ID |
Date |
Author |
Project |
Subject |
Run Number |
Events |
XYPosition |
Cooling |
Data |
72
|
Fri Apr 14 21:11:18 2017 |
Axel Koenig | ALiBaVa (IFX) | run_028_IFX_CB_100V.dat | 028 | 3000000 | 109, 4.5, 240 | No | Crap |
Remarks: Telescope run 551. Longrun over night. Run crashed. Events only up to 2288000 valid.
Parameter |
Value |
Alibava run |
run_028_IFX_CB_100V.dat |
Telescope run |
551 |
Triggers |
3000000 |
Alibava trigger eff. |
~60 % |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-100 |
Latency |
132 |
Table x |
109 mm |
Table z |
4.5 mm |
Tabe rot |
240 deg
|
|
73
|
Sat Apr 15 09:57:51 2017 |
Axel Koenig | ALiBaVa (IFX) | run_029_IFX_CB_130V.dat | 029 | 175000 | 109, 4.5, 240 | No | Unknown |
3Remarks: Telescope run 557. Bias scan 130 V.
Parameter |
Value |
Alibava run |
run_029_IFX_CB_130V.dat |
Telescope run |
557 |
Triggers |
175000 |
Alibava trigger eff. |
~60% |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-130 |
Latency |
132 |
Table x |
109 mm |
Table z |
4.5 mm |
Tabe rot |
240 deg (parallel to tele. planes)
|
|
74
|
Sat Apr 15 12:12:00 2017 |
Axel Koenig | ALiBaVa (IFX) | run_030_IFX_CB_100V.dat | 030 | 1000000 | 109, 4.5, 240 | No | Unknown |
3Remarks: Telescope run 558. Bias scan 100 V. Telescope chiller turned off. Could be the root cause of crashing runs.
Parameter |
Value |
Alibava run |
run_030_IFX_CB_100V.dat |
Telescope run |
558 |
Triggers |
1000000 |
Alibava trigger eff. |
~60% |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-100 |
Latency |
132 |
Table x |
109 mm |
Table z |
4.5 mm |
Tabe rot |
240 deg (parallel to tele. planes)
|
|
75
|
Sat Apr 15 16:52:06 2017 |
Axel Koenig | ALiBaVa (IFX) | run_001_Baby_100V.dat | 001 | 250000 | 102.5, 22.5, 240 | No | Unknown |
Remarks: Telescope run 562. Telescope chiller turned off (could be the root cause of runs crashing).
Position: 102.5, 22.5, 240 (should be the center of the sensor).
Parameter |
Value |
Alibava run |
run_001_Baby_100V.dat |
Telescope run |
562 |
Triggers |
250000 |
Alibava trigger eff. |
~93% |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-100 |
Latency |
132 |
Table x |
102.5 mm |
Table z |
22.5 mm |
Tabe rot |
240 deg (parallel to tele. planes)
|
|
77
|
Sat Apr 15 17:38:58 2017 |
Axel Koenig | ALiBaVa (IFX) | run_002_Baby_100V.dat | 002 | 250000 | 102.5, 12.5, 240 | No | Unknown |
Remarks: Telescope run 563. Telescope chiller turned off (could be the root cause of runs crashing).
Parameter |
Value |
Alibava run |
run_002_Baby_100V.dat |
Telescope run |
563 |
Triggers |
250000 |
Alibava trigger eff. |
~93% |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-100 |
Latency |
132 |
Table x |
102.5 mm |
Table z |
12.5 mm |
Tabe rot |
240 deg (parallel to tele. planes)
|
|
78
|
Sat Apr 15 18:31:14 2017 |
Axel Koenig | ALiBaVa (IFX) | run_003_Baby_100V.dat | 003 | 250000 | 102.5, 32.5, 240 | No | Unknown |
Remarks: Telescope run 564. Telescope chiller turned off (could be the root cause of runs crashing).
Parameter |
Value |
Alibava run |
run_003_Baby_100V.dat |
Telescope run |
564 |
Triggers |
250000 |
Alibava trigger eff. |
~93% |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-100 |
Latency |
132 |
Table x |
102.5 mm |
Table z |
32.5 mm |
Tabe rot |
240 deg (parallel to tele. planes)
|
|
79
|
Sat Apr 15 19:20:34 2017 |
Axel Koenig | ALiBaVa (IFX) | run_004_Baby_100V.dat | 004 | 250000 | 102.5, 22.5, 250 | No | Unknown |
Remarks: Telescope run 565. Telescope chiller turned off (could be the root cause of runs crashing).
