Entry  Wed Sep 29 14:03:55 2010, Markus Friedl, BELLE, cal, cal004, 83800, 29.09.2010 14:02:25, 29.09.2010 14:45:05 

Belle stack: Origami-Micron-Micron-Origami
HV=100V, Ibias=100µA, beam off
no cooling

Temp. at modules: 
Origami 1: 50°C
Micron 1: 46°C
Micron 2: 48°C
Origami 2: 48°C

standard I2C settings
centered

Entry  Wed Sep 29 17:38:19 2010, Markus Friedl, Micron Baby, cal, cal005, 83800, 29.09.2010 17:36:38,  

micron baby stack
HV=100V I=6.5uA
Beam off

Entry  Thu Sep 30 15:06:28 2010, Markus Friedl, Micron Baby, cal, cal006, 83800, 30.09.2010 15:01:56,  

micron baby: 4 DUTs - angle scan
HV=100V I=6.8uA
Beam off (turned on at ~8k events)
 

Entry  Sat Oct 2 17:27:14 2010, Thomas Bergauer, ITE, cal, cal007, 83800, 02.10.2010 17:11:18 ,  intcal_ITE10PAD_W7.pngintcal_ITE10PAD_W9.png

ITE: 2 DUTs
2 x SILC + 2 x ITE
HV=100V

Beam on 

one short only per sensor

Entry  Sun Oct 3 08:14:39 2010, Markus Friedl, ITE, cal, cal008, 83800, 03.10.2010 08:13:06,  

6 x SILC + 2 x ITE
HV=100V
telescope position: 119

no beam (technical problem)

Entry  Sun Oct 3 14:35:12 2010, Marko Dragicevic, CNM, cal, cal009, 83800, 03.10.2010 13:50:54,  cal009_cal_apv6.pngcal009_cal_apv7.pngslowcontrol_CNM_cal009.gif

CMN-SILC stack
HV_SILC (Katze)=75V, 0.25uA (6 SiLC)
HV_CNM (Keithley)=30V, CNM_1=84nA, CNM_2=4.3uA
DUT outside of beam

calibration started after 1 hour warmup

Entry  Sun Oct 3 18:23:27 2010, Christian Irmler, Pt, cal, cal010, 83800, 03.10.2010 18:26:49 ,  slowcontrol_pt_cal010.gif

pt modules angle scan
h01-Pt and h02-Pt
currents: h01=23.1µA h02=10.4µA
HV=200V
beam on

modules flushed with dry air

 

Entry  Mon Oct 4 20:16:19 2010, Christian Irmler, BELLE, cal, cal011, 83800, 04.10.2010 20:15:00,  

angle scan Origami 1 and  Micron wedge 1
HV = 100V
chiller set to 14°C

HV connection of Origami 1 repaired

Entry  Tue Oct 5 17:50:23 2010, Christian Irmler, ITE, cal, cal012, 83800, 05.10.2010 17:50:26, 05.10.2010 18:09:48  

6 x SILC + 2 x ITE
HV=100V
telescope position: 124/48

DUTs and SILC4/5 swapped
new module order (downstream): SILC0 - SILC3 - SILC1 - W7 - W9 - SILC2 - SILC4 -SILC5
SILC1 and SILC2 are horizontal

beam on

modules flushed with dry air (black tube)
rel. humidity: ~0% @ 26°C

Entry  Wed Oct 6 11:03:59 2010, Marko Dragicevic, Micron Baby, cal, cal013, 83800, 06.10.2010 11:03:12,  

8 x Micron Baby (3 irradiated), no cooling
HV=100V I=115uA (start of run)
beam off

# VADJ/VPSP increased to 48 (was 30) for irradiated DUTs to restore baseline position (higher after irradiation)

ADC delay timing did not change after irradiation (verified)
 

Entry  Wed Oct 6 23:03:15 2010, Marko Dragicevic, BELLE, cal, cal014, 83800, 06.10.2010 23:01:56, 06.10.2010 23:34:07 

