Thomas BergauerAIDAintcal AIDA Stackcal03483.8k08.01.2014 12:32:2208.01.2014 13:01:30Unknown HV=100V  I=4.5uA
position adapted to center beam
 
  
  42   Wed Jan 8 14:10:07 2014 Thomas BergauerAIDArunrun03521168108.01.2014 14:01:2508.01.2014 17:03:55Unknown HV=100V  I=4.2uA
position adapted to center beam
 
accidentally
hit reset button on PC which ended this run
  
  50   Thu Jan 9 14:47:21 2014 Thomas Bergauercommonposition of DESY Table    UnknownBEAM mode:

    height (z): -346 (initially:
-357)
    horizontal (x): -255 

OFF
  
  1   Sat Jan 4 16:30:46 2014 Markus FriedlInfineoncalcal0018380004.01.2014 16:16:5104.01.2014 16:25:46GoodHV=300V  I=9.3uA  (total)   
  2   Sun Jan 5 15:08:09 2014 Markus FriedlAIDAcalcal0028380005.01.2014 15:01:3205.01.2014 15:30:36GoodHV=100V  I=6uA   
  3   Mon Jan 6 16:33:13 2014 Markus FriedlInfineonrunrun001200k06.01.2014 16:28:5806.01.2014 17:10:56GoodHV=300V  I=11.1uA
position centered horizontally (+5mm)
DESY table z=-346 x=-255
  
  4   Mon Jan 6 17:17:47 2014 Markus FriedlInfineonrunrun002200k06.01.2014 17:14:3906.01.2014 18:00:34GoodHV=100V  I=8.5uA
position centered horizontally (+5mm)
DESY table z=-346 x=-255
  
  5   Mon Jan 6 18:06:16 2014 Markus FriedlInfineonrunrun003200k06.01.2014 18:04:1406.01.2014 18:44:15GoodHV=200V  I=9.3uA
position centered horizontally (+5mm)
DESY table z=-346 x=-255
  
  6   Mon Jan 6 18:49:00 2014 Markus FriedlInfineonrunrun004200k06.01.2014 18:47:3806.01.2014 19:27:33GoodHV=250V  I=9.5uA
position centered horizontally (+5mm)
DESY table z=-346 x=-255
  
  8   Mon Jan 6 19:31:44 2014 Markus FriedlAIDArunrun005200k06.01.2014 19:29:5206.01.2014 20:09:59GoodHV=230V  I=9.9uA
position centered horizontally (+5mm)
DESY table z=-346 x=-255
  
  9   Mon Jan 6 20:14:21 2014 Markus FriedlInfineonrunrun006200k06.01.2014 20:12:2006.01.2014 20:54:34GoodHV=240V  I=9.4uA
position centered horizontally (+5mm)
DESY table z=-346 x=-255
  
  10   Mon Jan 6 20:57:38 2014 Markus FriedlInfineonrunrun007200k06.01.2014 20:56:0806.01.2014 21:36:54GoodHV=220V  I=9.1uA
position centered horizontally (+5mm)
DESY table z=-346 x=-255
  
  11   Mon Jan 6 21:40:29 2014 Markus FriedlInfineonrunrun008200k06.01.2014 21:38:4606.01.2014 22:20:13GoodHV=210V  I=9.0uA
position centered horizontally (+5mm)
DESY table z=-346 x=-255
  
  12   Mon Jan 6 22:23:39 2014 Markus FriedlInfineonrunrun009200k06.01.2014 22:22:1006.01.2014 23:03:23GoodHV=260V  I=9.3uA
position centered horizontally (+5mm)
DESY table z=-346 x=-255
  
  13   Mon Jan 6 23:06:51 2014 Markus FriedlInfineonrunrun010200k06.01.2014 23:03:2306.01.2014 23:47:02 GoodHV=270V  I=9.6uA
position centered horizontally (+5mm)
DESY table z=-346 x=-255
  
