Sat Oct 27 00:43:55 2012, Christian Irmler, MicronBaby, cal, cal029, 83800, 27.10.2012 00:10:15, 27.10.2012 00:36:55, Good
|
After irradiation
Micron baby stack after gamma irrdiation
HV = -100 V
I = -58 uA |
Sat Oct 27 19:36:24 2012, Markus Friedl, MicronOrigami_gross, cal, cal031, 83800, 27.10.2012 19:59:50, 27.10.2012 20:44:38, Good
|
Belle stack 2 after irradiation
HV = +50V -50V
I = 112 µA
CO2 cooling |
Sun Oct 28 15:46:07 2012, Markus Friedl, MicronBaby, cal, cal034, 83800, 28.10.2012 15:19:10, 28.10.2012 15:40:47, Good
|
Micron Baby stack 2 after irradiation (intermediate strips)
HV = -100 V
I = 33 uA |
Mon Oct 29 11:36:40 2012, Markus Friedl, Infineon, cal, cal037, 83800, 29.10.2012 11:35:12, 29.10.2012 11:58:02 , Good
|
Infineon stack Version 2 after irradiation
HV = +300 V
I = 137 A
|
Mon Oct 29 22:36:52 2012, Markus Friedl, HPK_baby, cal, cal038, 83800, 29.10.2012 22:17:34, 29.10.2012 22:47:47, Good
|
HPK baby stack after irradiation with partial cooling (HPK neutron & HPK neutron/gamma)
HV = +75 Vm -75V
I = 226A
Temp sensors configuration:
11 HPK Neutron n in
12 HPK Neutron n -> p
13 ----
14 HPK Neutron p -> Gamma p
15 Gamma p -> n
16 Neutron Gamma n out |
Tue Oct 30 10:37:19 2012, Markus Friedl, HPK_baby, cal, cal039, 83800, 30.10.2012 10:35:35 , 30.10.2012 11:02:06 , Good
|
Turned off cooling, now taking warm data
HPK baby stack after irradiation - no cooling
I = 200A and growing (warming up)
positioned centered on p and n
|
Tue Oct 30 17:24:01 2012, Markus Friedl, MicronOrigami_gross, cal, cal040, 83800, 30.10.2012 17:20:28 , 30.10.2012 17:54:09, Unknown
|
Cooling for 2DSSD established after some struggle.
2DSSD fully read out with cooling
HV= +40 V -40 V I = 2 uA
Beam position: none
Positions of thermal sensors:
11 Kapillare Anfang
12 Kapillare Ende
13 ---
14 unmittelbar nach Kapillare
15 Origami in
16 Origami out
17 Origami -z
18 Origami CE
Strip order:
n-Seite: rechts (Salev) nach links (Jura)
p-Seite: oben nach unten
APV chips with lower noise:
-z ce
p-side #5 #10 (= pos 4)
n-side #13 (= pos1) #19 (= pos 3)
|
Fri Oct 12 23:51:35 2012, Markus Friedl, common, Zimmernummer & Telefonnummern, , , , , Unknown
|
2nd week:
Markus: 38 - 420 (phone: 70 420)
Thomas: 38 - 431 (phone: 70 431)
Chrisu: 38 - 428 (phone: 70 428)
Manni: 38 - 434 (phone: 70 434)
1st week:
Markus: 41 - 130 (phone: 60 130)
Annekathrin: 41 - 206 (phone: 60 206)
Thomas: 39 - 108 (phone: 79 108)
Chrisu: 41 - 224 (phone: 60 224)
Manni: 39 - 317 (phone: 79 317)
Marko: 39 - 206 (phone: 79 206)
HEPHYs CERN Handy: 162471
Bat. 887 HNA-451: C3568Y
Bat. 887 HNA-447: 12345 (westseitige Tür) |
Wed Oct 17 14:06:24 2012, Markus Friedl, MicronOrigami_gross, Stack für stealth parasitic mode, , , , , Good
|
Dieser Stack wird jetzt quasi als Pausenkasperl eingebaut und läuft im "stealth parasitic" mode während der Irradiation-Woche. Gut, dass wir alles auch von Wien aus fernsteuern können :-)
|
Sat Oct 13 14:44:47 2012, Manfred Valentan, MicronOrigami_gross, Stack configuration, Belle Stack 3, , , , , Unknown
|
Module in Beamrichtung:
2-DSSD Origami -z
2-DSSD Origami ce
1 Micron Wedge New
