  Alibava testbeam at DESY in Oct. 2017 (Dominic, Peter), Page 2 of 7  Not logged in ELOG logo
New entries since:Thu Jan 1 01:00:00 1970
ID Date Author DUT Subject Run Number Eventsdown XYPosition Cooling Data
  51   Sat Oct 21 10:12:42 2017 Peter PaulitschIFX_200_3e14run00022322330000075.93, 65.30, -70.55°YesGood
run settings
Parameter Value
Alibava filename  run000223.dat
Telescope filename  run000223.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -800 V
Dark current  
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 300000
Telescope triggers 300000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 32.3%

 

  57   Sat Oct 21 10:20:39 2017 Peter PaulitschHPK_3e14run00023523530000075.93, 65.30, -70.55°YesGood
run settings
Parameter Value
Alibava filename  run000235.dat
Telescope filename  run000235.raw
Pedestal filename KIT_HPK_200_3e14.ped
Bias voltage -800 V
Dark current 40 uA but thermal runaway
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 310000
Telescope triggers 300000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency  200
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 46.1%

 

  93   Tue Oct 24 18:27:22 2017 Peter PaulitschHPK_3e14run0003083083000075.50, 66.00, -40°YesGood
Parameter Value
Alibava filename  run000308.dat
Telescope filename  run000308.raw
Pedestal filename KIT_pedestal_600V_22102017.ped
Bias voltage -600 V
Dark current  
Sensor temperature  
Latency 140 [a.u.]
Alibava triggers 30000
Telescope triggers 30000
Data synced yes
Table position x/y 75.50, 66.00
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency  
  97   Tue Oct 24 18:29:50 2017 Peter PaulitschHPK_3e14run0003153153000075.50, 66.00, -40°YesGood
Parameter Value
Alibava filename  run000315.dat
Telescope filename  run000315.raw
Pedestal filename KIT_pedestal_600V_22102017.ped
Bias voltage -600 V
Dark current  
Sensor temperature  
Latency 140 [a.u.]
Alibava triggers 30000
Telescope triggers 30000
Data synced yes
Table position x/y 75.50, 66.00
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency  
  129   Mon Oct 30 12:00:43 2017 Peter PaulitschIFX_200_1e15run000383383270999976.50, 66.00, 0°YesGood
Parameter Value
Alibava filename  run000383.dat
Telescope filename  run000383.raw
Pedestal filename KIT_pedestal_600V_26102017.ped
Bias voltage -600 V
Dark current  
Sensor temperature  
Latency 140 [a.u.]
Alibava triggers 2709999
Telescope triggers 2709950
Data synced yes
Table position x/y 76.50, 66.00
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency  
  5   Tue Oct 17 15:15:32 2017 Dominic BloechIFX_CenterBiasrun000083 83260000122.9, 26, -3.3°NoUnknown
run settings
Parameter Value
Alibava filename  run000083.dat
Telescope filename  run000083.raw
Pedestal filename pedestal_17102017
Bias voltage -100 V
Dark current -162 nA
Sensor temperature 22 °C
Latency 140 [a.u.]
Alibava triggers 260000
Telescope triggers  
Data synced yes
Table position x/y 122.9, 26
Region of interest all
Trigger frequency  220 Hz
Meas. time 90 min
Beam energy 5.6 GeV
Cooling no
Alibava efficiency 44,2%

 

  49   Sat Oct 21 09:57:58 2017 Peter PaulitschIFX_200_3e14run00022122125000075.93, 65.30, -70.55°YesGood
run settings
Parameter Value
Alibava filename  run000221.dat
Telescope filename  run000221.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -800 V
Dark current  
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 250000
Telescope triggers 250000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.4%

 

  50   Sat Oct 21 10:04:21 2017 Peter PaulitschIFX_200_3e14run00022222225000075.93, 65.30, -70.55°YesGood
run settings
Parameter Value
Alibava filename  run000222.dat
Telescope filename  run000222.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -800 V
Dark current  
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 250000
Telescope triggers 250000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.4%

 

  122   Mon Oct 30 11:48:12 2017 Peter PaulitschHPK_200_1e15run00036436425000076.50, 66.00, 5°YesGood
Parameter Value
Alibava filename  run000364.dat
Telescope filename  run000364.raw
Pedestal filename KIT_pedestal_600V_23102017.ped
Bias voltage -600 V
Dark current  
Sensor temperature  
Latency 140 [a.u.]
Alibava triggers 250000
Telescope triggers 250000
Data synced yes
Table position x/y 76.50, 66.00
Region of interest bias line
Trigger frequency  135 Hz
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency  
  29   Fri Oct 20 14:50:03 2017 Peter PaulitschIFX_200_3e14run00017917920000075.93, 65.30, -70.55°YesGood
run settings
Parameter Value
Alibava filename  run000179.dat
Telescope filename  run000179.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -700 V
Dark current 40 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 200000
Telescope triggers 200000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.2%

 

  99   Tue Oct 24 18:33:15 2017 Peter PaulitschHPK_3e14run00031731720000075.50, 66.00, -40°YesGood
Parameter Value
Alibava filename  run000317.dat
Telescope filename  run000317.raw
Pedestal filename KIT_pedestal_600V_22102017.ped
Bias voltage -600 V
Dark current  
Sensor temperature  
Latency 140 [a.u.]
Alibava triggers 200000
Telescope triggers 200000
Data synced yes
Table position x/y 75.50, 66.00
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency  
  55   Sat Oct 21 10:18:38 2017 Peter PaulitschHPK_3e14run00023323315000075.93, 65.30, -70.55°YesGood
run settings
Parameter Value
Alibava filename  run000233.dat
Telescope filename  run000233.raw
Pedestal filename KIT_HPK_200_3e14.ped
Bias voltage -800 V
Dark current 40 uA thermal runaway
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 150000
Telescope triggers 150000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency  200
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 46.1%

