ID |
Date |
Author |
Project |
Subject |
Run Number |
Events |
StartTime |
EndTime |
26
|
Tue Sep 28 22:09:39 2010 |
Markus Friedl | BELLE | Temperature sensors Belle stack | | | | |
PicoLogic TC-08 temperature sensor assignment Belle stack:
Sensor type: K
channel 1: Origami_mod1_in
channel 2: Origami_mod1_out
channel 3: Origami_mod2_out (swapped with 4) - label not changed
channel 4: Origami_mod2_in (swapped with 3) - label not changed
channel 5: Micron_wedge_mod1_p
channel 6: Micron_wedge_mod1_n
channel 7: Micron_wedge_mod2_p
channel 8: Micron_wedge_mod2_n
The cooling pipes of the two Origami modules are connected in series, respectively. |
27
|
Tue Sep 28 23:23:47 2010 |
Markus Friedl | BELLE | run | run006 | 2003 | 28.09.2010 23:13:22 | 28.09.2010 23:22:26 |
channels swapped -> stopped |
28
|
Tue Sep 28 23:24:35 2010 |
Markus Friedl | BELLE | run | run007 | 1460 | 28.09.2010 23:23:29 | 28.09.2010 23:27:54 |
Belle stack: Origami-Micron-Micron-Origami
HV=100V, Ibias=38A, beam on
cooling: Origami only, chiller: -25C, Module: -10C
modified I2C settings for cooled modules
centered again
telescope run 20082
sensor config bug (only relevant for sensor display) found -> stopped |
29
|
Tue Sep 28 23:29:50 2010 |
Markus Friedl | BELLE | run | run008 | 140000 | 28.09.2010 23:29:01 | 29.09.2010 09:36:58 |
Belle stack: Origami-Micron-Micron-Origami
HV=100V, Ibias=38A, beam on
cooling: Origami only, chiller: -25C, Module: -10C
modified I2C settings for cooled modules
centered again
telescope run 20083 |
Attachment 1: run008_sensordisplay.jpg
|
|
Attachment 2: run008_sensordisplay2.jpg
|
|
30
|
Wed Sep 29 09:39:33 2010 |
Markus Friedl | BELLE | cal | cal003 | 83800 | 29.09.2010 09:38:00 | |
same as cal002, but now with corrected readout order |
31
|
Wed Sep 29 10:30:59 2010 |
Markus Friedl | BELLE | run | run009 | 30000 | 29.09.2010 10:27:27 | 29.09.2010 11:06:35 |
zero suppressed (HIT4) @ FADC
Belle stack: Origami-Micron-Micron-Origami - HIT4
HV=100V, Ibias=28A, beam on
cooling: Origami only, chiller: -25C, Module: -10C
modified I2C settings for cooled modules
centered in both axes
telescope run 20084 |
32
|
Wed Sep 29 11:09:10 2010 |
Markus Friedl | BELLE | run | run010 | 30000 | 29.09.2010 11:08:03 | 29.09.2010 11:50:07 |
zero suppressed (HIT5) @ FADC
Belle stack: Origami-Micron-Micron-Origami - HIT5
HV=100V, Ibias=28A, beam on
cooling: Origami only, chiller: -25C, Module: -10C
modified I2C settings for cooled modules
centered in both axes
telescope run 20085 |
33
|
Wed Sep 29 11:53:16 2010 |
Erik Huemer | BELLE | run | run011 | 19003 | 29.09.2010 12:19:57 | 29.09.2010 14:00:57 |
Belle stack: Origami-Micron-Micron-Origami
HV=100V, Ibias=83.22µA, beam on
no cooling
Temp. at modules:
Origami 1: 50°C
Micron 1: 46°C
Micron 2: 48°C
Origami 2: 48°C
standard I2C settings
centered
telescope run 20085 (continued) |
34
|
Wed Sep 29 14:03:55 2010 |
Markus Friedl | BELLE | cal | cal004 | 83800 | 29.09.2010 14:02:25 | 29.09.2010 14:45:05 |
Belle stack: Origami-Micron-Micron-Origami
HV=100V, Ibias=100µA, beam off
no cooling
Temp. at modules:
Origami 1: 50°C
Micron 1: 46°C
Micron 2: 48°C
Origami 2: 48°C
standard I2C settings
centered |
81
|
Mon Oct 4 20:16:19 2010 |
Christian Irmler | BELLE | cal | cal011 | 83800 | 04.10.2010 20:15:00 | |
angle scan Origami 1 and Micron wedge 1
HV = 100V
