109
|
Sun Oct 10 15:45:04 2010 |
Thomas Bergauer | common | end of success | | | 09.10.2010 19:44:23 | NaN |
;-)
genau das gleiche wollte ich auch gerade schreiben! |
108
|
Sat Oct 9 19:44:36 2010 |
Markus Friedl | common | end of success | | | 09.10.2010 19:44:23 | NaN |
|
107
|
Sat Oct 9 18:56:35 2010 |
Thomas Bergauer | common | path to telescope data | | | | |
eudetmac001.cern.ch
path is /Volumes/Data/eudet/Data/cern_2010_sept_silc_apvdaq (Vienna system)
and /Volumes/Data/eudet/Data/cern_2010_sept_silc_alibava (Santander system) |
106
|
Sat Oct 9 15:29:09 2010 |
Markus Friedl | BELLE | run | run058 | 30000 | 09.10.2010 15:26:50 | 09.10.2010 16:21:37 |
Origami1 (n-side only) + Micron1 (non-irrad)
angle scan: 73 - now with multi6 (previous run was single)
HV=100V / 35uA (beginning of run) 34uA (end of run)
cooling: chiller set to 15C
telescope run 20201
Origami p-side does not receive triggers -> disabled
origami flex straight part needed to be folded due to severe rotation angle -> nanonics connector problem...
|
|
105
|
Sat Oct 9 12:25:36 2010 |
Thomas Bergauer | BELLE | run | run057 | 276475 | 09.10.2010 12:22:21 | 09.10.2010 15:08:46 |
Origami1 (n-side only) + Micron1 (non-irrad)
angle scan: 73 degree
HV=100V / 36 uA (beginning), 35uA (end of run)
cooling: chiller set to 15C
PROBLEM: wrong config file (single hit, not multi6)
telescope run 20200
Origami p-side does not receive triggers -> disabled
origami flex straight part needed to be folded due to severe rotation angle -> nanonics connector problem...
|
|
104
|
Sat Oct 9 12:00:55 2010 |
Christian Irmler | BELLE | cal | cal017 | 83800 | 09.10.2010 11:59:37 | |
Origami1 (n-side only) + Micron1 (non-irrad)
angle scan: 73°
HV=100V / 40 µA
cooling: chiller set to 15°C
Origami p-side does not receive triggers -> disabled
origami flex straight part needed to be folded due to severe rotation angle -> nanonics connector problem... |
103
|
Sat Oct 9 00:17:02 2010 |
Markus Friedl | BELLE | run | run056 | 225k | 09.10.2010 00:15:53 | 09.10.2010 09:54:28 |
Belle stack w/o Origami2 after irradiation
HV=80V / I_start=255 µA, I_stop=243 µA
cooling: Origami only, chiller set to -25°C,
2 suspicious APVs on n side of Micron (possibly due to burnt strip)
same position as in previous run
telescope run 20198 - now with correct telescope threshold settings
a lot of "Error detected in APV 21, subevent 2. Event discarded, reset sent." errors started at 09.10.2010 09:00:25 (after whole night without any error) |
102
|
Fri Oct 8 23:43:07 2010 |
Christian Irmler | BELLE | run | run055 | 12316 | 08.10.2010 23:42:51 | 09.10.2010 00:14:56 |
Belle stack w/o Origami2 after irradiation
HV=80V / 255 µA
cooling: Origami only, chiller set to -25°C,
2 suspicious APVs on n side of Micron due to burnt strip
same position as in previous run
telescope run 20197 - wrong config file @ telescope -> new run with correct settings |
101
|
Fri Oct 8 23:22:32 2010 |
Manfred Valentan | BELLE | Schnittwinkel barrel vs. slanted | | | | |
17deg acceptance => Schnittwinkel barrel option = 90-17 = 73deg -----> rotary table 80.5
layer zusätzlich slanted um Winkel alpha => Schnittwinkel = 73 - alpha
slant angles: layer 4: 11.9deg => Schnittwinkel 61.1deg -----> rotary table 68.6
layer 5: 17.2deg => Schnittwinkel 55.8deg -----> rotary table 63.3
layer 6: 21.1deg => Schnittwinkel 51.9deg -----> rotary table 59.4
Im Vergleich mit senkrechtem Durchgang der Teilchen erwarte ich eine Vergrößerung der cluster width um den Faktor tan(Schnittwinkel):
slant layer 4: 1.81
slant layer 5: 1.47
slant layer 6: 1.28
barrel: 3.27 |
100
|
Fri Oct 8 23:05:03 2010 |
