ID Date Author Project Subjectup Run Number Events XYPosition Cooling Data Text Attachments
  36   Thu Apr 13 21:23:11 2017 Axel KoenigALiBaVa (IFX)run_018_IFX_CB_120V.dat018175000109, 4.5, 240NoUnknownRemarks: Telescope run 533. Bias scan 120
V.

  
  145   Fri Apr 21 08:15:03 2017 Axel KoenigALiBaVa (IFX)run_018_Irrad_500V.dat01820000096, 64, 360NoUnknownRemarks: Telescope run 650. Bias voltage
-500V, central position of sensor (DUT parallel
to telescope planes at 360 deg due to different
  
  96   Mon Apr 17 10:05:57 2017 Axel KoenigALiBaVa (IFX)run_019_Baby_110V.dat019112000102.5, 22.5, 240NoUnknownRemarks: Telescope run 596. Bias scan 110
V.

Telescope chiller turned off (could
  
  146   Fri Apr 21 08:38:17 2017 Axel KoenigALiBaVa (IFX)run_019_Irrad_500V.dat01920000096, 64, 360NoUnknownRemarks: Telescope run 651. Bias voltage
-500V, central position of sensor (DUT parallel
to telescope planes at 360 deg due to different
  
  99   Mon Apr 17 10:36:22 2017 Axel KoenigALiBaVa (IFX)run_020_Baby_120V.dat020112000102.5, 22.5, 240NoUnknownRemarks: Telescope run 597. Bias scan 120
V.

Telescope chiller turned off (could
  
  59   Fri Apr 14 08:59:26 2017 Axel KoenigALiBaVa (IFX)run_020_IFX_CB_100V.dat020190000109, 4.5, 240NoCrapRemarks: Telescope run 535. Bias scan 100
V. Run crashed over night. Telescope sync
only up to event 190000 (planned 2500000
  
  147   Fri Apr 21 08:57:30 2017 Axel KoenigALiBaVa (IFX)run_020_Irrad_500V.dat02020000096, 64, 360NoUnknownRemarks: Telescope run 652. Bias voltage
-500V, central position of sensor (DUT parallel
to telescope planes at 360 deg due to different
  
  100   Mon Apr 17 11:04:40 2017 Axel KoenigALiBaVa (IFX)run_021_Baby_130V.dat021112000102.5, 22.5, 240NoUnknownRemarks: Telescope run 598. Bias scan 130
V.

Telescope chiller turned off (could
  
  148   Fri Apr 21 09:14:45 2017 Axel KoenigALiBaVa (IFX)run_021_Irrad_500V.dat02120000096, 64, 360NoUnknownRemarks: Telescope run 653. Bias voltage
-500V, central position of sensor (DUT parallel
to telescope planes at 360 deg due to different
  
  101   Mon Apr 17 11:34:55 2017 Axel KoenigALiBaVa (IFX)run_022_Baby_140V.dat022112000102.5, 22.5, 240NoUnknownRemarks: Telescope run 599. Bias scan 140
V.

Telescope chiller turned off (could
  
  65   Fri Apr 14 13:06:30 2017 Axel KoenigALiBaVa (IFX)run_022_IFX_CB_100V.dat022500000109, 10.5, 240NoUnknownRemarks: Telescope run 545. Bias voltage
100 V. Beam position on only one chip.

  
  104   Mon Apr 17 12:06:11 2017 Axel KoenigALiBaVa (IFX)run_023_Baby_150V.dat023112000102.5, 22.5, 240NoUnknownRemarks: Telescope run 600. Bias scan 150
V.

Telescope chiller turned off (could
  
  68   Fri Apr 14 14:49:09 2017 Axel KoenigALiBaVa (IFX)run_023_IFX_CB_100V.dat023385000109, 0, 240NoCrapRemarks: Telescope run 546. Bias voltage
100 V. Beam position on only one chip. Run
destroyed. Intended trigger count 500k
  
  60   Fri Apr 14 09:25:09 2017 Axel KoenigALiBaVa (IFX)run_024_IFX_CB_140V.dat024175000109, 4.5, 240NoUnknown3Remarks: Telescope run 547. Bias scan
140 V.

  
  69   Fri Apr 14 17:22:52 2017 Axel KoenigALiBaVa (IFX)run_025_IFX_CB_150V.dat025175000109, 4.5, 240NoUnknown3Remarks: Telescope run 548. Bias scan
150 V.

  
  70   Fri Apr 14 18:22:58 2017 Axel KoenigALiBaVa (IFX)run_026_IFX_CB_100V.dat026350000109, 4.5, 230NoUnknownRemarks: Telescope run 549. Angle shift
230 deg.

  
  71   Fri Apr 14 19:43:13 2017 Axel KoenigALiBaVa (IFX)run_027_IFX_CB_100V.dat027350000109, 4.5, 250NoUnknownRemarks: Telescope run 550. Angle shift
250 deg.

  
  72   Fri Apr 14 21:11:18 2017 Axel KoenigALiBaVa (IFX)run_028_IFX_CB_100V.dat0283000000109, 4.5, 240NoCrapRemarks: Telescope run 551. Longrun over
night. Run crashed. Events only up to 2288000
valid.
  
  73   Sat Apr 15 09:57:51 2017 Axel KoenigALiBaVa (IFX)run_029_IFX_CB_130V.dat029175000109, 4.5, 240NoUnknown3Remarks: Telescope run 557. Bias scan
130 V.

  
  74   Sat Apr 15 12:12:00 2017 Axel KoenigALiBaVa (IFX)run_030_IFX_CB_100V.dat0301000000109, 4.5, 240NoUnknown3Remarks: Telescope run 558. Bias scan
100 V. Telescope chiller turned off. Could
be the root cause of crashing runs.
  
ELOG V3.1.5-fc6679b