Entry  Tue Nov 18 12:35:32 2014, Axel König, common, Infineon (4 modules) TLU Settings, , , , , Unknown 
  1.  VNC to 192.168.0.18
  2. In command line: cd C:\Users\telescope\eudaq\bin
  3. TLUControl.exe.exe -d 2 -a 11 - i RJ45

Operating with AIDA telescope: TLUControl.exe.exe -d 3 -a 11 - i RJ45

Dut Position (left to right):  11, 06, 05, 07

Entry  Tue Nov 18 18:29:55 2014, Axel König, Infineon, calibration run, Infineon_run002, , , , Unknown 

Infineon_run002 is just a calibration run for the vertical positioned sensors. 

data file in L:\cern14\infineon\Infineon Run vert
config file l:\cern14\infineon\Infineon Run horiz\cern14_infineon_single.cfg
Entry  Tue Nov 18 18:33:31 2014, Axel König, Infineon, hardware run, Infineon_run003, 300000, , , Unknown 

Infineon new stack - batch 4

single hit

HV=300V, total I = 4.0044 µA

strips vertical

Table Position: H: 96mm V: 101mm

data file in L:\cern14\infineon\Infineon Run vert

config file l:\cern14\infineon\Infineon Run horiz\cern14_infineon_single.cfg

 

--Stopped 18.11 / 20:20---

Reason: Enough events, adjusting new DUT position for next run

Events: 300000

Entry  Tue Nov 18 23:04:34 2014, Axel König, Infineon, hardware run, Infineon_run004, 400000, , , Unknown 

Infineon new stack - batch 4 

single hit
 
HV=300V, total I = 4.0044 µA
 
strips vertical
 
Table Position: H: 96mm V: 81mm
 
data file in L:\cern14\infineon\Infineon Run vert
 
config file l:\cern14\infineon\Infineon Run horiz\cern14_infineon_single.cfg
 
--stopped 18.11 // 23:05---
Reason: Enough events, next run with the same position but now multi six
Events: 400000
 
Entry  Wed Nov 19 00:58:27 2014, Axel König, Infineon, hardware run, Infineon_run005, 87000, , , Unknown 

Infineon new stack - batch 4 

multi hit
 
HV=300V, total I = 4.0200 µA
 
strips vertical
 
Table Position: H: 96mm V: 81mm
 
data file in L:\cern14\infineon\Infineon Run vert
 
config file l:\cern14\infineon\Infineon Run horiz\cern14_infineon_multi6.cfg
 
--stopped 19.11 // 01:00 --
Reason: Enough events, new dut postion for next multi run
Events: 87000
Entry  Wed Nov 19 09:26:39 2014, Axel König, Infineon, hardware run, Infineon_run006, 1500, , , Unknown 

Infineon new stack - batch 4

multi hit
 
HV=300V, total I = 4.0200 µA
 
strips vertical
 
Table Position: H: 96mm V: 91mm
 
data file in L:\cern14\infineon\Infineon Run vert
 
config file l:\cern14\infineon\Infineon Run horiz\cern14_infineon_multi6.cfg
 
-- stopped --
1500 Events only --> no trigger signals from TU (01:30 Uhr)
Entry  Wed Nov 19 16:24:25 2014, Axel König, common, Running EUDAQ, , , , , Unknown 

Windows PC at Telescope:

0.) JTAG Programming: Chose a threshold file (recommended threshold 5) File: M26TB_Cern_2011.thresh5.mcf--> press start

1.) C:\Users\telescope\Desktop\users\BelleII --> start Mimosa 26 Telescope --> select all six Mimosa's

2.) C:\Users\telescope\Desktop\users\BelleII --> start startEudaq_fullWindows.bat

3.) Use configuration file hephy1 (location: C:\eudaq\conf). Change it according to your TLU settings (DUT mask, AND mask, etc.)

4.) configure

5.) start

x) Online monitor in preparation

Files: Are saved in C:\eudaq\data with internal numbering "run000000xxx.raw" (look at the date and time). [You can connect locally to this folder at HEPHY-SVD PC]

Entry  Wed Nov 19 18:20:29 2014, Axel König, Infineon, calibration run, Infineon_run007_cal, , , , Unknown 
 
Entry  Wed Nov 19 21:56:44 2014, Axel König, Infineon, hardware run, Infineon_run008, 244167, , , Unknown 

Infineon new stack - batch 4 

single hit
 
HV=300V, total I = 4.0758 µA
 
strips vertical
 
Table Position: H: 92.5mm V: 81mm
 
data file in L:\cern14\infineon\Infineon Run vert
 
config file l:\cern14\infineon\Infineon Run horiz\cern14_infineon_single.cfg
 
--stopped 19.11 // 21:53 --
Reason: Want to run the eudaq for the next run
Events: 244167
Entry  Wed Nov 19 22:15:41 2014, Axel König, Infineon, hardware run, Infineon_run010, 1498709, , , Unknown 