Belle stack without Origami2
HV=80V I=227uA

beam on

Entry  Fri Oct 8 17:26:36 2010, Markus Friedl, BELLE, cal, cal015, 83800, 08.10.2010 17:22:44,  

Belle stack without Origami2
HV=11V I=32uA

beam on

2 suspicious APVs on n side of Micron due to burnt strip

trying to identify problematic Micron n side channels with low bias voltage (APV22 baseline saturates at higher bias)

Entry  Fri Oct 8 23:05:03 2010, Markus Friedl, BELLE, cal, cal016, 83800, 08.10.2010 23:03:16,  

Belle stack w/o Origami2 after irradiation
HV=80V / 258 µA
cooling: Origami only, chiller set to -25°C,
2 suspicious APVs on n side of Micron due to burnt strip
 

Bias connection of Origami repaired --> lower noise on p-side

Entry  Sat Oct 9 12:00:55 2010, Christian Irmler, BELLE, cal, cal017, 83800, 09.10.2010 11:59:37,  

Origami1 (n-side only) + Micron1 (non-irrad)
angle scan: 73°
HV=100V / 40 µA
cooling: chiller set to 15°C

Origami p-side does not receive triggers -> disabled

origami flex straight part needed to be folded due to severe rotation angle -> nanonics connector problem...

Entry  Thu Mar 31 10:28:19 2011, Manfred Valentan, Micron Baby, DUT5, source run, post irrad, pre zupf, cal100, , 31.03.2011 10:24:59, 31.03.2011 10:33:15 

DUT5
cal-file: micronbaby_single_irrad_cal100.cfg


HV=+-50V
90Sr 1mCi ~centered
I_start=99muA
I_end=103muA

 

Entry  Thu Mar 31 12:11:04 2011, Manfred Valentan, Micron Baby, DUT4, source run, post irrad, pre zupf, cal101, , 31.03.2011 12:09:36, 31.03.2011 12:17:52 

DUT4
config file: micronbaby_single_irrrad_cal101.cfg


HV=+-50V
90Sr 1mCi ~centered
I_start=128µA
I_end=140µA

 

Entry  Mon Apr 11 18:25:35 2011, Manfred Valentan, Micron Baby, DUT5, source run, post irrad, pre zupf, cal104, 83800, 11.04.2011 18:12:38, 11.04.2011 18:20:50 

DUT5 (atoll)
config file: D:/cern10/micronbaby/data_lab/DUT5/micronbaby_single_irrrad.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=26µA
I_end=48µA

 

Entry  Wed Apr 13 12:13:09 2011, Manfred Valentan, Micron Baby, DUT5, source run, post irrad, post zupf ** DO NOT USE **, cal105, 83800, 13.04.2011 12:02:41, 13.04.2011 12:10:55  

DUT5 (atoll), each second bond removed to simulate intermediate strips
config file: D:/cern10/micronbaby/data_lab/DUT5/micronbaby_single_irrrad.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=18µA
I_end=21.5µA

data recorded with new switching power supplies and old intcal, which doesn't ignore bad strips!!!!!

Entry  Wed May 11 16:28:20 2011, Manfred Valentan, Micron Baby, DUT5, source run, post irrad, post zupf, cal106, 100k, 11.05.2011 16:00, 11.05.2011 16:25 

DUT5 (atoll)
config file: D:/cern10/micronbaby/data_lab/DUT5/micronbaby_multi6_irrrad.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=52µA
I_end=120µA

 DAQ modified so that only strips marked as good are considered when calculating the common mode correction.

Entry  Tue May 17 13:02:34 2011, Manfred Valentan, Micron Baby, TRK1, source run, unirrad, prezupf, cal107, 83800, 17.05.2011 12:20, 17.05.2011 12:35 

TRK1: B2Micron_2826_1_baby2 (common p-stop)
config file: D:/cern10/micronbaby/data_lab/TRK1/prezupf/micronbaby_single.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=1.1µA
I_end=2µA