  Alibava testbeam at DESY in Oct. 2017 (Dominic, Peter), Page 7 of 7  Not logged in ELOG logo
Entry  Fri Oct 20 14:57:30 2017, Peter Paulitsch, IFX_200_3e14, run000182, 182, 100000, 75.93, 65.30, -115.55°, Yes, Good 
run settings
Parameter Value
Alibava filename  run000182.dat
Telescope filename  run000182.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current 40 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 100000
Telescope triggers 100000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.6%

 

Entry  Fri Oct 20 14:58:56 2017, Peter Paulitsch, IFX_200_3e14, run000184, 184, 100000, 75.93, 65.30, -110.55°, Yes, Good 
run settings
Parameter Value
Alibava filename  run000184.dat
Telescope filename  run000184.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current 40 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 100000
Telescope triggers 100000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.5%

 

Entry  Fri Oct 20 15:00:11 2017, Peter Paulitsch, IFX_200_3e14, run000188, 188, 100000, 75.93, 65.30, -100.55°, Yes, Good 
run settings
Parameter Value
Alibava filename  run000188.dat
Telescope filename  run000188.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current 40 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 100000
Telescope triggers 100000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.2%

 

Entry  Fri Oct 20 15:02:42 2017, Peter Paulitsch, IFX_200_3e14, run000190, 190, 100000, 75.93, 65.30, -100.55°, Yes, Good 
run settings
Parameter Value
Alibava filename  run000190.dat
Telescope filename  run000190.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current 40 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 100000
Telescope triggers 100000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.9%

 

Entry  Fri Oct 20 15:06:38 2017, Peter Paulitsch, IFX_200_3e14, run000193, 193, 100000, 75.93, 65.30, -100.55°, Yes, Good 
run settings
Parameter Value
Alibava filename  run000193.dat
Telescope filename  run000193.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current 40 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 100000
Telescope triggers 100000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.3%

 

Entry  Fri Oct 20 15:26:26 2017, Peter Paulitsch, IFX_200_3e14, run000201, 201, 100000, 75.93, 65.30, -90.55°, Yes, Good 
run settings
Parameter Value
Alibava filename  run000201.dat
Telescope filename  run000201.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current 40 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 100000
Telescope triggers 100000
Data synced no
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.7%

 

Entry  Sat Oct 21 10:15:13 2017, Peter Paulitsch, HPK_3e14, run000228, 228, 100000, 75.93, 65.30, -70.55°, Yes, Good 
run settings
Parameter Value
Alibava filename  run000228.dat
Telescope filename  run000228.raw
Pedestal filename KIT_HPK_200_3e14.ped
Bias voltage -600 V
Dark current 40 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 100000
Telescope triggers 100000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency  200
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 46.1%

 

Entry  Sat Oct 21 10:17:03 2017, Peter Paulitsch, HPK_3e14, run000231, 231, 100000, 75.93, 65.30, -70.55°, Yes, Good 
run settings
Parameter Value
Alibava filename  run000231.dat
Telescope filename  run000231.raw
Pedestal filename KIT_HPK_200_3e14.ped
Bias voltage -600 V
Dark current 40 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 100000
Telescope triggers 100000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency  200
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 46.1%

 

Entry  Mon Oct 23 12:19:57 2017, Peter Paulitsch, IFX_200_3e14, run000186, 188, 100000, 75.93, 65.30, -100.55°, Yes, Good 
run settings
Parameter Value
Alibava filename  run000186.dat
Telescope filename  run000186.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current 40 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 100000
Telescope triggers 100000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.1%

 

Entry  Mon Oct 30 12:04:14 2017, Peter Paulitsch, IFX_200_1e15, run000389, 389, 100000, 76.50, 66.00, 0°, Yes, Unknown 
Parameter Value
Alibava filename  run000389.dat
Telescope filename  run000389.raw
Pedestal filename KIT_pedestal_600V_26102017.ped
Bias voltage -600 V
Dark current  
Sensor temperature  
Latency 140 [a.u.]
Alibava triggers 100000
Telescope triggers 100000
Data synced yes
Table position x/y 76.50, 66.00
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency  
Entry  Wed Oct 18 12:28:08 2017, Peter Paulitsch, IFX_CenterBias, run000101, 101, 10000, 122.9, 23.8, -3.3°, No, Good 
run settings
Parameter Value
Alibava filename  run000101.dat
Telescope filename  run000101.raw
Pedestal filename pedestal_17102017
Bias voltage -100 V
Dark current -160 nA
Sensor temperature 24.6 °C
Latency 140 [a.u.]
Alibava triggers 10000
Telescope triggers 14500
Data synced yes
Table position x/y 122.9, 23.8
Region of interest bias line
Trigger frequency  185 Hz
Meas. time 2 min
Beam energy 5.6 GeV
Cooling no
Alibava efficiency 52.3%

 

Entry  Thu Oct 12 10:06:31 2017, Dominic Bloech, common, Packlist, , , , No, Unknown ALiBaVa-System_klein.jpgStuff_klein.jpg

# Testbeam Packliste


### Alibava stuff

* Alibava Mainboard
* Alibava power supply 
* Alibava Modules: IFX_CenterBias, HPK_CenterBias, IFX_baby
* Flatband cable 2x
* Alibava resetbox + Cable
* USB AtoB 2x
* BIAS Cable for BeetleChip
* 2410 SMU + Power + USBtoRS232
* Alibava mounting stuff


### Tools and Stuff

* Labeling machine
* Duct tape
* 2 pairs of wool gloves
* Seitenschneider klein
* USBtoRS232 2x
* Lan cable
* Schiebelehre
* Imbuschlüssel
* Screwdriver set
* Special screwdriver set
* Cutter 2x
* labeling cartridge
* Lemo 16 ns
* Lemo 8 ns 2x
* Lemo 4 ns
* Lemo 2 ns

 

 

### PC and stuff

* DAQ PC + Cables

* Monitor + Cables
* Mouse
* Keyboard

 

### Data Storage
* Toshiba 1TB USB3
* Maxtor 4TB USB3

Entry  Mon Oct 16 15:01:21 2017, Dominic Bloech, common, Preparations, , , , No, Unknown IMG_20171016_093005514.jpgIMG_20171016_150215807.jpgIMG_20171016_150230304.jpgIMG_20171016_150238370.jpg

### General

* Setup was set-up without problems. Pedestral and Calibration runs with ALIBAVA successfully.

** Warning: Trigger Out signal (busy signal) for the TLU is a TTL signal. Whereas the Trigger In signal can be a NIM signal. Either converting signal or adept in the TLU (we adepted the TLU matrix)

** TLU double triggered -> discriminated TLU signal beforehand 

*  Longterm BIAS Voltage for the IFX CenterBias to -100 V, slow control records IV-behavior in 10 sec steps.

* Security lectures taken

* 3 coincident measurements: 2 PM and the FEI4-Plane are connected to an "and" in the TLU

* Latency of Alibava without FEI4-Plane 133 [a.u] with FEI4 140 [a.u].

** Warning: This is due to the converter from the FEI4 which inverts the signal of the FEI4, this signal is then very noisy and would generate false positive triggers. -> Discriminate -> Long cables and latencies

### Other

* NI-Crate was broken (replaced by DESY)