Home CMS Production Clean room MedAustron HEPHY testbeams old
testbeam-MA2018 testbeam-MA2019 testbeam-MA2020 testbeam-MA2021 testbeam-MA2023 testbeam-MA2024
  Testbeams at MedAustron in 2024, Page 11 of 11  Not logged in ELOG logo
Entry  Sun Jun 9 00:10:38 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser, 20, TOF-Cal GSI, 145.4 MeV, Low Flux High, Proton, Good, 23:31, 00:00, , , ,  
5 cm plastic water (in isocentre)
2 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 330 kHz on LGAD

14 min break with beam off (DAQ failed, had to power cycle TRB3s and all DIRICH boards-->out of synch DIRICH)
Entry  Sun Jun 9 01:22:48 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Knochen, 22, TOF-Cal GSI, 145.4 MeV, Low Flux High, Proton, Good, 00:14, 00:23, , , ,  
1 cm dense bone (in isocentre)
2.8 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 600 kHz on LGAD

WET measurement
Entry  Sun Jun 9 01:26:56 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Knochen, 24, TOF-Cal GSI, 145.4 MeV, Low Flux High, Proton, Good, 01:02, 01:14, , , ,  
2 cm dense bone (in isocentre)
2.1 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 315 kHz on LGAD

WET measurement
Entry  Sun Jun 9 01:32:51 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Knochen, 26, TOF-Cal GSI, 145.4 MeV, Low Flux High, Proton, Good, 01:30, 01:45, , , ,  
3 cm dense bone (in isocentre)
1.7 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 230 kHz on LGAD

WET measurement
Entry  Sun Jun 9 02:02:16 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Knochen, 28, TOF-Cal GSI, 145.4 MeV, Low Flux High, Proton, Good, 02:00, 02:14, , , ,  
4 cm dense bone (in isocentre)
1.4 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 175 kHz on LGAD

WET measurement
Entry  Sat Dec 14 12:53:28 2024, Felix Ulrich-Pur, Shift 23.11.24 low flux He test with RaMon, 4, RaMon, 124.4 MeV/u, Low Flux Medium, Helium, Good, 06:33, 06:33, 1, , ,  
Mean rate per spill~ 2-8E5 p/s (online estimation without proper analysis)
Entry  Sat Dec 14 12:58:35 2024, Felix Ulrich-Pur, Shift 23.11.24 low flux He test with RaMon, 11, RaMon, 124.4 MeV/u, Low Flux High, Helium, Good, 06:43, 06:43, 1, , ,  
Mean rate per spill~ 4Ee6 p/s (online estimation without proper analysis)
Entry  Sat Dec 14 13:35:59 2024, Felix Ulrich-Pur, Energy vs TOF calibration, 23, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 124.4 MeV/u, Low Flux High, Helium, Good, 11:24, 11:37, 13, , ,  
TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

5MHz with Ramon and only 500kHz TRB
Entry  Sat Jun 8 20:08:55 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs Calibrationcurve, 6, TOF-Cal GSI, 115.7MeV, Low Flux Medium, Proton, Good, 18:24, 19:00, , , ,  
400 khz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 65kHz on LGAD

TOF Calibration Curve
Entry  Sat Jun 8 20:09:25 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs Calibrationcurve, 7, TOF-Cal GSI, 100.4 MeV, Low Flux Medium, Proton, Good, 19:00, 19:42, , , ,  
300 khz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 65kHz on LGAD

TOF Calibration Curve
Entry  Sun Aug 25 12:39:45 2024, tb-crew, only base, 70, RD50-MPW4, 100.4 MeV, Low Flux Very Low, Proton, Unknown, 25.08.2024 12:24:36, 25.08.2024 12:40:12, 15, Normal, ,  peary_config.cfgmatrix_mpw4_base.txtmatrix_mpw4_piggy.txtmpw4.conf
automatic log for run 70
Comment:
E-scan
Entry  Sat Mar 30 06:34:39 2024, Bernhard Pilsl, And a new testbeam starts :D, , , , , , , , , , , ,  
 
Entry  Sat Mar 30 08:05:59 2024, Christian Irmler, RD50-MPW4, 999, , , , , , , , , , ,  
Tracker calibration run: mpw4-tj-monopix2_00-000999.xml
Generated global xml file for MPW4 shift: mpw4-tj-monopix2_TLU_00-000999.xml 

Previous local runs and xml files (mpw4-tj-monopix2_00-000212.xml) are unusable, because sensor L5.1.1.0 had a loose cotact and thus low gain.
Entry is currently edited by RD50 Generic User on 193.170.243.131  Entry  Sun Aug 25 10:02:35 2024, RD50 Generic User, init_geo_file, , , , , , , , , , , ,  detectors.geo
position of duts needs to be switched
Entry  Mon Aug 26 09:18:25 2024, RD50 Generic User, geo, , , , , , , , , , , ,  detectors_new.geo
new geo file with adjusted dut position
Entry  Sat Dec 14 13:02:29 2024, Felix Ulrich-Pur, Shift 23.11.24 low flux Carbon test with RaMon (did not work), 15, RaMon, , Low Flux High, Carbon, Crap, 06:55, 07:01, 6, , ,  
Low flux carbon did not work

1µs EFE, 30s Extraction, Deg100
we tried different energies (402.MeV/u)-> 120 MeV/u, but unfortunately we did not get any beam
Entry  Sat Dec 14 13:04:57 2024, Felix Ulrich-Pur, Shift 23.11.24 low flux Carbon test with RaMon (did not work), 16, RaMon, , Low Flux High, Carbon, Crap, 07:02, 07:20, 18, , ,  
Low flux carbon did not work

5µs EFE, 10s Extraction, Deg100
we tried different energies (402.MeV/u)-> 238 MeV/u, but unfortunately we did not get any beam
ELOG V3.1.5-fc6679b