  14   Mon Jan 6 23:50:59 2014 Markus FriedlInfineonrunrun011200k06.01.2014 23:49:0907.01.2014 00:30:38 GoodHV=280V  I=9.6uA
position centered horizontally (+5mm)
DESY table z=-346 x=-255
  
  15   Tue Jan 7 00:36:03 2014 Markus FriedlInfineonrunrun012190k07.01.2014 00:33:5807.01.2014 02:12:32 GoodHV=150V  I=8.0uA
position centered horizontally (+5mm)
DESY table z=-346 x=-255
  
  18   Tue Jan 7 02:18:26 2014 Markus FriedlInfineonrunrun013215k07.01.2014 02:14:5007.01.2014 08:31:25CrapHV=150V  I=7.90uA
position centered horizontally (+5mm)
DESY table z=-346 x=-255
  
  24   Tue Jan 7 11:50:34 2014 Markus FriedlInfineonscanrun018x10k07.01.2014 11:xx:xx07.01.2014 11:xx:xxGoodDelay scan with single sample
Several runs
  
  25   Tue Jan 7 11:52:27 2014 Markus FriedlInfineonrunrun019650k07.01.2014 11:47:2407.01.2014 13:15:21GoodHV=300V  I=9.2uA
position adapted to center beam better
(+10mm horizontal)
  
  26   Tue Jan 7 13:20:48 2014 Markus FriedlInfineonrunrun020300k07.01.2014 13:18:5207.01.2014 14:35:49GoodHV=300V  I=8.5uA
position adapted to center beam better
(+10mm horizontal)
  
  27   Tue Jan 7 14:47:38 2014 Markus FriedlInfineonrunrun021200k07.01.2014 14:46:0807.01.2014 15:13:23GoodHV=100V  I=7.2uA
position adapted to center beam better
(+10mm horizontal)
  
  29   Tue Jan 7 15:16:11 2014 Markus FriedlInfineonrunrun022200k07.01.2014 15:15:0907.01.2014 15:42:17GoodHV=150V  I=7.3uA
position adapted to center beam better
(+10mm horizontal)
  
  30   Tue Jan 7 15:54:22 2014 Markus FriedlInfineonrunrun023200k07.01.2014 15:53:0607.01.2014 16:25:26GoodHV=200V  I=8.0uA
position adapted to center beam better
(+10mm horizontal)
  
  31   Tue Jan 7 16:36:49 2014 Markus FriedlInfineonrunrun024200k07.01.2014 16:35:4407.01.2014 17:03:10GoodHV=210V  I=8.1uA
position adapted to center beam better
(+10mm horizontal)
  
  32   Tue Jan 7 17:09:56 2014 Markus FriedlInfineonrunrun025200k07.01.2014 17:08:5407.01.2014 17:36:12GoodHV=220V  I=8.1uA
position adapted to center beam better
(+10mm horizontal)
  
  33   Tue Jan 7 17:39:20 2014 Markus FriedlInfineonrunrun026200k07.01.2014 17:38:0107.01.2014 18:05:11GoodHV=230V  I=8.2uA
position adapted to center beam better
(+10mm horizontal)
  
  34   Tue Jan 7 18:11:52 2014 Markus FriedlInfineonrunrun027200k07.01.2014 18:10:4507.01.2014 18:37:42GoodHV=240V  I=8.2uA
position adapted to center beam better
(+10mm horizontal)
  
  35   Tue Jan 7 18:41:55 2014 Markus FriedlInfineonrunrun028200k07.01.2014 18:40:4007.01.2014 19:07:32GoodHV=250V  I=8.3uA
position adapted to center beam better
(+10mm horizontal)
  
  36   Tue Jan 7 19:12:07 2014 Markus FriedlInfineonrunrun029200k07.01.2014 19:11:0507.01.2014 19:38:00GoodHV=260V  I=8.7uA
position adapted to center beam better
(+10mm horizontal)
  
