Tue Nov 18 13:13:53 2014, Thomas Bergauer, Infineon, hardware run, infineon_run001, 225508, , , Unknown
|
Infineon new stack - batch 4
single hit
HV=300V, total I = 4.2475 µA
strips horizontal
data file in L:\cern14\infineon\Infineon Run horiz
config file l:\cern14\infineon\Infineon Run horiz\cern14_infineon_single.cfg
STOPPED by main User in H6B (18:11, 15:00)
Events: 225508
Next step: rearange modules vertically
|
Sat Nov 22 17:00:43 2014, Axel König, Infineon, Infineon old, damaged strixel sensors, strange IV behaviour, , , , Unknown
|
1.) No events recorded for strixel sensors (run001). They seem to be damaged like they were at DESY.
2.) Very strange IV behaviour for every old Infineon module (especially TB12_baby1). |
Thu Nov 20 18:15:48 2014, Axel König, Infineon, hardware run, Infineon_run013_voltage, 51273, , , Unknown
|
Infineon new stack - batch 4
single hit + eudaq
HV=100V, total I = 3.0326 µA
strips vertical
Table Position: H: 89mm V: 81mm
data file in L:\cern14\infineon\Infineon Run vert
config file l:\cern14\infineon\Infineon Run horiz\cern14_infineon_single.cfg
eudaq file: run000959.raw
--stopped 20.11 // 18:10 --
Eudaq: 54928
Events: 51273 |
Thu Nov 20 15:46:21 2014, Axel König, Infineon, hardware run, Infineon_run012, 263470, , , Unknown
|
Infineon new stack - batch 4
single hit
HV=300V, total I = ? µA
strips vertical
Table Position: H: 89mm V: 81mm
data file in L:\cern14\infineon\Infineon Run vert
config file l:\cern14\infineon\Infineon Run horiz\cern14_infineon_single.cfg
--stopped 20.11 // 15:30 --
Eudaq: No Events build --> searching for failure
Events: 263470 |
Thu Nov 20 15:47:56 2014, Axel König, Infineon, hardware run, Infineon_run011, , , , Unknown
|
Infineon new stack - batch 4
faulty!!! |
Wed Nov 19 22:15:41 2014, Axel König, Infineon, hardware run, Infineon_run010, 1498709, , , Unknown
|
Infineon new stack - batch 4
single hit
HV=300V, total I = 4.0200 µA
strips vertical
Table Position: H: 92.5mm V: 81mm
EUDAQ data: Run951 and Run952
data file in L:\cern14\infineon\Infineon Run vert
config file l:\cern14\infineon\Infineon Run horiz\cern14_infineon_multi6.cfg
-- stopped 20.11 / 09:35--
SVD3DAQ Events: 1498709
EUDAQ Events: 2777
EUDAQ stopped counting events --> log entry: "01:38 6-ERROR unexpected trigger number 277714 (expecting 277713)" (TLU Producer) |
Wed Nov 19 21:56:44 2014, Axel König, Infineon, hardware run, Infineon_run008, 244167, , , Unknown
|
Infineon new stack - batch 4
single hit
HV=300V, total I = 4.0758 µA
strips vertical
Table Position: H: 92.5mm V: 81mm
data file in L:\cern14\infineon\Infineon Run vert
config file l:\cern14\infineon\Infineon Run horiz\cern14_infineon_single.cfg
--stopped 19.11 // 21:53 --
Reason: Want to run the eudaq for the next run
Events: 244167 |
Wed Nov 19 18:20:29 2014, Axel König, Infineon, calibration run, Infineon_run007_cal, , , , Unknown
|
|
Wed Nov 19 09:26:39 2014, Axel König, Infineon, hardware run, Infineon_run006, 1500, , , Unknown
|
Infineon new stack - batch 4
multi hit
HV=300V, total I = 4.0200 µA
strips vertical
Table Position: H: 96mm V: 91mm
data file in L:\cern14\infineon\Infineon Run vert
config file l:\cern14\infineon\Infineon Run horiz\cern14_infineon_multi6.cfg
-- stopped --
1500 Events only --> no trigger signals from TU (01:30 Uhr) |
Wed Nov 19 00:58:27 2014, Axel König, Infineon, hardware run, Infineon_run005, 87000, , , Unknown
|
Infineon new stack - batch 4
multi hit
HV=300V, total I = 4.0200 µA
strips vertical
Table Position: H: 96mm V: 81mm
data file in L:\cern14\infineon\Infineon Run vert
config file l:\cern14\infineon\Infineon Run horiz\cern14_infineon_multi6.cfg
--stopped 19.11 // 01:00 --
Reason: Enough events, new dut postion for next multi run
Events: 87000 |
Tue Nov 18 23:04:34 2014, Axel König, Infineon, hardware run, Infineon_run004, 400000, , , Unknown
|
Infineon new stack - batch 4
single hit
HV=300V, total I = 4.0044 µA
strips vertical
Table Position: H: 96mm V: 81mm
data file in L:\cern14\infineon\Infineon Run vert
config file l:\cern14\infineon\Infineon Run horiz\cern14_infineon_single.cfg
--stopped 18.11 // 23:05---
Reason: Enough events, next run with the same position but now multi six
Events: 400000
|
Tue Nov 18 18:33:31 2014, Axel König, Infineon, hardware run, Infineon_run003, 300000, , , Unknown
|
Infineon new stack - batch 4
single hit
HV=300V, total I = 4.0044 µA
strips vertical
Table Position: H: 96mm V: 101mm
data file in L:\cern14\infineon\Infineon Run vert
config file l:\cern14\infineon\Infineon Run horiz\cern14_infineon_single.cfg
--Stopped 18.11 / 20:20---
Reason: Enough events, adjusting new DUT position for next run
Events: 300000 |
Tue Nov 18 18:29:55 2014, Axel König, Infineon, calibration run, Infineon_run002, , , , Unknown
|
Infineon_run002 is just a calibration run for the vertical positioned sensors.
