  Alibava testbeam at DESY in Oct. 2017 (Dominic, Peter), Page 5 of 7  Not logged in ELOG logo
New entries since:Thu Jan 1 01:00:00 1970
ID Date Author DUT Subject Run Number Eventsdown XYPosition Cooling Data
  4   Tue Oct 17 15:13:42 2017 Dominic BloechIFX_CenterBiasrun000081811010000122.9, 26, -3.3°NoGood
run settings
Parameter Value
Alibava filename  run000081.dat
Telescope filename  run000081.raw
Pedestal filename pedestal_17102017
Bias voltage -100 V
Dark current -162 nA
Sensor temperature 22 °C
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y 122.9, 26
Region of interest all
Trigger frequency  220 Hz
Meas. time 90 min
Beam energy 5.6 GeV
Cooling no
Alibava efficiency 44.1%

 

  6   Wed Oct 18 11:48:51 2017 Peter PaulitschIFX_CenterBiasrun000085851010000122.9, 26, -3.3°NoGood
run settings
Parameter Value
Alibava filename  run000085.dat
Telescope filename  run000085.raw
Pedestal filename pedestal_17102017
Bias voltage -100 V
Dark current -165 nA
Sensor temperature 24.5 °C
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y 122.9, 26
Region of interest all
Trigger frequency  185 Hz
Meas. time 90 min
Beam energy 5.6 GeV
Cooling no
Alibava efficiency 44.1%

 

  7   Wed Oct 18 11:59:30 2017 Peter PaulitschIFX_CenterBiasrun000088881010000122.9, 31, -3.3°NoGood
run settings
Parameter Value
Alibava filename  run000088.dat
Telescope filename  run000088.raw
Pedestal filename pedestal_17102017
Bias voltage -100 V
Dark current -165 nA
Sensor temperature 24.5 °C
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y 122.9, 31
Region of interest upper part
Trigger frequency  185 Hz
Meas. time 90 min
Beam energy 5.6 GeV
Cooling no
Alibava efficiency 45.4%

 

  8   Wed Oct 18 12:12:55 2017 Peter PaulitschIFX_CenterBiasrun000091911010000122.9, 21, -3.3°NoGood
run settings
Parameter Value
Alibava filename  run000091.dat
Telescope filename  run000091.raw
Pedestal filename pedestal_17102017
Bias voltage -100 V
Dark current -165 nA
Sensor temperature 24.6 °C
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y 122.9, 21
Region of interest lower part
Trigger frequency  185 Hz
Meas. time 90 min
Beam energy 5.6 GeV
Cooling no
Alibava efficiency 44.7%

 

  9   Wed Oct 18 12:19:14 2017 Peter PaulitschIFX_CenterBiasrun000093931010000122.9, 26, -3.3°NoGood
run settings
Parameter Value
Alibava filename  run000093.dat
Telescope filename  run000093.raw
Pedestal filename pedestal_17102017
Bias voltage -100 V
Dark current -165 nA
Sensor temperature 24.5 °C
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y 122.9, 26
Region of interest all
Trigger frequency  185 Hz
Meas. time 90 min
Beam energy 5.6 GeV
Cooling no
Alibava efficiency 44.1%

 

  28   Fri Oct 20 14:35:48 2017 Peter PaulitschIFX_200_3e14run000177177101000075.93, 65.30, -70.55°YesGood
run settings
Parameter Value
Alibava filename  run000177.dat
Telescope filename  run000177.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -400 V
Dark current 40 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.8%

 

  42   Fri Oct 20 15:30:18 2017 Peter PaulitschIFX_200_3e14run000206206101000075.93, 65.30, -90.55°YesGood
run settings
Parameter Value
Alibava filename  run000206.dat
Telescope filename  run000206.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current 40 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.1%

 

  44   Fri Oct 20 15:33:44 2017 Peter PaulitschIFX_200_3e14run000210210101000075.93, 65.30, -80.55°YesGood
run settings
Parameter Value
Alibava filename  run000210.dat
Telescope filename  run000210.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current 40 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.9%