 

  118   Mon Oct 30 11:44:03 2017 Peter PaulitschHPK_200_1e15run00035435415000076.50, 66.00, 0°YesGood
Parameter Value
Alibava filename  run000354.dat
Telescope filename  run000354.raw
Pedestal filename KIT_pedestal_600V_23102017.ped
Bias voltage -400 V
Dark current  
Sensor temperature  
Latency 140 [a.u.]
Alibava triggers 150000
Telescope triggers 150000
Data synced yes
Table position x/y 76.50, 66.00
Region of interest bias line
Trigger frequency  135 Hz
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency  
  128   Mon Oct 30 11:55:04 2017 Peter PaulitschIFX_200_1e15run00037937915000076.50, 66.00, 0°YesGood
Parameter Value
Alibava filename  run000379.dat
Telescope filename  run000379.raw
Pedestal filename KIT_pedestal_600V_26102017.ped
Bias voltage -400 V
Dark current  
Sensor temperature  
Latency 140 [a.u.]
Alibava triggers 150000
Telescope triggers 150000
Data synced yes
Table position x/y 76.50, 66.00
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency  
  10   Wed Oct 18 12:22:03 2017 Peter PaulitschIFX_CenterBiasrun00009999110000122.9, 23.5, -3.3°NoGood
run settings
Parameter Value
Alibava filename  run000099.dat
Telescope filename  run000099.raw
Pedestal filename pedestal_17102017
Bias voltage -100 V
Dark current -160 nA
Sensor temperature 24.6 °C
Latency 140 [a.u.]
Alibava triggers 110000
Telescope triggers 100000
Data synced yes
Table position x/y 122.9, 23.5
Region of interest bias line
Trigger frequency  185 Hz
Meas. time 17 min
Beam energy 5.6 GeV
Cooling no
Alibava efficiency 52.3%

 

  12   Wed Oct 18 13:29:39 2017 Peter PaulitschIFX_CenterBiasrun000102102110000122.9, 23.2, -3.3°NoGood
run settings
Parameter Value
Alibava filename  run000102.dat
Telescope filename  run000102.raw
Pedestal filename pedestal_17102017
Bias voltage -100 V
Dark current -160 nA
Sensor temperature 24.6 °C
Latency 140 [a.u.]
Alibava triggers 110000
Telescope triggers 100000
Data synced yes
Table position x/y 122.9, 23.8
Region of interest bias line
Trigger frequency  185 Hz
Meas. time 17 min
Beam energy 5.6 GeV
Cooling no
Alibava efficiency 52.3%

 

  13   Wed Oct 18 13:30:58 2017 Peter PaulitschIFX_CenterBiasrun000104104110000122.9, 23.5, -3.3°NoUnknown
run settings
Parameter Value
Alibava filename  run000104.dat
Telescope filename  run000104.raw
Pedestal filename pedestal_17102017
Bias voltage -100 V
Dark current -160 nA
Sensor temperature 24.6 °C
Latency 140 [a.u.]
Alibava triggers 110000
Telescope triggers 100000
Data synced yes
Table position x/y 122.9, 23.5
Region of interest bias line
Trigger frequency  200 Hz
Meas. time 15 min
Beam energy 5.6 GeV
Cooling no
Alibava efficiency 52.3%

 

  15   Wed Oct 18 13:34:54 2017 Peter PaulitschIFX_CenterBiasrun000107107110000122.9, 23.5, -3.3°NoGood
run settings
Parameter Value
Alibava filename  run000107.dat
Telescope filename  run000107.raw
Pedestal filename pedestal_17102017
Bias voltage -100 V
Dark current -160 nA
Sensor temperature 24.6 °C
Latency 140 [a.u.]
Alibava triggers 110000
Telescope triggers 100000
Data synced yes
Table position x/y 122.9, 23.5
Region of interest bias line
Trigger frequency  200 Hz
Meas. time 15 min
Beam energy 5.6 GeV
Cooling no
Alibava efficiency 53.4%

 

  16   Wed Oct 18 13:36:50 2017 Peter PaulitschIFX_CenterBiasrun000109109110000122.9, 25.75, -3.3°NoGood
run settings
Parameter Value
Alibava filename  run000109.dat
Telescope filename  run000109.raw
Pedestal filename pedestal_17102017
Bias voltage -100 V
Dark current -160 nA
Sensor temperature 24.6 °C
Latency 140 [a.u.]
Alibava triggers 110000
Telescope triggers 100000
Data synced yes
Table position x/y 122.9, 25.75
Region of interest bias line
Trigger frequency  200 Hz
Meas. time 15 min
Beam energy 5.6 GeV
Cooling no
Alibava efficiency 53.4%

 

  18   Fri Oct 20 12:15:47 2017 Peter PaulitschIFX_200_3e14run00011611611000075.93, 63.00, -72.55°YesGood
run settings
Parameter Value
Alibava filename  run000116.dat
Telescope filename  run000116.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current 39 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 110000
Telescope triggers 100000
Data synced yes
Table position x/y 75.93, 63.00
Region of interest bias line
Trigger frequency  210 Hz
Meas. time 15 min
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.2%

 

ELOG V3.1.5-fc6679b