Markus Friedl | BELLE | cal | cal016 | 83800 | 08.10.2010 23:03:16 | |
Belle stack w/o Origami2 after irradiation
HV=80V / 258 µA
cooling: Origami only, chiller set to -25°C,
2 suspicious APVs on n side of Micron due to burnt strip
Bias connection of Origami repaired --> lower noise on p-side |
99
|
Fri Oct 8 18:12:43 2010 |
Markus Friedl | BELLE | run | run054 | 90000 | 08.10.2010 18:08:48 | 08.10.2010 21:49:36 |
Belle stack w/o Origami2 after irradiation
HV=80V I=224uA (beginning of run)
2 suspicious APVs on Micron n side
xy-table position 7 / 30 / 277.5 (shoots at the smaller edge of the trapezoidal sensor)
telescope run 20195 |
98
|
Fri Oct 8 18:12:15 2010 |
Christian Irmler | BELLE | run | run053 | 98000 | 08.10.2010 00:33:14 | 08.10.2010 13:36:57 |
Belle stack w/o Origami2 after irradiation
HV=80V
Telescope position 55/30/277.5
NIM trigger
temperature of chiller set to 15°C
Analog signal of APVs #22 and #23 has been bound to ground after 1200 events. |
97
|
Fri Oct 8 17:26:36 2010 |
Markus Friedl | BELLE | cal | cal015 | 83800 | 08.10.2010 17:22:44 | |
Belle stack without Origami2
HV=11V I=32uA
beam on
2 suspicious APVs on n side of Micron due to burnt strip
trying to identify problematic Micron n side channels with low bias voltage (APV22 baseline saturates at higher bias) |
96
|
Wed Oct 6 23:48:49 2010 |
Markus Friedl | BELLE | run | run052 | 165789 | 06.10.2010 23:47:12 | 07.10.2010 08:43:42 |
Belle stack without Origami2
HV=80V I=217uA
telescope setting 103 / 30 / 277.5
telescope run 20184
temperature of chiller set to 15°C
beam stopped at 7.10. 08:00 sharp, no beam until 8.10 - 8:00 according to SPS Page 1. We keep DAQ and telescope running and cross our fingers for at least some spills of particles.
Irradiated micron module 2, n-side: excessive hit count on channels at the edges of the APV chips - no such thing on the p-side |
95
|
Wed Oct 6 23:03:15 2010 |
Marko Dragicevic | BELLE | cal | cal014 | 83800 | 06.10.2010 23:01:56 | 06.10.2010 23:34:07 |
Belle stack without Origami2
HV=80V I=227uA
beam on |
94
|
Wed Oct 6 23:02:09 2010 |
Marko Dragicevic | common | Beam Configuration | | | | |
Beam Settings and Profile on 5.10.2010 - 17:00h
Saved as: H6B.801 |
Attachment 1: BeamProfile.gif
|
|
Attachment 2: BeamMagnets.gif
|
|
Attachment 3: BeamColli.gif
|
|
93
|
Wed Oct 6 16:33:55 2010 |
Markus Friedl | Micron Baby | run | run051 | 90000 | 06.10.2010 16:32:03 | 06.10.2010 20:59:58 |
8 x Micron Baby (3 irradiated), no cooling
HV=100V I=209uA (start of run, still rising); 203uA (end of run)
beam on, NIM trigger from TLU (no handshake)
telescope position 131 / 53.5 / 7.5
no telescope run - being tuned (some intervals during data taking are possible)
|
92
|
Wed Oct 6 11:59:29 2010 |
Markus Friedl | Micron Baby | run | run050 | 90000 | 06.10.2010 11:57:55 | 06.10.2010 15:30:37 |
8 x Micron Baby (3 irradiated), no cooling
HV=100V I=141uA (start of run)
beam on
telescope position 118 / 41.1 / 7.5
no telescope run |
91
|
Wed Oct 6 11:51:25 2010 |
Thomas Bergauer | Micron Baby | run | run049 | 4489 | 06.10.2010 11:38:21 | 06.10.2010 11:49:23 |
8 x Micron Baby (3 irradiated), no cooling
HV=100V I=141uA (start of run)
beam on
not yet optimized XY-table positions (111/43.5/7.5) |
90
|
Wed Oct 6 11:03:59 2010 |
Marko Dragicevic | Micron Baby | cal | cal013 | 83800 | 06.10.2010 11:03:12 | |
8 x Micron Baby (3 irradiated), no cooling
HV=100V I=115uA (start of run)
beam off
# VADJ/VPSP increased to 48 (was 30) for irradiated DUTs to restore baseline position (higher after irradiation)
ADC delay timing did not change after irradiation (verified)
|