  Alibava/Mapsa testbeam at DESY in April 2017 (Axel, Grossi), Page 3 of 7  Not logged in ELOG logo
New entries since:Thu Jan 1 01:00:00 1970
ID Date Authorup Project Subject Run Number Events XYPosition Cooling Data
  93   Mon Apr 17 09:01:09 2017 Axel KoenigALiBaVa (IFX)run_017_Baby_80V.dat017112000102.5, 22.5, 240NoUnknown

Remarks: Telescope run 594. Bias scan 80 V.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_017_Baby_80V.dat
Telescope run 594
Triggers 112000
Alibava trigger eff. ~90%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -80
Latency 132
Table x 102.5 mm
Table z 22.5 mm
Tabe rot

240 deg

  94   Mon Apr 17 09:04:02 2017 Axel KoenigALiBaVa (IFX)run_006_Baby_100V.dat006k.a.102.5, 22.5, 230NoUnknown

Remarks: Telescope run 570. Telescope chiller turned off (could be the root cause of runs crashing).

Forgot ro record data in elog.

Parameter Value
Alibava run run_006_Baby_100V.dat
Telescope run 570
Triggers k.a.
Alibava trigger eff. ~93%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -100
Latency 132
Table x 102.5 mm
Table z 22.5 mm
Tabe rot

230 deg

  95   Mon Apr 17 09:29:40 2017 Axel KoenigALiBaVa (IFX)run_018_Baby_90V.dat018112000102.5, 22.5, 240NoUnknown

Remarks: Telescope run 595. Bias scan 90 V.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_018_Baby_90V.dat
Telescope run 595
Triggers 112000
Alibava trigger eff. ~90%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -90
Latency 132
Table x 102.5 mm
Table z 22.5 mm
Tabe rot

240 deg

  96   Mon Apr 17 10:05:57 2017 Axel KoenigALiBaVa (IFX)run_019_Baby_110V.dat019112000102.5, 22.5, 240NoUnknown

Remarks: Telescope run 596. Bias scan 110 V.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_019_Baby_110V.dat
Telescope run 596
Triggers 112000
Alibava trigger eff. ~90%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -110
Latency 132
Table x 102.5 mm
Table z 22.5 mm
Tabe rot

240 deg

  98   Mon Apr 17 10:14:15 2017 Axel KoenigcommonRun summary Baby   NoUnknown
Alibava run Telescope run File existence check Run type Triggers Remarks
001 562 yes first run 250000 none
002 563 yes central position shifted by -1 cm vertically 250000 none
003 564 yes central position shifted by + 1cm 250000 none
004 565 yes central position angle +10 deg 250000 none
005 568 yes central position angle -10 deg 250000 none
006 570 yes unknown (maybe long run over night that crashed) ? data may be corrupted
007 579 yes bias scan -10V 450000 run crashed data only valid up to 274915
008 583 yes bias scan -20V 175000 none
009 586 yes bias scan -30V 135000 none
010 587 yes bias scan -40V 135000 none
011 588 yes bias scan -50V 135000 none
012 589 yes bias scan -60V 135000 none
013 590 yes bias scan -65V 112000 none
014 591 yes bias scan -70V 112000 none
015 592 yes long run over night 2500000 none
016 593 yes bias scan -75V 112000 none
017 594 yes bias scan -80V 112000 none
018 595 yes bias scan -90V 112000 none
019 596 yes bias scan -110V 112000 none
020 597 yes bias scan -120V 112000 none
021 598 yes bias scan -130V 112000 none
022 599 yes bias scan -140V 112000 none
023 600 yes bias scan -150V 112000 none

Central beam position: x = 102 mm, y = 22.5 mm, rot = 240 deg

Current was stable for all runs i.e. 500 +- 10 nA at 100 V

Trigger efficiency ~ 90% for all runs (except bias scan).

Temperature in area 20+-1 deg.

Telescope chiller off for all runs.

  99   Mon Apr 17 10:36:22 2017 Axel KoenigALiBaVa (IFX)run_020_Baby_120V.dat020112000102.5, 22.5, 240NoUnknown

Remarks: Telescope run 597. Bias scan 120 V.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_020_Baby_120V.dat
Telescope run 597
Triggers 112000
Alibava trigger eff. ~90%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -120
Latency 132
Table x 102.5 mm
Table z 22.5 mm
Tabe rot

240 deg

  100   Mon Apr 17 11:04:40 2017 Axel KoenigALiBaVa (IFX)run_021_Baby_130V.dat021112000102.5, 22.5, 240NoUnknown

Remarks: Telescope run 598. Bias scan 130 V.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_021_Baby_130V.dat
Telescope run 598
Triggers 112000
Alibava trigger eff. ~90%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -130
Latency 132
Table x 102.5 mm
Table z 22.5 mm
Tabe rot

240 deg

  101   Mon Apr 17 11:34:55 2017 Axel KoenigALiBaVa (IFX)run_022_Baby_140V.dat022112000102.5, 22.5, 240NoUnknown

Remarks: Telescope run 599. Bias scan 140 V.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_022_Baby_140V.dat
Telescope run 599
Triggers 112000
Alibava trigger eff. ~90%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -140
Latency 132
Table x 102.5 mm
Table z 22.5 mm
Tabe rot

240 deg

  104   Mon Apr 17 12:06:11 2017 Axel KoenigALiBaVa (IFX)run_023_Baby_150V.dat023112000102.5, 22.5, 240NoUnknown

Remarks: Telescope run 600. Bias scan 150 V.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_023_Baby_150V.dat
Telescope run 600
Triggers 112000
Alibava trigger eff. ~90%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -150
Latency 132
Table x 102.5 mm
Table z 22.5 mm
Tabe rot