  37   Tue Jan 7 19:44:45 2014 Markus FriedlInfineonrunrun030200k07.01.2014 19:43:3507.01.2014 20:11:03GoodHV=270V  I=8.6uA
position adapted to center beam better
(+10mm horizontal)
  
  38   Tue Jan 7 20:26:50 2014 Markus FriedlInfineonrunrun031200k07.01.2014 20:25:3207.01.2014 20:58:47GoodHV=280V  I=8.4uA
position adapted to center beam better
(+10mm horizontal)
  
  39   Tue Jan 7 21:02:58 2014 Markus FriedlInfineonrunrun032200k07.01.2014 21:01:4207.01.2014 21:29:03GoodHV=290V  I=8.3uA
position adapted to center beam better
(+10mm horizontal)
  
  40   Tue Jan 7 21:34:41 2014 Markus FriedlInfineonrunrun0334.117.024 (4M)07.01.2014 21:33:1208.01.2014 08:52:37GoodHV=300V  I=8.3uA
position adapted to center beam better
(+10mm horizontal)
  
  43   Wed Jan 8 17:20:40 2014 Markus FriedlAIDArunrun036320k08.01.2014 17:18:5108.01.2014 22:14:58Unknown HV=100V  I=4.3uA
centered position
5GeV beam
  
  44   Wed Jan 8 22:19:11 2014 Markus FriedlAIDArunrun0371.2M08.01.2014 22:18:3009.01.2014 10:08:14GoodHV=100V  I=4.4uA
centered position
5GeV beam, single sample
  
  45   Thu Jan 9 10:10:31 2014 Markus FriedlAIDAcalcal0388380009.01.2014 10:09:2709.01.2014 10:38:23GoodHV=100V  I=4.2uA
centered position
5GeV beam, single sample
  
  46   Thu Jan 9 11:02:14 2014 Markus FriedlAIDAinfo    UnknownCreated bad strip masks for the AIDA config
files
  
  47   Thu Jan 9 11:04:34 2014 Markus FriedlAIDArunrun039100k09.01.2014 11:02:3609.01.2014 12:28:41GoodHV=10V  I=2.1uA
centered position, HV scan
5GeV beam, 6 samples
  
  48   Thu Jan 9 12:31:29 2014 Markus FriedlAIDArunrun040100k09.01.2014 12:30:3209.01.2014 13:55:47GoodHV=20V  I=2.9uA
centered position, HV scan
5GeV beam, 6 samples
  
  49   Thu Jan 9 14:07:31 2014 Markus FriedlAIDArunrun041100k09.01.2014 14:06:3209.01.2014 15:33:13GoodHV=30V  I=3.3uA
centered position, HV scan
5GeV beam, 6 samples
  
  51   Thu Jan 9 16:36:30 2014 Markus FriedlAIDArunrun042100k09.01.2014 16:35:4309.01.2014 18:02:17GoodHV=40V  I=3.4uA
centered position, HV scan
5GeV beam, 6 samples
  
  52   Thu Jan 9 18:09:40 2014 Markus FriedlAIDArunrun044100k09.01.2014 18:08:1909.01.2014 19:35:21GoodHV=50V  I=3.6uA
centered position, HV scan
5GeV beam, 6 samples
  
  53   Thu Jan 9 19:37:33 2014 Markus FriedlAIDArunrun045100k09.01.2014 19:36:2109.01.2014 21:01:31 GoodHV=60V  I=3.8uA
centered position, HV scan
5GeV beam, 6 samples
  
  54   Thu Jan 9 21:41:20 2014 Markus FriedlAIDArunrun046100k09.01.2014 21:40:2909.01.2014 23:07:16GoodHV=70V  I=3.8uA
centered position, HV scan
5GeV beam, 6 samples
  
  55   Thu Jan 9 23:27:27 2014 Markus FriedlAIDArunrun047100k09.01.2014 23:26:3510.01.2014 00:52:56GoodHV=80V  I=3.8uA
centered position, HV scan
5GeV beam, 6 samples
  
