ID Date Author Subject Run Number DUT Beam Energy Rate Settingdown Particle Type Data t-Start t-Stop Duration_min Frontend Phantom Phantom thickness Text Attachments
  96   Sun Jun 9 01:27:38 2024 Felix Ulrich-PurTOF CAL HADES LGADs WET Messung Knochen25TOF-Cal GSI194.3 MeVLow Flux HighProtonGood01:1601:30    3 cm dense bone (in isocentre)
2.8 Mhz on Ramon (after 5 Layers of
Silicon and close to second iso centre) and
  
  97   Sun Jun 9 01:32:51 2024 Felix Ulrich-PurTOF CAL HADES LGADs WET Messung Knochen26TOF-Cal GSI145.4 MeVLow Flux HighProtonGood01:3001:45    3 cm dense bone (in isocentre)
1.7 Mhz on Ramon (after 5 Layers of
Silicon and close to second iso centre) and
  
  98   Sun Jun 9 01:48:51 2024 Felix Ulrich-PurTOF CAL HADES LGADs WET Messung Knochen27TOF-Cal GSI194.3 MeVLow Flux HighProtonGood01:4801:59    4 cm dense bone (in isocentre)
2 Mhz on Ramon (after 5 Layers of
Silicon and close to second iso centre) and
  
  100   Sun Jun 9 02:02:16 2024 Felix Ulrich-PurTOF CAL HADES LGADs WET Messung Knochen28TOF-Cal GSI145.4 MeVLow Flux HighProtonGood02:0002:14    4 cm dense bone (in isocentre)
1.4 Mhz on Ramon (after 5 Layers of
Silicon and close to second iso centre) and
  
  101   Sun Jun 9 02:18:55 2024 Felix Ulrich-PurTOF CAL HADES LGADs WET Messung Knochen29TOF-Cal GSI194.3 MeVLow Flux HighProtonGood02:18     5 cm dense bone (in isocentre)
1.8 Mhz on Ramon (after 5 Layers of
Silicon and close to second iso centre) and
  
  147   Sat Dec 14 12:56:05 2024 Felix Ulrich-PurShift 23.11.24 low flux He test with RaMon8RaMon205.2 MeV/uLow Flux HighHeliumGood06:3906:401   Mean rate per spill~ 5e6 p/s (online estimation
without proper analysis)
  
  148   Sat Dec 14 12:56:38 2024 Felix Ulrich-PurShift 23.11.24 low flux He test with RaMon9RaMon177.8 MeV/uLow Flux HighHeliumGood06:4006:411   Mean rate per spill~ 4.5e6 p/s (online
estimation without proper analysis)
  
  149   Sat Dec 14 12:57:13 2024 Felix Ulrich-PurShift 23.11.24 low flux He test with RaMon10RaMon147.2 MeV/uLow Flux HighHeliumGood06:4206:421   Mean rate per spill~ 3..5e6 p/s (online
estimation without proper analysis)
  
  150   Sat Dec 14 12:58:35 2024 Felix Ulrich-PurShift 23.11.24 low flux He test with RaMon11RaMon124.4 MeV/uLow Flux HighHeliumGood06:4306:431   Mean rate per spill~ 4Ee6 p/s (online estimation
without proper analysis)
  
  151   Sat Dec 14 12:59:33 2024 Felix Ulrich-PurShift 23.11.24 low flux He test with RaMon12RaMon97.2MeV/uLow Flux HighHeliumGood06:4306:441   Mean rate per spill~ 6Ee6 p/s (online estimation
without proper analysis)
  
  152   Sat Dec 14 13:00:26 2024 Felix Ulrich-PurShift 23.11.24 low flux He test with RaMon13RaMon73.2 MeV/uLow Flux HighHeliumGood06:4406:451   Mean rate per spill~ 4Ee6 p/s (online estimation
without proper analysis)
  
  153   Sat Dec 14 13:01:14 2024 Felix Ulrich-PurShift 23.11.24 low flux He test with RaMon14RaMon63.3 MeV/uLow Flux HighHeliumGood06:4606:461   Mean rate per spill~ 5Ee6 p/s (online estimation
without proper analysis)
  
  154   Sat Dec 14 13:02:29 2024 Felix Ulrich-PurShift 23.11.24 low flux Carbon test with RaMon (did not work)15RaMon Low Flux HighCarbonCrap06:5507:016   Low flux carbon did not work

1µs EFE, 30s Extraction,
Deg100
  
  155   Sat Dec 14 13:04:57 2024 Felix Ulrich-PurShift 23.11.24 low flux Carbon test with RaMon (did not work)16RaMon Low Flux HighCarbonCrap07:0207:2018   Low flux carbon did not work

5µs EFE, 10s Extraction,
Deg100
  
  160   Sat Dec 14 13:31:49 2024 Felix Ulrich-PurEnergy vs TOF calibration21TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon147.2 MeV/uLow Flux HighHeliumGood10:4110:5615   TRB Setup: 260V bias and 250mA pre-amps,
threshold 1000

3.5MHz with Ramon and only 600kHz
  
  161   Sat Dec 14 13:34:18 2024 Felix Ulrich-PurEnergy vs TOF calibration21TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon97.2MeV/uLow Flux HighHeliumGood10:5811:2325   TRB Setup: 260V bias and 250mA pre-amps,
threshold 1000

3MHz with Ramon and only 300kHz
  
  162   Sat Dec 14 13:35:59 2024 Felix Ulrich-PurEnergy vs TOF calibration23TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon124.4 MeV/uLow Flux HighHeliumGood11:2411:3713   TRB Setup: 260V bias and 250mA pre-amps,
threshold 1000

5MHz with Ramon and only 500kHz
  
  163   Sat Dec 14 13:37:26 2024 Felix Ulrich-PurEnergy vs TOF calibration24TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon73.2 MeV/uLow Flux HighHeliumGood11:4212:0018   TRB Setup: 260V bias and 250mA pre-amps,
threshold 1000

2MHz with Ramon and only 200kHz
  
  164   Sat Dec 14 13:39:15 2024 Felix Ulrich-PurEnergy vs TOF calibration25TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon63.3 MeV/uLow Flux HighHeliumGood12:0512:2318   TRB Setup: 260V bias and 250mA pre-amps,
threshold 1000

380 kHz with Ramon and only 180kHz
  
  165   Sat Dec 14 13:40:47 2024 Felix Ulrich-Purthreshold and preamp tests26TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon63.3 MeV/uLow Flux HighHeliumGood12:3512:4914   TRB Setup: 260V bias and 250mA pre-amps,
threshold 10000 (changed threshold)

Diamond: 500mA, threshold 50
  
ELOG V3.1.5-fc6679b