Infineon new stack - batch 4 

 
single hit
 
HV=300V, total I = 4.0200 µA
 
strips vertical
 
Table Position: H: 92.5mm V: 81mm
 
EUDAQ data: Run951 and Run952
 
data file in L:\cern14\infineon\Infineon Run vert
 
config file l:\cern14\infineon\Infineon Run horiz\cern14_infineon_multi6.cfg
 
-- stopped 20.11 / 09:35-- 
 
SVD3DAQ Events: 1498709
EUDAQ Events: 2777
 
EUDAQ stopped counting events --> log entry: "01:38 6-ERROR unexpected trigger number 277714 (expecting 277713)" (TLU Producer)
Entry  Thu Nov 20 15:46:21 2014, Axel König, Infineon, hardware run, Infineon_run012, 263470, , , Unknown 

Infineon new stack - batch 4 

single hit
 
HV=300V, total I = ? µA
 
strips vertical
 
Table Position: H: 89mm V: 81mm
 
data file in L:\cern14\infineon\Infineon Run vert
 
config file l:\cern14\infineon\Infineon Run horiz\cern14_infineon_single.cfg
 
--stopped 20.11 // 15:30 --
Eudaq: No Events build --> searching for failure
Events: 263470
Entry  Thu Nov 20 15:47:56 2014, Axel König, Infineon, hardware run, Infineon_run011, , , , Unknown 

 Infineon new stack - batch 4 

 faulty!!!
Entry  Thu Nov 20 17:16:56 2014, Axel König, common, how to run the EUDAQ offline monitor, , , , , Unknown 

YOU need

1.) Folder EUDAQ OFFLINE Monitor

2.) ROOT

 

Starting: 

1.) open command line and chose the path where OfflineMonitor is

2.)Type -i "patch of raw file" -o test.root 100

3.) helpfile typing -h

Entry  Thu Nov 20 18:15:48 2014, Axel König, Infineon, hardware run, Infineon_run013_voltage, 51273, , , Unknown 

Infineon new stack - batch 4 

single hit + eudaq
 
HV=100V, total I = 3.0326 µA
 
strips vertical
 
Table Position: H: 89mm V: 81mm
 
data file in L:\cern14\infineon\Infineon Run vert
 
config file l:\cern14\infineon\Infineon Run horiz\cern14_infineon_single.cfg
 
eudaq file: run000959.raw
 
--stopped 20.11 // 18:10 --
Eudaq: 54928
Events: 51273
Entry  Thu Nov 20 22:12:21 2014, Axel König, common, SPS Logbook / SPS Online Monitor, , , , , Unknown 

Logbook: https://ab-dep-op-elogbook.web.cern.ch/ab-dep-op-elogbook/elogbook/secure/eLogbook.php?shiftId=1059325

 

Monitor:  http://op-webtools.web.cern.ch/op-webtools/vistar/vistars.php?usr=SPS1

Entry  Fri Nov 21 01:43:54 2014, Axel König, AIDA, calibration run, AIDA_run001_calibration, , , , Unknown 

Calibration run with AIDA DUT.

Single hit

Entry  Fri Nov 21 11:10:08 2014, Axel König, AIDA, hardware run, AIDA_run002, 211362, , , Unknown 

AIDA

multi hit
Voltage=100V, I=4.1688uA
(new table calibration): V:56mm, H:96mm
 
EUDAQ File: Run000962, Run000963, Run000964
 
config: L:\cern14\AIDA
data: L:\cern14\AIDA
 
--Stopped 21.11 / 10:30--
DaqEvents: 211362
EUDAQ EVENTS: ca. 41000
Entry  Fri Nov 21 12:14:46 2014, Axel König, AIDA, hardware run, AIDA_run003, 20617, , , Unknown 

AIDA

 

single hit

Voltage=100V, I=5.4240uA

(new table calibration): V:66mm, H:74mm

EUDAQ file:  Run000965

config: L:\cern14\AIDA

data: L:\cern14\AIDA

 

--Stopped 21.11 / 12:12--

DaqEvents: 20617 (-600 ped)

EUDAQ EVENTS: 20019 (2 more than DAQ [+2 offset]) 

Entry  Fri Nov 21 15:37:30 2014, Axel König, AIDA, hardware run, AIDA_run004, 7057, , , Unknown 

AIDA

 

single hit

Voltage=100V, I=5.4357uA

(new table calibration): V:71mm, H:80mm

 

config: L:\cern14\AIDA

data: L:\cern14\AIDA

EUDAQ File: Run000977

--Stopped 21.11 / 12:12--

DaqEvents: 7057 (-600 ped)

EUDAQ EVENTS: 4658 (less than Daq) 

Entry  Fri Nov 21 18:20:47 2014, Axel König, AIDA, hardware run, AIDA_run005, 20045, , , Unknown 

AIDA

single hit

Voltage=100V, I=5.4357uA

(new table calibration): V:71mm, H:80mm

config: L:\cern14\AIDA

data: L:\cern14\AIDA

EUDAQ File: Run000979

--Stopped 21.11 / 12:12--

DaqEvents: 20045 (-600 ped)

EUDAQ EVENTS: ~18k

ELOG V3.1.5-fc6679b