Home CMS Production Clean room MedAustron HEPHY testbeams old
testbeam-RD50-DESY-Jul2023 testbeam-RD50-CERN-Oct2022 testbeam-DESY-Oct2017 testbeam-DESY-April2017 testbeam-SPS2015 testbeam-SPS2014 testbeam-DESY14 testbeam-SPS12 testbeam-SPS11 testbeam-SPS10 testbeam-SPS09 testbeam-SPS08 LP-TPC
  Belle testbeam at SPS in June 2015, Page 1 of 6  Not logged in ELOG logo
ID Date Author Project Subject Run Number Events XYPosition Cooling Datadown
  1   Tue May 12 10:27:39 2015 Benedikt WürknerSVD3 (H6B)Setup information    Unknown

Information concerning the SVD3 setup during a beam test

In the beam area setup the whole crate and the power supply with the usual spider cable and the two Keithly 2410 high voltage supplies. Monitoring of the 2410 happens via a NI MAX interface that connects to the GPIB-Ethernet interface and needs to be configured there. The IP and password are noted on the device. The software can be found on heros in medialib/LONGTERM/longterm R3.1_doubleSided as a normal *.vi file. 

Default adresses 16 and 17 as defined in labview, need to figure out on site which is which, if the voltage selected (e.g. 100V) is applied as +50/-50 or if it is applied as +100/-100 before connecting the cables. 

Take a run without dut as well to check the telescope integrity. Without dut in beamline!!!!!

Attachment 1: spider-config.jpg
spider-config.jpg
  2   Wed Jun 3 10:48:42 2015 Benedikt WürknercommonNetworking    Unknown

WLAN Router:

SSID: Hephy

password: 20supersecret14

DHCP Range: 192.168.0.200 to 239

Netmask: /24

DNS Server: 192.168.0.254

Gateway: 192.168.0.254

external IP: 128.141.174.109 (für diverse Port Forwardings)

 

Manuell vergebene IP Adressen bitte unbedingt Benni mitteilen oder/und in der folgenden Tabelle eintragen! 

IP/MAC Zuweisung:

IP Machine/Name intern Port Forwarding (aussen)
192.168.0.124 SVD Kenwood  VNC(5900)  VNC(9000)
192.168.0.11 SVD3 DAQ  VNC(5900)  VNC(9001)
192.168.0.12 TUX DAQ  VNC(5900), VNC(5901), SSH(22)  VNC(9009), VNC(9002), SSH(9003),
192.168.0.13 Belle XY Tisch    
192.168.0.14 TUXOA  SSH(22), VNC(5900)  SSH(9004), VNC(9012)
192.168.0.20 JTAG Server    
192.168.0.148 alixcool    
192.168.0.16 slowcontrol  VNC(5900)  VNC(9005)
192.168.0.144 CAEN Power supply    
192.168.0.17 zenpixelwiki  SSH(22), VNC(5900)  SSH(9006), VNC(9008)
192.168.0.18 TLU (pcx25042, XY)  VNC(5900)  VNC(9007)
192.168.0.19 Cooling control PC VNC(5900) VNC(9011)
192.168.0.240 NI GPIB - ENET / 1000    
192.168.0.150 Printserver  VNC(5900)  VNC(9010)
       
137.138.113.254 Aida PC   vnc(5900)

 

Admin Zugang Wlan Router:

IP: 192.168.0.254

Username: admin

password: Hephy2014

 

  3   Wed Jun 3 17:23:07 2015 Thomas BergauerBelle II FADC (H6A)XY stage positions    Unknown

x-Achse: positiv Richtung Saleve (counting room)

y-Achse: positiv Richtung UNTEN!!