1 Micron Wedge Old
n-Seite: Beam abgewand
p-Seite: Beam zugewand
Ausleserichtung Origami:
n-Seite aufsteigend links -> rechts
p-Seite aufsteigend oben -> unten
Ausleserichtung Wedge:
p-Seite: unten -> oben
n-Seite: rechts -> links
Breite Seite der Wedge Sensoren links
Auslese der p-Seite mit 4 APVs:
Origami: Obere 4 APVs (letzte 2 APVs nicht ausgelesen, unten)
Micron: APVs 4,5 nicht ausgelesen (2 obersten)
Micron new: APV0 auf n-Seite I²C - Fehler, APV3 auf p-Seite hat 1/2 Signal
|
Mon Oct 15 18:52:18 2012, Marko Dragicevic, MicronOrigami_gross, Stack configuration, Belle Stack 2, , , , , Unknown
|
Belle stack 2:
Modules:
- Origami #4
- Origami #3
- Wedge new 1
- Wedge old
n-Seite: Beam abgewand
p-Seite: Beam zugewand
Ausleserichtung Origami und Wedge:
- n-Seite: rechts -> links
- p-Seite: unten -> oben
Breite Seite der Wedge Sensoren links
Auslese der p-Seiten mit 4 APVs:
Origami: die beiden obersten APVs (#4,5) nicht ausgelesen
Micron: die beiden obersten APVs (#4,5) nicht ausgelesen
Micron new: APV0 auf n-Seite I²C - Fehler
Temperatur sensors:
- T11: Origami #4 in
- T12: Origami #4 out
- T13: Origami #3 in (wrong values!)
- T14: Origami #3 out
- T15: Wedge new 1 out
- T16: Wedge old out
- T17: Wedge new 1 cooling block
- T18: Wedge old cooling block
|
Fri Oct 12 23:52:57 2012, Markus Friedl, common, Sozialisierung, , , , , Unknown
|
Prost, Bergi!
(Das ist nur ein Test, ob du wirklich all die Elog-Emails liest.) |
Tue Oct 16 23:03:09 2012, Markus Friedl, common, Sozialisierung, , , , , Unknown
|
So steuert man einen Testbeam!
|
Fri Oct 12 15:04:55 2012, Marko Dragicevic, common, Network configuration, , , , , Unknown
|
Router:
IP 137.138.113.230
Hostname ROUTERHEPHY.cern.ch
connected on outlet 0887-1:4602/02
Devices
DAQPC: 192.168.0.36
Slowcontrol (REINRAUMLAPTOP): 192.168.0.37
AnneKathrinPC (AOH): 192.168.0.40
GPIB->Ethernet: 192.168.0.240
Coolingbox: 192.168.0.148
XY Table: 192.168.0.13
VAIO (CONF1): 192.168.0.42
BeamControl: 137.138.113.210 CWO-HNA447-H6A.CERN.CH
VMheros am DAQPC: 192.168.0.47
Port Forwards:
DAQPC (192.168.0.36): 5901
Slowcontrol (192.168.0.37): 5902
AnneKathrinPC (192.168.0.40): 5903
VAIO (CONF1) (192.168.0.42): 5904
VMheros am DAQPC (192.168.0.47): 5022
Comments:
Remote Control von CESAR:
$ x11vnc -bg -forever |
Thu Oct 18 09:14:07 2012, Manfred Valentan, common, Fahrt nach Mol, , , , , Unknown
|
Liebe Kollegen,
wir brechen. auf. Wünscht uns Glück! |
Wed Oct 31 15:54:59 2012, Christian Irmler, common, Ethernetkabel zu H6A, , , , , Unknown
|
Ethernetkabel HNA447 -> H6A:
Das obere Ende im Counting Hut liegt im Zwischenboden unterhalb der orangenen Steckdose beim Eingang mit dem elektronischen Zahlenschloß.
Das untere Ende in der Zone hängt aufgerollt an einem Pfeiler der Kabeltasse, unweit der Stelle an der das Kabel von oben hinunter gelassen wurden. Das ist ewta dort wo die Stiege zu PPE146 endet.
siehe Fotos :) |
Fri Oct 12 19:42:36 2012, Christian Irmler, MicronBaby, Configuration of normal stack, , , , , Unknown
|
Micron baby stack with all strips connected to APV chips
n-side faces the beam, p-side not read out.