  56   Fri Jan 10 00:58:46 2014 Markus FriedlAIDArunrun048500k10.01.2014 00:57:4810.01.2014 08:17:57GoodHV=90V  I=3.9uA
centered position, HV scan
5GeV beam, 6 samples
  
  57   Fri Jan 10 08:21:46 2014 Markus FriedlAIDArunrun049100k10.01.2014 08:20:49 10.01.2014 09:49:46GoodHV=100V  I=4.2uA
centered position, HV scan
5GeV beam, 6 samples
  
  58   Fri Jan 10 09:54:11 2014 Markus FriedlAIDArunrun050100k10.01.2014 09:53:1310.01.2014 11:21:12GoodHV=110V  I=4.1uA
centered position, HV scan
5GeV beam, 6 samples
  
  59   Fri Jan 10 11:02:31 2014 Markus FriedlBelle SVDPower supply  10.01.2014 11:00:00 Unknown   KenWood100114.png 
  60   Fri Jan 10 11:26:54 2014 Markus FriedlAIDArunrun051100k10.01.2014 11:26:0810.01.2014 12:53:12GoodHV=120V  I=4.6uA
centered position, HV scan
5GeV beam, 6 samples
  
  61   Fri Jan 10 13:04:49 2014 Markus FriedlAIDArunrun052100k10.01.2014 13:04:0510.01.2014 14:30:18GoodHV=130V  I=4.6uA
centered position, HV scan
5GeV beam, 6 samples
  
  62   Sat Jan 11 20:56:29 2014 Markus FriedlAIDArunrun053100k11.01.2014 20:55:2911.01.2014 22:25:43GoodHV=140V  I=7.1uA
centered position, HV scan
5GeV beam, 6 samples
  
  63   Sat Jan 11 22:32:19 2014 Markus FriedlAIDArunrun054800k11.01.2014 22:30:4412.01.2014 10:13:10GoodHV=150V  I=5.3uA
centered position, HV scan
5GeV beam, 6 samples
  
  64   Sun Jan 12 10:36:37 2014 Markus FriedlAIDArunrun05531091312.01.2014 10:34:4412.01.2014 15:42:47GoodHV=100V  I=3.6uA
position = 23 / -35 mm relative to
before
  
  65   Sun Jan 12 18:04:18 2014 Markus FriedlAIDArunrun056510k12.01.2014 17:59:5713.01.2014 01:47:50GoodHV=100V  I=3.7uA
position = -25 / 0 mm relative to
initial
  
  66   Mon Jan 13 01:58:14 2014 Markus FriedlAIDArunrun057900k13.01.2014 01:56:2613.01.2014 10:48:22GoodHV=100V  I=3.7uA
position = 0 / -30 mm relative to
initial (but again centered)
  
  67   Mon Jan 13 12:02:40 2014 Markus FriedlcommonKontakt    UnknownRichard Thalmeier   Tel. 4801
Hao Yin   
  
  69   Thu Jan 23 22:47:49 2014 Hao Yin AIDACalibrationRun0018380023.01.2014 13:35:5523.01.2014 02:08:15Goodtaken with tuxdaq...
HV=75V  I=?3.7uA?
position = 0 / -30 mm relative to
  
  70   Fri Jan 24 00:21:16 2014 Hao Yin AIDAHW_AIDA00140k24.02.2014 00:18 UnknownHV=100V  I=3.7uA
position = 0 / -30 mm relative to
initial (but again centered) no idea ---
  
  72   Fri Jan 24 11:23:15 2014 Hao Yin AIDAHW_AIDA0037635624.02.2014 10:3024.02.2014 11:22UnknownHV=100V  I=3.7uA
position = 0 / -30 mm relative to
initial (but again centered) no idea ---
  
  68   Sun Jan 19 18:01:46 2014 Christian IrmlercommonAIDA setup operated with TuxDAQ    UnknownNew SIS driver version sis1100-2.13-3 installed
on DAQPC. 
To install the driver call as root:
  