Parameter |
Value |
Alibava run |
run_004_Baby_100V.dat |
Telescope run |
565 |
Triggers |
250000 |
Alibava trigger eff. |
~93% |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-100 |
Latency |
132 |
Table x |
102.5 mm |
Table z |
22.5 mm |
Tabe rot |
250 deg
|
|
80
|
Sat Apr 15 20:12:41 2017 |
Axel Koenig | ALiBaVa (IFX) | run_005_Baby_100V.dat | 005 | 250000 | 102.5, 22.5, 230 | No | Unknown |
Remarks: Telescope run 568. Telescope chiller turned off (could be the root cause of runs crashing).
Parameter |
Value |
Alibava run |
run_005_Baby_100V.dat |
Telescope run |
568 |
Triggers |
250000 |
Alibava trigger eff. |
~93% |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-100 |
Latency |
132 |
Table x |
102.5 mm |
Table z |
22.5 mm |
Tabe rot |
230 deg
|
|
81
|
Sat Apr 15 21:18:26 2017 |
Axel Koenig | ALiBaVa (IFX) | run_007_Baby_10V.dat | 007 | 450000 | 102.5, 22.5, 240 | No | Crap |
Remarks: Telescope run 579. Bias scan 10 V. run crashed. Data vaild up to 274915.
Telescope chiller turned off (could be the root cause of runs crashing).
Parameter |
Value |
Alibava run |
run_007_Baby_10V.dat |
Telescope run |
579 |
Triggers |
450000 |
Alibava trigger eff. |
~24% |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-10 |
Latency |
132 |
Table x |
102.5 mm |
Table z |
22.5 mm |
Tabe rot |
240 deg
|
|
82
|
Sun Apr 16 13:15:55 2017 |
Axel Koenig | ALiBaVa (IFX) | run_008_Baby_20V.dat | 008 | 175000 | 102.5, 22.5, 240 | No | Unknown |
Remarks: Telescope run 583. Bias scan 20 V. run crashed.
Telescope chiller turned off (could be the root cause of runs crashing).
Parameter |
Value |
Alibava run |
run_008_Baby_20V.dat |
Telescope run |
583 |
Triggers |
175000 |
Alibava trigger eff. |
~60% |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-20 |
Latency |
132 |
Table x |
102.5 mm |
Table z |
22.5 mm |
Tabe rot |
240 deg
|
|
83
|
Sun Apr 16 15:44:35 2017 |
Axel Koenig | ALiBaVa (IFX) | run_009_Baby_30V.dat | 009 | 135000 | 102.5, 22.5, 240 | No | Unknown |
Remarks: Telescope run 586. Bias scan 30 V.
Telescope chiller turned off (could be the root cause of runs crashing).
Parameter |
Value |
Alibava run |
run_009_Baby_30V.dat |
Telescope run |
586 |
Triggers |
135000 |
Alibava trigger eff. |
~60% |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-30 |
Latency |
132 |
Table x |
102.5 mm |
Table z |
22.5 mm |
Tabe rot |
240 deg
|
|
84
|
Sun Apr 16 17:42:06 2017 |
Axel Koenig | ALiBaVa (IFX) | run_010_Baby_40V.dat | 010 | 135000 | 102.5, 22.5, 240 | No | Unknown |
Remarks: Telescope run 587. Bias scan 40 V.
Telescope chiller turned off (could be the root cause of runs crashing).
Parameter |
Value |
Alibava run |
run_010_Baby_40V.dat |
Telescope run |
587 |
Triggers |
135000 |
Alibava trigger eff. |
~60% |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-40 |
Latency |
132 |
Table x |
102.5 mm |
Table z |
22.5 mm |
Tabe rot |
240 deg
|
|
85
|
Sun Apr 16 18:25:59 2017 |
Axel Koenig | ALiBaVa (IFX) | run_011_Baby_50V.dat | 011 | 135000 | 102.5, 22.5, 240 | No | Unknown |
Remarks: Telescope run 588. Bias scan 50 V.
Telescope chiller turned off (could be the root cause of runs crashing).
Parameter |
Value |
Alibava run |
run_011_Baby_50V.dat |
Telescope run |
588 |
Triggers |
135000 |
Alibava trigger eff. |
~60% |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-50 |
Latency |
132 |
Table x |
102.5 mm |
Table z |
22.5 mm |
Tabe rot |
240 deg
|
|
86
|
Sun Apr 16 19:04:41 2017 |
Axel Koenig | ALiBaVa (IFX) | run_012_Baby_60V.dat | 012 | 135000 | 102.5, 22.5, 240 | No | Unknown |
Remarks: Telescope run 589. Bias scan 60 V.