 

  47   Fri Oct 20 15:41:56 2017 Peter PaulitschIFX_200_3e14run000219219101000075.93, 65.30, -70.55°YesGood
run settings
Parameter Value
Alibava filename  run000219.dat
Telescope filename  run000219.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -500 V
Dark current 40 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.5%

 

  60   Mon Oct 23 09:24:23 2017 Dominic BloechHPK_3e14run000240240101000075.93, 65.30, -70.55°YesCrap
Parameter Value
Alibava filename  run000240.dat
Telescope filename  run000240.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -100 V
Dark current  
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 3010000
Telescope triggers 3000000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency  200
Meas. time 80min
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 29.7%
  62   Mon Oct 23 12:10:15 2017 Peter PaulitschIFX_200_3e14run000175175101000075.93, 65.30, -70.55°YesGood
run settings
Parameter Value
Alibava filename  run000177.dat
Telescope filename  run000177.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current 40 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.1%

 

  65   Tue Oct 24 17:50:04 2017 Peter PaulitschHPK_3e14run000242242101000075.93, 65.30, -70.55°YesGood
Parameter Value
Alibava filename  run000242.dat
Telescope filename  run000242.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current  
Sensor temperature -7
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 29.7%
  66   Tue Oct 24 17:51:18 2017 Peter PaulitschHPK_3e14run000246246101000075.93, 65.30, -100.55°YesGood
Parameter Value
Alibava filename  run000246.dat
Telescope filename  run000246.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current  
Sensor temperature -7
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency  
  68   Tue Oct 24 17:56:36 2017 Peter PaulitschHPK_3e14run000249249101000075.93, 65.30, -90.55°YesGood
Parameter Value
Alibava filename  run000249.dat
Telescope filename  run000249.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current  
Sensor temperature -7
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency  
  71   Tue Oct 24 17:58:43 2017 Peter PaulitschHPK_3e14run000257257101000075.93, 65.30, -80.55°YesGood
Parameter Value
Alibava filename  run000257.dat
Telescope filename  run000257.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current  
Sensor temperature -7
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency  
  73   Tue Oct 24 18:01:13 2017 Peter PaulitschHPK_3e14run000261261101000075.93, 65.30, -70.55°YesGood
Parameter Value
Alibava filename  run000261.dat
Telescope filename  run000261.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -500 V
Dark current  
Sensor temperature -7
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency  
  79   Tue Oct 24 18:08:57 2017 Peter PaulitschHPK_200run0002752751010000NoGood
Parameter Value
Alibava filename  run000275.dat
Telescope filename  run000275.raw
Pedestal filename HEPHY_pedestal_HPK_CB_2.ped
Bias voltage -200 V
Dark current  
Sensor temperature  
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y  
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling no
Alibava efficiency  
  80   Tue Oct 24 18:09:24 2017 Peter PaulitschHPK_200run0002772771010000NoGood
Parameter Value
Alibava filename  run000277.dat
Telescope filename  run000277.raw
Pedestal filename HEPHY_pedestal_HPK_CB_2.ped
Bias voltage -200 V
Dark current  
Sensor temperature  
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y  
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling no
Alibava efficiency  
  83   Tue Oct 24 18:10:39 2017 Peter PaulitschHPK_200run0002832831010000NoGood
Parameter Value
Alibava filename  run000283.dat
Telescope filename  run000283.raw
Pedestal filename HEPHY_pedestal_HPK_CB_2.ped
Bias voltage -200 V
Dark current  
Sensor temperature  
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y  
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling no
Alibava efficiency  
  84   Tue Oct 24 18:11:13 2017 Peter PaulitschHPK_200run0002852851010000NoGood
Parameter Value
Alibava filename  run000285.dat
Telescope filename  run000285.raw
Pedestal filename HEPHY_pedestal_HPK_CB_2.ped
Bias voltage -200 V
Dark current  
Sensor temperature  
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y  
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling no
Alibava efficiency  