Home CMS Production Clean room MedAustron HEPHY testbeams old
testbeam-MA2018 testbeam-MA2019 testbeam-MA2020 testbeam-MA2021 testbeam-MA2023 testbeam-MA2024
  Testbeams at MedAustron in 2024, Page 5 of 11  Not logged in ELOG logo
ID Date Author Subject Run Numberdown DUT Beam Energy Rate Setting Particle Type Data t-Start t-Stop Duration_min Frontend Phantom Phantom thickness
  143   Sat Dec 14 12:53:28 2024 Felix Ulrich-PurShift 23.11.24 low flux He test with RaMon4RaMon124.4 MeV/uLow Flux MediumHeliumGood06:3306:331   

Mean rate per spill~ 2-8E5 p/s (online estimation without proper analysis)

  178   Sat Dec 14 15:17:57 2024 Felix Ulrich-PurWET measurement slabs39TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon205.2 MeV/uLow Flux HighHeliumGood17:0017:1413 Dense Bone Slabs1cm

1cm dense bone

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

4MHz Ramon 700 kHz TRB

  177   Sat Dec 14 15:17:10 2024 Felix Ulrich-PurWET measurement slabs38TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon147.2 MeV/uLow Flux HighHeliumGood16:4216:5513 Plastic Water Slabs5cm

5cm plastic water

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

2.5MHz Ramon 200 kHz TRB

  176   Sat Dec 14 15:15:46 2024 Felix Ulrich-PurWET measurement slabs37TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon205.2 MeV/uLow Flux HighHeliumGood16:2816:4212 Plastic Water Slabs5cm

5cm plastic water

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

3-5MHz Ramon 500 kHz TRB

  175   Sat Dec 14 15:15:04 2024 Felix Ulrich-PurWET measurement slabs36TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon147.2 MeV/uLow Flux HighHeliumGood16:1116:2312 Plastic Water Slabs4cm

4cm plastic water

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

3MHz Ramon 300 kHz TRB

  174   Sat Dec 14 15:12:46 2024 Felix Ulrich-PurWET measurement slabs35TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon205.2 MeV/uLow Flux HighHeliumGood15:5316:1017 Plastic Water Slabs4cm

4cm plastic water

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

4MHz Ramon 600 kHz TRB

  173   Sat Dec 14 15:12:01 2024 Felix Ulrich-PurWET measurement slabs34TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon147.2 MeV/uLow Flux HighHeliumGood15:3715:4811 Plastic Water Slabs3cm

3cm plastic water

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

3MHz Ramon 300 kHz TRB

  172   Sat Dec 14 15:10:35 2024 Felix Ulrich-PurWET measurement slabs33TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon205.2 MeV/uLow Flux HighHeliumGood15:2415:3612 Plastic Water Slabs3cm

3cm plastic water

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

3-4MHz Ramon 600 kHz TRB

  171   Sat Dec 14 15:09:34 2024 Felix Ulrich-PurWET measurement slabs32TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon147.2 MeV/uLow Flux HighHeliumGood15:0615:1913 Plastic Water Slabs2cm

2cm plastic water

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

 

 

  170   Sat Dec 14 13:56:16 2024 Felix Ulrich-PurWET measurement slabs31TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon205.2 MeV/uLow Flux HighHeliumGood14:4815:0618 Plastic Water Slabs2cm

2cm plastic water

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

4MHz Ramon 1.4MHz TRB

  169   Sat Dec 14 13:49:20 2024 Felix Ulrich-PurWET measurement slabs30TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon147.2 MeV/uLow Flux HighHeliumGood14:0214:4543 Plastic Water Slabs1cm

1cm plastic water

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

4MHz Ramon 500 kHz TRB

  74   Sat Jun 8 19:57:39 2024 Felix Ulrich-PurTOF CAL HADES LGADs Calibrationcurve3TOF-Cal GSI194.3 MeVLow Flux MediumProtonGood16:5117:28    

600 khz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 235kHz on LGAD

TOF Calibration Curve

  142   Sat Dec 14 12:52:57 2024 Felix Ulrich-PurShift 23.11.24 low flux He test with RaMon3RaMon147.2 MeV/uLow Flux MediumHeliumGood06:3206:331   

Mean rate per spill~ 2.5-4-5E5 p/s (online estimation without proper analysis)

  101   Sun Jun 9 02:18:55 2024 Felix Ulrich-PurTOF CAL HADES LGADs WET Messung Knochen29TOF-Cal GSI194.3 MeVLow Flux HighProtonGood02:18     

5 cm dense bone (in isocentre)
1.8 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 350 kHz on LGAD

WET measurement

  168   Sat Dec 14 13:46:11 2024 Felix Ulrich-PurWET measurement slabs29TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon205.2 MeV/uLow Flux HighHeliumGood13:3414:0329 Plastic Water Slabs1cm

1cm Plastic Water

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000 (changed threshold)

Diamond: 900mA, threshold 1000 offset

DOGMA lost DiRICH after restarting Wave (happened several times)

  100   Sun Jun 9 02:02:16 2024 Felix Ulrich-PurTOF CAL HADES LGADs WET Messung Knochen28TOF-Cal GSI145.4 MeVLow Flux HighProtonGood02:0002:14    

4 cm dense bone (in isocentre)
1.4 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 175 kHz on LGAD

WET measurement

  167   Sat Dec 14 13:45:34 2024 Felix Ulrich-Purthreshold and preamp tests 28TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon63.3 MeV/uLow Flux HighHeliumGood13:2013:322   

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000 (changed threshold)

Diamond: 900mA, threshold 1000 offset

  98   Sun Jun 9 01:48:51 2024 Felix Ulrich-PurTOF CAL HADES LGADs WET Messung Knochen27TOF-Cal GSI194.3 MeVLow Flux HighProtonGood01:4801:59    

4 cm dense bone (in isocentre)
2 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 470 kHz on LGAD

WET measurement

  166   Sat Dec 14 13:42:52 2024 Felix Ulrich-Purthreshold and preamp tests 27TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon63.3 MeV/uLow Flux HighHeliumGood12:5213:1624   

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000 (changed threshold)

Diamond: 900mA, threshold 50

  97   Sun Jun 9 01:32:51 2024 Felix Ulrich-PurTOF CAL HADES LGADs WET Messung Knochen26TOF-Cal GSI145.4 MeVLow Flux HighProtonGood01:3001:45    

3 cm dense bone (in isocentre)
1.7 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 230 kHz on LGAD

WET measurement

ELOG V3.1.5-fc6679b