 

position of laddertestbox xy stages:

Rev. Pos (BWD Sensor, BWD, Oben) x= 82.3, y=110.6

Mitte Sensor:

Pos 1: (mitte BWD sensor): x=144.74, y=80.8

Pos 2: (mitte Origami -z): x=269.62, y=80.8

Pos 3: (mitte Origami ce): x=394.42, y=80.8

Pos 4 (mitte FWD Micron Sensor): x=517, y=80.8

Ecke:

Pos 5 (Ecke BWD sensor): x=92.3, y=105.6

Pos 6 (Ecke Origami -z): x=217.18, y=105.6

Pos 7 (Ecke Origami ce): x=342.06, y=105.6

Pos 8 (Ecke FWD sensor): x=467,62, y=56,3 

Pos 9 (Ecke FWD sensor): x=568,33 y=64,48

Attachment 1: L5_Class_B-.PNG
L5_Class_B-.PNG
  9   Thu Jun 4 18:49:03 2015 Benedikt WürknerSVD3 (H6B)Setup    Unknown
Attachment 1: IMG_20150604_184442.jpg
IMG_20150604_184442.jpg
Attachment 2: IMG_20150604_184451.jpg
IMG_20150604_184451.jpg
Attachment 3: IMG_20150604_184516.jpg
IMG_20150604_184516.jpg
  14   Fri Jun 5 19:55:00 2015 Hao YinBelle II FADC (H6A)Sixlet Calibration003   Unknown

lat = 98

room temp

pedestalrun 003

  16   Fri Jun 5 23:38:51 2015 Hao YinBelle II FADC (H6A)Pedestal00510000  Unknown

Readjust after cooling box.

room temp, firrun002
 

  20   Sat Jun 6 16:14:36 2015 Benedikt WürknerSVD3 (H6B)calibration run and setup informationcal01   Unknown

Calibration Run with +- 50V Bias with higher compliance since the p-side draws more current than expected (23muA by the MAMBO)

 

  22   Sat Jun 6 16:18:48 2015 Benedikt WürknerSVD3 (H6B)hardware runhw00350000  Unknown

Table: 100:70

Bias: +-50V

Run to confirm trigger delay

  24   Sat Jun 6 21:00:15 2015 Benedikt WürknerSVD3 (H6B)first hardware run with TLU and Telescopehw004117171  Unknown

Table: 100:70

Bias: +-50V

600 pedestal

0 offset (at the beginning)

Started APVDAQ and waited till 600 were taken then started EUDAQ

EUDAQ Run 21

Mimosa Thresholds lvl 6

  27   Sun Jun 7 10:36:38 2015 Hao YinBelle II FADC (H6A)Pedestal00610000  Unknown

Cooled, Temp see attachment
FirRun002
 

Attachment 1: Pos1.PNG
Pos1.PNG
  30   Sun Jun 7 12:10:24 2015 Benedikt WürknerSVD3 (H6B)Telescope Threshold8b test run 004614399  Unknown

Telescope in combination with SVD3

  31   Sun Jun 7 12:11:42 2015 Benedikt WürknerSVD3 (H6B)Telescope Threshold7 test run 004738175  Unknown

Telescope alone

  32   Sun Jun 7 12:12:00 2015 Benedikt WürknerSVD3 (H6B)Telescope Threshold7 test run 004853356  Unknown

Telescope alone

  36   Sun Jun 7 14:44:29 2015 Thomas BergauerSVD3 (H6B)XY Stage and DUT Positions    Unknown

Coordinate system:

pos x-Direction = Jura side

pos y-Direction = Down

Zero-point is in the top right corner in Beam direction

 

absolute Position of PI Stages:

Pos 1: x=100 mm, y=79 mm (Hits in APV 3 und 8-9 in p-Seite und 14 und 17 in n-Seite )

Pos 2: x=69.3 mm, y=79 mm (Hits in APV 3-4 und 8-9 in p-Seite und 13 und 18 in n-Seite) 

Pos 3: x=38.6 mm, y=84 mm

Pos 4: x=130.7 mm, y=84 mm

Pos 5: x=30.5 mm, y= 99.74 mm (wide side top)

Pos 6: x=136.08 mm, y= 91.49 mm (narrow side top)

Pos 7: x= 30.5 mm, y= 79 mm (wide side center)

Pos 8: x=136.08 mm, y= 79 mm (narrow side center)

