28
|
Tue Sep 28 23:24:35 2010 |
Markus Friedl | BELLE | run | run007 | 1460 | 28.09.2010 23:23:29 | 28.09.2010 23:27:54 |
Belle stack: Origami-Micron-Micron-Origami
HV=100V, Ibias=38A, beam on
cooling: Origami only, chiller: -25C, Module: -10C
modified I2C settings for cooled modules
centered again
telescope run 20082
sensor config bug (only relevant for sensor display) found -> stopped |
27
|
Tue Sep 28 23:23:47 2010 |
Markus Friedl | BELLE | run | run006 | 2003 | 28.09.2010 23:13:22 | 28.09.2010 23:22:26 |
channels swapped -> stopped |
26
|
Tue Sep 28 22:09:39 2010 |
Markus Friedl | BELLE | Temperature sensors Belle stack | | | | |
PicoLogic TC-08 temperature sensor assignment Belle stack:
Sensor type: K
channel 1: Origami_mod1_in
channel 2: Origami_mod1_out
channel 3: Origami_mod2_out (swapped with 4) - label not changed
channel 4: Origami_mod2_in (swapped with 3) - label not changed
channel 5: Micron_wedge_mod1_p
channel 6: Micron_wedge_mod1_n
channel 7: Micron_wedge_mod2_p
channel 8: Micron_wedge_mod2_n
The cooling pipes of the two Origami modules are connected in series, respectively. |
25
|
Tue Sep 28 22:07:16 2010 |
Markus Friedl | BELLE | cal | cal002 DO NOT USE -> go to cal003 | 83800 | 28.09.2010 22:01:55 | |
Belle stack (Origami, Micron, Micron, Origami)
now cooling only for Origami, Micron are uncooled
chiller set point: -25°C, measured values on Origami: -10°C
HV=100V, Ibias=38µA, beam off
modified I2C settings for cold modules
temperature monitoring: Origami2 pipes in<->out swapped (to achieve shorter tubes)
PROBLEM: Origami n channels are swapped -> USE cal003 INSTEAD!!! |
24
|
Tue Sep 28 16:50:31 2010 |
Thomas Bergauer | common | serial connection chiller | | | | |
Der chiller braucht ein gekreuztes serielles Kabel!! |
23
|
Tue Sep 28 15:09:16 2010 |
Markus Friedl | BELLE | run | run005 | 40000 | 28.09.2010 15:07:08 | 28.09.2010 20:20:12 |
previous run stopped due to beam problem (shifted, broad, etc).
now recovered, restarting with same settings.
telescope run: 20080 |
22
|
Tue Sep 28 14:48:00 2010 |
Markus Friedl | BELLE | run | run004 | 1755 | 28.09.2010 14:48:28 | 28.09.2010 15:06:31 |
same as before, but now pos 1 is located on the other edge (-48mm from center)
telescope 7 / 30 / 277.5
telescope run 20079 |
21
|
Tue Sep 28 14:38:24 2010 |
Christian Irmler | BELLE | Temperature sensors Belle stack | | | | |
PicoLogic TC-08 temperature sensor assignment Belle stack:
Sensor type: K
channel 1: Origami_mod1_in
channel 2: Origami_mod1_out
channel 3: Origami_mod2_in
channel 4: Origami_mod2_out
channel 5: Micron_wedge_mod1_p
channel 6: Micron_wedge_mod1_n
channel 7: Micron_wedge_mod2_p
channel 8: Micron_wedge_mod2_n
The cooling pipes of the two Origami modules and that of the Micron modules are connected in series, respectively. |
20
|
Tue Sep 28 13:04:02 2010 |
Markus Friedl | BELLE | run | run003 | 25000 | 28.09.2010 13:00:05 | 28.09.2010 14:41:14 |
now for real (previous run was at wrong position)
all same as run001, except position now still centered on p-strips, but on high edge (last APVs) for n side
position: 103 / 30 / 277.5
corresponds to telescope run 20078 |
19
|
Tue Sep 28 12:54:33 2010 |
Markus Friedl | BELLE | run | run002 | | 28.09.2010 12:46:54 | |
