ID Date Author Project Subjectdown Run Number Events XYPosition Cooling Data
  67   Tue Jun 9 11:20:25 2015 Benedikt WürknerSVD3 (H6B)Vertical Beam Profile Run65500002NoUnknown

Trying to get a nice plot of the beam profile by only using the vertical PMTs (number 0 and 2) see http://elog.hephy.at/testbeam-SPS2015/66 for details. 

Pos 2: x=69.3 mm, y=79 mm (Hits in APV 3-4 und 8-9 in p-Seite und 13 und 18 in n-Seite) 

Producer.TLU Config Change:

  • AndMask=0x5
  69   Tue Jun 9 14:33:50 2015 Benedikt WürknerSVD3 (H6B)Vertical Beam Profile Run67500002NoUnknown

Trying to get a nice plot of the beam profile by only using the vertical PMT 0 (upstream vertical, the long one)  see http://elog.hephy.at/testbeam-SPS2015/66 for details. 

Pos 2: x=69.3 mm, y=79 mm (Hits in APV 3-4 und 8-9 in p-Seite und 13 und 18 in n-Seite) 

Producer.TLU Config Change:

  • AndMask=0x1 
  95   Fri Jun 12 16:52:52 2015 Benedikt WürknerSVD3 (H6B)Telescope without DUT76~30k NoGood

Threshold 6

Testrun to see how the telescope behaves without a DUT to test if there is an increased amount of multiple scattering after the aluminum planes of the FW and BW module. 

 

  98   Fri Jun 12 17:08:23 2015 Benedikt WürknerSVD3 (H6B)Telescope with FEI480~30k NoGood

Threshold 6

Testrun to see how the telescope behaves without a DUT to test if there is an increased amount of multiple scattering after the aluminum planes of the FW and BW module. 

This time with the upstream FEI4 plane

  58   Mon Jun 8 18:08:49 2015 Thomas BergauerSVD3 (H6B)Telescope online monitor    Unknown

a working version of the EUDET online monitor (which works also with the FE-I4) is at the following location

~/BELLETEST/eudaq-neu/build/monitors/onlinemon$ ./OnlineMon.exe -sc 0 -f /home/telescope/eudaq/data/run000060.raw

  30   Sun Jun 7 12:10:24 2015 Benedikt WürknerSVD3 (H6B)Telescope Threshold8b test run 004614399  Unknown

Telescope in combination with SVD3

  31   Sun Jun 7 12:11:42 2015 Benedikt WürknerSVD3 (H6B)Telescope Threshold7 test run 004738175  Unknown

Telescope alone

  32   Sun Jun 7 12:12:00 2015 Benedikt WürknerSVD3 (H6B)Telescope Threshold7 test run 004853356  Unknown

Telescope alone

  66   Tue Jun 9 10:01:03 2015 Lukas BullaSVD3 (H6B)TLU setup info    Unknown

Channels:

  • Upstream vertical -> channel 0; (bitmask 1)
  • Upstream horizontal -> channel 1; (bitmask 2)
  • Downstream vertical -> channel 2; (bitmask 4)
  • Downstream horizontal ->channel 3; (bitmask 8)

 

TLU DuT interface:

DUT Interface Bitmask DAQ
1 1 Mimosa (EUDET Telescope)
2 2 leer
3 4 SVD3
4 8 FE-I4

 

  12   Fri Jun 5 12:02:29 2015 Hao YinBelle II FADC (H6A)Sixlet Calibration001   Good

room temp max subevent at 2. with lat = 96

  13   Fri Jun 5 12:03:34 2015 Hao YinBelle II FADC (H6A)Sixlet Calibration002   Good

lat = 97 to a greater range of lat scan

  14   Fri Jun 5 19:55:00 2015 Hao YinBelle II FADC (H6A)Sixlet Calibration003   Unknown

lat = 98

room temp

pedestalrun 003

  75   Wed Jun 10 10:40:24 2015 Hao YinBelle II FADC (H6A)Sixlet Calibration004  NoGood

room temp
pedestalrun 008
FIRRun004

max frame pos = 0. frame

[six]
lvl = 20, 30
lat = 95
sam = 20
grp = 8

  76   Wed Jun 10 11:19:33 2015 Hao YinBelle II FADC (H6A)Sixlet Calibration005  NoGood

room temp
pedestalrun 008
FIRRun004

max frame pos = 0. frame

[six]
lvl = 20, 30
lat = 96
sam = 20
grp = 8

  77   Wed Jun 10 12:03:53 2015 Hao YinBelle II FADC (H6A)Sixlet Calibration006  NoGood

room temp
pedestalrun 008
FIRRun004

max frame pos = 1-2-3. frame n side still shifted

[six]
lvl = 20, 30
lat = 97
sam = 20
grp = 8

  78   Wed Jun 10 12:04:27 2015 Hao YinBelle II FADC (H6A)Sixlet Calibration007  NoGood

room temp
pedestalrun 008
FIRRun004

max frame pos = 1-2-3. frame n side still shifted

[six]
lvl = 20, 30
lat = 98
sam = 20
grp = 8

  79   Wed Jun 10 13:12:33 2015 Hao YinBelle II FADC (H6A)Sixlet Calibration010  YesGood

cooled
pedestalrun 008
FIRRun004

max frame pos = 1-2-3. frame n side still shifted

[six]
lvl = 15, 35
lat = 96
sam = 50
grp = 8

  86   Thu Jun 11 11:41:50 2015 Hao YinBelle II FADC (H6A)Sixlet Calibration011  YesGood

cooled
pedestalrun 008
FIRRun004

max frame pos = 1-2-3. frame n side still shifted

[six]
lvl = 15, 35
lat = 96
sam = 50
grp = 8

  87   Thu Jun 11 11:42:39 2015 Hao YinBelle II FADC (H6A)Sixlet Calibration012  YesGood

cooled until about 400k after that room temp
pedestalrun 009
FIRRun005

sixlet calibration run with only ladders

[six]
lvl = 0, 255
lat = 97
sam = 50
grp = 8

  1   Tue May 12 10:27:39 2015 Benedikt WürknerSVD3 (H6B)Setup information    Unknown

Information concerning the SVD3 setup during a beam test

In the beam area setup the whole crate and the power supply with the usual spider cable and the two Keithly 2410 high voltage supplies. Monitoring of the 2410 happens via a NI MAX interface that connects to the GPIB-Ethernet interface and needs to be configured there. The IP and password are noted on the device. The software can be found on heros in medialib/LONGTERM/longterm R3.1_doubleSided as a normal *.vi file. 

Default adresses 16 and 17 as defined in labview, need to figure out on site which is which, if the voltage selected (e.g. 100V) is applied as +50/-50 or if it is applied as +100/-100 before connecting the cables. 

Take a run without dut as well to check the telescope integrity. Without dut in beamline!!!!!

ELOG V3.1.5-fc6679b