Channels counting from bottom to top
Modules as seen by the beam:
Module name z-position [mm] Sensor ID
COMM-1 0.0 B2Micron_2965_10_baby2
COMM-2 24.7 B2Micron_2965_14_baby2
SPRAY-1 49.4 B2Micron_2825_05_baby3
SPRAY-2 75.5 B2Micron_2825_05_baby4
COMB-1 100.9 B2Micron_2965_10_baby4
COMB-2 135.0 B2Micron_2965_14_baby4
ATOLL-1 159.0 B2Micron_2965_14_baby3
ATOLL-2 183.7 B2Micron_2965_17_baby3 |
Sun Oct 14 02:00:12 2012, Manfred Valentan, MicronBaby, Configuration of intermediate stack, , , , , Unknown
|
Micron baby stack with every other strip connected to APV chips
n-side faces the beam, p-side not read out.
Channels counting from bottom to top
Modules as seen by the beam:
Module name z-position [mm] Sensor ID
COMM-I-1 0.0 B2Micron_2965_16_baby2
COMM-I-2 26.0 B2Micron_2965_17_baby2
SPRAY-I-1 50.4 B2Micron_2825_15_baby2 (known issue: I2C-error of APV3)
SPRAY-I-2 76.2 B2Micron_2825_15_baby3
COMB-I-1 101.8 B2Micron_2965_16_baby4
COMB-I-2 136.8 B2Micron_2965_17_baby4
ATOLL-I-1 161.0 B2Micron_2965_10_baby3
ATOLL-I-2 185.3 B2Micron_2965_13_baby3 |
Sat Oct 13 19:04:23 2012, Markus Friedl, MicronOrigami_gross, CO2 cooling, , , , , Unknown
|
Wir haben ja schon seit langem vermutet, dass man mit reiner Gasphase nicht kühlen kann. Jetzt erbringen wir den experimentellen Beweis dafür.
Expandierendes Gas kühlt zwar auch, aber nicht so tief runter und vor allem sind beträchtliche Temperaturdifferenzen zwischen den einzelnen Modulen zu beobachten. Das erste Origami bekommt etwa -13°C aber bis zum letzten Micron sind wir auf Raumtemperatur.
Warum das Ganze? Weil Bergi eine CO2-Batterie ohne Steigrohr bestellt hat. Annekathrin ist zwar froh, dass wir das herausgefunden hat und nicht ihre Anlage schuld ist, aber auch traurig, weil dieser Testbeam natürlich die Krönung ihrer Diplomarbeit hätte sein sollen.
Bestellung 27.9.2011 (EDH: Material Request 4732023): SCEM 60.04.15.150.4 = CO2 30 BK 12 CYL. 450 KG (TP); Price: 671 CHF -- das "TP" steht für tube plongeur, zu deutsch: Steigrohr
Bestellung 15.9.2012 (EDH: Material Request 5033041): SCEM 60.04.15.110.3 = CO2 30 BK 12 CYL. - 450 KG; Price: 391 CHF -- das ist erstens eine Andere SCEM (die von 2011 gibt es nicht mehr) und es steht kein "TP". Allein die andere Nummer hätte eine Rückfrage bei Annekathrin implizieren müssen.
Bergi: Das kostet dich ein großes "sumimasen".
Derzeit laufen wir mit Gaskühlung, aber natürlich denken wir schon an einen Plan B.
Über den offiziellen Weg am CERN kriegen wir niemals rechtzeitig CO2-Flaschen, weil die im Store gar nicht lagernd sind.
Montag früh könnten wir mit dem großen Auto zu einem Gashändler in der Nähe fahren und hoffen, dass dort 3 große Gasflaschen mit Steigrohr auf Lager sind. Batterie ist natürlich hoffnungslos, das können wir nicht ohne fremde Hilfe transportieren, einzelne Flaschen hingegen schon.
Eine andere Frage wäre, ob Hans Postema et al. ein paar richtige Gasflaschen auf Lager haben, die wir uns kurzfristig ausborgen könnten. |
Mon Oct 15 19:21:56 2012, Marko Dragicevic, common, Beam settings and profile, , , , , Unknown
|
Beam setting file:
H6A.804: FM HAD (H8=+180, H6=120@0mrad) PARALL IN H6A; RD42
Kommentar (Marko):
Das alte Beam Setting File war optimiert fuer H6B, sprich der Beam Focus war dort. Dieses Setting ist neuer und hat den Focus in H6B. |
|