Telescope chiller turned off (could be the root cause of runs crashing).
Parameter |
Value |
Alibava run |
run_012_Baby_60V.dat |
Telescope run |
589 |
Triggers |
135000 |
Alibava trigger eff. |
~60% |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-60 |
Latency |
132 |
Table x |
102.5 mm |
Table z |
22.5 mm |
Tabe rot |
240 deg
|
|
87
|
Sun Apr 16 19:46:01 2017 |
Axel Koenig | ALiBaVa (IFX) | run_013_Baby_65V.dat | 013 | 112000 | 102.5, 22.5, 240 | No | Unknown |
Remarks: Telescope run 590. Bias scan 65 V.
Telescope chiller turned off (could be the root cause of runs crashing).
Parameter |
Value |
Alibava run |
run_013_Baby_65V.dat |
Telescope run |
590 |
Triggers |
112000 |
Alibava trigger eff. |
~60% |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-65 |
Latency |
132 |
Table x |
102.5 mm |
Table z |
22.5 mm |
Tabe rot |
240 deg
|
|
88
|
Sun Apr 16 20:22:27 2017 |
Axel Koenig | ALiBaVa (IFX) | run_014_Baby_70V.dat | 014 | 250000 | 102.5, 22.5, 240 | No | Unknown |
Remarks: Telescope run 591. Bias scan 70 V.
Telescope chiller turned off (could be the root cause of runs crashing).
Parameter |
Value |
Alibava run |
run_014_Baby_70V.dat |
Telescope run |
591 |
Triggers |
250000 |
Alibava trigger eff. |
~60% |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-70 |
Latency |
132 |
Table x |
102.5 mm |
Table z |
22.5 mm |
Tabe rot |
240 deg
|
|
90
|
Sun Apr 16 21:43:15 2017 |
Axel Koenig | ALiBaVa (IFX) | run_015_Baby_100V.dat | 015 | 2500000 | 102.5, 22.5, 240 | No | Unknown |
Remarks: Telescope run 592. Long run, bias voltage 100 V. Long run finished without any problems.
Telescope chiller turned off (could be the root cause of runs crashing).
Parameter |
Value |
Alibava run |
run_015_Baby_100V.dat |
Telescope run |
592 |
Triggers |
2500000 |
Alibava trigger eff. |
~60% |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-100 |
Latency |
132 |
Table x |
102.5 mm |
Table z |
22.5 mm |
Tabe rot |
240 deg
|
|
92
|
Mon Apr 17 08:14:24 2017 |
Axel Koenig | ALiBaVa (IFX) | run_016_Baby_75V.dat | 016 | 112000 | 102.5, 22.5, 240 | No | Unknown |
Remarks: Telescope run 593. Bias scan 75 V.
Telescope chiller turned off (could be the root cause of runs crashing).
Parameter |
Value |
Alibava run |
run_016_Baby_75V.dat |
Telescope run |
593 |
Triggers |
112000 |
Alibava trigger eff. |
~60% |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-75 |
Latency |
132 |
Table x |
102.5 mm |
Table z |
22.5 mm |
Tabe rot |
240 deg
|
|
93
|
Mon Apr 17 09:01:09 2017 |
Axel Koenig | ALiBaVa (IFX) | run_017_Baby_80V.dat | 017 | 112000 | 102.5, 22.5, 240 | No | Unknown |
Remarks: Telescope run 594. Bias scan 80 V.
Telescope chiller turned off (could be the root cause of runs crashing).
Parameter |
Value |
Alibava run |
run_017_Baby_80V.dat |
Telescope run |
594 |
Triggers |
112000 |
Alibava trigger eff. |
~90% |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-80 |
Latency |
132 |
Table x |
102.5 mm |
Table z |
22.5 mm |
Tabe rot |
240 deg
|
|
94
|
Mon Apr 17 09:04:02 2017 |
Axel Koenig | ALiBaVa (IFX) | run_006_Baby_100V.dat | 006 | k.a. | 102.5, 22.5, 230 | No | Unknown |
Remarks: Telescope run 570. Telescope chiller turned off (could be the root cause of runs crashing).
Forgot ro record data in elog.
Parameter |
Value |
Alibava run |
run_006_Baby_100V.dat |
Telescope run |
570 |
Triggers |
k.a. |
Alibava trigger eff. |
~93% |
Beam energy |
5.6 GeV |
Telescope threshold |
6 |
Bias voltage |
-100 |
Latency |
132 |
Table x |
102.5 mm |
Table z |
22.5 mm |
Tabe rot |
230 deg
|
|