3 cm plastic water (in isocentre) 4.3 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 1MHz on LGAD
WET measurement
3 cm plastic water (in isocentre) 2.6 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 400 kHz on LGAD
wrong energy don't use
4 cm plastic water (in isocentre) 3.8 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 900 kHz on LGAD
4 cm plastic water (in isocentre) 2 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 500 kHz on LGAD
5 cm plastic water (in isocentre) 3.7 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 795 kHz on LGAD
5 cm plastic water (in isocentre) 2 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 330 kHz on LGAD
14 min break with beam off (DAQ failed, had to power cycle TRB3s and all DIRICH boards-->out of synch DIRICH)
1 cm dense bone (in isocentre) 4.5 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 1.2 MHz on LGAD
1 cm dense bone (in isocentre) 2.8 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 600 kHz on LGAD
2 cm dense bone (in isocentre) - Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 1 MHz on LGAD
interlock at 00:31 till 00:50 --> 19min break
2 cm dense bone (in isocentre) 2.1 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 315 kHz on LGAD
3 cm dense bone (in isocentre) 2.8 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 600 kHz on LGAD
3 cm dense bone (in isocentre) 1.7 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 230 kHz on LGAD
4 cm dense bone (in isocentre) 2 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 470 kHz on LGAD
4 cm dense bone (in isocentre) 1.4 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 175 kHz on LGAD
5 cm dense bone (in isocentre) 1.8 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 350 kHz on LGAD
position of duts needs to be switched
new geo file with adjusted dut position
Mean rate per spill~ 4-6E5 p/s (online estimation without proper analysis)
Mean rate per spill~ 6-8E5 p/s (online estimation without proper analysis)