Home CMS Production Clean room MedAustron HEPHY testbeams old
testbeam-MA2018 testbeam-MA2019 testbeam-MA2020 testbeam-MA2021 testbeam-MA2023 testbeam-MA2024
  Testbeams at MedAustron in 2024, Page 8 of 11  Not logged in ELOG logo
Entry  Sat Dec 14 13:34:18 2024, Felix Ulrich-Pur, Energy vs TOF calibration, 21, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 10:58, 11:23, 25, , ,  
TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

3MHz with Ramon and only 300kHz with TRB-44ps single channel res prae calib
Entry  Sat Dec 14 13:35:59 2024, Felix Ulrich-Pur, Energy vs TOF calibration, 23, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 124.4 MeV/u, Low Flux High, Helium, Good, 11:24, 11:37, 13, , ,  
TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

5MHz with Ramon and only 500kHz TRB
Entry  Sat Dec 14 13:37:26 2024, Felix Ulrich-Pur, Energy vs TOF calibration, 24, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 73.2 MeV/u, Low Flux High, Helium, Good, 11:42, 12:00, 18, , ,  
TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

2MHz with Ramon and only 200kHz TRB
Entry  Sat Dec 14 13:39:15 2024, Felix Ulrich-Pur, Energy vs TOF calibration, 25, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 63.3 MeV/u, Low Flux High, Helium, Good, 12:05, 12:23, 18, , ,  
TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

380 kHz with Ramon and only 180kHz TRB, 42ps single channel res prae calib
Entry  Sat Dec 14 13:40:47 2024, Felix Ulrich-Pur, threshold and preamp tests, 26, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 63.3 MeV/u, Low Flux High, Helium, Good, 12:35, 12:49, 14, , ,  
TRB Setup: 260V bias and 250mA pre-amps, threshold 10000 (changed threshold)

Diamond: 500mA, threshold 50
Entry  Sat Dec 14 13:42:52 2024, Felix Ulrich-Pur, threshold and preamp tests , 27, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 63.3 MeV/u, Low Flux High, Helium, Good, 12:52, 13:16, 24, , ,  
TRB Setup: 260V bias and 250mA pre-amps, threshold 1000 (changed threshold)

Diamond: 900mA, threshold 50
Entry  Sat Dec 14 13:45:34 2024, Felix Ulrich-Pur, threshold and preamp tests , 28, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 63.3 MeV/u, Low Flux High, Helium, Good, 13:20, 13:32, 2, , ,  
TRB Setup: 260V bias and 250mA pre-amps, threshold 1000 (changed threshold)

Diamond: 900mA, threshold 1000 offset
Entry  Sat Dec 14 13:46:11 2024, Felix Ulrich-Pur, WET measurement slabs, 29, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 205.2 MeV/u, Low Flux High, Helium, Good, 13:34, 14:03, 29, , Plastic Water Slabs, 1cm 
1cm Plastic Water

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000 (changed threshold)

Diamond: 900mA, threshold 1000 offset
Entry  Sat Dec 14 13:49:20 2024, Felix Ulrich-Pur, WET measurement slabs, 30, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 147.2 MeV/u, Low Flux High, Helium, Good, 14:02, 14:45, 43, , Plastic Water Slabs, 1cm 
1cm plastic water

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset
Entry  Sat Dec 14 13:56:16 2024, Felix Ulrich-Pur, WET measurement slabs, 31, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 205.2 MeV/u, Low Flux High, Helium, Good, 14:48, 15:06, 18, , Plastic Water Slabs, 2cm 
2cm plastic water

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset
Entry  Sat Dec 14 15:09:34 2024, Felix Ulrich-Pur, WET measurement slabs, 32, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 147.2 MeV/u, Low Flux High, Helium, Good, 15:06, 15:19, 13, , Plastic Water Slabs, 2cm 
2cm plastic water

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset
Entry  Sat Dec 14 15:10:35 2024, Felix Ulrich-Pur, WET measurement slabs, 33, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 205.2 MeV/u, Low Flux High, Helium, Good, 15:24, 15:36, 12, , Plastic Water Slabs, 3cm 
3cm plastic water

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset
Entry  Sat Dec 14 15:12:01 2024, Felix Ulrich-Pur, WET measurement slabs, 34, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 147.2 MeV/u, Low Flux High, Helium, Good, 15:37, 15:48, 11, , Plastic Water Slabs, 3cm 
3cm plastic water

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset
Entry  Sat Dec 14 15:12:46 2024, Felix Ulrich-Pur, WET measurement slabs, 35, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 205.2 MeV/u, Low Flux High, Helium, Good, 15:53, 16:10, 17, , Plastic Water Slabs, 4cm 
4cm plastic water

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset
Entry  Sat Dec 14 15:15:04 2024, Felix Ulrich-Pur, WET measurement slabs, 36, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 147.2 MeV/u, Low Flux High, Helium, Good, 16:11, 16:23, 12, , Plastic Water Slabs, 4cm 
4cm plastic water

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset
Entry  Sat Dec 14 15:15:46 2024, Felix Ulrich-Pur, WET measurement slabs, 37, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 205.2 MeV/u, Low Flux High, Helium, Good, 16:28, 16:42, 12, , Plastic Water Slabs, 5cm 
5cm plastic water

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset
Entry  Sat Dec 14 15:17:10 2024, Felix Ulrich-Pur, WET measurement slabs, 38, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 147.2 MeV/u, Low Flux High, Helium, Good, 16:42, 16:55, 13, , Plastic Water Slabs, 5cm 
5cm plastic water

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset
Entry  Sat Dec 14 15:17:57 2024, Felix Ulrich-Pur, WET measurement slabs, 39, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 205.2 MeV/u, Low Flux High, Helium, Good, 17:00, 17:14, 13, , Dense Bone Slabs, 1cm 
1cm dense bone

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset
Entry  Sat Dec 14 15:18:48 2024, Felix Ulrich-Pur, WET measurement slabs, 40, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 147.2 MeV/u, Low Flux High, Helium, Good, 17:14, 17:28, 14, , Dense Bone Slabs, 1cm 
1cm dense bone

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset
Entry  Sat Dec 14 15:19:57 2024, Felix Ulrich-Pur, WET measurement slabs, 41, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 205.2 MeV/u, Low Flux High, Helium, Good, 17:32, 17:44, 12, , Dense Bone Slabs, 2cm 
2cm dense bone

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset
ELOG V3.1.5-fc6679b