chiller set to 14°C
HV connection of Origami 1 repaired |
82
|
Mon Oct 4 21:01:25 2010 |
Christian Irmler | BELLE | run | run43 | 30000 | 04.10.2010 20:59:48 | 04.10.2010 23:04:24 |
angle scan Origami 1 and Micron wedge 1
HV = 100V
chiller set to 14°C
angle: 60°
axis 1: 92.4 (horizontal)
axis 2: 35 (vertical)
axis 3: -52.5 (rotary table)
telescope run: 20149 |
83
|
Mon Oct 4 23:12:16 2010 |
Manfred Valentan | BELLE | run | run44 | 30000 | 04.10.2010 23:?:? | 04.10.2010 23.57.23 |
angle scan Origami 1 and Micron wedge 1
HV = 100V
chiller set to 14°C
angle: 30°
axis 1: 92.4 (horizontal)
axis 2: 35 (vertical)
axis 3: -22.5 (rotary table)
telescope run: 20150 |
84
|
Mon Oct 4 23:59:05 2010 |
Manfred Valentan | BELLE | run | run45 | | 04.10.2010 ??:??:?? | 05.10.2010 00:56:48 |
angle scan Origami 1 and Micron wedge 1
HV = 100V
chiller set to 14°C
angle: 45°
axis 1: 92.4 (horizontal)
axis 2: 35 (vertical)
axis 3: -37.5 (rotary table)
telescope run: 20151 |
85
|
Tue Oct 5 10:20:30 2010 |
Manfred Valentan | BELLE | run | run46 | 350000 | 05.10.2010 01:00:00 | 05.10.2010 10:16:07 |
angle scan Origami 1 and Micron wedge 1
HV = 100V
chiller set to 14°C
angle: 15°
axis 1: 92.4 (horizontal)
axis 2: 35 (vertical)
axis 3: -7.5 (rotary table)
telescope run: 20152 |
86
|
Tue Oct 5 10:23:09 2010 |
Manfred Valentan | BELLE | run | run47 | 30000 | 05.10.2010 10:21:20 | 05.10.2010 |
angle scan Origami 1 and Micron wedge 1
HV = 100V
chiller set to 14°C
angle: 0°
axis 1: 92.4 (horizontal)
axis 2: 35 (vertical)
axis 3: +7.5 (rotary table)
telescope run: 20153 |
95
|
Wed Oct 6 23:03:15 2010 |
Marko Dragicevic | BELLE | cal | cal014 | 83800 | 06.10.2010 23:01:56 | 06.10.2010 23:34:07 |
Belle stack without Origami2
HV=80V I=227uA
beam on |
96
|
Wed Oct 6 23:48:49 2010 |
Markus Friedl | BELLE | run | run052 | 165789 | 06.10.2010 23:47:12 | 07.10.2010 08:43:42 |
Belle stack without Origami2
HV=80V I=217uA
telescope setting 103 / 30 / 277.5
telescope run 20184
temperature of chiller set to 15°C
beam stopped at 7.10. 08:00 sharp, no beam until 8.10 - 8:00 according to SPS Page 1. We keep DAQ and telescope running and cross our fingers for at least some spills of particles.
Irradiated micron module 2, n-side: excessive hit count on channels at the edges of the APV chips - no such thing on the p-side |
97
|
Fri Oct 8 17:26:36 2010 |
Markus Friedl | BELLE | cal | cal015 | 83800 | 08.10.2010 17:22:44 | |
Belle stack without Origami2
HV=11V I=32uA
beam on
2 suspicious APVs on n side of Micron due to burnt strip
trying to identify problematic Micron n side channels with low bias voltage (APV22 baseline saturates at higher bias) |
98
|
Fri Oct 8 18:12:15 2010 |
Christian Irmler | BELLE | run | run053 | 98000 | 08.10.2010 00:33:14 | 08.10.2010 13:36:57 |
Belle stack w/o Origami2 after irradiation
HV=80V
Telescope position 55/30/277.5
NIM trigger
temperature of chiller set to 15°C
Analog signal of APVs #22 and #23 has been bound to ground after 1200 events. |
99
|
Fri Oct 8 18:12:43 2010 |
Markus Friedl | BELLE | run | run054 | 90000 | 08.10.2010 18:08:48 | 08.10.2010 21:49:36 |
Belle stack w/o Origami2 after irradiation
HV=80V I=224uA (beginning of run)
2 suspicious APVs on Micron n side
xy-table position 7 / 30 / 277.5 (shoots at the smaller edge of the trapezoidal sensor)
telescope run 20195 |