ID Date Author Subject Run Number DUT Beam Energy Rate Setting Particle Type Data t-Start t-Stop Duration_min Frontend Phantomdown Phantom thickness
  154   Sat Dec 14 13:02:29 2024 Felix Ulrich-PurShift 23.11.24 low flux Carbon test with RaMon (did not work)15RaMon Low Flux HighCarbonCrap06:5507:016   

Low flux carbon did not work

1µs EFE, 30s Extraction, Deg100
we tried different energies (402.MeV/u)-> 120 MeV/u, but unfortunately we did not get any beam

  155   Sat Dec 14 13:04:57 2024 Felix Ulrich-PurShift 23.11.24 low flux Carbon test with RaMon (did not work)16RaMon Low Flux HighCarbonCrap07:0207:2018   

Low flux carbon did not work

5µs EFE, 10s Extraction, Deg100
we tried different energies (402.MeV/u)-> 238 MeV/u, but unfortunately we did not get any beam

  156   Sat Dec 14 13:07:15 2024 Felix Ulrich-PurEnergy vs TOF calibration17TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon205.2 MeV/uLow Flux MediumHeliumGood08:4609:4256   

before run, setups had to be aligned (first slab in slabholder was put in isocentre. However, in this run, no slab was used). Room was closed at 08:10

Beam->Nozzle-> Slabholder->TOF-CAL (TRB)-> DOGMA Setup(Diamond+LGADs)-> RaMon

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

measured 400kHz with Ramon and only 100kHz with LGADs (third LGAD) and 60kHz accepted trrigger rate

beam was shifted vertically by x? cm? (need to lookup again). So TRB setup was shifted with motors such that the He beam was hitting the centre of the LGADs

  157   Sat Dec 14 13:16:33 2024 Felix Ulrich-PurEnergy vs TOF calibration +Alignment of DOGMA setup18TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon205.2 MeV/uLow Flux MediumHeliumGood09:4209:486   

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

DOGMA setup had to be shifted manually to align it with the beam

  158   Sat Dec 14 13:26:29 2024 Felix Ulrich-PurEnergy vs TOF calibration19TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon177.8 MeV/uLow Flux MediumHeliumGood10:0810:2517   

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

750kHz with Ramon and only 160kHz with TRB, 44ps single channel res preacalib

  159   Sat Dec 14 13:28:57 2024 Felix Ulrich-PurEnergy vs TOF calibration20TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon147.2 MeV/uLow Flux MediumHeliumGood10:3310:385   

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

300kHz with Ramon and only 40kHz with TRB--> decision to go to low flux high was made

  160   Sat Dec 14 13:31:49 2024 Felix Ulrich-PurEnergy vs TOF calibration21TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon147.2 MeV/uLow Flux HighHeliumGood10:4110:5615   

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

3.5MHz with Ramon and only 600kHz with TRB-52ps single channel res prae calib

  161   Sat Dec 14 13:34:18 2024 Felix Ulrich-PurEnergy vs TOF calibration21TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon97.2MeV/uLow Flux HighHeliumGood10:5811:2325   

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

3MHz with Ramon and only 300kHz with TRB-44ps single channel res prae calib

  162   Sat Dec 14 13:35:59 2024 Felix Ulrich-PurEnergy vs TOF calibration23TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon124.4 MeV/uLow Flux HighHeliumGood11:2411:3713   

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

5MHz with Ramon and only 500kHz TRB

  163   Sat Dec 14 13:37:26 2024 Felix Ulrich-PurEnergy vs TOF calibration24TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon73.2 MeV/uLow Flux HighHeliumGood11:4212:0018   

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

2MHz with Ramon and only 200kHz TRB

  164   Sat Dec 14 13:39:15 2024 Felix Ulrich-PurEnergy vs TOF calibration25TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon63.3 MeV/uLow Flux HighHeliumGood12:0512:2318   

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

380 kHz with Ramon and only 180kHz TRB, 42ps single channel res prae calib

  165   Sat Dec 14 13:40:47 2024 Felix Ulrich-Purthreshold and preamp tests26TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon63.3 MeV/uLow Flux HighHeliumGood12:3512:4914   

TRB Setup: 260V bias and 250mA pre-amps, threshold 10000 (changed threshold)

Diamond: 500mA, threshold 50

  166   Sat Dec 14 13:42:52 2024 Felix Ulrich-Purthreshold and preamp tests 27TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon63.3 MeV/uLow Flux HighHeliumGood12:5213:1624   

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000 (changed threshold)

Diamond: 900mA, threshold 50

  167   Sat Dec 14 13:45:34 2024 Felix Ulrich-Purthreshold and preamp tests 28TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon63.3 MeV/uLow Flux HighHeliumGood13:2013:322   

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000 (changed threshold)

Diamond: 900mA, threshold 1000 offset

  188   Sat Dec 14 15:28:46 2024 Felix Ulrich-Purthreshold and preamp tests49TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon97.2MeV/uLow Flux HighHeliumGood20:0420:095   

TRB Setup: 260V bias and 250mA pre-amps, threshold 10000 (was too high)

Diamond: 900mA, threshold 1000 offset

3.5MHz Ramon

  189   Sat Dec 14 15:30:23 2024 Felix Ulrich-Purthreshold and preamp tests50TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon97.2MeV/uLow Flux HighHeliumGood20:1920:3516   

TRB Setup: 260V bias and 250mA pre-amps, threshold 5000 (looks good, clustersize max at ~3)

Diamond: 900mA, threshold 1000 offset

4MHz Ramon 370kHz TRB

  218   Sat Dec 14 15:50:19 2024 Felix Ulrich-PurLGAD+Diamond test with clinical carbon beam80TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon402MeV/u CarbonGood05:1505:5035   

Carbon Deg10 (clinical rate)

TRB Setup: 260V bias and 250mA pre-amps, threshold 5000

Diamond: 900mA, threshold 1000 offset