Home CMS Production Clean room MedAustron HEPHY testbeams old
testbeam-MA2018 testbeam-MA2019 testbeam-MA2020 testbeam-MA2021 testbeam-MA2023 testbeam-MA2024
  MedAustron 2021, Page 5 of 18  Not logged in ELOG logo
ID Date Author Subject Run Number DUT Beam Energy Rate Setting Particle Type Events Data Beam Betatron Beam Chopper Beam Degrader Beam EFE Beam Orbdown Beam QuadScan PhantomType PhantomAngle
  81   Sat Feb 27 09:40:15 2021 Felix Ulrich-PurRate verification7SiC+Si+RaMon62.4I(EFE+SID0)  Good        

Start 07:40
rate looks fine, however operators mentioned, that CPD data might have been loaded incorrectly-->pause during run and reloaded (07:48), rate seems unchanged

 

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.475V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x: 0
y: 0 
 

  82   Sat Feb 27 09:41:06 2021 Felix Ulrich-PurWaveforms SiC+Si8SiC+Si+RaMon252.7I(EFE+SID0)  Good        

08:34 room request, PC for SiC had to be restarted, trigger were connected as prior to shift

Start 08:53
rate looks fine

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.475V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

verändere Position, ab 7:49 bei Aufnahme von Waveforms Position:

x = 0
y = 0

7:51 verändere Position

7:58 neue Position:
x = 8
y = 68

 

FPGA shutdown --> Datenaufzeichnung bis etwa 8:27, ab dann Suchen des Fehlers

 

  83   Sat Feb 27 09:42:44 2021 Felix Ulrich-PurWaveforms SiC+Si9SiC+Si+RaMon145.4I(EFE+SID0)  Good        

Start 09:37

SiC setrup had to be rewired (wr0ng polarity somewhere (insert here))

 

Umstellen:
channel 2: LVDS-Signal vertauschen
channel 2 Referenz: 2.100 V 

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.475V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x = 8
y = 68

  84   Sat Feb 27 18:36:15 2021 Felix Ulrich-PurWaveforms SiC+Si10SiC+Si+RaMon145.4I(EFE+SID0)  Good        

Start 10:10
polarity now correct

Wiederholung von vorher, aber keine Aufnahme von neuen Waveforms

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.100 V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x = 8
y = 68

  85   Sat Feb 27 18:38:09 2021 Felix Ulrich-PurWaveforms SiC+Si11SiC+Si+RaMon62.4I(EFE+SID0)  Good        

Start 10:24
polarity now correct
no signal in sic --> energy too low due to scintis

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.100 V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x = 8
y = 68

  86   Sat Feb 27 18:38:54 2021 Felix Ulrich-PurWaveforms SiC+Si12SiC+Si+RaMon252.7I(EFE+SID0)  Good        

Start 10:28
polarity now correct

now signal in SiC

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.100 V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x = 8
y = 68

  87   Sat Feb 27 18:42:13 2021 Felix Ulrich-PurWaveforms SiC+Si13SiC+Si+RaMon252.7III(EFE+SID2)  Good        

Start 10:38
no signal in SiC

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.100 V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x = 8
y = 68

  88   Sat Feb 27 18:43:12 2021 Felix Ulrich-PurWaveforms SiC+Si14SiC+Si+RaMon145.4III(EFE+SID2)  Good        

Start 10:47

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.100 V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x = 8
y = 68

  89   Sat Feb 27 18:43:35 2021 Felix Ulrich-PurWaveforms SiC+Si15SiC+Si+RaMon100.4III(EFE+SID2)  Good        

Start 11:09

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.100 V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x = 8
y = 68

  90   Sat Feb 27 18:44:06 2021 Felix Ulrich-PurWaveforms SiC+Si16SiC+Si+RaMon100.4I(EFE+SID0)  Good        

Start 11:36

schöne Signale

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.100 V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x = 8
y = 68

  91   Sat Feb 27 18:44:23 2021 Felix Ulrich-PurWaveforms SiC+Si17SiC+Si+RaMon800VII  Good        

Start 11:55
Stopped at 12:05-->acc failure

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.100 V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x = 8
y = 68

  92   Sat Feb 27 18:49:06 2021 Felix Ulrich-PurWaveforms SiC+Si18SiC+Si+RaMon800VII  Good        

Start 12:39

Neustart der vorherigen Messung

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.100 V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x = 8
y = 68

  93   Sat Feb 27 18:50:16 2021 Felix Ulrich-PurWaveforms SiC+Si19SiC+Si+RaMon252.7nominal  Good  100     

Start 13:40

Degrader 100 
Messung aufgenommen, keine Screenshots

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.100 V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x = 8
y = 68

  94   Sat Feb 27 18:52:12 2021 Felix Ulrich-PurWaveforms SiC+Si20SiC+Si+RaMon252.7nominalProton Good  10     

Start 13:42

Degrader 10


channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.100 V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x = 8
y = 68

  95   Sat Feb 27 18:56:11 2021 Felix Ulrich-PurWaveforms SiC+Si21SiC+Si+RaMon252.7nominalProton Good  50     

Start 13:56

Degrader 50


channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.100 V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x = 8
y = 68

 

  96   Sat Feb 27 18:56:58 2021 Felix Ulrich-PurWaveforms SiC+Si22SiC+Si+RaMon252.7nominalProton Good  100     

Start 14:08

Degrader 100


channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.100 V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x = 8
y = 68

  97   Sat Feb 27 19:01:26 2021 Felix Ulrich-PurWaveforms SiC+Si23SiC+Si+RaMon400.8nominalCarbon Good  10     

Start 14:18 

14:30 enter room request: abbau and shielding was activated, called RP, removed SiC+Si Setup+Al shielding of DSSDs

Degrader 10, Carbon
Messung bis 14:24

danach neue Aufzeichnung (Hintergrundstrahlung) Start ca 14:25


channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.100 V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x = 8
y = 68

  98   Sat Feb 27 19:06:07 2021 Felix Ulrich-PurLocal Runs DSSDs+SiC50-56SiC+DSSD trackerNoisenominal  Good        

Start 16:15:
Final MT=1, final BSY length =500 (2kHz)

  99   Sat Feb 27 19:07:42 2021 Felix Ulrich-PurHV Scan DSSD+SiC57SiC+DSSD tracker100.4II(BETA+EFE)Proton Good        

Start 16:46:
Final MT=1, final BSY length =500 (2kHz), Trigger latency 40 (not optimal)

  100   Sat Feb 27 19:08:51 2021 Felix Ulrich-PurHV Scan DSSD+SiC58SiC+DSSD tracker100.4II(BETA+EFE)Proton Good        

Start 16:50:
Final MT=1, final BSY length =500 (2kHz), Trigger latency 41 optimal

HV 100V

ELOG V3.1.5-fc6679b