data file in L:\cern14\infineon\Infineon Run vert
config file l:\cern14\infineon\Infineon Run horiz\cern14_infineon_single.cfg |
Sun Nov 23 09:32:48 2014, Axel König, Infineon, hardware run, Infineon_old_run007, 63913, , , Unknown
|
Infineon old
multi hit
Voltage=300V, I=10.4028uA
(new table calibration): V:75, H:96.5
config: L:\cern14\AIDA
data: L:\cern14\AIDA
EUDAQ File: run000999
DaqEvents: 63913 (-600 ped)
EUDAQ EVENTS: daq events - 600 (+2 offset) |
Sun Nov 23 09:36:27 2014, Axel König, Infineon, hardware run, Infineon_old_run006, 526562, , , Unknown
|
Infineon old
multi hit
Voltage=300V, I=10.4028uA
(new table calibration): V:75, H:95
config: L:\cern14\AIDA
data: L:\cern14\AIDA
EUDAQ File: run000998
DaqEvents: 526562 (-600 ped)
EUDAQ EVENTS: daq events - 600 (+2 offset) |
Sun Nov 23 00:51:07 2014, Axel König, Infineon, hardware run, Infineon_old_run005, 26569, , , Unknown
|
Infineon old
multi hit
Voltage=300V, I=10.4609uA
(new table calibration): V:75, H:94.5
config: L:\cern14\AIDA
data: L:\cern14\AIDA
EUDAQ File: run000997
DaqEvents: 26569 (-600 ped)
EUDAQ EVENTS: daq events - 600 (+2 offset)
|
|
Sun Nov 23 00:24:15 2014, Axel König, Infineon, hardware run, Infineon_old_run004, 168725, , , Unknown
|
Infineon old
multi hit
Voltage=300V, I=10.6529uA
(new table calibration): V:81, H:88
config: L:\cern14\AIDA
data: L:\cern14\AIDA
EUDAQ File: run000996
DaqEvents: 168725 (-600 ped)
EUDAQ EVENTS: daq events - 600 (+2 offset) |
Sun Nov 23 00:20:41 2014, Axel König, Infineon, hardware run, Infineon_old_run003, 30000, , , Unknown
|
Infineon old
multi hit
Voltage=300V, I=10.9797uA
(new table calibration): V:?, H:?
config: L:\cern14\AIDA
data: L:\cern14\AIDA
EUDAQ File: run000994
DaqEvents: 30000 (-600 ped)
EUDAQ EVENTS: daq events - 600 (+2 offset) |
Sat Nov 22 18:08:03 2014, Axel König, Infineon, hardware run, Infineon_old_run002, 55000, , , Unknown
|
Infineon old
multi hit
Voltage=300V, I=10.9797uA
(new table calibration): V:81mm, H:102mm
config: L:\cern14\AIDA
data: L:\cern14\AIDA
EUDAQ File: run000993
DaqEvents: 55000 (-600 ped)
EUDAQ EVENTS: daq events - 600 (+2 offset) |
Sat Nov 22 16:53:39 2014, Axel König, Infineon, Infineon old, Infineon_old_run001, 55000 , , , Unknown
|
Infineon old
multi hit
Voltage=300V, I=10.9797uA
(new table calibration): V:81mm, H:97mm
config: L:\cern14\AIDA
data: L:\cern14\AIDA
EUDAQ File: run000992
DaqEvents: 55000 (-600 ped)
EUDAQ EVENTS: daq events - 600 (+2 offset) |
|