Pos 9: x=30.5 mm, y= 60.67 mm (wide side down)

Pos 10: x= 136.08mm, y=68.9 mm (narrow side down)

Attachment 1: FWD-BWD.PNG
FWD-BWD.PNG
Attachment 2: apv-numbering.jpg
apv-numbering.jpg
Attachment 3: FWD-BWD_Bemaßung.PNG
FWD-BWD_Bemaßung.PNG
  38   Sun Jun 7 15:59:05 2015 Hao YinBelle II FADC (H6A)Hardware01250000  Unknown

cooled (see attachment)

FIRRun002, PedestalRun006

target: CE, pos 2

cooling box emergency break ... ie about the last 1-10 k not useable.

Attachment 1: Pos3.PNG
Pos3.PNG
  39   Sun Jun 7 20:02:40 2015 Hao YinBelle II FADC (H6A)Cooling Box Emergency break    Unknown

Something triggered a emergency break. HWRun012 only last 1-10k event has to be discarded.

  50   Mon Jun 8 11:33:24 2015 Thomas Bergauercommonbeam profile    Unknown

120GeV hadrons, beam config H6A.BELLE.003

Attachment 1: Beamprofile.png
Beamprofile.png
  53   Mon Jun 8 15:03:53 2015 Benedikt WürknerSVD3 (H6B)hardware run with TLU and FEI4 WORKING5919000  Unknown

Testrun with the FEI4 working for the first time. 

Lessons learned about the configuration:

EnableDUTVeto needs to be set for the FEI4 to allow it to work

HandShakeMode needs to be enabled for the FEI4 as well

"Enable RJ45 input during scans" needs to be enabled for the USB-controller

  54   Mon Jun 8 15:49:23 2015 Benedikt WürknerSVD3 (H6B)Hardware FEI4605000010NoUnknown

Hardware run with FEI4, Telescope and FW&BW Modules

Pos 10: x= 136.08mm, y=68.9 mm (narrow side down)

 

Bilder: Hitmap FE-I4 und Correlation zw. Mimosa und FE-I4

Attachment 1: Screenshot_2015-06-08_17.59.50.png
Screenshot_2015-06-08_17.59.50.png
Attachment 2: Screenshot_2015-06-08_17.41.03.png
Screenshot_2015-06-08_17.41.03.png
  56   Mon Jun 8 16:53:13 2015 Benedikt WürknerSVD3 (H6B)EUTelescope Fitter not working investigation    Unknown

Apparently every problem can be attributed to a misalignment problem. The telescope has troubles with heavily misaligned DUTs. Was resolved by creating different gear-files for each run with pretty precise coordinates of the DUT position calculated for the telescope coordinate system. 

Additionally one has to watch out for misaligned planes of the telescope itself. It was required to allow for more than 2mm offset for the telescope planes behind the DUTs. This can be easily checked by looking at the PreAligner plots in the -hitmaker-histo.root file.

It look something like this:

and not like this:

 

Run number offset value Comment
49 -1  
50 -1 Expected default value. EORE=49401
51 -1 Expected default value. EORE=49400
52 -1 Checked -1 EORE= 49401
53 -1 Checked -1 EORE= 49401
54 -1 Checked -1 EORE= 119401, 
55 -1 Checked -1 EORE= 409388
56 -1 Checked -1 EORE= 49392
57 -1 Checked -1 EORE= 49377
58 -1 Checked -1 EORE= 49397

 

Run Position x-Table x-telescope FW x-telescope BW
49 1 100 -13,7 -16,7
50 2 69,3 17 14
51 3 38,6 47,7 44,7
52 4 130,7 -44,4 -47,4
53 5 30,5 55,8 52,8
54 6 136,8 -50,5 -53,5
55 7 30,5 55,8 52,8
56 8 136,8 -50,5 -53,5
57 9 30,5 55,8 52,8
58 10 136,8 -50,5 -53,5
Attachment 2: bad-alignment.png
bad-alignment.png
ELOG V3.1.5-fc6679b