Sat Dec 14 13:02:29 2024, Felix Ulrich-Pur, Shift 23.11.24 low flux Carbon test with RaMon (did not work), 15, RaMon, , Low Flux High, Carbon, Crap, 06:55, 07:01, 6, , ,
|
Low flux carbon did not work
1µs EFE, 30s Extraction, Deg100
we tried different energies (402.MeV/u)-> 120 MeV/u, but unfortunately we did not get any beam |
Sat Dec 14 13:04:57 2024, Felix Ulrich-Pur, Shift 23.11.24 low flux Carbon test with RaMon (did not work), 16, RaMon, , Low Flux High, Carbon, Crap, 07:02, 07:20, 18, , ,
|
Low flux carbon did not work
5µs EFE, 10s Extraction, Deg100
we tried different energies (402.MeV/u)-> 238 MeV/u, but unfortunately we did not get any beam |
Sat Dec 14 13:07:15 2024, Felix Ulrich-Pur, Energy vs TOF calibration, 17, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 205.2 MeV/u, Low Flux Medium, Helium, Good, 08:46, 09:42, 56, , ,
|
before run, setups had to be aligned (first slab in slabholder was put in isocentre. However, in this run, no slab was used). Room was closed at 08:10
Beam->Nozzle-> Slabholder->TOF-CAL (TRB)-> DOGMA Setup(Diamond+LGADs)-> RaMon |
Sat Dec 14 13:16:33 2024, Felix Ulrich-Pur, Energy vs TOF calibration +Alignment of DOGMA setup, 18, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 205.2 MeV/u, Low Flux Medium, Helium, Good, 09:42, 09:48, 6, , ,
|
TRB Setup: 260V bias and 250mA pre-amps, threshold 1000
DOGMA setup had to be shifted manually to align it with the beam |
Sat Dec 14 13:26:29 2024, Felix Ulrich-Pur, Energy vs TOF calibration, 19, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 177.8 MeV/u, Low Flux Medium, Helium, Good, 10:08, 10:25, 17, , ,
|
TRB Setup: 260V bias and 250mA pre-amps, threshold 1000
750kHz with Ramon and only 160kHz with TRB, 44ps single channel res preacalib |
Sat Dec 14 13:28:57 2024, Felix Ulrich-Pur, Energy vs TOF calibration, 20, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 147.2 MeV/u, Low Flux Medium, Helium, Good, 10:33, 10:38, 5, , ,
|
TRB Setup: 260V bias and 250mA pre-amps, threshold 1000
300kHz with Ramon and only 40kHz with TRB--> decision to go to low flux high was made |
Sat Dec 14 13:31:49 2024, Felix Ulrich-Pur, Energy vs TOF calibration, 21, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 147.2 MeV/u, Low Flux High, Helium, Good, 10:41, 10:56, 15, , ,
|
TRB Setup: 260V bias and 250mA pre-amps, threshold 1000
3.5MHz with Ramon and only 600kHz with TRB-52ps single channel res prae calib |
Sat Dec 14 13:34:18 2024, Felix Ulrich-Pur, Energy vs TOF calibration, 21, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 10:58, 11:23, 25, , ,
|
TRB Setup: 260V bias and 250mA pre-amps, threshold 1000
3MHz with Ramon and only 300kHz with TRB-44ps single channel res prae calib |
Sat Dec 14 13:35:59 2024, Felix Ulrich-Pur, Energy vs TOF calibration, 23, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 124.4 MeV/u, Low Flux High, Helium, Good, 11:24, 11:37, 13, , ,
|
TRB Setup: 260V bias and 250mA pre-amps, threshold 1000
5MHz with Ramon and only 500kHz TRB |
Sat Dec 14 13:37:26 2024, Felix Ulrich-Pur, Energy vs TOF calibration, 24, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 73.2 MeV/u, Low Flux High, Helium, Good, 11:42, 12:00, 18, , ,
|
TRB Setup: 260V bias and 250mA pre-amps, threshold 1000
2MHz with Ramon and only 200kHz TRB |
Sat Dec 14 13:39:15 2024, Felix Ulrich-Pur, Energy vs TOF calibration, 25, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 63.3 MeV/u, Low Flux High, Helium, Good, 12:05, 12:23, 18, , ,
|
TRB Setup: 260V bias and 250mA pre-amps, threshold 1000
380 kHz with Ramon and only 180kHz TRB, 42ps single channel res prae calib |
Sat Dec 14 13:40:47 2024, Felix Ulrich-Pur, threshold and preamp tests, 26, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 63.3 MeV/u, Low Flux High, Helium, Good, 12:35, 12:49, 14, , ,
|
TRB Setup: 260V bias and 250mA pre-amps, threshold 10000 (changed threshold)
Diamond: 500mA, threshold 50 |
Sat Dec 14 13:42:52 2024, Felix Ulrich-Pur, threshold and preamp tests , 27, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 63.3 MeV/u, Low Flux High, Helium, Good, 12:52, 13:16, 24, , ,
|
TRB Setup: 260V bias and 250mA pre-amps, threshold 1000 (changed threshold)
Diamond: 900mA, threshold 50 |
Sat Dec 14 13:45:34 2024, Felix Ulrich-Pur, threshold and preamp tests , 28, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 63.3 MeV/u, Low Flux High, Helium, Good, 13:20, 13:32, 2, , ,
|
TRB Setup: 260V bias and 250mA pre-amps, threshold 1000 (changed threshold)
Diamond: 900mA, threshold 1000 offset |
Sat Dec 14 15:28:46 2024, Felix Ulrich-Pur, threshold and preamp tests, 49, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 20:04, 20:09, 5, , ,
|
TRB Setup: 260V bias and 250mA pre-amps, threshold 10000 (was too high)
Diamond: 900mA, threshold 1000 offset
3.5MHz Ramon |
Sat Dec 14 15:30:23 2024, Felix Ulrich-Pur, threshold and preamp tests, 50, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 20:19, 20:35, 16, , ,
|
TRB Setup: 260V bias and 250mA pre-amps, threshold 5000 (looks good, clustersize max at ~3)
Diamond: 900mA, threshold 1000 offset
4MHz Ramon 370kHz TRB |
Sat Dec 14 15:50:19 2024, Felix Ulrich-Pur, LGAD+Diamond test with clinical carbon beam, 80, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 402MeV/u, , Carbon, Good, 05:15, 05:50, 35, , ,
|
Carbon Deg10 (clinical rate)
TRB Setup: 260V bias and 250mA pre-amps, threshold 5000
Diamond: 900mA, threshold 1000 offset |
|