Entry  Mon Sep 27 15:39:33 2010, Thomas Bergauer, common, PI-XY Table, , , ,  

M-521.dd: (links/rechts)

downstream rechts => pos. limit [0]

downstream links => neg. limit [207]

----------------------

M-511.dd: (oben/unten)

oben [102]

unten [0]

----------------------

M-060.dg: (rotation)

parallel beam [7,5]

 

-------------------------------------------

Remote Control:

lokal ssh tunnel starten:
ssh -g -C -L5959:zenpixelw:5900 eudet@eudetmac001.cern.ch

danach mit "Chicken of the VNC" (von lokalem Mac) VNC auf localhost (port 5959). 

Entry  Mon Sep 27 15:41:28 2010, Thomas Bergauer, common, Szintillatoren, , , ,  

1: breit  (1)

2: schmal (2)

3: breit  (4)

4: schmal  (8)

 

AND aus beiden breiten: 5

AND aus beiden schmalen: 10

 

Entry  Mon Sep 27 15:42:09 2010, Thomas Bergauer, common, TLU: wir stecken auf DUT 3, , , ,  

TLUControl.exe ist auf zenpixell3 unter /home/eudet/eudaq/bin

TLUControl.exe -d 14 -a 12
-dXX Mask for DUT:
1   abgeschnittenes Dummykabel
2   EUDRB hinten
4   EUDRB vorne
8   DUT3(wir)
-t100 internal trigger
-aXX Mask for Szintis:
1   4 mm (front)
2   4 mm (front)
4   9 mm (back)
8   9 mm (back)
Entry  Mon Sep 27 18:54:47 2010, Thomas Bergauer, BELLE, XY Position, , , ,  

center position: 55/40/277,5

Entry  Tue Sep 28 16:50:31 2010, Thomas Bergauer, common, serial connection chiller, , , ,  

 Der chiller braucht ein gekreuztes serielles Kabel!!

Entry  Wed Sep 29 16:28:17 2010, Thomas Bergauer, common, EUDET telescope pedestal run, 20089, 122850 (events built), 29.09.2010 16:27, 29.09.2010 17:33 

 EUDAQ config eudet2-test-6boards-in-2crates

rate: ~745.5 Hz

 

Entry  Wed Sep 29 17:38:15 2010, Thomas Bergauer, common, EUDET telescope pedestal run, 20097, 118125, 29.09.2010 17:37, 29.09.2010 18:09 

new pedestal run of EUDET telescope with new JTAG settings of plane 3 (sensor 11) [according to Ingrid]:


VRef2    98
VRef1A  87
VRef1B  143
VRef1C  108
VRef1D   153

 

Entry  Wed Sep 29 18:53:05 2010, Thomas Bergauer, Micron Baby, XY Positions, , , ,  

center position: 113/47/7.5

top left (direction jura): 121/53.5/7.5

bottom right (direction saleve): 105/40.5/7.5

Entry  Thu Sep 30 14:45:16 2010, Thomas Bergauer, Micron Baby, run, run015, 22.5k, 30.09.2010 13:18:11, 30.09.2010 14:38:01 

micron baby stack angle scan 15 grad mit 3 modulen

HV=100V, I=3.8uA

Beam on

 

telescope position: 129/76/187.5

telescope run 20109

Entry  Thu Sep 30 23:09:29 2010, Thomas Bergauer, Micron Baby, run, run019, 162000, 30.09.2010 23:08:03, 01.10.2010 9:33:38 run019_sensordisplay1.jpgEudet-X_Cluster_Corr-run020115.jpgEudet-Num_Clusters-run020115.jpg

micron baby: 4 DUTs
HV=100V I=12uA I_stop=15.4uA
Beam on
45° rotation

 

telescope run  20115 (after JTAG programming of telescope chips due to config errors)


Entry  Fri Oct 1 23:50:28 2010, Thomas Bergauer, CNM, run, run023, 2275, 01.10.2010 23:49:29, 01.10.2010 23:53:12 