same as run001 after shifting x by 48mm
103 / 30 / 277.5 ? unsure after crash of xy stage PC |
18
|
Tue Sep 28 03:06:31 2010 |
Markus Friedl | BELLE | run | run001 | 119516 | 28.09.2010 03:06:04 | 28.09.2010 12:18:33 |
first run with complete Belle stack: 2 x Origami, 2 x Micron
HV=100V I=49uA
beam is on
chiller is set to 15°C (module temperature is ~17°C)
both Origamis and both Microns are serially cooled
position:centered in both axes 55 / 30 / 277.5
telescope: run20076 (in zero suppressed mode) |
17
|
Tue Sep 28 01:13:57 2010 |
Markus Friedl | BELLE | cal | cal001 | 83800 | 28.09.2010 01:11:20 | |
first intcal with complete Belle stack: 2 x Origami, 2 x Micron
HV=100V I=49uA
beam is on
chiller is set to 15°C (module temperature is ~17°C)
both Origamis and both Microns are serially cooled |
16
|
Mon Sep 27 18:54:47 2010 |
Thomas Bergauer | BELLE | XY Position | | | | |
center position: 55/40/277,5 |
15
|
Mon Sep 27 15:42:09 2010 |
Thomas Bergauer | common | TLU: wir stecken auf DUT 3 | | | | |
TLUControl.exe ist auf zenpixell3 unter /home/eudet/eudaq/bin
TLUControl.exe -d 14 -a 12
-dXX Mask for DUT:
1 abgeschnittenes Dummykabel
2 EUDRB hinten
4 EUDRB vorne
8 DUT3(wir)
-t100 internal trigger
-aXX Mask for Szintis:
1 4 mm (front)
2 4 mm (front)
4 9 mm (back)
8 9 mm (back) |
14
|
Mon Sep 27 15:41:28 2010 |
Thomas Bergauer | common | Szintillatoren | | | | |
1: breit (1)
2: schmal (2)
3: breit (4)
4: schmal (8)
AND aus beiden breiten: 5
AND aus beiden schmalen: 10
|
13
|
Mon Sep 27 15:39:33 2010 |
Thomas Bergauer | common | PI-XY Table | | | | |
M-521.dd: (links/rechts)
downstream rechts => pos. limit [0]
downstream links => neg. limit [207]
----------------------
M-511.dd: (oben/unten)
oben [102]
unten [0]
----------------------
M-060.dg: (rotation)
parallel beam [7,5]
-------------------------------------------
Remote Control:
lokal ssh tunnel starten:
ssh -g -C -L5959:zenpixelw:5900 eudet@eudetmac001.cern.ch
danach mit "Chicken of the VNC" (von lokalem Mac) VNC auf localhost (port 5959).
|
12
|
Fri Sep 24 12:09:32 2010 |
Manfred Valentan | BELLE | gds-file of Micron wafer | | | | |
|
11
|
Thu Sep 23 16:23:07 2010 |
Manfred Valentan | Micron Baby | Modul B2Micron_2826_1_Baby2, DUT (eingeschraenkt) & TRACKING | | | | |
p-side: Hybrid 43
APV3: Streifen 9 schlecht
n-side: Hybrid 33
APV4: 0-10 andere Bond-Parameter?, 28&29 erhoehtes noise
APV5: 111-128: 4 schlechte Streifen + common mode correction?
ALS DUT GEEIGNET!
FUER TRACKING GEEIGNET!
|
10
|
Thu Sep 23 16:22:33 2010 |
Manfred Valentan | Micron Baby | Modul B2Micron_2826_1_Baby3, TRACKING | | | | |
p-side: Hybrid 44
alles ok
n-side: Hybrid 34
APV4: zur Gaenze schlecht, APV-Fehler?
APV5: beide Randstreifen schlecht
FUER TRACKING GEEIGNET!
|
9
|
Thu Sep 23 16:19:54 2010 |
Manfred Valentan | Micron Baby | Modul B2Micron_2826_1_Baby4, DUT EINGESCHRAENKT, TRACKING | | | | |
p-side: Hybrid 45
APV0: beide Randstreifen schlecht
APV1: Streifen 0 schlecht
APV3: beide Randstreifen etwas schlechter, nicht schlimm
n-side: Hybrid 36
APV4: 0, 9, 40, 79-85, 97, 102, 124 Sensorfehler?
APV5: 0-30 vereinzelt schlechte Streifen, 115-128 Bonds beruehren
ALS DUT AUGENZUDRUECKEN GEEIGNET!
FUER TRACKING GEEIGNET!
|