240 deg

  105   Mon Apr 17 13:30:01 2017 Axel KoenigALiBaVa (IFX)run_001_HPK_CB_200V.dat001450000107, 0, 240NoUnknown

Remarks: Telescope run 604. Bias scan 200 V. Only one chip.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_001_HPK_CB_200V.dat
Telescope run 604
Triggers 450000
Alibava trigger eff. ~64%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -200
Latency 132
Table x 107 mm
Table z 0 mm
Tabe rot

240 deg

  106   Mon Apr 17 16:36:56 2017 Axel KoenigALiBaVa (IFX)run_002_HPK_CB_200V.dat002450000107, 5.8, 240NoUnknown

Remarks: Telescope run 605. Bias scan 200 V. Only one chip.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_002_HPK_CB_200V.dat
Telescope run 605
Triggers 450000
Alibava trigger eff. ~64%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -200
Latency 132
Table x 107 mm
Table z 5.8 mm
Tabe rot

240 deg

  107   Mon Apr 17 17:02:34 2017 Axel KoenigALiBaVa (IFX)run_003_HPK_CB_200V.dat003250000107, 2.6, 230NoUnknown

Remarks: Telescope run 606. Angle scan.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_003_HPK_CB_200V.dat
Telescope run 606
Triggers 250000
Alibava trigger eff. ~62%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -200
Latency 132
Table x 107 mm
Table z 2.6 mm
Tabe rot

230 deg

  108   Mon Apr 17 17:25:24 2017 Axel KoenigcommonTelescope geometry ALiBaVa   NoUnknown

The plane distances are measured at the center of the planes.

The distance of the DUT to the neighbouring planes is measured from the position of the sensor to the center of the planes.

   
Attachment 1: IMG_1522.JPG
IMG_1522.JPG
Attachment 2: beam.pdf
Attachment 3: beam1.pdf
Attachment 4: beam2.pdf
  109   Mon Apr 17 17:46:38 2017 Axel KoenigALiBaVa (IFX)run_005_HPK_CB_50V.dat005150000107, 2.6, 240NoUnknown

Remarks: Telescope run 609. Bias scan 50 V. Bias scan.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_005_HPK_CB_50V.dat
Telescope run 609
Triggers 150000
Alibava trigger eff. ~62%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -50
Latency 132
Table x 107 mm
Table z 2.6 mm
Tabe rot

240 deg

  111   Mon Apr 17 19:19:14 2017 Axel KoenigALiBaVa (IFX)run_006_HPK_CB_100V.dat006150000107, 2.6, 240NoUnknown

Remarks: Telescope run 610. Bias scan 100 V. Bias scan.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_006_HPK_CB_100V.dat
Telescope run 610
Triggers 150000
Alibava trigger eff. ~62%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -100
Latency 132
Table x 107 mm
Table z 2.6 mm
Tabe rot

240 deg

  113   Mon Apr 17 19:54:08 2017 Axel KoenigALiBaVa (IFX)run_007_HPK_CB_150V.dat007300000107, 2.6, 240NoUnknown

Remarks: Telescope run 611. Bias scan 150 V. Bias scan. Run crashed at 294000 triggers!

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_007_HPK_CB_150V.dat
Telescope run 611
Triggers 300000
Alibava trigger eff. ~62%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -150
Latency 132
Table x 107 mm
Table z 2.6 mm
Tabe rot

240 deg

  114   Mon Apr 17 21:52:45 2017 Axel KoenigALiBaVa (IFX)run_008_HPK_CB_250V.dat008150000107, 2.6, 240NoUnknown

Remarks: Telescope run 613. Bias scan 250 V.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_008_HPK_CB_250V.dat
Telescope run 613
Triggers 150000
Alibava trigger eff. ~62%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -250
Latency 132
Table x 107 mm
Table z 2.6 mm
Tabe rot

240 deg

  115   Mon Apr 17 22:27:44 2017 Axel KoenigALiBaVa (IFX)run_009_HPK_CB_200V.dat0092400000107, 2.6, 240NoUnknown

Remarks: Telescope run 614. Long run over night. run crashed over night!

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_009_HPK_CB_200V.dat
Telescope run 614
Triggers 2400000
Alibava trigger eff. ~62%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -200
Latency 132
Table x 107 mm
Table z 2.6 mm
Tabe rot

240 deg

  118   Tue Apr 18 08:04:39 2017 Axel KoenigALiBaVa (IFX)run_004_HPK_CB_200V.dat004250000107, 2.6, 250NoUnknown

Remarks: Telescope run 607. Angle scan.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_004_HPK_CB_200V.dat
Telescope run 607
Triggers 250000
Alibava trigger eff. ~62%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -200
Latency 132
Table x 107 mm
Table z 2.6 mm
Tabe rot

250 deg

  119   Tue Apr 18 08:07:15 2017 Axel KoenigALiBaVa (IFX)run_010_HPK_CB_200V.dat010350000107, 2.6, 240NoUnknown

Remarks: Telescope run 616. Short run central position. Repeated due to crash over night.

No beam during run.Only ~10000 events.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_010_HPK_CB_200V.dat
Telescope run 616
Triggers 350000
Alibava trigger eff. ~62%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -200
Latency 132
Table x 107 mm
Table z 2.6 mm
Tabe rot

240 deg