CMN-SILC stack
HV_SILC (Katze)=75V, 0.25uA (6 SiLC)
HV_CNM (Keithley)=30V, CNM_1=84nA, CNM_2=4.3uA

too much optimized alignment (XY-table to 133/74/187.5)   ;-)

telescope run 20119

Entry  Fri Oct 1 23:55:12 2010, Thomas Bergauer, CNM, run, run024, 320k, 01.10.2010 23:53:12, 02.10.2010 09:47:37 cnm_2d_2.pngcnm_2d_1.png

CMN-SILC stack
HV_SILC (Katze)=75V, 0.25uA (6 SiLC)
HV_CNM (Keithley)=30V, CNM_1=84nA, CNM_2=4.3uA

optimized alignment (XY-table to 133/76/187.5)

telescope run 20120

Entry  Sat Oct 2 09:57:20 2010, Thomas Bergauer, CNM, run, run025, 150k, 02.10.2010 09:54:28, 02.10.2010 14:21:20 run20122.pngslowcontrol_cnm.pngcnm_2d_1.pngcnm_2d_2.png

 CMN-SILC stack

HV_SILC (Katze)=75V, 0.25uA (6 SiLC)
HV_CNM (Keithley)=30V, CNM_1=84nA, CNM_2=4.3uA

new position along the CNM strips (XY-table to 143/76/187.5)

telescope run 20122

 

Entry  Sat Oct 2 17:27:14 2010, Thomas Bergauer, ITE, cal, cal007, 83800, 02.10.2010 17:11:18 ,  intcal_ITE10PAD_W7.pngintcal_ITE10PAD_W9.png

ITE: 2 DUTs
2 x SILC + 2 x ITE
HV=100V

Beam on 

one short only per sensor

Entry  Wed Oct 6 11:51:25 2010, Thomas Bergauer, Micron Baby, run, run049, 4489, 06.10.2010 11:38:21, 06.10.2010 11:49:23 
8 x Micron Baby (3 irradiated), no cooling
HV=100V I=141uA (start of run)

beam on 

not yet optimized XY-table positions (111/43.5/7.5)

Entry  Sat Oct 9 12:25:36 2010, Thomas Bergauer, BELLE, run, run057, 276475, 09.10.2010 12:22:21, 09.10.2010 15:08:46 

 

Origami1 (n-side only) + Micron1 (non-irrad)

angle scan: 73 degree

HV=100V / 36 uA  (beginning), 35uA (end of run)

cooling: chiller set to 15C

 PROBLEM: wrong config file (single hit, not multi6)

telescope run 20200

Origami p-side does not receive triggers -> disabled

origami flex straight part needed to be folded due to severe rotation angle -> nanonics connector problem...

 

Entry  Sat Oct 9 18:56:35 2010, Thomas Bergauer, common, path to telescope data, , , ,  

eudetmac001.cern.ch

path is /Volumes/Data/eudet/Data/cern_2010_sept_silc_apvdaq (Vienna system)

and /Volumes/Data/eudet/Data/cern_2010_sept_silc_alibava (Santander system)

    Reply  Sun Oct 10 15:45:04 2010, Thomas Bergauer, common, end of success, , , 09.10.2010 19:44:23, NaN 

 ;-) 

genau das gleiche wollte ich auch gerade schreiben!

Entry  Wed May 4 15:14:31 2011, Thomas Bergauer, Pt, info about module, , , ,  

pt-Modul hat jeweils zwei OB1 Sensoren von STM (mit 122um Pitch)

im Abstand von 2mm uebereinander

Entry  Tue Sep 28 01:13:57 2010, Markus Friedl, BELLE, cal, cal001, 83800, 28.09.2010 01:11:20,  

first intcal with complete Belle stack: 2 x Origami, 2 x Micron

HV=100V  I=49uA

beam is on

chiller is set to 15°C (module temperature is ~17°C)

both Origamis and both Microns are serially cooled